Ivenix Infusion System recall (Fresenius Kabi USA, LLC)

Medical deviceClass IOngoing

Summary: Fresenius Kabi USA, LLC is recalling Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004. Reason: Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

What is being recalled

Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Why

Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.

How serious is it

Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.

Where it was sold

Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI; (states named: California, Colorado, Florida, Georgia, Idaho, Illinois, Maryland, Michigan, Minnesota, Missouri, Mississippi, Nebraska, New Jersey, Nevada, Oklahoma, Oregon, South Carolina, Texas, Virginia, Washington, Wisconsin)

How to tell if yours is affected

Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2224-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass I
StatusOngoing
Recall startedMay 6, 2026
ReportedJune 10, 2026
CompanyFresenius Kabi USA, LLC
Company locationNorth Andover, MA, United States
Initiated byVoluntary: Firm initiated
Quantity18,444