Medline Convenience Kits: 1) LVAD DRIVELINE TRAY recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) LVAD DRIVELINE TRAY, Model Number: DM1035A; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What is being recalled
Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) LVAD DRIVELINE TRAY, Model Number: DM1035A; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
Why
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution.
How to tell if yours is affected
1) DM1035, UDI-DI: 10653160327752(each), 653160327755(case), Lot Number: 2020042290; 2) DM1035, UDI-DI: 10653160327752(each), 653160327755(case), Lot Number: 2020052290; 3) DM1035, UDI-DI: 10653160327752(each), 653160327755(case), Lot Number: 2020072290; 4) DM1035, UDI-DI: 10653160327752(each), 653160327755(case), Lot Number: 2020082690; 5) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2022030390; 6) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2022072590; 7) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2023062390; 8) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2023101290; 9) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2025031190; 10) DM1035A, UDI-DI: 10653160345251(each), 653160345254(case), Lot Number: 2025051290; 11) DT8670A, UDI-DI: 653160177473(each), 10653160177470(case), Lot Number: 2020051390
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2125-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 7, 2026 |
| Reported | May 13, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 9720 units |
More recalls from Medline Industries, LP · all medical device recalls · recalls from January 2026