1. DBD-MTS recall (Medline Industries, LP)

Medical deviceClass IOngoing

Summary: Medline Industries, LP is recalling 1. DBD-MTS,3V,MAN,KIT, Medline Kit Number/SKU 60010195; 2. MTS-LEFT HEART,KIT, Medline Kit Number/SKU 60033123; 3. MTS,3 VALVE,KIT, Medline Kit Number/SKU 60050183; 4. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60080349; 5. MTS,3 VALVE,LEFT HEART,KIT, Medline Kit Number/SKU 60100338; 6. MT. Reason: Medline has identified the presence of particulate within the fluid path of the Manifolds.

What is being recalled

1. DBD-MTS,3V,MAN,KIT, Medline Kit Number/SKU 60010195; 2. MTS-LEFT HEART,KIT, Medline Kit Number/SKU 60033123; 3. MTS,3 VALVE,KIT, Medline Kit Number/SKU 60050183; 4. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60080349; 5. MTS,3 VALVE,LEFT HEART,KIT, Medline Kit Number/SKU 60100338; 6. MTS LHK, Medline Kit Number/SKU 60101017; 7. DBD-MTS,3 VALVE,KIT, Medline Kit Number/SKU 60130506; 8. MTS,LEFT,HRT,KIT, Medline Kit Number/SKU 60131324; 9. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60132238; 10. MTS,3,PORT,KIT, Medline Kit Number/SKU 60138301; 11. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60140589; 12. MTS,3 VALVE,W/, Medline Kit Number/SKU 60192012; 13. DBD-MTS,4V,MANIFOLD,KIT, Medline Kit Number/SKU 60220488; 14. MTS,LHK, Medline Kit Number/SKU 60232091; 15. MTO,3 VALVE,MANIFOLD,KIT, Medline Kit Number/SKU 65010037; 16. MTO,LEFT HEART,KIT, Medline Kit Number/SKU 65012819; 17. DBD-MTO,2V,LEFT HEART,KIT, Medline Kit Number/SKU 65040421; 18. DBD-MTO,LEFT HEART,KIT, Medline Kit Number/SKU 65183482; 19. MTS KIM, Medline Kit Number/SKU 65660872; 20. LHK,-, Medline Kit Number/SKU 601006269; 21. DBD-MTS,MANIFOLD,KIT, Medline Kit Number/SKU 601203616; 22. MTO,4V,KIT, Medline Kit Number/SKU 650104611; 23. MTO,MANIFOLD,KIT, Medline Kit Number/SKU 651304116; 24. KIMAL 3 PORT RH OFF WITH CC PG, Medline Kit Number/SKU 656206837; 25. KIM MANIFOLD SET VER ONE - PG, Medline Kit Number/SKU 656608724; 26. CARDIAC MAN PG, Medline Kit Number/SKU 656608737; 27. MTO MANIFOLD, Medline Kit Number/SKU 656608739; 28. CARDIAC MANIFOLD SET, Medline Kit Number/SKU 656608749; 29. CARDIAC MAN SET, Medline Kit Number/SKU 656608753; 30. KIMAL 3 PORT MAN KIT, Medline Kit Number/SKU 656608755; 31. DBD-MTS,LHK, Medline Kit Number/SKU 6010236231; 32. LHK, Medline Kit Number/SKU 6561068101; 33. LHK, Medline Kit Number/SKU 60022681B; 34. CARDIAC CATH PACK-LF, Medline Kit Number/SKU DYNJ0376288O; 35. CARDIAC CATH PACK-LF, Medline Kit Number/SKU DYNJ0376288P; 36. CARDIAC CATH PACK, Medline Kit Number/SKU DYNJ04636M; 37. CATH LAB PACK-LF, Medline Kit Number/SKU DYNJ0771922AA; 38. CATH LAB PACK-LF, Medline Kit Number/SKU DYNJ0771922Y; 39. CATH PACK, Medline Kit Number/SKU DYNJ21810L; 40. CATH PACK, Medline Kit Number/SKU DYNJ21810M; 41. CATH PACK, Medline Kit Number/SKU DYNJ21810N; 42. CARDIAC CATH PACK W/ADD ON, Medline Kit Number/SKU DYNJ22553F; 43. CATH LAB TRAY, Medline Kit Number/SKU DYNJ24250F; 44. HEART CATH, Medline Kit Number/SKU DYNJ27274J; 45. HEART CATH, Medline Kit Number/SKU DYNJ27274K; 46. HEART CATH, Medline Kit Number/SKU DYNJ27274L; 47. CARDIAC CATH PACK, Medline Kit Number/SKU DYNJ32376G; 48. CARDIAC CATH PACK, Medline Kit Number/SKU DYNJ32376I; 49. CATH/ANGIO PACK, Medline Kit Number/SKU DYNJ38087G; 50. CATH LAB PACK-LF, Medline Kit Number/SKU DYNJ38333C; 51. CATH PACK, Medline Kit Number/SKU DYNJ44473C; 52. CATH PACK, Medline Kit Number/SKU DYNJ44473D; 53. CATH LAB ANGIOGRAPHY PACK, Medline Kit Number/SKU DYNJ50779F; 54. ANGIO DRAPE CARDIAC PACK, Medline Kit Number/SKU DYNJ51615N; 55. ANGIO DRAPE CARDIAC PACK, Medline Kit Number/SKU DYNJ51615O; 56. ANGIO DRAPE CARDIAC PACK, Medline Kit Number/SKU DYNJ51615Q; 57. CARDIAC CATH IMF 86040- LF, Medline Kit Number/SKU DYNJ53947I; 58. CARDIAC CATH IMF 86040- LF, Medline Kit Number/SKU DYNJ53947J; 59. CARDIAC CATH LAB, Medline Kit Number/SKU DYNJ57748B; 60. PK CUST CATH LAB, Medline Kit Number/SKU DYNJ60584F; 61. PK CUST CATH LAB, Medline Kit Number/SKU DYNJ60584G; 62. PK CUST CATH LAB, Medline Kit Number/SKU DYNJ60584I; 63. HEART ABLATION DRAPE PACK, Medline Kit Number/SKU DYNJ61984B; 64. ANGIOGRAPHY PACK, Medline Kit Number/SKU DYNJ64525B; 65. HMSL ANGIOGRAPHY PACK, Medline Kit Number/SKU DYNJ64525C; 66. CATH LAB PACK, Medline Kit Number/SKU DYNJ64548D; 67. CATH LAB PACK, Medline Kit Number/SKU DYNJ64548F; 68. CATH LAB & E. P. PACK, Medline Kit Number/SKU DYNJ646

Why

Medline has identified the presence of particulate within the fluid path of the Manifolds.

How serious is it

Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.

Where it was sold

US (nationwide) and AE, AU, CA, JP, KR, NL, SG, SK. (states named: California)

How to tell if yours is affected

Medline Kit Number/SKU 60010195: UDI/DI each 10193489040234, UDI/DI case 40193489040235, Lot Number: 24CBF820; Medline Kit Number/SKU 60010195: UDI/DI each 10193489040234, UDI/DI case 40193489040235, Lot Number: 24DBR149; Medline Kit Number/SKU 60010195: UDI/DI each 10193489040234, UDI/DI case 40193489040235, Lot Number: 24HBO437; Medline Kit Number/SKU 60010195: UDI/DI each 10193489040234, UDI/DI case 40193489040235, Lot Number: 24IBJ827; Medline Kit Number/SKU 60010195: UDI/DI each 10193489040234, UDI/DI case 40193489040235, Lot Number: 25ABI082; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25BBI613; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25BBR862; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25DBC934; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25DBT579; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25FBF145; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25GBJ123; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25HBQ032; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25IBF219; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25JBS750; Medline Kit Number/SKU 60033123: UDI/DI each 10193489042436, UDI/DI case 40193489042437, Lot Number: 25LBD357; Medline Kit Number/SKU 60050183: UDI/DI each 10193489043273, UDI/DI case 40193489043274, Lot Number: 25ABW077; Medline Kit Number/SKU 60050183: UDI/DI each 10193489043273, UDI/DI case 40193489043274, Lot Number: 25CBO600; Medline Kit Number/SKU 60050183: UDI/DI each 10193489043273, UDI/DI case 40193489043274, Lot Number: 25DBR987; Medline Kit Number/SKU 60050183: UDI/DI each 10193489043273, UDI/DI case 40193489043274, Lot Number: 25FBH978; Medline Kit Number/SKU 60050183: UDI/DI each 10193489043273, UDI/DI case 40193489043274, Lot Number: 25IBF270; Medline Kit Number/SKU 60050183: UDI/DI each 10193489043273, UDI/DI case 40193489043274, Lot Number: 25KBJ395; Medline Kit Number/SKU 60080349: UDI/DI each 10193489044539, UDI/DI case 40193489044530, Lot Number: 24LBN469; Medline Kit Number/SKU 60080349: UDI/DI each 10193489044539, UDI/DI case 40193489044530, Lot Number: 25CBA408; Medline Kit Number/SKU 60080349: UDI/DI each 10193489044539, UDI/DI case 40193489044530, Lot Number: 25CBK717; Medline Kit Number/SKU 60080349: UDI/DI each 10193489044539, UDI/DI case 40193489044530, Lot Number: 25EBF843; Medline Kit Number/SKU 60080349: UDI/DI each 10193489044539, UDI/DI case 40193489044530, Lot Number: 25IBF307; Medline Kit Number/SKU 60080349: UDI/DI each 10193489044539, UDI/DI case 40193489044530, Lot Number: 25KBA752; Medline Kit Number/SKU 60080349: UDI/DI each 10193489044539, UDI/DI case 40193489044530, Lot Number: 26ABT007; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 24LBT840; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 25ABU494; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 25BBR313; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 25DBR517; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 25EBB825; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 25EBS427; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 25FBT465; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 25IBJ517; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 25LBC673; Medline Kit Number/SKU 60100338: UDI/DI each 10193489045062, UDI/DI case 40193489045063, Lot Number: 26BBF073; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 23HBX390; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 23IBD647; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 23LBC207; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 24ABK892; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 24DBA937; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 24DBJ585; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 24EBE043; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 24IBN995; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 24IBO100; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 24LBO436; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 25CBH999; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 25CBT160; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 25DBR989; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 25FBD599; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 25HBN775; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 25JBM965; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 25LBH604; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 25LBS225; Medline Kit Number/SKU 60101017: UDI/DI each 10193489045383, UDI/DI case 40193489045384, Lot Number: 26BBJ154; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 23LBD402; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 23LBS016; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 24ABD855; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 24BBG034; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 24BBM144; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 24CBO038; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 24CBO214; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 24DBK642; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 24KBN940; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 25BBS815; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 25CBG643; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 25CBU110; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 25EBC252; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 25FBF616; Medline Kit Number/SKU 60130506: UDI/DI each 10193489046458, UDI/DI case 40193489046459, Lot Number: 25FBK959; Medline Kit Number/SKU 60131324: UDI/DI each 10193489046809, UDI/DI case 40193489046800, Lot Number: 24KBN354; Medline Kit Number/SKU 60131324: UDI/DI each 10193489046809, UDI/DI case 40193489046800, Lot Number: 24KBN941; Medline Kit Number/SKU 60131324: UDI/DI each 10193489046809, UDI/DI case 40193489046800, Lot Number: 24LBT120; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 24BBR657; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 24CBA189; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 24FBM887; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25ABG502; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25ABV176; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25BBR314; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25CBI001; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25CBL913; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25CBO416; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25DBV071; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25EBC051; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25EBE549; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25IBT687; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25IBU966; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 25KBB948; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 26ABM807; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 26BBL416; Medline Kit Number/SKU 60132238: UDI/DI each 10193489047233, UDI/DI case 40193489047234, Lot Number: 26CBH011; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 25ABF936; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 25ABR903; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 25BBI319; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 25BBQ787; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 25DBO069; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 25EBP376; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 25HBC010; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 25JBA192; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 25LBA490; Medline Kit Number/SKU 60138301: UDI/DI each 10193489047639, UDI/DI case 40193489047630, Lot Number: 26BBJ144; Medline Kit Number/SKU 60140589: UDI/DI each 10193489048032, UDI/DI case 40193489048033, Lot Number: 24LBQ695; Medline Kit Number/SKU 60140589: UDI/DI each 10193489048032, UDI/DI case 40193489048033, Lot Number: 25ABQ318; Medline Kit Number/SKU 60140589: UDI/DI each 10193489048032, UDI/DI case 40193489048033, Lot Number: 25EBB629; Medline Kit Number/SKU 60140589: UDI/DI each 10193489048032, UDI/DI case 40193489048033, Lot Number: 25FBE410; Medline Kit Number/SKU 60140589: UDI/DI each 10193489048032, UDI/DI case 40193489048033, Lot Number: 25FBO675; Medline Kit Number/SKU 60140589: UDI/DI each 10193489048032, UDI/DI case 40193489048033, Lot Number: 25HBL092; Medline Kit Number/SKU 60140589: UDI/DI each 10193489048032, UDI/DI case 40193489048033, Lot Number: 25IBM684; Medline Kit Number/SKU 60140589: UDI/DI each 10193489048032, UDI/DI case 40193489048033, Lot Number: 25KBJ830; Medline Kit Number/SKU 60140589: UDI/DI each 10193489048032, UDI/DI case 40193489048033, Lot Number: 26ABN648; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 24LBL633; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 24LBT384; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 25CBQ237; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 25EBH401; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 25GBX825; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 25HBL094; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 26ABC178; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 26ABF962; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 26BBN954; Medline Kit Number/SKU 60192012: UDI/DI each 10193489050929, UDI/DI case 40193489050920, Lot Number: 26CBB706; Medline Kit Number/SKU 60220488: UDI/DI each 10193489052527, UDI/DI case 40193489052528, Lot Number: 23LBC519; Medline Kit Number/SKU 60220488: UDI/DI each 10193489052527, UDI/DI case 40193489052528, Lot Number: 23LBC520; Medline Kit Number/SKU 60220488: UDI/DI each 10193489052527, UDI/DI case 40193489052528, Lot Number: 23LBD120; Medline Kit Number/SKU 60220488: UDI/DI each 10193489052527, UDI/DI case 40193489052528, Lot Number: 24BBL703; Medline Kit Number/SKU 60220488: UDI/DI each 10193489052527, UDI/DI case 40193489052528, Lot Number: 24CBM350; Medline Kit Number/SKU 60220488: UDI/DI each 10193489052527, UDI/DI case 40193489052528, Lot Number: 24CBM351; Medline Kit Number/SKU 60220488: UDI/DI each 10193489052527, UDI/DI case 40193489052528, Lot Number: 24HBS750; Medline Kit Number/SKU 60220488: UDI/DI each 10193489052527, UDI/DI case 40193489052528, Lot Number: 24JBC085; Medline Kit Number/SKU 60232091: UDI/DI each 10193489053258, UDI/DI case 40193489053259, Lot Number: 25FBE411; Medline Kit Number/SKU 60232091: UDI/DI each 10193489053258, UDI/DI case 40193489053259, Lot Number: 25GBS552; Medline Kit Number/SKU 60232091: UDI/DI each 10193489053258, UDI/DI case 40193489053259, Lot Number: 25IBN368; Medline Kit Number/SKU 60232091: UDI/DI each 10193489053258, UDI/DI case 40193489053259, Lot Number: 25JBI441; Medline Kit Number/SKU 60232091: UDI/DI each 10193489053258, UDI/DI case 40193489053259, Lot Number: 25JBX630; Medline Kit Number/SKU 60232091: UDI/DI each 10193489053258, UDI/DI case 40193489053259, Lot Number: 25LBC407; Medline Kit Number/SKU 60232091: UDI/DI each 10193489053258, UDI/DI case 40193489053259, Lot Number: 26BBB138; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25CBD801; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25CBK055; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25CBP471; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25DBR961; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25EBD424; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25FBE633; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25FBJ631; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25GBC837; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25GBN114; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25GBW454; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25HBD429; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25IBF794; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25JBL011; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25KBO033; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25KBT052; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25LBC283; Medline Kit Number/SKU 65010037: UDI/DI each 10193489059502, UDI/DI case 40193489059503, Lot Number: 25LBF571; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 25CBL361; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 25DBQ337; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 25EBN624; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 25FBE412; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 25FBN083; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 25GBD061; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 25IBG953; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 25KBO724; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 25KBS323; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 26ABC753; Medline Kit Number/SKU 65012819: UDI/DI each 10193489059687, UDI/DI case 40193489059688, Lot Number: 26ABR022; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 24KBT591; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 24LBE978; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25DBQ343; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25DBT925; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25EBJ331; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25EBN550; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25EBU247; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25FBE413; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25FBF761; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25HBA587; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25JBL132; Medline Kit Number/SKU 65040421: UDI/DI each 10193489060416, UDI/DI case 40193489060417, Lot Number: 25JBR927; Medline Kit Number/SKU 65183482: UDI/DI each 10193489062823, UDI/DI case 40193489062824, Lot Number: 23LBV434; Medline Kit Number/SKU 65183482: UDI/DI each 10193489062823, UDI/DI case 40193489062824, Lot Number: 24CBB755; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 23KBJ803; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 23LBC925; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 23LBP705; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24BBK267; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24DBU989; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24GBT179; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24IBL496; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24JBA690; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24KBI563; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24KBN943; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24KBQ269; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24LBJ407; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 24LBO500; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25ABB054; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25BBI520; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25CBL739; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25CBT917; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25DBA597; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25FBQ181; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25FBS611; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25HBL493; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25IBI773; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25IBT228; Medline Kit Number/SKU 65660872: UDI/DI each 10193489989557, UDI/DI case 40193489989558, Lot Number: 25LBI277; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 23KBW790; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24BBG011; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24BBV062; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24EBH524; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24FBD353; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24GBT409; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24GBT418; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24JBX239; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24KBL357; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24KBL665; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24KBR053; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 24LBB805; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 25CBN927; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 25EBA296; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 25FBN589; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 25IBF118; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 25IBV472; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 26ABA296; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 26ABM374; Medline Kit Number/SKU 601006269: UDI/DI each 10193489189582, UDI/DI case 40193489189583, Lot Number: 26BBO252; Medline Kit Number/SKU 601203616: UDI/DI each 10193489070774, UDI/DI case 40193489070775, Lot Number: 23LBN507; Medline Kit Number/SKU 601203616: UDI/DI each 10193489070774, UDI/DI case 40193489070775, Lot Number: 24BBD883; Medline Kit Number/SKU 601203616: UDI/DI each 10193489070774, UDI/DI case 40193489070775, Lot Number: 24CBP951; Medline Kit Number/SKU 601203616: UDI/DI each 10193489070774, UDI/DI case 40193489070775, Lot Number: 24FBS128; Medline Kit Number/SKU 601203616: UDI/DI each 10193489070774, UDI/DI case 40193489070775, Lot Number: 24HBL889; Medline Kit Number/SKU 601203616: UDI/DI each 10193489070774, UDI/DI case 40193489070775, Lot Number: 24HBW911; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 24LBO506; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25ABL641; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25BBM277; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25CBL017; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25CBN644; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25CBS021; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25EBG284; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25FBI559; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25GBC062; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25GBN092; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25HBG211; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25IBA166; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25IBR553; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25JBA722; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25JBJ009; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25JBL024; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25JBX186; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25KBH488; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25LBM779; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 25LBM780; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26ABG265; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26ABR490; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26ABR491; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26ABR492; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26ABU629; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26BBF384; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26BBI348; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26BBL421; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26BBT130; Medline Kit Number/SKU 650104611: UDI/DI each 10193489074802, UDI/DI case 40193489074803, Lot Number: 26CBA497; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25BBD354; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25BBI530; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25CBD811; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25CBL374; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25DBQ359; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25EBJ265; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25EBU808; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25EBV383; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25GBG209; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25GBH353; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25GBN115; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25GBV470; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25GBW455; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25GBX436; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25JBU837; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25KBR043; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 25LBT349; Medline Kit Number/SKU 651304116: UDI/DI each 10193489075892, UDI/DI case 40193489075893, Lot Number: 26ABE723; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 23JBN268; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 23KBJ916; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 23LBC919; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 23LBI119; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 24ABI603; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 24BBI553; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 24DBK464; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 24GBT178; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 24HBI912; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 40193489991377, Lot Number: 24IBA808; Medline Kit Number/SKU 656206837: UDI/DI each 10193489991376, UDI/DI case 4019348999 1377, Lot Number: 24IBD546; Medline Kit Number/SKU 656608724: UDI/DI each 10193489991291, UDI/DI case 40193489991292, Lot Number: 23GBA808; Medline Kit Number/SKU 656608724: UDI/DI each 10193489991291, UDI/DI case 40193489991292, Lot Number: 23GBG163; Medline Kit Number/SKU 656608724: UDI/DI each 10193489991291, UDI/DI case 40193489991292, Lot Number: 23GBI378; Medline Kit Number/SKU 656608724: UDI/DI each 10193489991291, UDI/DI case 40193489991292, Lot Number: 23GBL315; Medline Kit Number/SKU 656608724: UDI/DI each 10193489991291, UDI/DI case 40193489991292, Lot Number: 23GBV232; Medline Kit Number/SKU 656608724: UDI/DI each 10193489991291, UDI/DI case 40193489991292, Lot Number: 23HBA773; Medline Kit Number/SKU 656608724: UDI/DI each 10193489991291, UDI/DI case 40193489991292, Lot Number: 23HBF670; Medline Kit Number/SKU 656608737: UDI/DI each 10193489991383, UDI/DI case 40193489991384, Lot Number: 23FBU942; Medline Kit Number/SKU 656608737: UDI/DI each 10193489991383, UDI/DI case 40193489991384, Lot Number: 23GBI382; Medline Kit Number/SKU 656608737: UDI/DI each 10193489991383, UDI/DI case 40193489991384, Lot Number: 23HBK089; Medline Kit Number/SKU 656608739: UDI/DI each 10193489991390, UDI/DI case 40193489991391, Lot Number: 24JBA691; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24ABI614; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24BBK268; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24CBI159; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24CBM852; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24CBT018; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24EBH534; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24EBO034; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24FBD257; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24GBT180; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24IBA807; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24JBA692; Medline Kit Number/SKU 656608749: UDI/DI each 10193489991123, UDI/DI case 40193489991124, Lot Number: 24JBN963; Medline Kit Number/SKU 656608753: UDI/DI each 10193489991482, UDI/DI case 40193489991483, Lot Number: 24CBI161; Medline Kit Number/SKU 656608753: UDI/DI each 10193489991482, UDI/DI case 40193489991483, Lot Number: 24FBA789; Medline Kit Number/SKU 656608753: UDI/DI each 10193489991482, UDI/DI case 40193489991483, Lot Number: 24FBB313; Medline Kit Number/SKU 656608753: UDI/DI each 10193489991482, UDI/DI case 40193489991483, Lot Number: 24HBE699; Medline Kit Number/SKU 656608753: UDI/DI each 10193489991482, UDI/DI case 40193489991483, Lot Number: 25ABB328; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 23KBK009; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 23LBP715; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24ABI609; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24BBI554; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24BBM932; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24CBT017; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24EBO035; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24FBD619; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24GBT181; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24HBI916; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24HBJ811; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24IBA812; Medline Kit Number/SKU 656608755: UDI/DI each 10193489991147, UDI/DI case 40193489991148, Lot Number: 24IBG270; Medline Kit Number/SKU 6010236231: UDI/DI each 10193489099423, UDI/DI case 40193489099424, Lot Number: 24CBR608; Medline Kit Number/SKU 6010236231: UDI/DI each 10193489099423, UDI/DI case 40193489099424, Lot Number: 24GBO598; Medline Kit Number/SKU 6010236231: UDI/DI each 10193489099423, UDI/DI case 40193489099424, Lot Number: 24JBR853; Medline Kit Number/SKU 6010236231: UDI/DI each 10193489099423, UDI/DI case 40193489099424, Lot Number: 25ABR099; Medline Kit Number/SKU 6010236231: UDI/DI each 10193489099423, UDI/DI case 40193489099424, Lot Number: 25DBQ922; Medline Kit Number/SKU 6010236231: UDI/DI each 10193489099423, UDI/DI case 40193489099424, Lot Number: 25EBB576; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 23GBU202; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 23HBM328; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 23JBH162; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 23KBE496; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 24ABI618; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 24ABM576; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 24CBU036; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 24EBT264; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 24FBC202; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 24GBG798; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 24HBJ828; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 25CBD800; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 25EBN593; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 25FBE418; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 25FBG927; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 25GBG204; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 25IBC554; Medline Kit Number/SKU 6561068101: UDI/DI each 10193489990935, UDI/DI case 40193489990936, Lot Number: 25JBU766; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 23LBA250; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24ABA804; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24BBG031; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24FBN874; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24FBQ965; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24HBJ130; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24HBK019; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24HBW999; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24KBL890; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24KBR052; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 24LBB981; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 25ABS656; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 25DBM315; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 25FBJ157; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 25HBA355; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 25HBL267; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 25IBV424; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 25LBD366; Medline Kit Number/SKU 60022681B: UDI/DI each 10197344009299, UDI/DI case 40197344009290, Lot Number: 26BBF629; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 24CMF949; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 24CMI112; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 24EMF179; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 24GMG910; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 24HMJ763; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 24JMB965; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 24KMB099; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25AMG187; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25BME027; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25BMF926; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25CMH544; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25EMF693; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25FMB279; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25GMH129; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25HMA941; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25IMJ033; Medline Kit Number/SKU DYNJ0376288O: UDI/DI each 10195327601454, UDI/DI case 40195327601455, Lot Number: 25KMH462; Medline Kit Number/SKU DYNJ0376288P: UDI/DI each 10198459619373, UDI/DI case 40198459619374, Lot Number: 26BMC186; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24BMB876; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24BMD750; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24CMF244; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24DMG629; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24EMI143; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24FMF489; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24GMI275; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24JMA465; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24JMJ459; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24KMF145; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 24LMF579; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 25BMA856; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 25BMI967; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 25EMD048; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 25FMC559; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 25FMH403; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 25GMK011; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 25IME291; Medline Kit Number/SKU DYNJ04636M: UDI/DI each 10195327558048, UDI/DI case 40195327558049, Lot Number: 26AMK555; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 24GMI024; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 24HMD068; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 24JMB505; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 24KMA550; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 24KMF042; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 24LMG721; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 25BMA048; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 25BMG528; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 25DMB976; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 25EMA202; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 25FME481; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 25HMC806; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 25HMH013; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 25KMB617; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 25KMD947; Medline Kit Number/SKU DYNJ0771922AA: UDI/DI each 10198459051029, UDI/DI case 40198459051020, Lot Number: 26AMC941; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 23GBQ140; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 23HBA939; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 23HBQ366; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 23HBW031; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 23IBH451; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 23KBP468; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 23LBH140; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 24ABK612; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 24CBL891; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 24FMA381; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 24FMH344; Medline Kit Number/SKU DYNJ0771922Y: UDI/DI each 10195327069322, UDI/DI case 40195327069323, Lot Number: 24GME044; Medline Kit Number/SKU DYNJ21810L: UDI/DI each 10888277971400, UDI/DI case 40888277971401, Lot Number: 23GBK136; Medline Kit Number/SKU DYNJ21810L: UDI/DI each 10888277971400, UDI/DI case 40888277971401, Lot Number: 23GBU247; Medline Kit Number/SKU DYNJ21810L: UDI/DI each 10888277971400, UDI/DI case 40888277971401, Lot Number: 23IBJ801; Medline Kit Number/SKU DYNJ21810L: UDI/DI each 10888277971400, UDI/DI case 40888277971401, Lot Number: 23JBV890; Medline Kit Number/SKU DYNJ21810M: UDI/DI each 10195327517397, UDI/DI case 40195327517398, Lot Number: 23LBG347; Medline Kit Number/SKU DYNJ21810M: UDI/DI each 10195327517397, UDI/DI case 40195327517398, Lot Number: 23LBI873; Medline Kit Number/SKU DYNJ21810M: UDI/DI each 10195327517397, UDI/DI case 40195327517398, Lot Number: 24BBD622; Medline Kit Number/SKU DYNJ21810M: UDI/DI each 10195327517397, UDI/DI case 40195327517398, Lot Number: 24CBS441; Medline Kit Number/SKU DYNJ21810M: UDI/DI each 10195327517397, UDI/DI case 40195327517398, Lot Number: 24EBJ595; Medline Kit Number/SKU DYNJ21810M: UDI/DI each 10195327517397, UDI/DI case 40195327517398, Lot Number: 24GBQ250; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 24IBD183; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 24LBB632; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 25ABQ969; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 25DBM587; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 25EBG601; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 25EBP472; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 25HBR963; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 25HBU257; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 25KBP287; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 25KBP985; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 26BBE689; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 26BBJ088; Medline Kit Number/SKU DYNJ21810N: UDI/DI each 10198459087462, UDI/DI case 40198459087463, Lot Number: 26CBI735; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 23IBB004; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 23JBU876; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 23KBA334; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 23KBQ360; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 24ABO900; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 24ABR318; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 24CBB023; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 24DBT793; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 24EBQ985; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 24FBN183; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 24HBA008; Medline Kit Number/SKU DYNJ22553F: UDI/DI each 10195327329334, UDI/DI case 40195327329335, Lot Number: 24IBS346; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 23HBR680; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 23IBO441; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 23JBO248; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 23KBF054; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24ABW269; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24CBH708; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24DBJ179; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24DBK002; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24FBM355; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24HBI925; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24IBB180; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24IBB401; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24IBP420; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 24KBS901; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 25ABB309; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 25BBJ742; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 25CBJ890; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 25DBR206; Medline Kit Number/SKU DYNJ24250F: UDI/DI each 10193489857566, UDI/DI case 40193489857567, Lot Number: 25EBB121; Medline Kit Number/SKU DYNJ27274J: UDI/DI each 10195327360078, UDI/DI case 40195327360079, Lot Number: 23HMD020; Medline Kit Number/SKU DYNJ27274J: UDI/DI each 10195327360078, UDI/DI case 40195327360079, Lot Number: 23IMH614; Medline Kit Number/SKU DYNJ27274J: UDI/DI each 10195327360078, UDI/DI case 40195327360079, Lot Number: 23JMB187; Medline Kit Number/SKU DYNJ27274J: UDI/DI each 10195327360078, UDI/DI case 40195327360079, Lot Number: 23LMD025; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 23LMC918; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24AMD273; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24AMG998; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24BMH689; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24DMB975; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24EMF238; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24FMH662; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24FMI006; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24HMG371; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24IMJ030; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24JMH567; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 24LMB005; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 25BMC222; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 25DMC833; Medline Kit Number/SKU DYNJ27274K: UDI/DI each 10195327527709, UDI/DI case 40195327527700, Lot Number: 25GMA065; Medline Kit Number/SKU DYNJ27274L: UDI/DI each 10198459445828, UDI/DI case 40198459445829, Lot Number: 25IMA053; Medline Kit Number/SKU DYNJ27274L: UDI/DI each 10198459445828, UDI/DI case 40198459445829, Lot Number: 25JMH189; Medline Kit Number/SKU DYNJ27274L: UDI/DI each 10198459445828, UDI/DI case 40198459445829, Lot Number: 25LMB027; Medline Kit Number/SKU DYNJ32376G: UDI/DI each 10195327574420, UDI/DI case 40195327574421, Lot Number: 24AMJ267; Medline Kit Number/SKU DYNJ32376G: UDI/DI each 10195327574420, UDI/DI case 40195327574421, Lot Number: 24DMG093; Medline Kit Number/SKU DYNJ32376G: UDI/DI each 10195327574420, UDI/DI case 40195327574421, Lot Number: 24GMA682; Medline Kit Number/SKU DYNJ32376G: UDI/DI each 10195327574420, UDI/DI case 40195327574421, Lot Number: 24HMC264; Medline Kit Number/SKU DYNJ32376G: UDI/DI each 10195327574420, UDI/DI case 40195327574421, Lot Number: 24HMF960; Medline Kit Number/SKU DYNJ32376G: UDI/DI each 10195327574420, UDI/DI case 40195327574421, Lot Number: 24KMB733; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 24LMG585; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 25BMH316; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 25CMA831; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 25DMF124; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 25EMI648; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 25FMI828; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 25HMB559; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 25IMD227; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 25KME051; Medline Kit Number/SKU DYNJ32376I: UDI/DI each 10198459195624, UDI/DI case 40198459195625, Lot Number: 26AMB979; Medline Kit Number/SKU DYNJ38087G: UDI/DI each 10198459193989, UDI/DI case 40198459193980, Lot Number: 25KMI514; Medline Kit Number/SKU DYNJ38087G: UDI/DI each 10198459193989, UDI/DI case 40198459193980, Lot Number: 26BMH799; Medline Kit Number/SKU DYNJ38333C: UDI/DI each 10193489487794, UDI/DI case 40193489487795, Lot Number: 23GMD240; Medline Kit Number/SKU DYNJ38333C: UDI/DI each 10193489487794, UDI/DI case 40193489487795, Lot Number: 23HME792; Medline Kit Number/SKU DYNJ38333C: UDI/DI each 10193489487794, UDI/DI case 40193489487795, Lot Number: 23HMH289; Medline Kit Number/SKU DYNJ38333C: UDI/DI each 10193489487794, UDI/DI case 40193489487795, Lot Number: 23HMH584; Medline Kit Number/SKU DYNJ38333C: UDI/DI each 10193489487794, UDI/DI case 40193489487795, Lot Number: 23IMF926; Medline Kit Number/SKU DYNJ38333C: UDI/DI each 10193489487794, UDI/DI case 40193489487795, Lot Number: 23JMB248; Medline Kit Number/SKU DYNJ38333C: UDI/DI each 10193489487794, UDI/DI case 40193489487795, Lot Number: 23LMH195; Medline Kit Number/SKU DYNJ38333C: UDI/DI each 10193489487794, UDI/DI case 40193489487795, Lot Number: 24AMG536; Medline Kit Number/SKU DYNJ38333C: UDI/DI each 10193489487794, UDI/DI case 40193489487795, Lot Number: 24GMF740; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 23GDB073; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 23IDA974; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 23IDB421; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 23LDA810; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 23LDB342; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 24ADB345; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 24BDA511; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 24CDB995; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 24EDB639; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 24GDB406; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 24IDA224; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 24IDA486; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 24IDA913; Medline Kit Number/SKU DYNJ44473C: UDI/DI each 10193489905250, UDI/DI case 40193489905251, Lot Number: 24IMH683; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 24KMF525; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 25BMF533; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 25DMK041; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 25FMD153; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 25HMA046; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 25HMD550; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 25HMF357; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 25JME621; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 26AMC383; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 26AMH118; Medline Kit Number/SKU DYNJ44473D: UDI/DI each 10198459182013, UDI/DI case 40198459182014, Lot Number: 26BMJ704; Medline Kit Number/SKU DYNJ50779F: UDI/DI each 10193489436068, UDI/DI case 40193489436069, Lot Number: 23FMJ773; Medline Kit Number/SKU DYNJ50779F: UDI/DI each 10193489436068, UDI/DI case 40193489436069, Lot Number: 23JMB530; Medline Kit Number/SKU DYNJ50779F: UDI/DI each 10193489436068, UDI/DI case 40193489436069, Lot Number: 23LMH087; Medline Kit Number/SKU DYNJ50779F: UDI/DI each 10193489436068, UDI/DI case 40193489436069, Lot Number: 24BMF696; Medline Kit Number/SKU DYNJ50779F: UDI/DI each 10193489436068, UDI/DI case 40193489436069, Lot Number: 24CMJ355; Medline Kit Number/SKU DYNJ50779F: UDI/DI each 10193489436068, UDI/DI case 40193489436069, L ot Number: 24DMA986; Medline Kit Number/SKU DYNJ51615N: UDI/DI each 10195327418946, UDI/DI case 40195327418947, Lot Number: 23GMB033; Medline Kit Number/SKU DYNJ51615N: UDI/DI each 10195327418946, UDI/DI case 40195327418947, Lot Number: 23HMA279; Medline Kit Number/SKU DYNJ51615N: UDI/DI each 10195327418946, UDI/DI case 40195327418947, Lot Number: 23IMH204; Medline Kit Number/SKU DYNJ51615N: UDI/DI each 10195327418946, UDI/DI case 40195327418947, Lot Number: 23KMA476; Medline Kit Number/SKU DYNJ51615N: UDI/DI each 10195327418946, UDI/DI case 40195327418947, Lot Number: 24CMB390; Medline Kit Number/SKU DYNJ51615O: UDI/DI each 10195327693756, UDI/DI case 40195327693757, Lot Number: 24DMG646; Medline Kit Number/SKU DYNJ51615O: UDI/DI each 10195327693756, UDI/DI case 40195327693757, Lot Number: 24EMD666; Medline Kit Number/SKU DYNJ51615O: UDI/DI each 10195327693756, UDI/DI case 40195327693757, Lot Number: 24HMF566; Medline Kit Number/SKU DYNJ51615O: UDI/DI each 10195327693756, UDI/DI case 40195327693757, Lot Number: 24IME313; Medline Kit Number/SKU DYNJ51615O: UDI/DI each 10195327693756, UDI/DI case 40195327693757, Lot Number: 24JMG871; Medline Kit Number/SKU DYNJ51615Q: UDI/DI each 10198459298127, UDI/DI case 40198459298128, Lot Number: 25CMH554; Medline Kit Number/SKU DYNJ51615Q: UDI/DI each 10198459298127, UDI/DI case 40198459298128, Lot Number: 25DMB241; Medline Kit Number/SKU DYNJ51615Q: UDI/DI each 10198459298127, UDI/DI case 40198459298128, Lot Number: 25DMD724; Medline Kit Number/SKU DYNJ51615Q: UDI/DI each 10198459298127, UDI/DI case 40198459298128, Lot Number: 25EMA585; Medline Kit Number/SKU DYNJ51615Q: UDI/DI each 10198459298127, UDI/DI case 40198459298128, Lot Number: 25FMB571; Medline Kit Number/SKU DYNJ53947I: UDI/DI each 10195327507350, UDI/DI case 40195327507351, Lot Number: 23KLA450; Medline Kit Number/SKU DYNJ53947I: UDI/DI each 10195327507350, UDI/DI case 40195327507351, Lot Number: 23KLB150; Medline Kit Number/SKU DYNJ53947I: UDI/DI each 10195327507350, UDI/DI case 40195327507351, Lot Number: 24ALA971; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 24ELA396; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 24ELA650; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 24GLA955; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 24HLB182; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25ALA448; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25ALA801; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25BLA350; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25BLA406; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25DLB012; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25ELA058; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25FLA819; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25GLA626; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25HLA565; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 25KLA585; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 26BLA436; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 26BLA524; Medline Kit Number/SKU DYNJ53947J: UDI/DI each 10198459010545, UDI/DI case 40198459010546, Lot Number: 26CLA293; Medline Kit Number/SKU DYNJ57748B: UDI/DI each 10198459218323, UDI/DI case 40198459218324, Lot Number: 26BMD521; Medline Kit Number/SKU DYNJ57748B: UDI/DI each 10198459218323, UDI/DI case 40198459218324, Lot Number: 26BMF704; Medline Kit Number/SKU DYNJ60584F: UDI/DI each 10195327432867, UDI/DI case 40195327432868, Lot Number: 23HBU057; Medline Kit Number/SKU DYNJ60584F: UDI/DI each 10195327432867, UDI/DI case 40195327432868, Lot Number: 23IBM917; Medline Kit Number/SKU DYNJ60584F: UDI/DI each 10195327432867, UDI/DI case 40195327432868, Lot Number: 23JBV286; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 23KBF153; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 23KBM011; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 23LBE274; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24BBE523; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24BBV486; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24DBK225; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24EBA252; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24FBD782; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24GBF266; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24GBO263; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24HBM894; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24IBE073; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24JBJ390; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24JBP090; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24KBM289; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 24LBJ022; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 25ABT699; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 25CBC232; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 25DBG628; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 25EBP227; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 25EBT448; Medline Kit Number/SKU DYNJ60584G: UDI/DI each 10195327547707, UDI/DI case 40195327547708, Lot Number: 25GBI773; Medline Kit Number/SKU DYNJ60584I: UDI/DI each 10198459497391, UDI/DI case 40198459497392, Lot Number: 25IBK188; Medline Kit Number/SKU DYNJ60584I: UDI/DI each 10198459497391, UDI/DI case 40198459497392, Lot Number: 25IBR007; Medline Kit Number/SKU DYNJ60584I: UDI/DI each 10198459497391, UDI/DI case 40198459497392, Lot Number: 25KBE810; Medline Kit Number/SKU DYNJ60584I: UDI/DI each 10198459497391, UDI/DI case 40198459497392, Lot Number: 26ABD930; Medline Kit Number/SKU DYNJ60584I: UDI/DI each 10198459497391, UDI/DI case 40198459497392, Lot Number: 26BBQ989; Medline Kit Number/SKU DYNJ61984B: UDI/DI each 10193489962161, UDI/DI case 40193489962162, Lot Number: 23GBK917; Medline Kit Number/SKU DYNJ61984B: UDI/DI each 10193489962161, UDI/DI case 40193489962162, Lot Number: 23HBK786; Medline Kit Number/SKU DYNJ61984B: UDI/DI each 10193489962161, UDI/DI case 40193489962162, Lot Number: 23IBN871; Medline Kit Number/SKU DYNJ61984B: UDI/DI each 10193489962161, UDI/DI case 40193489962162, Lot Number: 23KBS842; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 23GBK785; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 23GBT353; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 23HBK725; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 23IBE961; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 23JBG437; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 23LBA330; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 24BBN031; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 24CBP403; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 24EBA810; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 24FBE993; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 24GBP015; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 24IBH886; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 24IBH954; Medline Kit Number/SKU DYNJ64525B: UDI/DI each 10195327009267, UDI/DI case 40195327009268, Lot Number: 24LBA784; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 25BBH144; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 25BBI209; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 25EBK891; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 25HBE964; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 25IBG813; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 25IBL495; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 25KBN046; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 25LBG439; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 26BBD434; Medline Kit Number/SKU DYNJ64525C: UDI/DI each 10198459253393, UDI/DI case 40198459253394, Lot Number: 26BBJ117; Medline Kit Number/SKU DYNJ64548D: UDI/DI each 10195327427795, UDI/DI case 40195327427796, Lot Number: 23HBC766; Medline Kit Number/SKU DYNJ64548D: UDI/DI each 10195327427795, UDI/DI case 40195327427796, Lot Number: 23HBI072; Medline Kit Number/SKU DYNJ64548D: UDI/DI each 10195327427795, UDI/DI case 40195327427796, Lot Number: 24BBE317; Medline Kit Number/SKU DYNJ64548D: UDI/DI each 10195327427795, UDI/DI case 40195327427796, Lot Number: 24CBN029; Medline Kit Number/SKU DYNJ64548D: UDI/DI each 10195327427795, UDI/DI case 40195327427796, Lot Number: 24DBH607; Medline Kit Number/SKU DYNJ64548D: UDI/DI each 10195327427795, UDI/DI case 40195327427796, Lot Number: 24EBC502; Medline Kit Number/SKU DYNJ64548D: UDI/DI each 10195327427795, UDI/DI case 40195327427796, Lot Number: 24GBD761; Medline Kit Number/SKU DYNJ64548D: UDI/DI each 10195327427795, UDI/DI case 40195327427796, Lot Number: 24GBW862; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 24IBM407; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 24JBJ978; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 24LBA232; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 24LBQ496; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 25BBP659; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 25EBE998; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 25GBI681; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 25GBI687; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 25GBI791; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 25KBS593; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 25LBT133; Medline Kit Number/SKU DYNJ64548F: UDI/DI each 10198459116438, UDI/DI case 40198459116439, Lot Number: 26ABA698; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 23FBU846; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 23GBO963; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 23IBR519; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 23JBL760; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 23LBG554; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24CBB871; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24DBQ844; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24EBH330; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24FBO069; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24GBS987; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24HBV812; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24JBH168; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24JBI059; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24JBV114; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 24LBS172; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 25BBI896; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 25DBG678; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 25FBD216; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 25IBI788; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 25JBS272; Medline Kit Number/SKU DYNJ64626B: UDI/DI each 10195327196196, UDI/DI case 40195327196197, Lot Number: 25KBP062; Medline Kit Number/SKU DYNJ65174D: UDI/DI each 10193489962154, UDI/DI case 40193489962155, Lot Number: 23IBN826; Medline Kit Number/SKU DYNJ65174D: UDI/DI each 10193489962154, UDI/DI case 40193489962155, Lot Number: 23LBF175; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24BBF854; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24BBS884; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24CBL206; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24EBV445; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24FBN026; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24IBP110; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24KBI588; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24KBL514; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24LBA234; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24LBL857; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 24LBM317; Medline Kit Number/SKU DYNJ65174F: UDI/DI each 10195327597276, UDI/DI case 40195327597277, Lot Number: 25BBV027; Medline Kit Number/SKU DYNJ65185A: UDI/DI each 10195327648060, UDI/DI case 40195327648061, Lot Number: 24DBD207; Medline Kit Number/SKU DYNJ65185A: UDI/DI each 10195327648060, UDI/DI case 40195327648061, Lot Number: 24EBC831; Medline Kit Number/SKU DYNJ65185A: UDI/DI each 10195327648060, UDI/DI case 40195327648061, Lot Number: 24EBV254; Medline Kit Number/SKU DYNJ65185A: UDI/DI each 10195327648060, UDI/DI case 40195327648061, Lot Number: 24GBP711; Medline Kit Number/SKU DYNJ65185A: UDI/DI each 10195327648060, UDI/DI case 40195327648061, Lot Number: 24IBT429; Medline Kit Number/SKU DYNJ65185B: UDI/DI each 10198459164347, UDI/DI case 40198459164348, Lot Number: 24KBB438; Medline Kit Number/SKU DYNJ65185C: UDI/DI each 10198459200533, UDI/DI case 40198459200534, Lot Number: 25ABB181; Medline Kit Number/SKU DYNJ65185D: UDI/DI each 10198459269509, UDI/DI case 40198459269500, Lot Number: 25CBA241; Medline Kit Number/SKU DYNJ65902C: UDI/DI each 10195327458737, UDI/DI case 40195327458738, Lot Number: 23GBK658; Medline Kit Number/SKU DYNJ65902C: UDI/DI each 10195327458737, UDI/DI case 40195327458738, Lot Number: 23HBK672; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 23IBG651; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 23KBG457; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 23KBW483; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 24ABW303; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 24EBJ302; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 24FBB743; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 24GBW454; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 24HBP013; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 24IBA778; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 24JBV205; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 24KBQ761; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 24LBR216; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 25ABS609; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 25DBL595; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 25EBN113; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 25HBK280; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 25HBM865; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 25IBQ857; Medline Kit Number/SKU DYNJ65902D: UDI/DI each 10195327497712, UDI/DI case 40195327497713, Lot Number: 25JBK935; Medline Kit Number/SKU DYNJ65902F: UDI/DI each 10198459628870, UDI/DI case 40198459628871, Lot Number: 25LBJ533; Medline Kit Number/SKU DYNJ65902F: UDI/DI each 10198459628870, UDI/DI case 40198459628871, Lot Number: 26ABU700; Medline Kit Number/SKU DYNJ66343: UDI/DI each 10193489340358, UDI/DI case 40193489340359, Lot Number: 23FBU418; Medline Kit Number/SKU DYNJ66693A: UDI/DI each 10193489861136, UDI/DI case 40193489861137, Lot Number: 23HBW456; Medline Kit Number/SKU DYNJ66693A: UDI/DI each 10193489861136, UDI/DI case 40193489861137, Lot Number: 23IBF813; Medline Kit Number/SKU DYNJ66693A: UDI/DI each 10193489861136, UDI/DI case 40193489861137, Lot Number: 23KBH509; Medline Kit Number/SKU DYNJ66693B: UDI/DI each 10195327588717, UDI/DI case 40195327588718, Lot Number: 23LBR442; Medline Kit Number/SKU DYNJ66693B: UDI/DI each 10195327588717, UDI/DI case 40195327588718, Lot Number: 24DBR718; Medline Kit Number/SKU DYNJ66693B: UDI/DI each 10195327588717, UDI/DI case 40195327588718, Lot Number: 24DMK589; Medline Kit Number/SKU DYNJ66693B: UDI/DI each 10195327588717, UDI/DI case 40195327588718, Lot Number: 24FMD201; Medline Kit Number/SKU DYNJ66693B: UDI/DI each 10195327588717, UDI/DI case 40195327588718, Lot Number: 24HMG727; Medline Kit Number/SKU DYNJ66693B: UDI/DI each 10195327588717, UDI/DI case 40195327588718, Lot Number: 24JMI173; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 23JBC058; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 23LBC662; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 24ABX667; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 24DBD603; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 24EBT986; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 24HBO317; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 24IBG768; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 24IBM385; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 24LBB262; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 24LBQ683; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 25DBN219; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 25EBA527; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 25GBM917; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 25IBL617; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 25KBF881; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 25LBU332; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 26BBB990; Medline Kit Number/SKU DYNJ69674A: UDI/DI each 10195327125400, UDI/DI case 40195327125401, Lot Number: 26BBH614; Medline Kit Number/SKU DYNJ81911B: UDI/DI each 10198459524523, UDI/DI case 40198459524524, Lot Number: 25JBP347; Medline Kit Number/SKU DYNJ81911B: UDI/DI each 10198459524523, UDI/DI case 40198459524524, Lot Number: 25LBJ585; Medline Kit Number/SKU DYNJ81911B: UDI/DI each 10198459524523, UDI/DI case 40198459524524, Lot Number: 26ABM116; Medline Kit Number/SKU DYNJ82971B: UDI/DI each 10198459595806, UDI/DI case 40198459595807, Lot Number: 25LMD112; Medline Kit Number/SKU DYNJ82971B: UDI/DI each 10198459595806, UDI/DI case 40198459595807, Lot Number: 26AMC552; Medline Kit Number/SKU DYNJ83412A: UDI/DI each 10195327465377, UDI/DI case 40195327465378, Lot Number: 23GBN981; Medline Kit Number/SKU DYNJ83412A: UDI/DI each 10195327465377, UDI/DI case 40195327465378, Lot Number: 23JBT478; Medline Kit Number/SKU DYNJ83412B: UDI/DI each 10195327562748, UDI/DI case 40195327562749, Lot Number: 23KBT841; Medline Kit Number/SKU DYNJ83412C: UDI/DI each 10195327570187, UDI/DI case 40195327570188, Lot Number: 24ABX696; Medline Kit Number/SKU DYNJ83412C: UDI/DI each 10195327570187, UDI/DI case 40195327570188, Lot Number: 24CBP472; Medline Kit Number/SKU DYNJ83412C: UDI/DI each 10195327570187, UDI/DI case 40195327570188, Lot Number: 24DBS736; Medline Kit Number/SKU DYNJ83412C: UDI/DI each 10195327570187, UDI/DI case 40195327570188, Lot Number: 24FBF826; Medline Kit Number/SKU DYNJ83412C: UDI/DI each 10195327570187, UDI/DI case 40195327570188, Lot Number: 24GBC751; Medline Kit Number/SKU DYNJ83412C: UDI/DI each 10195327570187, UDI/DI case 40195327570188, Lot Number: 24HBK795; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 24JBL374; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 24KBH903; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 24LBL884; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 25ABN962; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 25DBC435; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 25DBQ206; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 25GBT085; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 25HBQ142; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 25IBI399; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 25IMB496; Medline Kit Number/SKU DYNJ83412D: UDI/DI each 10198459140297, UDI/DI case 40198459140298, Lot Number: 26AMF300; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 23IBB691; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 23JBO999; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 23LBP160; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 24ABK547; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 24CBR498; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 24EBT390; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 24GBC841; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 24IBJ312; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 24JBR302; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 24KBK326; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 25ABV473; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 25CBU488; Medline Kit Number/SKU DYNJ84341: UDI/DI each 10195327359546, UDI/DI case 40195327359547, Lot Number: 25DBA530; Medline Kit Number/SKU DYNJ86750: UDI/DI each 10195327672621, UDI/DI case 40195327672622, Lot Number: 24FBA632; Medline Kit Number/SKU DYNJ86750: UDI/DI each 10195327672621, UDI/DI case 40195327672622, Lot Number: 24IBH880; Medline Kit Number/SKU DYNJ86750: UDI/DI each 10195327672621, UDI/DI case 40195327672622, Lot Number: 24JBA311; Medline Kit Number/SKU DYNJ88038: UDI/DI each 10198459089152, UDI/DI case 40198459089153, Lot Number: 24HBW931; Medline Kit Number/SKU DYNJ88038: UDI/DI each 10198459089152, UDI/DI case 40198459089153, Lot Number: 25CBF573; Medline Kit Number/SKU DYNJ88038: UDI/DI each 10198459089152, UDI/DI case 40198459089153, Lot Number: 25FBG188; Medline Kit Number/SKU DYNJ88091: UDI/DI each 10198459098154, UDI/DI case 40198459098155, Lot Number: 24IMF597; Medline Kit Number/SKU DYNJ88091: UDI/DI each 10198459098154, UDI/DI case 40198459098155, Lot Number: 25BMD403; Medline Kit Number/SKU DYNJ88091A: UDI/DI each 10198459267512, UDI/DI case 40198459267513, Lot Number: 25CME058; Medline Kit Number/SKU DYNJ88091A: UDI/DI each 10198459267512, UDI/DI case 40198459267513, Lot Number: 25DMG688; Medline Kit Number/SKU DYNJ88091A: UDI/DI each 10198459267512, UDI/DI case 40198459267513, Lot Number: 25GMI619; Medline Kit Number/SKU DYNJ88091B: UDI/DI each 10198459488733, UDI/DI case 40198459488734, Lot Number: 25IMC845; Medline Kit Number/SKU DYNJ88091B: UDI/DI each 10198459488733, UDI/DI case 40198459488734, Lot Number: 25KMH998; Medline Kit Number/SKU DYNJ88091B: UDI/DI each 10198459488733, UDI/DI case 40198459488734, Lot Number: 26CMB904; Medline Kit Number/SKU DYNJ89372: UDI/DI each 10198459271281, UDI/DI case 40198459271282, Lot Number: 25DMD483; Medline Kit Number/SKU DYNJ89372: UDI/DI each 10198459271281, UDI/DI case 40198459271282, Lot Number: 25EMA204; Medline Kit Number/SKU DYNJ89372: UDI/DI each 10198459271281, UDI/DI case 40198459271282, Lot Number: 25GMH683; Medline Kit Number/SKU DYNJ89372: UDI/DI each 10198459271281, UDI/DI case 40198459271282, Lot Number: 25IMB440; Medline Kit Number/SKU DYNJ89372: UDI/DI each 10198459271281, UDI/DI case 40198459271282, Lot Number: 26AMH888; Medline Kit Number/SKU DYNJ912124: UDI/DI each 10198459635984, UDI/DI case 40198459635985, Lot Number: 26ABS012; Medline Kit Number/SKU DYNJT2164L: UDI/DI each 10193489991345, UDI/DI case 40193489991346, Lot Number: 23GBM001; Medline Kit Number/SKU DYNJT5674: UDI/DI each 10198459492174, UDI/DI case 40198459492175, Lot Number: 25IBN116; Medline Kit Number/SKU DYNJT6659: UDI/DI each 10198459573521, UDI/DI case 40198459573522, Lot Number: 25KBT390; Medline Kit Number/SKU DYNJV0044AF: UDI/DI each 10198459394638, UDI/DI case 40198459394639, Lot Number: 25IMB924; Medline Kit Number/SKU DYNJV0044AF: UDI/DI each 10198459394638, UDI/DI case 40198459394639, Lot Number: 25IMC139; Medline Kit Number/SKU DYNJV0044AF: UDI/DI each 10198459394638, UDI/DI case 40198459394639, Lot Number: 25LMI817; Medline Kit Number/SKU NAM0017: UDI/DI each 10195327386566, UDI/DI case 40195327386567, Lot Number: 23IBS523; Medline Kit Number/SKU NAM0017: UDI/DI each 10195327386566, UDI/DI case 40195327386567, Lot Number: 23JBT185; Medline Kit Number/SKU NAM0017: UDI/DI each 10195327386566, UDI/DI case 40195327386567, Lot Number: 23KBJ926; Medline Kit Number/SKU NAM0017: UDI/DI each 10195327386566, UDI/DI case 40195327386567, Lot Number: 24BBN433; Medline Kit Number/SKU NAM0029: UDI/DI each 10195327689582, UDI/DI case 40195327689583, Lot Number: 24FBH082; Medline Kit Number/SKU NAM0029: UDI/DI each 10195327689582, UDI/DI case 40195327689583, Lot Number: 24HBE143; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 24KBL664; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 24LBO182; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 25ABB321; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 25ABQ460; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147 457, Lot Number: 25BBN806; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 25CBL645; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 25CBQ900; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 25DBK491; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 25DBP291; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 25EBD435; Medline Kit Number/SKU NAM0029A: UDI/DI each 10198459147456, UDI/DI case 40198459147457, Lot Number: 25EBJ261; Medline Kit Number/SKU NAM0029B: UDI/DI each 10198459598500, UDI/DI case 40198459598501, Lot Number: 26CBG949; Medline Kit Number/SKU NAM0041: UDI/DI each 10198459060120, UDI/DI case 40198459060121, Lot Number: 24IBG189; Medline Kit Number/SKU NAM0041: UDI/DI each 10198459060120, UDI/DI case 40198459060121, Lot Number: 24JBS180; Medline Kit Number/SKU NAM0041: UDI/DI each 10198459060120, UDI/DI case 40198459060121, Lot Number: 25BBD379; Medline Kit Number/SKU NAM0041: UDI/DI each 10198459060120, UDI/DI case 40198459060121, Lot Number: 25BBI537; Medline Kit Number/SKU NAM0045: UDI/DI each 10198459066276, UDI/DI case 40198459066277, Lot Number: 24IBD550; Medline Kit Number/SKU NAM0045: UDI/DI each 10198459066276, UDI/DI case 40198459066277, Lot Number: 24JBK932; Medline Kit Number/SKU NAM0045: UDI/DI each 10198459066276, UDI/DI case 40198459066277, Lot Number: 24LBM784; Medline Kit Number/SKU NAM0045: UDI/DI each 10198459066276, UDI/DI case 40198459066277, Lot Number: 25ABV072; Medline Kit Number/SKU NAM0045: UDI/DI each 10198459066276, UDI/DI case 40198459066277, Lot Number: 25BBD380; Medline Kit Number/SKU NAM0045: UDI/DI each 10198459066276, UDI/DI case 40198459066277, Lot Number: 25EBD436; Medline Kit Number/SKU NAM0045: UDI/DI each 10198459066276, UDI/DI case 40198459066277, Lot Number: 25EBN603; Medline Kit Number/SKU NAM0072: UDI/DI each 10198459179082, UDI/DI case 40198459179083, Lot Number: 25ABV075; Medline Kit Number/SKU NAM0072: UDI/DI each 10198459179082, UDI/DI case 40198459179083, Lot Number: 25BBD389; Medline Kit Number/SKU NAM0072: UDI/DI each 10198459179082, UDI/DI case 40198459179083, Lot Number: 25DBU083; Medline Kit Number/SKU SAMPC0701: UDI/DI each 10195327468774, UDI/DI case 40195327468775, Lot Number: 23HDA684; Medline Kit Number/SKU SAMPC0703: UDI/DI each 10195327480950, UDI/DI case 40195327480951, Lot Number: 23HDA846; Medline Kit Number/SKU SAMPC0708: UDI/DI each 10198459088445, UDI/DI case 40198459088446, Lot Number: 24HDA755; Medline Kit Number/SKU SAMPD152837: UDI/DI each 10198459058561, UDI/DI case 40198459058562, Lot Number: 24GBM271; Medline Kit Number/SKU VASC1073A: UDI/DI each 10889942781881, UDI/DI case 40889942781882, Lot Number: 25KME586; Medline Kit Number/SKU VASC1175B: UDI/DI each 10195327491154, UDI/DI case 40195327491155, Lot Number: 23JBL570; Medline Kit Number/SKU VASC1175B: UDI/DI each 10195327491154, UDI/DI case 40195327491155, Lot Number: 24ABH521; Medline Kit Number/SKU VASC1175B: UDI/DI each 10195327491154, UDI/DI case 40195327491155, Lot Number: 24BBL253; Medline Kit Number/SKU VASC1175B: UDI/DI each 10195327491154, UDI/DI case 40195327491155, Lot Number: 24EBP053; Medline Kit Number/SKU VASC1175B: UDI/DI each 10195327491154, UDI/DI case 40195327491155, Lot Number: 24JBX442; Medline Kit Number/SKU VASC1175B: UDI/DI each 10195327491154, UDI/DI case 40195327491155, Lot Number: 24LBI063; Medline Kit Number/SKU VASC1175B: UDI/DI each 10195327491154, UDI/DI case 40195327491155, Lot Number: 25EBD026; Medline Kit Number/SKU VASC1175B: UDI/DI each 10195327491154, UDI/DI case 40195327491155, Lot Number: 25JBV611; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 23GBO413; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 23JBS340; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 23KBJ698; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 24ABC832; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 24BBR007; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 24EBG473; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 24FBS827; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 24JBJ738; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 24KBR249; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 24LBB676; Medline Kit Number/SKU VASC1307: UDI/DI each 10195327057855, UDI/DI case 40195327057856, Lot Number: 24LBF951; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 25BBS813; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 25CBU102; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 25DBT093; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 25FBF141; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 25FBU483; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 25IBO955; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 25JBA394; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 25KBI436; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 25KBS220; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 26ABW699; Medline Kit Number/SKU VASC1307A: UDI/DI each 10198459281211, UDI/DI case 40198459281212, Lot Number: 26BBN614; Medline Kit Number/SKU VASC1309: UDI/DI each 10195327067861, UDI/DI case 40195327067862, Lot Number: 24HBK223; Medline Kit Number/SKU VASC1330: UDI/DI each 10195327140267, UDI/DI case 40195327140268, Lot Number: 23GBG557; Medline Kit Number/SKU VASC1330: UDI/DI each 10195327140267, UDI/DI case 40195327140268, Lot Number: 23JBA868; Medline Kit Number/SKU VASC1330: UDI/DI each 10195327140267, UDI/DI case 40195327140268, Lot Number: 23KBO634; Medline Kit Number/SKU VASC1330: UDI/DI each 10195327140267, UDI/DI case 40195327140268, Lot Number: 23LBC259; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 23GBG636; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 23IBM148; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 23JBQ166; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 23KBX574; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 24BBQ341; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 24DBL151; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 24EBC605; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 24FBS828; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 24HBN733; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 24IBV678; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 24KBI833; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 24LBC227; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 25ABM013; Medline Kit Number/SKU VASC1336: UDI/DI each 10195327148560, UDI/DI case 40195327148561, Lot Number: 25CBU503; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 23GBL954; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 23HBJ412; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 23JBU630; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 23KBM761; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 23LBI879; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 24BBI218; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 24FBS126; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 24GBK942; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 24JBP862; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 24JBQ832; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 25BBI016; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 25CBG059; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 25DBI492; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 25EBG580; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 25FBP248; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 25GBX869; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 25HBO331; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 25KBJ331; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 26ABE855; Medline Kit Number/SKU VASC1341: UDI/DI each 10195327163228, UDI/DI case 40195327163229, Lot Number: 26BBM371; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 23GBF251; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 23HBW617; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 23JBS631; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 23KBJ693; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 24ABC841; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 24BBG070; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 24FBN902; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 24GBJ936; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 24GBJ937; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 24JBP863; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 24KBR198; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 24LBC213; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 25BBE728; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 25BBK464; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 25CBG060; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 25CBV014; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 25EBB515; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 25FBP249; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 25GBX607; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 25IBO066; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 25KBF126; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 26ABA326; Medline Kit Number/SKU VASC1345: UDI/DI each 10195327179342, UDI/DI case 40195327179343, Lot Number: 26CBB784; Medline Kit Number/SKU VASC1376: UDI/DI each 10195327265908, UDI/DI case 40195327265909, Lot Number: 25LBD347; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 24ABF474; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 24BBQ185; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 24FBM433; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 24HBO427; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 25ABK652; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 25CBQ238; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 25IBP316; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 25KBT042; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 25LBO217; Medline Kit Number/SKU VASC1378A: UDI/DI each 10195327397784, UDI/DI case 40195327397785, Lot Number: 26BBN602; Medline Kit Number/SKU VASC1395: UDI/DI each 10195327336707, UDI/DI case 40195327336708, Lot Number: 23IBH009; Medline Kit Number/SKU VASC1395: UDI/DI each 10195327336707, UDI/DI case 40195327336708, Lot Number: 24CBV165; Medline Kit Number/SKU VASC1418: UDI/DI each 10195327410049, UDI/DI case 40195327410040, Lot Number: 23HBJ408; Medline Kit Number/SKU VASC1418: UDI/DI each 10195327410049, UDI/DI case 40195327410040, Lot Number: 24EBK828; Medline Kit Number/SKU VASC1418: UDI/DI each 10195327410049, UDI/DI case 40195327410040, Lot Number: 24FBR525; Medline Kit Number/SKU VASC1432: UDI/DI each 10195327436285, UDI/DI case 40195327436286, Lot Number: 23GBP189; Medline Kit Number/SKU VASC1432: UDI/DI each 10195327436285, UDI/DI case 40195327436286, Lot Number: 24ABE724; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 23JBQ888; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 24BBO469; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 24CBK102; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 24GBJ664; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 24IBJ614; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 24JBD638; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 24LBO430; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25ABS945; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25CBB738; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25CBG062; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25CBK197; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25DBK917; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25EBK746; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25GBH482; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25HBL992; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25HBV179; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 25KBB710; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 26ABB466; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 26ABU509; Medline Kit Number/SKU VASC1445: UDI/DI each 10195327469122, UDI/DI case 40195327469123, Lot Number: 26BBO249; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 23IBW308; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 23JBW780; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 23KBD474; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 23KBT735; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 24GBG265; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 24JBP854; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 25BBU939; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 25CBH997; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 25HBB956; Medline Kit Number/SKU VASC1454: UDI/DI each 10195327497194, UDI/DI case 40195327497195, Lot Number: 26CBA661; Medline Kit Number/SKU VASC1458: UDI/DI each 10195327507305, UDI/DI case 40195327507306, Lot Number: 23KBG748; Medline Kit Number/SKU VASC1458: UDI/DI each 10195327507305, UDI/DI case 40195327507306, Lot Number: 24DBS588; Medline Kit Number/SKU VASC1458: UDI/DI each 10195327507305, UDI/DI case 40195327507306, Lot Number: 24DBT477; Medline Kit Number/SKU VASC1458: UDI/DI each 10195327507305, UDI/DI case 40195327507306, Lot Number: 24EBP054; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 24FBC400; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 24GBQ830; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 24HBM889; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 24KBR203; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 24LBB555; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 25ABG248; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 25ABR181; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 25CBQ246; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 25EBC788; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 25GBN426; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 25IBG362; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 25LBD363; Medline Kit Number/SKU VASC1458A: UDI/DI each 10198459030444, UDI/DI case 40198459030445, Lot Number: 26ABV296; Medline Kit Number/SKU VASC1468: UDI/DI each 10195327535964, UDI/DI case 40195327535965, Lot Number: 23KBT472; Medline Kit Number/SKU VASC1468: UDI/DI each 10195327535964, UDI/DI case 40195327535965, Lot Number: 23LBH014; Medline Kit Number/SKU VASC1468: UDI/DI each 10195327535964, UDI/DI case 40195327535965, Lot Number: 24IBW117; Medline Kit Number/SKU VASC1468: UDI/DI each 10195327535964, UDI/DI case 40195327535965, Lot Number: 24JBU059; Medline Kit Number/SKU VASC1468: UDI/DI each 10195327535964, UDI/DI case 40195327535965, Lot Number: 25ABK642; Medline Kit Number/SKU VASC1468: UDI/DI each 10195327535964, UDI/DI case 40195327535965, Lot Number: 25IBK771; Medline Kit Number/SKU VASC1468: UDI/DI each 10195327535964, UDI/DI case 40195327535965, Lot Number: 25JBO724; Medline Kit Number/SKU VASC1468: UDI/DI each 10195327535964, UDI/DI case 40195327535965, Lot Number: 25JBV916; Medline Kit Number/SKU VASC1468: UDI/DI each 10195327535964, UDI/DI case 40195327535965, Lot Number: 25JBW936; Medline Kit Number/SKU VASC1476: UDI/DI each 10195327555450, UDI/DI case 40195327555451, Lot Number: 24ABI604; Medline Kit Number/SKU VASC1476: UDI/DI each 10195327555450, UDI/DI case 40195327555451, Lot Number: 24BBR685; Medline Kit Number/SKU VASC1476: UDI/DI each 10195327555450, UDI/DI case 40195327555451, Lot Number: 24CBL307; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 24BBM518; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 24EBP055; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 24GBO567; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 24HBO359; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 24IBD026; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 24IBD173; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 24KBO173; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25ABD185; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25ABN409; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25BBS172; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25EBD514; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25EBK215; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25FBD752; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25HBI077; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25HBP690; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25HBQ603; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25IBR975; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25JBE669; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 25KBA765; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 26ABE861; Medline Kit Number/SKU VASC1495: UDI/DI each 10195327613532, UDI/DI case 40195327613533, Lot Number: 26BBE208; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 24FBO095; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 24HBE148; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 24HBO351; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 24JBS222; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 24LBO873; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 25EBP296; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 25GBL073; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 25HBP387; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 25IBJ406; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 25JBT230; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 25LBD358; Medline Kit Number/SKU VASC1524: UDI/DI each 10198459015540, UDI/DI case 40198459015541, Lot Number: 26ABM420; Medline Kit Number/SKU VASC1529: UDI/DI each 10198459024757, UDI/DI case 40198459024758, Lot Number: 24GBW722; Medline Kit Number/SKU VASC1529: UDI/DI each 10198459024757, UDI/DI case 40198459024758, Lot Number: 24KBF604; Medline Kit Number/SKU VASC1529: UDI/DI each 10198459024757, UDI/DI case 40198459024758, Lot Number: 24KBI922; Medline Kit Number/SKU VASC1529: UDI/DI each 10198459024757, UDI/DI case 40198459024758, Lot Number: 24LBE301; Medline Kit Number/SKU VASC1529: UDI/DI each 10198459024757, UDI/DI case 40198459024758, Lot Number: 25ABG688; Medline Kit Number/SKU VASC1529: UDI/DI each 10198459024757, UDI/DI case 40198459024758, Lot Number: 25BBA701; Medline Kit Number/SKU VASC1535: UDI/DI each 10198459039898, UDI/DI case 40198459039899, Lot Number: 24LBH402; Medline Kit Number/SKU VASC1535: UDI/DI each 10198459039898, UDI/DI case 40198459039899, Lot Number: 25BBR553; Medline Kit Number/SKU VASC1552: UDI/DI each 10198459054662, UDI/DI case 40198459054663, Lot Number: 24IBG423; Medline Kit Number/SKU VASC1552: UDI/DI each 10198459054662, UDI/DI case 40198459054663, Lot Number: 24JBA136; Medline Kit Number/SKU VASC1552: UDI/DI each 10198459054662, UDI/DI case 40198459054663, Lot Number: 25ABG220; Medline Kit Number/SKU VASC1552: UDI/DI each 10198459054662, UDI/DI case 40198459054663, Lot Number: 25ABT643; Medline Kit Number/SKU VASC1552A: UDI/DI each 10198459309274, UDI/DI case 40198459309275, Lot Number: 25CBT107; Medline Kit Number/SKU VASC1552A: UDI/DI each 10198459309274, UDI/DI case 40198459309275, Lot Number: 25FBK473; Medline Kit Number/SKU VASC1552A: UDI/DI each 10198459309274, UDI/DI case 40198459309275, Lot Number: 25FBT692; Medline Kit Number/SKU VASC1552B: UDI/DI each 10198459421679, UDI/DI case 40198459421670, Lot Number: 25GBL095; Medline Kit Number/SKU VASC1552B: UDI/DI each 10198459421679, UDI/DI case 40198459421670, Lot Number: 25GBW734; Medline Kit Number/SKU VASC1552B: UDI/DI each 10198459421679, UDI/DI case 40198459421670, Lot Number: 25IBG349; Medline Kit Number/SKU VASC1552B: UDI/DI each 10198459421679, UDI/DI case 40198459421670, Lot Number: 25JBR297; Medline Kit Number/SKU VASC1552B: UDI/DI each 10198459421679, UDI/DI case 40198459421670, Lot Number: 25JBW381; Medline Kit Number/SKU VASC1552B: UDI/DI each 10198459421679, UDI/DI case 40198459421670, Lot Number: 26ABM418; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 24JMJ373; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 25AMI133; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 25CMA605; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 25DMG953; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 25EMA592; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 25FMI627; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 25HMJ570; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 25KMC677; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 25LMF146; Medline Kit Number/SKU VASC1573: UDI/DI each 10198459087363, UDI/DI case 40198459087364, Lot Number: 26CMC102; Medline Kit Number/SKU VASC1587: UDI/DI each 10198459109171, UDI/DI case 40198459109172, Lot Number: 24IBU796; Medline Kit Number/SKU VASC1587: UDI/DI each 10198459109171, UDI/DI case 40198459109172, Lot Number: 25ABD085; Medline Kit Number/SKU VASC1587: UDI/DI each 10198459109171, UDI/DI case 40198459109172, Lot Number: 25ABE333; Medline Kit Number/SKU VASC1587: UDI/DI each 10198459109171, UDI/DI case 40198459109172, Lot Number: 25EBI346; Medline Kit Number/SKU VASC1587: UDI/DI each 10198459109171, UDI/DI case 40198459109172, Lot Number: 25HBB611; Medline Kit Number/SKU VASC1587: UDI/DI each 10198459109171, UDI/DI case 40198459109172, Lot Number: 25HBM085; Medline Kit Number/SKU VASC1595: UDI/DI each 10198459116292, UDI/DI case 40198459116293, Lot Number: 24JBD973; Medline Kit Number/SKU VASC1595: UDI/DI each 10198459116292, UDI/DI case 40198459116293, Lot Number: 24JBF683; Medline Kit Number/SKU VASC1595: UDI/DI each 10198459116292, UDI/DI case 40198459116293, Lot Number: 24JBK069; Medline Kit Number/SKU VASC1595: UDI/DI each 10198459116292, UDI/DI case 40198459116293, Lot Number: 24KBA325; Medline Kit Number/SKU VASC1595: UDI/DI each 10198459116292, UDI/DI case 40198459116293, Lot Number: 25ABN567; Medline Kit Number/SKU VASC1595: UDI/DI each 10198459116292, UDI/DI case 40198459116293, Lot Number: 25EBF873; Medline Kit Number/SKU VASC1595: UDI/DI each 10198459116292, UDI/DI case 40198459116293, Lot Number: 25HBB986; Medline Kit Number/SKU VASC1595: UDI/DI each 10198459116292, UDI/DI case 40198459116293, Lot Number: 25HBH329; Medline Kit Number/SKU VASC1595: UDI/DI each 10198459116292, UDI/DI case 40198459116293, Lot Number: 26CBB911; Medline Kit Number/SKU VASC1601A: UDI/DI each 10198459290893, UDI/DI case 40198459290894, Lot Number: 25CMD775; Medline Kit Number/SKU VASC1601A: UDI/DI each 10198459290893, UDI/DI case 40198459290894, Lot Number: 26BME893; Medline Kit Number/SKU VASC1601A: UDI/DI each 10198459290893, UDI/DI case 40198459290894, Lot Number: 26BMF480; Medline Kit Number/SKU VASC1614: UDI/DI each 10198459147463, UDI/DI case 40198459147464, Lot Number: 25HBF882; Medline Kit Number/SKU VASC1614: UDI/DI each 10198459147463, UDI/DI case 40198459147464, Lot Number: 25HBH102; Medline Kit Number/SKU VASC1621: UDI/DI each 10198459151699, UDI/DI case 40198459151690, Lot Number: 24JBR859; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 24KBD713; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25ABP307; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25DBB874; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25DBK489; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25DBS122; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25EBK745; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25GBJ696; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25HBL699; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25IBU553; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25JBD501; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25KBA698; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 25LBM073; Medline Kit Number/SKU VASC1628: UDI/DI each 10198459167492, UDI/DI case 40198459167493, Lot Number: 26ABQ183; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 25BBK945; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 25DBA347; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 25DBH403; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 25DBT073; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 25EBB516; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 25FBU210; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 25IBT715; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 25LBF337; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 25LBJ928; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 26ABM416; Medline Kit Number/SKU VASC1675: UDI/DI each 10198459239649, UDI/DI case 40198459239640, Lot Number: 26BBE214; Medline Kit Number/SKU VASC1677: UDI/DI each 10198459241239, UDI/DI case 40198459241230, Lot Number: 25BBE387; Medline Kit Number/SKU VASC1677: UDI/DI each 10198459241239, UDI/DI case 40198459241230, Lot Number: 25BBE706; Medline Kit Number/SKU VASC1677: UDI/DI each 10198459241239, UDI/DI case 40198459241230, Lot Number: 25BBI099; Medline Kit Number/SKU VASC1688: UDI/DI each 10198459248764, UDI/DI case 40198459248765, Lot Number: 25DMI539; Medline Kit Number/SKU VASC1688: UDI/DI each 10198459248764, UDI/DI case 40198459248765, Lot Number: 25FMD563; Medline Kit Number/SKU VASC1688: UDI/DI each 10198459248764, UDI/DI case 40198459248765, Lot Number: 25HME391; Medline Kit Number/SKU VASC1688: UDI/DI each 10198459248764, UDI/DI case 40198459248765, Lot Number: 25IMC893; Medline Kit Number/SKU VASC1688: UDI/DI each 10198459248764, UDI/DI case 40198459248765, Lot Number: 25KMD926; Medline Kit Number/SKU VASC1688: UDI/DI each 10198459248764, UDI/DI case 40198459248765, Lot Number: 25KMG797; Medline Kit Number/SKU VASC1688: UDI/DI each 10198459248764, UDI/DI case 40198459248765, Lot Number: 26BMG823; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 25ABR120; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 25CBS012; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 25DBC444; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 25DBT061; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 25EBG581; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 25EBM219; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 25FBR113; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 25JBO893; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 25LBU338; Medline Kit Number/SKU VASC1689: UDI/DI each 10198459249556, UDI/DI case 40198459249557, Lot Number: 26ABP243; Medline Kit Number/SKU VASC1707: UDI/DI each 10198459269349, UDI/DI case 40198459269340, Lot Number: 25CBQ372; Medline Kit Number/SKU VASC1707: UDI/DI each 10198459269349, UDI/DI case 40198459269340, Lot Number: 25CBS015; Medline Kit Number/SKU VASC1707: UDI/DI each 10198459269349, UDI/DI case 40198459269340, Lot Number: 25JBC829; Medline Kit Number/SKU VASC1707: UDI/DI each 10198459269349, UDI/DI case 40198459269340, Lot Number: 25JBN628; Medline Kit Number/SKU VASC1707: UDI/DI each 10198459269349, UDI/DI case 40198459269340, Lot Number: 25LBM018; Medline Kit Number/SKU VASC1707: UDI/DI each 10198459269349, UDI/DI case 40198459269340, Lot Number: 26ABQ078; Medline Kit Number/SKU VASC1707: UDI/DI each 10198459269349, UDI/DI case 40198459269340, Lot Number: 26BBH355; Medline Kit Number/SKU VASC1716: UDI/DI each 10198459288197, UDI/DI case 40198459288198, Lot Number: 25CBO782; Medline Kit Number/SKU VASC1716: UDI/DI each 10198459288197, UDI/DI case 40198459288198, Lot Number: 25EBI380; Medline Kit Number/SKU VASC1716: UDI/DI each 10198459288197, UDI/DI case 40198459288198, Lot Number: 25FBF136; Medline Kit Number/SKU VASC1716: UDI/DI each 10198459288197, UDI/DI case 40198459288198, Lot Number: 25GBJ090; Medline Kit Number/SKU VASC1716: UDI/DI each 10198459288197, UDI/DI case 40198459288198, Lot Number: 25HBV181; Medline Kit Number/SKU VASC1716: UDI/DI each 10198459288197, UDI/DI case 40198459288198, Lot Number: 25JBD320; Medline Kit Number/SKU VASC1716: UDI/DI each 10198459288197, UDI/DI case 40198459288198, Lot Number: 25KBT907; Medline Kit Number/SKU VASC1716: UDI/DI each 10198459288197, UDI/DI case 40198459288198, Lot Number: 26ABI826; Medline Kit Number/SKU VASC1716: UDI/DI each 10198459288197, UDI/DI case 40198459288198, Lot Number: 26CBB660; Medline Kit Number/SKU VASC1722: UDI/DI each 10198459290657, UDI/DI case 40198459290658, Lot Number: 25CBS091; Medline Kit Number/SKU VASC1722: UDI/DI each 10198459290657, UDI/DI case 40198459290658, Lot Number: 25HBB626; Medline Kit Number/SKU VASC1722: UDI/DI each 10198459290657, UDI/DI case 40198459290658, Lot Number: 25LBK313; Medline Kit Number/SKU VASC1722: UDI/DI each 10198459290657, UDI/DI case 40198459290658, Lot Number: 26BBL119; Medline Kit Number/SKU VASC1752: UDI/DI each 10198459337239, UDI/DI case 40198459337230, Lot Number: 25EBO391; Medline Kit Number/SKU VASC1752: UDI/DI each 10198459337239, UDI/DI case 40198459337230, Lot Number: 25FBK965; Medline Kit Number/SKU VASC1752: UDI/DI each 10198459337239, UDI/DI case 40198459337230, Lot Number: 25GBJ834; Medline Kit Number/SKU VASC1752: UDI/DI each 10198459337239, UDI/DI case 40198459337230, Lot Number: 25IBR988; Medline Kit Number/SKU VASC1752: UDI/DI each 10198459337239, UDI/DI case 40198459337230, Lot Number: 25JBO894; Medline Kit Number/SKU VASC1764: UDI/DI each 10198459354021, UDI/DI case 40198459354022, Lot Number: 26AME492; Medline Kit Number/SKU VASC1769: UDI/DI each 10198459357664, UDI/DI case 40198459357665, Lot Number: 25FBQ540; Medline Kit Number/SKU VASC1769: UDI/DI each 10198459357664, UDI/DI case 40198459357665, Lot Number: 26CBI351; Medline Kit Number/SKU VASC1823: UDI/DI each 10198459435423, UDI/DI case 40198459435424, Lot Number: 25IBU912; Medline Kit Number/SKU VASC1823: UDI/DI each 10198459435423, UDI/DI case 40198459435424, Lot Number: 25KBJ286; Medline Kit Number/SKU VASC1823: UDI/DI each 10198459435423, UDI/DI case 40198459435424, Lot Number: 25LBS157; Medline Kit Number/SKU VASC1823: UDI/DI each 10198459435423, UDI/DI case 40198459435424, Lot Number: 26ABW694; Medline Kit Number/SKU VASC1823: UDI/DI each 10198459435423, UDI/DI case 40198459435424, Lot Number: 26CBJ225; Medline Kit Number/SKU VASC1826: UDI/DI each 10198459437885, UDI/DI case 40198459437886, Lot Number: 25HBK292; Medline Kit Number/SKU VASC1826: UDI/DI each 10198459437885, UDI/DI case 40198459437886, Lot Number: 25KBU157; Medline Kit Number/SKU VASC1861: UDI/DI each 10198459533693, UDI/DI case 40198459533694, Lot Number: 26AMG783; Medline Kit Number/SKU VASC1861: UDI/DI each 10198459533693, UDI/DI case 40198459533694, Lot Number: 26CMC099; Medline Kit Number/SKU VASC1861: UDI/DI each 10198459533693, UDI/DI case 40198459533694, Lot Number: 26CME505; Medline Kit Number/SKU VASC1906: UDI/DI each 10198459612213, UDI/DI case 40198459612214, Lot Number: 26BMH550; Medline Kit Number/SKU VASC1909: UDI/DI each 10198459619779, UDI/DI case 40198459619770, Lot Number: 26BBC220; Medline Kit Number/SKU VASC1931: UDI/DI each 10198459667466, UDI/DI case 40198459667467, Lot Number: 26BMD155.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2144-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass I
StatusOngoing
Recall startedMarch 24, 2026
ReportedMay 20, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity322460 kits