Allura system recall (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.)
Medical deviceClass IIOngoing
Summary: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. is recalling Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (722002), Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper F. Reason: The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
What is being recalled
Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (722002), Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039), Cardio Vascular-Allura (722400);
Why
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen; (states named: Georgia)
How to tell if yours is affected
Code Information: Allura: 1. Allura Xper FD10C; Model: 722001; UDI-DI: N/A; Serial Numbers: 170, 242, 267, 142, 206, 316, 9, 306, 103, 33, 220; 2. Allura Xper FD10F; Model: 722002; UDI-DI: N/A; Serial Numbers: 0, 81, 47; 3. Allura Xper FD10; Model: 722003; UDI-DI: N/A; Serial Numbers: 208, 392, 109, 45, 1953, 1471, 1551, 485, 152, 928, 501, 90, 1435, 81, 939, 160, 472, 1349, 439, 1958, 1525, 72, 1187, 912, 368, 328A, 37, 1286, 1223, 487, 1061, 833, 688, 317, 157, 639, 375, 420, 225, 1338, 614, 214, 230, 736, 64, 1090, 1094, 1932, 1531, 421, 984, 373, 313, 918, 602, 374, 750, 1934, 1751, 1661, 961, 791, 454, 868, 416, 1329, 241; 4. Allura Xper FD10; Model: 722010; UDI: (01)00884838059030(21)921, (01)00884838059030(21)1440, (01)00884838059030(21)1328, (01)00884838059030(21)1292, (01)00884838059030(21)1428, (01)00884838059030(21)1396, (01)00884838059030(21)298, (01)00884838059030(21)1432, (01)00884838059030(21)1425, (01)00884838059030(21)1398, (01)00884838059030(21)1397, (01)00884838059030(21)1391, (01)00884838059030(21)1395, (01)00884838059030(21)314, (01)00884838059030(21)1297, (01)00884838059030(21)946, (01)00884838059030(21)1424, (01)00884838059030(21)1244, (01)00884838059030(21)1122, (01)00884838059030(21)1198, (01)00884838059030(21)1421, (01)00884838059030(21)1423, (01)00884838059030(21)1438, (01)00884838059030(21)697, (01)00884838059030(21)1192, (01)00884838059030(21)1326, (01)00884838059030(21)402, (01)00884838059030(21)1148, (01)00884838059030(21)1439, (01)00884838059030(21)1431, (01)00884838059030(21)1404, (01)00884838059030(21)1140, (01)00884838059030(21)1429, (01)00884838059030(21)1430, (01)00884838059030(21)1403, (01)00884838059030(21)1393, (01)00884838059030(21)1402, (01)00884838059030(21)406, (01)00884838059030(21)71, (01)00884838059030(21)1434, (01)00884838059030(21)1392, (01)00884838059030(21)1379, (01)00884838059030(21)1401, (01)00884838059030(21)1197, (01)00884838059030(21)1435, (01)00884838059030(21)1346, (01)00884838059030(21)232, (01)00884838059030(21)231, (01)00884838059030(21)248, (01)00884838059030(21)1137, (01)00884838059030(21)1446, (01)00884838059030(21)1422, (01)00884838059030(21)1433, (01)00884838059030(21)1131, (01)00884838059030(21)1218; 5. Allura Xper FD10; Model: 722026; UDI: (01)00884838054189(21)683, (01)00884838054189(21)1052, (01)00884838054189(21)1051, (01)00884838054189(21)786, (01)00884838054189(21)654, (01)00884838054189(21)1132, (01)00884838054189(21)1100, (01)00884838054189(21)770, (01)00884838054189(21)1338, (01)00884838054189(21)1068, (01)00884838054189(21)1072, (01)00884838054189(21)796, (01)00884838054189(21)1248, (01)00884838054189(21)1129, (01)00884838054189(21)781, (01)00884838054189(21)634, (01)00884838054189(21)908, (01)00884838054189(21)969, (01)00884838054189(21)1246, (01)00884838054189(21)877, (01)00884838054189(21)843, (01)00884838054189(21)1231, (01)00884838054189(21)1038, (01)00884838054189(21)1287, (01)00884838054189(21)878, (01)00884838054189(21)558, (01)00884838054189(21)633, (01)00884838054189(21)1050, (01)00884838054189(21)1000, (01)00884838054189(21)909, (01)00884838054189(21)841, (01)00884838054189(21)713, (01)00884838054189(21)731, (01)00884838054189(21)646, (01)00884838054189(21)728, (01)00884838054189(21)765, (01)00884838054189(21)959, (01)00884838054189(21)825, (01)00884838054189(21)942, (01)00884838054189(21)1192, (01)00884838054189(21)990, (01)00884838054189(21)774, (01)00884838054189(21)648, (01)00884838054189(21)966, (01)00884838054189(21)1268, (01)00884838054189(21)627, (01)00884838054189(21)604, (01)00884838054189(21)716, (01)00884838054189(21)1182, (01)00884838054189(21)965, (01)00884838054189(21)1189, (01)00884838054189(21)1126, (01)00884838054189(21)1210, (01)00884838054189(21)1239, (01)00884838054189(21)1215, (01)00884838054189(21)696, (01)00884838054189(21)664, (01)00884838054189(21)845, (01)00884838054189(21)778, (01)00884838054189(21)935, (01)00884838054189(21)782, (01)00884838054189(21)694, (01)00884838054189(21)838, (01)00884838054189(21)973, (01)00884838054189(21)985, (01)00884838054189(21)855, (01)00884838054189(21)612, (01)00884838054189(21)725, (01)00884838054189(21)805, (01)00884838054189(21)622, (01)00884838054189(21)1293, (01)00884838054189(21)1205, (01)00884838054189(21)617, (01)00884838054189(21)1101, (01)00884838054189(21)738, (01)00884838054189(21)957, (01)00884838054189(21)1340, (01)00884838054189(21)653, (01)00884838054189(21)663, (01)00884838054189(21)688, (01)00884838054189(21)704, (01)00884838054189(21)931, (01)00884838054189(21)967, (01)00884838054189(21)826, (01)00884838054189(21)998, (01)00884838054189(21)807, (01)00884838054189(21)986, (01)00884838054189(21)962, (01)00884838054189(21)961, (01)00884838054189(21)1049, (01)00884838054189(21)707, (01)00884838054189(21)866, (01)00884838054189(21)759, (01)00884838054189(21)1115, (01)00884838054189(21)830, (01)00884838054189(21)638, (01)00884838054189(21)1316, (01)00884838054189(21)748, (01)00884838054189(21)1278, (01)00884838054189(21)1104, (01)00884838054189(21)1368, (01)00884838054189(21)977, (01)00884838054189(21)1001, (01)00884838054189(21)806, (01)00884838054189(21)769, (01)00884838054189(21)771, (01)00884838054189(21)881, (01)00884838054189(21)581, (01)00884838054189(21)1307, (01)00884838054189(21)1128, (01)00884838054189(21)1127, (01)00884838054189(21)1092, (01)00884838054189(21)276, (01)00884838054189(21)722, (01)00884838054189(21)1188, (01)00884838054189(21)1326, (01)00884838054189(21)809, (01)00884838054189(21)729, (01)00884838054189(21)894, (01)00884838054189(21)1018, (01)00884838054189(21)978, (01)00884838054189(21)1201, (01)00884838054189(21)1202, (01)00884838054189(21)1204, (01)00884838054189(21)564, (01)00884838054189(21)1349, (01)00884838054189(21)1227, (01)00884838054189(21)1220, (01)00884838054189(21)1221, (01)00884838054189(21)650, (01)00884838054189(21)794, (01)00884838054189(21)681, (01)00884838054189(21)545, (01)00884838054189(21)1228, (01)00884838054189(21)1321, (01)00884838054189(21)1071, (01)00884838054189(21)1341, (01)00884838054189(21)897, (01)00884838054189(21)619, (01)00884838054189(21)1107, (01)00884838054189(21)1033, (01)00884838054189(21)804, (01)00884838054189(21)677, (01)00884838054189(21)678, (01)00884838054189(21)1271, (01)00884838054189(21)587, (01)00884838054189(21)1347, (01)00884838054189(21)1037, (01)00884838054189(21)709, (01)00884838054189(21)1203, (01)00884838054189(21)930, (01)00884838054189(21)609, (01)00884838054189(21)569, (01)00884838054189(21)979, (01)00884838054189(21)1313, (01)00884838054189(21)1093, (01)00884838054189(21)1054, (01)00884838054189(21)960, (01)00884838054189(21)928, (01)00884838054189(21)924, (01)00884838054189(21)1233, (01)00884838054189(21)721, (01)00884838054189(21)720, (01)00884838054189(21)1117, (01)00884838054189(21)1146, (01)00884838054189(21)858, (01)00884838054189(21)744, (01)00884838054189(21)745, (01)00884838054189(21)1272, (01)00884838054189(21)1249, (01)00884838054189(21)1091, (01)00884838054189(21)1064, (01)00884838054189(21)863, (01)00884838054189(21)697, (01)00884838054189(21)692, (01)00884838054189(21)756, (01)00884838054189(21)628, (01)00884838054189(21)749, (01)00884838054189(21)611, (01)00884838054189(21)1121, (01)00884838054189(21)1265, (01)00884838054189(21)1264, (01)00884838054189(21)1108, (01)00884838054189(21)662, (01)00884838054189(21)754, (01)00884838054189(21)1031, (01)00884838054189(21)997, (01)00884838054189(21)1356, (01)00884838054189(21)1305, (01)00884838054189(21)1355, (01)00884838054189(21)1362, (01)00884838054189(21)1329, (01)00884838054189(21)1360, (01)00884838054189(21)1364, (01)00884838054189(21)1363, (01)00884838054189(21)1366, (01)00884838054189(21)1199, (01)00884838054189(21)1123, (01)00884838054189(21)1113, (01)00884838054189(21)1026, (01)00884838054189(21)1029, (01)00884838054189(21)884, (01)00884838054189(21)902, (01)00884838054189(21)751, (01)00884838054189(21)753, (01)00884838054189(21)687, (01)00884838054189(21)1301, (01)00884838054189(21)944, (01)00884838054189(21)1196, (01)00884838054189(21)1147, (01)00884838054189(21)938, (01)00884838054189(21)686, (01)00884838054189(21)1019, (01)00884838054189(21)1103, (01)00884838054189(21)679, (01)00884838054189(21)1276, (01)00884838054189(21)927, (01)00884838054189(21)1245, (01)00884838054189(21)575, (01)00884838054189(21)1173, (01)00884838054189(21)589, (01)00884838054189(21)763, (01)00884838054189(21)594, (01)00884838054189(21)723, (01)00884838054189(21)1195, (01)00884838054189(21)656, (01)00884838054189(21)657, (01)00884838054189(21)658, (01)00884838054189(21)857, (01)00884838054189(21)803, (01)00884838054189(21)854, (01)00884838054189(21)779, (01)00884838054189(21)632, (01)00884838054189(21)1165, (01)00884838054189(21)1218, (01)00884838054189(21)1327, (01)00884838054189(21)730, (01)00884838054189(21)1286, (01)00884838054189(21)941, (01)00884838054189(21)915, (01)00884838054189(21)1229, (01)00884838054189(21)643, (01)00884838054189(21)726, (01)00884838054189(21)846, (01)00884838054189(21)637, (01)00884838054189(21)792, (01)00884838054189(21)810, (01)00884838054189(21)995, (01)00884838054189(21)1277, (01)00884838054189(21)1005, (01)00884838054189(21)1106, (01)00884838054189(21)620, (01)00884838054189(21)747, (01)00884838054189(21)783, (01)00884838054189(21)1275, (01)00884838054189(21)1118, (01)00884838054189(21)689, (01)00884838054189(21)1186, (01)00884838054189(21)1250, (01)00884838054189(21)1034, (01)00884838054189(21)631, (01)00884838054189(21)1230, (01)00884838054189(21)1372, (01)00884838054189(21)1046, (01)00884838054189(21)1047, (01)00884838054189(21)1306, (01)00884838054189(21)975, (01)00884838054189(21)626, (01)00884838054189(21)1255, (01)00884838054189(21)1184, (01)00884838054189(21)768, (01)00884838054189(21)917, (01)00884838054189(21)695, (01)00884838054189(21)780, (01)00884838054189(21)1179, (01)00884838054189(21)1208, (01)00884838054189(21)913, (01)00884838054189(21)1152, (01)00884838054189(21)1170, (01)00884838054189(21)840, (01)00884838054189(21)764, (01)00884838054189(21)936, (01)00884838054189(21)1053, (01)00884838054189(21)635, (01)00884838054189(21)606, (01)00884838054189(21)1371, (01)00884838054189(21)895, (01)00884838054189(21)630, (01)00884838054189(21)899, (01)00884838054189(21)734, (01)00884838054189(21)1308, (01)00884838054189(21)719, (01)00884838054189(21)593, (01)00884838054189(21)851, (01)00884838054189(21)1303, (01)00884838054189(21)727, (01)00884838054189(21)520, (01)00884838054189(21)1163, (01)00884838054189(21)903, (01)00884838054189(21)661, (01)00884838054189(21)1177, (01)00884838054189(21)1013, (01)00884838054189(21)1015, (01)00884838054189(21)1012, (01)00884838054189(21)1016, (01)00884838054189(21)871, (01)00884838054189(21)880, (01)00884838054189(21)1187, (01)00884838054189(21)1008, (01)00884838054189(21)940, (01)00884838054189(21)1191, (01)00884838054189(21)1216, (01)00884838054189(21)739, (01)00884838054189(21)624, (01)00884838054189(21)1185, (01)00884838054189(21)1353, (01)00884838054189(21)1296, (01)00884838054189(21)740, (01)00884838054189(21)1044, (01)00884838054189(21)1240, (01)00884838054189(21)625, (01)00884838054189(21)1190, (01)00884838054189(21)932, (01)00884838054189(21)1070, (01)00884838054189(21)1096, (01)00884838054189(21)943, (01)00884838054189(21)1243, (01)00884838054189(21)1097, (01)00884838054189(21)836, (01)00884838054189(21)1021, (01)00884838054189(21)1006, (01)00884838054189(21)1241, (01)00884838054189(21)1242, (01)00884838054189(21)761, (01)00884838054189(21)1359, (01)00884838054189(21)1194, (01)00884838054189(21)666, (01)00884838054189(21)1098, (01)00884838054189(21)649, (01)00884838054189(21)818, (01)00884838054189(21)958, (01)00884838054189(21)1048, (01)00884838054189(21)916, (01)00884838054189(21)703, (01)00884838054189(21)1289, (01)00884838054189(21)812, (01)00884838054189(21)1280, (01)00884838054189(21)1281, (01)00884838054189(21)951, (01)00884838054189(21)1171, (01)00884838054189(21)1175, (01)00884838054189(21)762, (01)00884838054189(21)1343, (01)00884838054189(21)610, (01)00884838054189(21)1145, (01)00884838054189(21)1065, (01)00884838054189(21)773, (01)00884838054189(21)1320, (01)00884838054189(21)912, (01)00884838054189(21)1138, (01)00884838054189(21)814, (01)00884838054189(21)1234, (01)00884838054189(21)691, (01)00884838054189(21)1174, (01)00884838054189(21)1136, (01)00884838054189(21)789, (01)00884838054189(21)672, (01)00884838054189(21)640, (01)00884838054189(21)641, (01)00884838054189(21)642, (01)00884838054189(21)1119, (01)00884838054189(21)873, (01)00884838054189(21)868, (01)00884838054189(21)816, (01)00884838054189(21)1086, (01)00884838054189(21)898, (01)00884838054189(21)551, (01)00884838054189(21)766, (01)00884838054189(21)1150, (01)00884838054189(21)574, (01)00884838054189(21)1155, (01)00884838054189(21)1213, (01)00884838054189(21)970, (01)00884838054189(21)1352, (01)00884838054189(21)1226, (01)00884838054189(21)988, (01)00884838054189(21)994, (01)00884838054189(21)1169, (01)00884838054189(21)752, (01)00884838054189(21)732, (01)00884838054189(21)659, (01)00884838054189(21)956, (01)00884838054189(21)1120, (01)00884838054189(21)1022, (01)00884838054189(21)901, (01)00884838054189(21)815, (01)00884838054189(21)1133, (01)00884838054189(21)1354, (01)00884838054189(21)735, (01)00884838054189(21)1267, (01)00884838054189(21)1318, (01)00884838054189(21)1261, (01)00884838054189(21)1035, (01)00884838054189(21)1010, (01)00884838054189(21)1315, (01)00884838054189(21)1314, (01)00884838054189(21)1116, (01)00884838054189(21)1217, (01)00884838054189(21)1282, (01)00884838054189(21)1009, (01)00884838054189(21)1302, (01)00884838054189(21)1076, (01)00884838054189(21)1346, (01)00884838054189(21)833, (01)00884838054189(21)1331, (01)00884838054189(21)1028, (01)00884838054189(21)1062, (01)00884838054189(21)1358, (01)00884838054189(21)1154, (01)00884838054189(21)989, (01)00884838054189(21)1336, (01)00884838054189(21)1319, (01)00884838054189(21)1310, (01)00884838054189(21)1309, (01)00884838054189(21)891, (01)00884838054189(21)1262, (01)00884838054189(21)1334, (01)00884838054189(21)1339, (01)00884838054189(21)1283, (01)00884838054189(21)1109, (01)00884838054189(21)1252, (01)00884838054189(21)1342, (01)00884838054189(21)708, (01)00884838054189(21)1335, (01)00884838054189(21)987, (01)00884838054189(21)1159, (01)00884838054189(21)1266, (01)00884838054189(21)1153, (01)00884838054189(21)1063, (01)00884838054189(21)1292, (01)00884838054189(21)981, (01)00884838054189(21)948, (01)00884838054189(21)859, (01)00884838054189(21)1007, (01)00884838054189(21)1328, (01)00884838054189(21)1332, (01)00884838054189(21)1351, (01)00884838054189(21)760, (01)00884838054189(21)1256, (01)00884838054189(21)984, (01)00884838054189(21)1161, (01)00884838054189(21)636, (01)00884838054189(21)1294, (01)00884838054189(21)1122, (01)00884838054189(21)939, (01)00884838054189(21)1274, (01)00884838054189(21)1176, (01)00884838054189(21)767, (01)00884838054189(21)1269, (01)00884838054189(21)1214, (01)00884838054189(21)1348, (01)00884838054189(21)1270, (01)00884838054189(21)1164, (01)00884838054189(21)1259, (01)00884838054189(21)1260, (01)00884838054189(21)1135, (01)00884838054189(21)1312, (01)00884838054189(21)1209, (01)00884838054189(21)992, (01)00884838054189(21)844, (01)00884838054189(21)1317, (01)00884838054189(21)621, (01)00884838054189(21)1219, (01)00884838054189(21)875, (01)00884838054189(21)937, (01)00884838054189(21)971, (01)00884838054189(21)714, (01)00884838054189(21)742, (01)00884838054189(21)1224, (01)00884838054189(21)1137, (01)00884838054189(21)651, (01)00884838054189(21)1168, (01)00884838054189(21)701, (01)00884838054189(21)777, (01)00884838054189(21)776, (01)00884838054189(21)1149, (01)00884838054189(21)1207, (01)00884838054189(21)1206, (01)00884838054189(21)1330, (01)00884838054189(21)919, (01)00884838054189(21)741, (01)00884838054189(21)1279, (01)00884838054189(21)1066, (01)00884838054189(21)922, (01)00884838054189(21)808, (01)00884838054189(21)1125, (01)00884838054189(21)338, (01)00884838054189(21)1156, (01)00884838054189(21)1222, (01)00884838054189(21)673, (01)00884838054189(21)835, (01)00884838054189(21)737, (01)00884838054189(21)811, (01)00884838054189(21)1144, (01)00884838054189(21)1143, (01)00884838054189(21)867, (01)00884838054189(21)869, (01)00884838054189(21)702, (01)00884838054189(21)918, (01)00884838054189(21)1140, (01)00884838054189(21)790, (01)00884838054189(21)660, (01)00884838054189(21)1370, (01)00884838054189(21)1045, (01)00884838054189(21)824, (01)00884838054189(21)712, (01)00884838054189(21)1027, (01)00884838054189(21)582, (01)00884838054189(21)1020, (01)00884838054189(21)684, (01)00884838054189(21)736, (01)00884838054189(21)652, (01)00884838054189(21)1225, (01)00884838054189(21)758, (01)00884838054189(21)926, (01)00884838054189(21)724, (01)00884838054189(21)950, (01)00884838054189(21)549, (01)00884838054189(21)819, (01)00884838054189(21)983, (01)00884838054189(21)921, (01)00884838054189(21)821, (01)00884838054189(21)1311, (01)00884838054189(21)1158, (01)00884838054189(21)1157, (01)00884838054189(21)772, (01)00884838054189(21)1023, (01)00884838054189(21)784, (01)00884838054189(21)1004, (01)00884838054189(21)886, (01)00884838054189(21)887, (01)00884838054189(21)1102, (01)00884838054189(21)1002, (01)00884838054189(21)308, (01)00884838054189(21)1058, (01)00884838054189(21)896, (01)00884838054189(21)710, (01)00884838054189(21)1151, (01)00884838054189(21)706, (01)00884838054189(21)443, (01)00884838054189(21)442, (01)00884838054189(21)547, (01)00884838054189(21)870, (01)00884838054189(21)1365, (01)00884838054189(21)889, (01)00884838054189(21)412, (01)00884838054189(21)862, (01)00884838054189(21)1369, (01)00884838054189(21)1090, (01)00884838054189(21)1273, (01)00884838054189(21)788, (01)00884838054189(21)1345, (01)00884838054189(21)1344, (01)00884838054189(21)827, (01)00884838054189(21)1030, (01)00884838054189(21)1178, (01)00884838054189(21)1131, (01)00884838054189(21)675, (01)00884838054189(21)1322, (01)00884838054189(21)623, (01)00884838054189(21)853, (01)00884838054189(21)1172, (01)00884838054189(21)892, (01)00884838054189(21)588, (01)00884838054189(21)1257, (01)00884838054189(21)743, (01)00884838054189(21)1162, (01)00884838054189(21)1323, (01)00884838054189(21)1325, (01)00884838054189(21)876, (01)00884838054189(21)655, (01)00884838054189(21)974, (01)00884838054189(21)793, (01)00884838054189(21)907, (01)00884838054189(21)1251, (01)00884838054189(21)669; 6. Allura Xper FD10/10; Model: 722005; UDI-DI: N/A: Serial Numbers: 51, 22, 259, 196, 530, 435, 233, 309, 89, 29, 399, 480, 6, 41, 71, 124, 458, 54, 55, 528, 80, 556, 74, 203, 60, 125, 426, 123, 64, 299; 7. Allura Xper FD10/10; Model: 722011; UDI: (01)00884838059047(21)49, (01)00884838059047(21)196, (01)00884838059047(21)157, (01)00884838059047(21)82, (01)00884838059047(21)83, (01)00884838059047(21)275, (01)00884838059047(21)101, (01)00884838059047(21)347, (01)00884838059047(21)198, (01)00884838059047(21)94, (01)00884838059047(21)176, (01)00884838059047(21)329, (01)00884838059047(21)337; 8. Allura Xper FD10/10; Model: 722027; UDI: (01)00884838054196(21)351, (01)00884838054196(21)332, (01)00884838054196(21)71, (01)00884838054196(21)240, (01)00884838054196(21)241, (01)00884838054196(21)298, (01)00884838054196(21)175, (01)00884838054196(21)280, (01)00884838054196(21)222, (01)00884838054196(21)183, (01)00884838054196(21)202, (01)00884838054196(21)164, (01)00884838054196(21)335, (01)00884838054196(21)288, (01)00884838054196(21)342, (01)00884838054196(21)306, (01)00884838054196(21)307, (01)00884838054196(21)265, (01)00884838054196(21)169, (01)00884838054196(21)203, (01)00884838054196(21)234, (01)00884838054196(21)267, (01)00884838054196(21)209, (01)00884838054196(21)284, (01)00884838054196(21)263, (01)00884838054196(21)174, (01)00884838054196(21)305, (01)00884838054196(21)246, (01)00884838054196(21)255, (01)00884838054196(21)178, (01)00884838054196(21)266, (01)00884838054196(21)187, (01)00884838054196(21)321, (01)00884838054196(21)328, (01)00884838054196(21)282, (01)00884838054196(21)184, (01)00884838054196(21)337, (01)00884838054196(21)357, (01)00884838054196(21)213, (01)00884838054196(21)293, (01)00884838054196(21)292, (01)00884838054196(21)346, (01)00884838054196(21)211, (01)00884838054196(21)144, (01)00884838054196(21)238, (01)00884838054196(21)247, (01)00884838054196(21)347, (01)00884838054196(21)188, (01)00884838054196(21)196, (01)00884838054196(21)339, (01)00884838054196(21)343, (01)00884838054196(21)256, (01)00884838054196(21)338, (01)00884838054196(21)334, (01)00884838054196(21)294, (01)00884838054196(21)345, (01)00884838054196(21)172, (01)00884838054196(21)323, (01)00884838054196(21)324, (01)00884838054196(21)189, (01)00884838054196(21)212, (01)00884838054196(21)268, (01)00884838054196(21)356, (01)00884838054196(21)244, (01)00884838054196(21)254, (01)00884838054196(21)362, (01)00884838054196(21)271, (01)00884838054196(21)186, (01)00884838054196(21)259, (01)00884838054196(21)236, (01)00884838054196(21)201, (01)00884838054196(21)275, (01)00884838054196(21)277, (01)00884838054196(21)349, (01)00884838054196(21)350, (01)00884838054196(21)348, (01)00884838054196(21)273, (01)00884838054196(21)249, (01)00884838054196(21)149, (01)00884838054196(21)148, (01)00884838054196(21)316, (01)00884838054196(21)274, (01)00884838054196(21)205, (01)00884838054196(21)270, (01)00884838054196(21)223, (01)00884838054196(21)181, (01)00884838054196(21)221, (01)00884838054196(21)245, (01)00884838054196(21)352, (01)00884838054196(21)358, (01)00884838054196(21)361, (01)00884838054196(21)228, (01)00884838054196(21)142, (01)00884838054196(21)214, (01)00884838054196(21)299, (01)00884838054196(21)114, (01)00884838054196(21)204, (01)00884838054196(21)225, (01)00884838054196(21)297, (01)00884838054196(21)182, (01)00884838054196(21)177, (01)00884838054196(21)300, (01)00884838054196(21)130, (01)00884838054196(21)210, (01)00884838054196(21)261, (01)00884838054196(21)260, (01)00884838054196(21)322, (01)00884838054196(21)360, (01)00884838054196(21)310, (01)00884838054196(21)291, (01)00884838054196(21)180, (01)00884838054196(21)287, (01)00884838054196(21)264, (01)00884838054196(21)272, (01)00884838054196(21)207, (01)00884838054196(21)239, (01)00884838054196(21)308, (01)00884838054196(21)243, (01)00884838054196(21)242, (01)00884838054196(21)218, (01)00884838054196(21)197, (01)00884838054196(21)235, (01)00884838054196(21)278, (01)00884838054196(21)250, (01)00884838054196(21)227, (01)00884838054196(21)233, (01)00884838054196(21)329, (01)00884838054196(21)296, (01)00884838054196(21)193, (01)00884838054196(21)219, (01)00884838054196(21)276, (01)00884838054196(21)326, (01)00884838054196(21)312, (01)00884838054196(21)304, (01)00884838054196(21)311, (01)00884838054196(21)199, (01)00884838054196(21)192, (01)00884838054196(21)289, (01)00884838054196(21)217, (01)00884838054196(21)355, (01)00884838054196(21)232, (01)00884838054196(21)315, (01)00884838054196(21)309, (01)00884838054196(21)171, (01)00884838054196(21)191, (01)00884838054196(21)269, (01)00884838054196(21)206, (01)00884838054196(21)330, (01)00884838054196(21)195, (01)00884838054196(21)215, (01)00884838054196(21)220, (01)00884838054196(21)198, (01)00884838054196(21)208, (01)00884838054196(21)286; 9. Allura Xper FD20; Model: 722006; UDI-DI: N/A; Serial Numbers: 2051, 1936, 518, 326, 66, 2587, 332, 249, 2274, 1496, 2183, 579, 1044, 2336, 443, 1913, 1259, 79, 939, 283, 405, 2161, 2068, 560, 1747, 1601, 479, 970, 1586, 878, 2537, 658, 384, 1469, 588, 328, 235, 381, 1536, 2504, 103, 826, 962, 169, 1197, 223, 595, 934, 1364, 972, 2567, 2042, 2380, 828, 528, 288, 1889, 1578, 1000, 942, 104, 2257, 1977, 1594, 118, 2003, 226, 1978, 333, 52, 53, 1242, 203, 521, 2427, 191, 1611, 1287, 250, 1582, 301, 1615, 1007, 01H041985, 2301, 1276, 234, 655, 914, 1302, 767, 781, 1038, 1306, 346, 8, 1524, 598, 639, 434, 687, 1580, 157, 491, 393, 621, 63, 2465, 2206, 585, 2487, 608, 1793, 884, 955, 1529, 1678, 1106, 946, 1650, 408, 1893, 380, 873, 102, 1232, 430, 1990, 1450, 957, 507, 1192, 136, 940, 274, 581, 2592, 2379, 890, 602, 280, 2493, 640, 2, 511; 10. Allura Xper FD20; Model: 722012; UDI: (01)00884838059054(21)2942, (01)00884838059054(21)2265, (01)00884838059054(21)2735, (01)00884838059054(21)2707, (01)00884838059054(21)135, (01)00884838059054(21)2740, (01)00884838059054(21)2843, (01)00884838059054(21)2835, (01)00884838059054(21)760, (01)00884838059054(21)735, (01)00884838059054(21)174, (01)00884838059054(21)2855, (01)00884838059054(21)2856, (01)00884838059054(21)2712, (01)00884838059054(21)606, (01)00884838059054(21)2721, (01)00884838059054(21)2085, (01)00884838059054(21)695, (01)00884838059054(21)2202, (01)00884838059054(21)2828, (01)00884838059054(21)601, (01)00884838059054(21)1207, (01)00884838059054(21)2898, (01)00884838059054(21)334, (01)00884838059054(21)1656, (01)00884838059054(21)483, (01)00884838059054(21)2305, (01)00884838059054(21)2896, (01)00884838059054(21)2726, (01)00884838059054(21)2722, (01)00884838059054(21)451, (01)00884838059054(21)2883, (01)00884838059054(21)2755, (01)00884838059054(21)2741, (01)00884838059054(21)1301, (01)00884838059054(21)2736, (01)00884838059054(21)782, (01)00884838059054(21)2890, (01)00884838059054(21)2829, (01)00884838059054(21)1675, (01)00884838059054(21)412, (01)00884838059054(21)320, (01)00884838059054(21)1224, (01)00884838059054(21)453, (01)00884838059054(21)1964, (01)00884838059054(21)325, (01)00884838059054(21)2742, (01)00884838059054(21)2889, (01)00884838059054(21)2713, (01)00884838059054(21)2899, (01)00884838059054(21)1272, (01)00884838059054(21)2857, (01)00884838059054(21)2886, (01)00884838059054(21)100, (01)00884838059054(21)2875, (01)00884838059054(21)2888, (01)00884838059054(21)71, (01)00884838059054(21)1807, (01)00884838059054(21)418, (01)00884838059054(21)1329, (01)00884838059054(21)2873, (01)00884838059054(21)1075, (01)00884838059054(21)2842, (01)00884838059054(21)2841, (01)00884838059054(21)2849, (01)00884838059054(21)2840, (01)00884838059054(21)2901, (01)00884838059054(21)1226, (01)00884838059054(21)2774, (01)00884838059054(21)1441, (01)00884838059054(21)2723, (01)00884838059054(21)2900, (01)00884838059054(21)316, (01)00884838059054(21)2892, (01)00884838059054(21)527, (01)00884838059054(21)2876, (01)00884838059054(21)2881, (01)00884838059054(21)525, (01)00884838059054(21)1806, (01)00884838059054(21)2844, (01)00884838059054(21)1712, (01)00884838059054(21)2760, (01)00884838059054(21)2868, (01)00884838059054(21)2869, (01)00884838059054(21)718, (01)00884838059054(21)1282, (01)00884838059054(21)1946, (01)00884838059054(21)2912, (01)00884838059054(21)1622, (01)00884838059054(21)2874, (01)00884838059054(21)351, (01)00884838059054(21)1655, (01)00884838059054(21)2282, (01)00884838059054(21)2891, (01)00884838059054(21)2764, (01)00884838059054(21)2880, (01)00884838059054(21)1171, (01)00884838059054(21)1921, (01)00884838059054(21)2838, (01)00884838059054(21)2887, (01)00884838059054(21)2567, (01)00884838059054(21)2703, (01)00884838059054(21)2759, (01)00884838059054(21)2945, (01)00884838059054(21)1747, (01)00884838059054(21)2879, (01)00884838059054(21)2335, (01)00884838059054(21)2913, (01)00884838059054(21)537, (01)00884838059054(21)2897, (01)00884838059054(21)288, (01)00884838059054(21)191, (01)00884838059054(21)2917, (01)00884838059054(21)2915, (01)00884838059054(21)2850, (01)00884838059054(21)2825, (01)00884838059054(21)2579, (01)00884838059054(21)2916, (01)00884838059054(21)2667, (01)00884838059054(21)2681, (01)00884838059054(21)2784, (01)00884838059054(21)2747, (01)00884838059054(21)2914, (01)00884838059054(21)1911, (01)00884838059054(21)2831, (01)00884838059054(21)2749, (01)00884838059054(21)2801, (01)00884838059054(21)2771, (01)00884838059054(21)2724, (01)00884838059054(21)2877, (01)00884838059054(21)2788, (01)00884838059054(21)2702, (01)00884838059054(21)465, (01)00884838059054(21)309, (01)00884838059054(21)2827, (01)00884838059054(21)2806, (01)00884838059054(21)1628, (01)00884838059054(21)2934, (01)00884838059054(21)2763, (01)00884838059054(21)2810, (01)00884838059054(21)2581, (01)00884838059054(21)2623, (01)00884838059054(21)2647, (01)00884838059054(21)2854, (01)00884838059054(21)2548, (01)00884838059054(21)2745, (01)00884838059054(21)2812, (01)00884838059054(21)2676, (01)00884838059054(21)2599, (01)00884838059054(21)2789, (01)00884838059054(21)2818, (01)00884838059054(21)2678, (01)00884838059054(21)2701, (01)00884838059054(21)2756, (01)00884838059054(21)1256, (01)00884838059054(21)2830, (01)00884838059054(21)982, (01)00884838059054(21)2565, (01)00884838059054(21)2752, (01)00884838059054(21)2811, (01)00884838059054(21)2852, (01)00884838059054(21)2671, (01)00884838059054(21)2649, (01)00884838059054(21)2698, (01)00884838059054(21)2661, (01)00884838059054(21)2690, (01)00884838059054(21)2775, (01)00884838059054(21)2773, (01)00884838059054(21)2834, (01)00884838059054(21)2720, (01)00884838059054(21)2626, (01)00884838059054(21)2823, (01)00884838059054(21)374, (01)00884838059054(21)2782, (01)00884838059054(21)2766, (01)00884838059054(21)2776, (01)00884838059054(21)2826, (01)00884838059054(21)2738, (01)00884838059054(21)2851, (01)00884838059054(21)2777, (01)00884838059054(21)2739, (01)00884838059054(21)2847, (01)00884838059054(21)2832, (01)00884838059054(21)2725, (01)00884838059054(21)2650, (01)00884838059054(21)2813, (01)00884838059054(21)998, (01)00884838059054(21)2744, (01)00884838059054(21)2797, (01)00884838059054(21)2229, (01)00884838059054(21)2228, (01)00884838059054(21)2781, (01)00884838059054(21)2853, (01)00884838059054(21)2716, (01)00884838059054(21)2808, (01)00884838059054(21)2837, (01)00884838059054(21)2768, (01)00884838059054(21)2673, (01)00884838059054(21)2674, (01)00884838059054(21)2803, (01)00884838059054(21)2732, (01)00884838059054(21)2798, (01)00884838059054(21)2786, (01)00884838059054(21)2778, (01)00884838059054(21)2559, (01)00884838059054(21)2746, (01)00884838059054(21)2770, (01)00884838059054(21)2691, (01)00884838059054(21)2762, (01)00884838059054(21)2729, (01)00884838059054(21)2750, (01)00884838059054(21)2804, (01)00884838059054(21)2751, (01)00884838059054(21)2640, (01)00884838059054(21)2730, (01)00884838059054(21)2562, (01)00884838059054(21)2795, (01)00884838059054(21)2718, (01)00884838059054(21)2682, (01)00884838059054(21)2574, (01)00884838059054(21)2779, (01)00884838059054(21)2711, (01)00884838059054(21)2758, (01)00884838059054(21)2836, (01)00884838059054(21)2733, (01)00884838059054(21)2824, (01)00884838059054(21)2715, (01)00884838059054(21)2704, (01)00884838059054(21)2734, (01)00884838059054(21)2793, (01)00884838059054(21)2792, (01)00884838059054(21)2566, (01)00884838059054(21)2757, (01)00884838059054(21)2765, (01)00884838059054(21)2785, (01)00884838059054(21)2767, (01)00884838059054(21)2794, (01)00884838059054(21)240, (01)00884838059054(21)2719, (01)00884838059054(21)2708, (01)00884838059054(21)2882, (01)00884838059054(21)2169, (01)00884838059054(21)2802, (01)00884838059054(21)2761, (01)00884838059054(21)2799, (01)00884838059054(21)2652, (01)00884838059054(21)2743, (01)00884838059054(21)2790, (01)00884838059054(21)2748, (01)00884838059054(21)2911, (01)00884838059054(21)2714, (01)00884838059054(21)2688, (01)00884838059054(21)2780, (01)00884838059054(21)2791, (01)00884838059054(21)1547, (01)00884838059054(21)2772, (01)00884838059054(21)657; 11. Allura Xper FD20; Model: 722028; UDI: (01)00884838054202(21)1756, (01)00884838054202(21)1569, (01)00884838054202(21)1956, (01)00884838054202(21)2206, (01)00884838054202(21)1233, (01)00884838054202(21)928, (01)00884838054202(21)2502, (01)00884838054202(21)1774, (01)00884838054202(21)869, (01)00884838054202(21)1017, (01)00884838054202(21)1957, (01)00884838054202(21)2261, (01)00884838054202(21)866, (01)00884838054202(21)1310, (01)00884838054202(21)1043, (01)00884838054202(21)1077, (01)00884838054202(21)995, (01)00884838054202(21)896, (01)00884838054202(21)877, (01)00884838054202(21)1439, (01)00884838054202(21)1343, (01)00884838054202(21)895, (01)00884838054202(21)1154, (01)00884838054202(21)1000, (01)00884838054202(21)1909, (01)00884838054202(21)1716, (01)00884838054202(21)1983, (01)00884838054202(21)2061, (01)00884838054202(21)1393, (01)00884838054202(21)1656, (01)00884838054202(21)1461, (01)00884838054202(21)1751, (01)00884838054202(21)1050, (01)00884838054202(21)1428, (01)00884838054202(21)1094, (01)00884838054202(21)2173, (01)00884838054202(21)1214, (01)00884838054202(21)52, (01)00884838054202(21)994, (01)00884838054202(21)1213, (01)00884838054202(21)1836, (01)00884838054202(21)982, (01)00884838054202(21)1646, (01)00884838054202(21)1396, (01)00884838054202(21)1394, (01)00884838054202(21)1650, (01)00884838054202(21)2275, (01)00884838054202(21)1058, (01)00884838054202(21)1353, (01)00884838054202(21)1466, (01)00884838054202(21)1340, (01)00884838054202(21)1978, (01)00884838054202(21)431, (01)00884838054202(21)1211, (01)00884838054202(21)1651, (01)00884838054202(21)2710, (01)00884838054202(21)1326, (01)00884838054202(21)759, (01)00884838054202(21)1232, (01)00884838054202(21)925, (01)00884838054202(21)961, (01)00884838054202(21)1736, (01)00884838054202(21)1619, (01)00884838054202(21)1567, (01)00884838054202(21)1130, (01)00884838054202(21)1570, (01)00884838054202(21)1100, (01)00884838054202(21)1385, (01)00884838054202(21)1267, (01)00884838054202(21)2103, (01)00884838054202(21)1395, (01)00884838054202(21)1408, (01)00884838054202(21)741, (01)00884838054202(21)1269, (01)00884838054202(21)1837, (01)00884838054202(21)1988, (01)00884838054202(21)1051, (01)00884838054202(21)854, (01)00884838054202(21)956, (01)00884838054202(21)2020, (01)00884838054202(21)2601, (01)00884838054202(21)1261, (01)00884838054202(21)1835, (01)00884838054202(21)967, (01)00884838054202(21)1483, (01)00884838054202(21)2745, (01)00884838054202(21)2249, (01)00884838054202(21)1311, (01)00884838054202(21)1103, (01)00884838054202(21)2645, (01)00884838054202(21)678, (01)00884838054202(21)1116, (01)00884838054202(21)1880, (01)00884838054202(21)911, (01)00884838054202(21)2192, (01)00884838054202(21)1171, (01)00884838054202(21)1498, (01)00884838054202(21)1156, (01)00884838054202(21)1279, (01)00884838054202(21)1708, (01)00884838054202(21)1710, (01)00884838054202(21)664, (01)00884838054202(21)906, (01)00884838054202(21)1155, (01)00884838054202(21)1085, (01)00884838054202(21)1946, (01)00884838054202(21)1834, (01)00884838054202(21)1652, (01)00884838054202(21)1011, (01)00884838054202(21)1508, (01)00884838054202(21)2293, (01)00884838054202(21)1890, (01)00884838054202(21)654, (01)00884838054202(21)1484, (01)00884838054202(21)1995, (01)00884838054202(21)1456, (01)00884838054202(21)1288, (01)00884838054202(21)1556, (01)00884838054202(21)1201, (01)00884838054202(21)1129, (01)00884838054202(21)976, (01)00884838054202(21)2641, (01)00884838054202(21)2348, (01)00884838054202(21)922, (01)00884838054202(21)973, (01)00884838054202(21)1068, (01)00884838054202(21)1069, (01)00884838054202(21)1931, (01)00884838054202(21)2253, (01)00884838054202(21)1870, (01)00884838054202(21)1095, (01)00884838054202(21)892, (01)00884838054202(21)2295, (01)00884838054202(21)2071, (01)00884838054202(21)1327, (01)00884838054202(21)1099, (01)00884838054202(21)1407, (01)00884838054202(21)1920, (01)00884838054202(21)2709, (01)00884838054202(21)2707, (01)00884838054202(21)2174, (01)00884838054202(21)1275, (01)00884838054202(21)1142, (01)00884838054202(21)1111, (01)00884838054202(21)1122, (01)00884838054202(21)2151, (01)00884838054202(21)1352, (01)00884838054202(21)888, (01)00884838054202(21)1755, (01)00884838054202(21)1855, (01)00884838054202(21)842, (01)00884838054202(21)840, (01)00884838054202(21)2310, (01)00884838054202(21)655, (01)00884838054202(21)2007, (01)00884838054202(21)2008, (01)00884838054202(21)968, (01)00884838054202(21)781, (01)00884838054202(21)774, (01)00884838054202(21)1850, (01)00884838054202(21)1522, (01)00884838054202(21)1448, (01)00884838054202(21)1413, (01)00884838054202(21)1538, (01)00884838054202(21)1930, (01)00884838054202(21)535, (01)00884838054202(21)938, (01)00884838054202(21)990, (01)00884838054202(21)1081, (01)00884838054202(21)808, (01)00884838054202(21)1634, (01)00884838054202(21)1387, (01)00884838054202(21)1630, (01)00884838054202(21)989, (01)00884838054202(21)1824, (01)00884838054202(21)1481, (01)00884838054202(21)887, (01)00884838054202(21)1202, (01)00884838054202(21)2407, (01)00884838054202(21)2635, (01)00884838054202(21)1809, (01)00884838054202(21)874, (01)00884838054202(21)1049, (01)00884838054202(21)1333, (01)00884838054202(21)1264, (01)00884838054202(21)981, (01)00884838054202(21)1447, (01)00884838054202(21)1189, (01)00884838054202(21)1188, (01)00884838054202(21)696, (01)00884838054202(21)966, (01)00884838054202(21)1875, (01)00884838054202(21)1482, (01)00884838054202(21)1405, (01)00884838054202(21)930, (01)00884838054202(21)932, (01)00884838054202(21)1001, (01)00884838054202(21)1041, (01)00884838054202(21)1054, (01)00884838054202(21)1032, (01)00884838054202(21)2252, (01)00884838054202(21)1472, (01)00884838054202(21)1266, (01)00884838054202(21)917, (01)00884838054202(21)2070, (01)00884838054202(21)971, (01)00884838054202(21)1014, (01)00884838054202(21)1200, (01)00884838054202(21)1374, (01)00884838054202(21)1499, (01)00884838054202(21)1222, (01)00884838054202(21)1705, (01)00884838054202(21)1514, (01)00884838054202(21)2110, (01)00884838054202(21)867, (01)00884838054202(21)2200, (01)00884838054202(21)1783, (01)00884838054202(21)1723, (01)00884838054202(21)1350, (01)00884838054202(21)1454, (01)00884838054202(21)1098, (01)00884838054202(21)983, (01)00884838054202(21)1856, (01)00884838054202(21)978, (01)00884838054202(21)1132, (01)00884838054202(21)1187, (01)00884838054202(21)1889, (01)00884838054202(21)1076, (01)00884838054202(21)952, (01)00884838054202(21)1441, (01)00884838054202(21)1457, (01)00884838054202(21)2284, (01)00884838054202(21)1424, (01)00884838054202(21)893, (01)00884838054202(21)879, (01)00884838054202(21)1854, (01)00884838054202(21)934, (01)00884838054202(21)924, (01)00884838054202(21)1004, (01)00884838054202(21)1373, (01)00884838054202(21)1846, (01)00884838054202(21)962, (01)00884838054202(21)1984, (01)00884838054202(21)2422, (01)00884838054202(21)1949, (01)00884838054202(21)918, (01)00884838054202(21)998, (01)00884838054202(21)1463, (01)00884838054202(21)1212, (01)00884838054202(21)1191, (01)00884838054202(21)1370, (01)00884838054202(21)1002, (01)00884838054202(21)2760, (01)00884838054202(21)1600, (01)00884838054202(21)1244, (01)00884838054202(21)1030, (01)00884838054202(21)1869, (01)00884838054202(21)2196, (01)00884838054202(21)1286, (01)00884838054202(21)1828, (01)00884838054202(21)1446, (01)00884838054202(21)901, (01)00884838054202(21)1351, (01)00884838054202(21)1349, (01)00884838054202(21)1074, (01)00884838054202(21)1677, (01)00884838054202(21)908, (01)00884838054202(21)1629, (01)00884838054202(21)991, (01)00884838054202(21)422, (01)00884838054202(21)1822, (01)00884838054202(21)900, (01)00884838054202(21)914, (01)00884838054202(21)882, (01)00884838054202(21)945, (01)00884838054202(21)1136, (01)00884838054202(21)1639, (01)00884838054202(21)909, (01)00884838054202(21)2083, (01)00884838054202(21)1820, (01)00884838054202(21)1331, (01)00884838054202(21)1566, (01)00884838054202(21)1598, (01)00884838054202(21)1826, (01)00884838054202(21)1368, (01)00884838054202(21)1649, (01)00884838054202(21)1549, (01)00884838054202(21)1845, (01)00884838054202(21)1980, (01)00884838054202(21)1568, (01)00884838054202(21)929, (01)00884838054202(21)861, (01)00884838054202(21)1035, (01)00884838054202(21)1256, (01)00884838054202(21)1024, (01)00884838054202(21)428, (01)00884838054202(21)1022, (01)00884838054202(21)1811, (01)00884838054202(21)1359, (01)00884838054202(21)1945, (01)00884838054202(21)1168, (01)00884838054202(21)940, (01)00884838054202(21)1657, (01)00884838054202(21)878, (01)00884838054202(21)1572, (01)00884838054202(21)905, (01)00884838054202(21)1039, (01)00884838054202(21)942, (01)00884838054202(21)1056, (01)00884838054202(21)2347, (01)00884838054202(21)1271, (01)00884838054202(21)1363, (01)00884838054202(21)2322, (01)00884838054202(21)2589, (01)00884838054202(21)1193, (01)00884838054202(21)1873, (01)00884838054202(21)1915, (01)00884838054202(21)1280, (01)00884838054202(21)1255, (01)00884838054202(21)1830, (01)00884838054202(21)1814, (01)00884838054202(21)1390, (01)00884838054202(21)713, (01)00884838054202(21)1583, (01)00884838054202(21)1709, (01)00884838054202(21)2194, (01)00884838054202(21)2273, (01)00884838054202(21)1718, (01)00884838054202(21)1914, (01)00884838054202(21)935, (01)00884838054202(21)2545, (01)00884838054202(21)2048, (01)00884838054202(21)884, (01)00884838054202(21)1400, (01)00884838054202(21)2314, (01)00884838054202(21)1329, (01)00884838054202(21)1462, (01)00884838054202(21)1235, (01)00884838054202(21)1021, (01)00884838054202(21)1416, (01)00884838054202(21)2475, (01)00884838054202(21)2021, (01)00884838054202(21)1295, (01)00884838054202(21)1844, (01)00884838054202(21)2155, (01)00884838054202(21)1810, (01)00884838054202(21)1635, (01)00884838054202(21)1636, (01)00884838054202(21)2178, (01)00884838054202(21)2002, (01)00884838054202(21)1726, (01)00884838054202(21)2421, (01)00884838054202(21)777, (01)00884838054202(21)1332, (01)00884838054202(21)972, (01)00884838054202(21)1911, (01)00884838054202(21)2172, (01)00884838054202(21)2093, (01)00884838054202(21)2350, (01)00884838054202(21)2352, (01)00884838054202(21)2098, (01)00884838054202(21)1786, (01)00884838054202(21)2568, (01)00884838054202(21)1165, (01)00884838054202(21)1955, (01)00884838054202(21)2482, (01)00884838054202(21)1321, (01)00884838054202(21)1877, (01)00884838054202(21)1571, (01)00884838054202(21)1551, (01)00884838054202(21)812, (01)00884838054202(21)1979, (01)00884838054202(21)1287, (01)00884838054202(21)1247, (01)00884838054202(21)1857, (01)00884838054202(21)2489, (01)00884838054202(21)511, (01)00884838054202(21)1061, (01)00884838054202(21)964, (01)00884838054202(21)1006, (01)00884838054202(21)997, (01)00884838054202(21)1242, (01)00884838054202(21)1682, (01)00884838054202(21)1060, (01)00884838054202(21)1059, (01)00884838054202(21)2427, (01)00884838054202(21)1325, (01)00884838054202(21)1173, (01)00884838054202(21)1314, (01)00884838054202(21)1157, (01)00884838054202(21)1861, (01)00884838054202(21)2538, (01)00884838054202(21)2090, (01)00884838054202(21)1831, (01)00884838054202(21)2547, (01)00884838054202(21)1199, (01)00884838054202(21)2594, (01)00884838054202(21)1924, (01)00884838054202(21)1821, (01)00884838054202(21)2230, (01)00884838054202(21)2232, (01)00884838054202(21)2025, (01)00884838054202(21)2107, (01)00884838054202(21)1621, (01)00884838054202(21)1540, (01)00884838054202(21)1120, (01)00884838054202(21)2532, (01)00884838054202(21)2212, (01)00884838054202(21)1679, (01)00884838054202(21)2086, (01)00884838054202(21)1789, (01)00884838054202(21)2364, (01)00884838054202(21)1688, (01)00884838054202(21)963, (01)00884838054202(21)1878, (01)00884838054202(21)1866, (01)00884838054202(21)1562, (01)00884838054202(21)937, (01)00884838054202(21)1421, (01)00884838054202(21)1604, (01)00884838054202(21)1713, (01)00884838054202(21)1389, (01)00884838054202(21)980, (01)00884838054202(21)986, (01)00884838054202(21)1133, (01)00884838054202(21)1893, (01)00884838054202(21)2217, (01)00884838054202(21)881, (01)00884838054202(21)2372, (01)00884838054202(21)1112, (01)00884838054202(21)2351, (01)00884838054202(21)2047, (01)00884838054202(21)2481, (01)00884838054202(21)2483, (01)00884838054202(21)1328, (01)00884838054202(21)2095, (01)00884838054202(21)1695, (01)00884838054202(21)1694, (01)00884838054202(21)1007, (01)00884838054202(21)794, (01)00884838054202(21)1148, (01)00884838054202(21)1293, (01)00884838054202(21)965, (01)00884838054202(21)1259, (01)00884838054202(21)2334, (01)00884838054202(21)1775, (01)00884838054202(21)1460, (01)00884838054202(21)1819, (01)00884838054202(21)1865, (01)00884838054202(21)984, (01)00884838054202(21)1952, (01)00884838054202(21)1815, (01)00884838054202(21)912, (01)00884838054202(21)993, (01)00884838054202(21)1386, (01)00884838054202(21)1925, (01)00884838054202(21)1186, (01)00884838054202(21)1730, (01)00884838054202(21)894, (01)00884838054202(21)2066, (01)00884838054202(21)1720, (01)00884838054202(21)846, (01)00884838054202(21)844, (01)00884838054202(21)845, (01)00884838054202(21)1895, (01)00884838054202(21)1458, (01)00884838054202(21)2079, (01)00884838054202(21)1493, (01)00884838054202(21)1382, (01)00884838054202(21)813, (01)00884838054202(21)1529, (01)00884838054202(21)2028, (01)00884838054202(21)2041, (01)00884838054202(21)1218, (01)00884838054202(21)1313, (01)00884838054202(21)1254, (01)00884838054202(21)1205, (01)00884838054202(21)1637, (01)00884838054202(21)2408, (01)00884838054202(21)1517, (01)00884838054202(21)2024, (01)00884838054202(21)1178, (01)00884838054202(21)1152, (01)00884838054202(21)2168, (01)00884838054202(21)2143, (01)00884838054202(21)1079, (01)00884838054202(21)1082, (01)00884838054202(21)1172, (01)00884838054202(21)1225, (01)00884838054202(21)1527, (01)00884838054202(21)2187, (01)00884838054202(21)2379, (01)00884838054202(21)2152, (01)00884838054202(21)1480, (01)00884838054202(21)1347, (01)00884838054202(21)1046, (01)00884838054202(21)2570, (01)00884838054202(21)2255, (01)00884838054202(21)1719, (01)00884838054202(21)2374, (01)00884838054202(21)2555, (01)00884838054202(21)1437, (01)00884838054202(21)2362, (01)00884838054202(21)1645, (01)00884838054202(21)2258, (01)00884838054202(21)1008, (01)00884838054202(21)1843, (01)00884838054202(21)728, (01)00884838054202(21)775, (01)00884838054202(21)1578, (01)00884838054202(21)1579, (01)00884838054202(21)1947, (01)00884838054202(21)2400, (01)00884838054202(21)1881, (01)00884838054202(21)2455, (01)00884838054202(21)2320, (01)00884838054202(21)2690, (01)00884838054202(21)2527, (01)00884838054202(21)2222, (01)00884838054202(21)1812, (01)00884838054202(21)2019, (01)00884838054202(21)1559, (01)00884838054202(21)1632, (01)00884838054202(21)2513, (01)00884838054202(21)939, (01)00884838054202(21)2328, (01)00884838054202(21)1391, (01)00884838054202(21)2126, (01)00884838054202(21)1953, (01)00884838054202(21)1558, (01)00884838054202(21)1070, (01)00884838054202(21)1781, (01)00884838054202(21)1089, (01)00884838054202(21)2383, (01)00884838054202(21)2000, (01)00884838054202(21)1693, (01)00884838054202(21)899, (01)00884838054202(21)1797, (01)00884838054202(21)1863, (01)00884838054202(21)2500, (01)00884838054202(21)1066, (01)00884838054202(21)2302, (01)00884838054202(21)1587, (01)00884838054202(21)1128, (01)00884838054202(21)2132, (01)00884838054202(21)2112, (01)00884838054202(21)1018, (01)00884838054202(21)1113, (01)00884838054202(21)1734, (01)00884838054202(21)1994, (01)00884838054202(21)1769, (01)00884838054202(21)1813, (01)00884838054202(21)2259, (01)00884838054202(21)1418, (01)00884838054202(21)1545, (01)00884838054202(21)2344, (01)00884838054202(21)2262, (01)00884838054202(21)1505, (01)00884838054202(21)933, (01)00884838054202(21)955, (01)00884838054202(21)1940, (01)00884838054202(21)1506, (01)00884838054202(21)1364, (01)00884838054202(21)1285, (01)00884838054202(21)1219, (01)00884838054202(21)1766, (01)00884838054202(21)1354, (01)00884838054202(21)1699, (01)00884838054202(21)2305, (01)00884838054202(21)2304, (01)00884838054202(21)2306, (01)00884838054202(21)2461, (01)00884838054202(21)1586, (01)00884838054202(21)2417, (01)00884838054202(21)2245, (01)00884838054202(21)1959, (01)00884838054202(21)2115, (01)00884838054202(21)2140, (01)00884838054202(21)2298, (01)00884838054202(21)1449, (01)00884838054202(21)2089, (01)00884838054202(21)1776, (01)00884838054202(21)1932, (01)00884838054202(21)2085, (01)00884838054202(21)954, (01)00884838054202(21)1125, (01)00884838054202(21)1317, (01)00884838054202(21)2237, (01)00884838054202(21)1459, (01)00884838054202(21)2195, (01)00884838054202(21)1894, (01)00884838054202(21)1337, (01)00884838054202(21)1433, (01)00884838054202(21)1515, (01)00884838054202(21)2292, (01)00884838054202(21)1612, (01)00884838054202(21)2614, (01)00884838054202(21)2058, (01)00884838054202(21)2038, (01)00884838054202(21)1597, (01)00884838054202(21)1601, (01)00884838054202(21)1246, (01)00884838054202(21)2391, (01)00884838054202(21)2311, (01)00884838054202(21)2087, (01)00884838054202(21)2001, (01)00884838054202(21)1294, (01)00884838054202(21)2045, (01)00884838054202(21)1848, (01)00884838054202(21)2553, (01)00884838054202(21)1477, (01)00884838054202(21)2511, (01)00884838054202(21)2177, (01)00884838054202(21)1605, (01)00884838054202(21)1616, (01)00884838054202(21)1141, (01)00884838054202(21)1231, (01)00884838054202(21)1610, (01)00884838054202(21)1618, (01)00884838054202(21)1417, (01)00884838054202(21)1479, (01)00884838054202(21)1626, (01)00884838054202(21)970, (01)00884838054202(21)1582, (01)00884838054202(21)1092, (01)00884838054202(21)1617, (01)00884838054202(21)921, (01)00884838054202(21)1404, (01)00884838054202(21)2426, (01)00884838054202(21)1525, (01)00884838054202(21)1399, (01)00884838054202(21)1608, (01)00884838054202(21)2059, (01)00884838054202(21)2603, (01)00884838054202(21)944, (01)00884838054202(21)1743, (01)00884838054202(21)2247, (01)00884838054202(21)2623, (01)00884838054202(21)985, (01)00884838054202(21)2153, (01)00884838054202(21)1210, (01)00884838054202(21)1487, (01)00884838054202(21)1488, (01)00884838054202(21)676, (01)00884838054202(21)1584, (01)00884838054202(21)1071, (01)00884838054202(21)1063, (01)00884838054202(21)1040, (01)00884838054202(21)2105, (01)00884838054202(21)1907, (01)00884838054202(21)1207, (01)00884838054202(21)1307, (01)00884838054202(21)2145, (01)00884838054202(21)1735, (01)00884838054202(21)2246, (01)00884838054202(21)1038, (01)00884838054202(21)1228, (01)00884838054202(21)2250, (01)00884838054202(21)1180, (01)00884838054202(21)1369, (01)00884838054202(21)1383, (01)00884838054202(21)2135, (01)00884838054202(21)2133, (01)00884838054202(21)857, (01)00884838054202(21)1411, (01)00884838054202(21)1717, (01)00884838054202(21)2444, (01)00884838054202(21)1181, (01)00884838054202(21)1029, (01)00884838054202(21)2248, (01)00884838054202(21)1384, (01)00884838054202(21)2324, (01)00884838054202(21)2355, (01)00884838054202(21)1898, (01)00884838054202(21)1042, (01)00884838054202(21)1388, (01)00884838054202(21)2534, (01)00884838054202(21)2460, (01)00884838054202(21)1027, (01)00884838054202(21)1622, (01)00884838054202(21)2121, (01)00884838054202(21)2416, (01)00884838054202(21)1576, (01)00884838054202(21)2433, (01)00884838054202(21)1580, (01)00884838054202(21)1047, (01)00884838054202(21)1902, (01)00884838054202(21)1577, (01)00884838054202(21)2215, (01)00884838054202(21)843, (01)00884838054202(21)2491, (01)00884838054202(21)820, (01)00884838054202(21)1065, (01)00884838054202(21)1624, (01)00884838054202(21)1623, (01)00884838054202(21)1425, (01)00884838054202(21)1687, (01)00884838054202(21)2084, (01)00884838054202(21)2218, (01)00884838054202(21)1772, (01)00884838054202(21)2043, (01)00884838054202(21)931, (01)00884838054202(21)2602, (01)00884838054202(21)2576, (01)00884838054202(21)1117, (01)00884838054202(21)1724, (01)00884838054202(21)1534, (01)00884838054202(21)2388, (01)00884838054202(21)2438, (01)00884838054202(21)2033, (01)00884838054202(21)1768, (01)00884838054202(21)2157, (01)00884838054202(21)1300, (01)00884838054202(21)2197, (01)00884838054202(21)2319, (01)00884838054202(21)2097, (01)00884838054202(21)2139, (01)00884838054202(21)2012, (01)00884838054202(21)838, (01)00884838054202(21)1511, (01)00884838054202(21)2630, (01)00884838054202(21)2291, (01)00884838054202(21)1969, (01)00884838054202(21)2312, (01)00884838054202(21)837, (01)00884838054202(21)1251, (01)00884838054202(21)2081, (01)00884838054202(21)2111, (01)00884838054202(21)2684, (01)00884838054202(21)1341, (01)00884838054202(21)2016, (01)00884838054202(21)2186, (01)00884838054202(21)2034, (01)00884838054202(21)1357, (01)00884838054202(21)1520, (01)00884838054202(21)1375, (01)00884838054202(21)2415, (01)00884838054202(21)1971, (01)00884838054202(21)1553, (01)00884838054202(21)1422, (01)00884838054202(21)2470, (01)00884838054202(21)1887, (01)00884838054202(21)2243, (01)00884838054202(21)1489, (01)00884838054202(21)2080, (01)00884838054202(21)1166, (01)00884838054202(21)1127, (01)00884838054202(21)2345, (01)00884838054202(21)1832, (01)00884838054202(21)2368, (01)00884838054202(21)2404, (01)00884838054202(21)1324, (01)00884838054202(21)1728, (01)00884838054202(21)1490, (01)00884838054202(21)2113, (01)00884838054202(21)1163, (01)00884838054202(21)2473, (01)00884838054202(21)2349, (01)00884838054202(21)2528, (01)00884838054202(21)1588, (01)00884838054202(21)2209, (01)00884838054202(21)2723, (01)00884838054202(21)2318, (01)00884838054202(21)1107, (01)00884838054202(21)2091, (01)00884838054202(21)2327, (01)00884838054202(21)1516, (01)00884838054202(21)988, (01)00884838054202(21)2221, (01)00884838054202(21)1316, (01)00884838054202(21)1732, (01)00884838054202(21)1662, (01)00884838054202(21)915, (01)00884838054202(21)948, (01)00884838054202(21)1970, (01)00884838054202(21)1496, (01)00884838054202(21)2117, (01)00884838054202(21)1026, (01)00884838054202(21)1973, (01)00884838054202(21)2487, (01)00884838054202(21)1252, (01)00884838054202(21)2515, (01)00884838054202(21)2063, (01)00884838054202(21)2480, (01)00884838054202(21)2591, (01)00884838054202(21)2213, (01)00884838054202(21)1900, (01)00884838054202(21)1753, (01)00884838054202(21)2296, (01)00884838054202(21)1376, (01)00884838054202(21)2051, (01)00884838054202(21)1805, (01)00884838054202(21)2120, (01)00884838054202(21)1939, (01)00884838054202(21)1891, (01)00884838054202(21)1745, (01)00884838054202(21)2183, (01)00884838054202(21)1478, (01)00884838054202(21)1543, (01)00884838054202(21)2129, (01)00884838054202(21)2216, (01)00884838054202(21)1410, (01)00884838054202(21)876, (01)00884838054202(21)1747, (01)00884838054202(21)2108, (01)00884838054202(21)2339, (01)00884838054202(21)2036, (01)00884838054202(21)2100, (01)00884838054202(21)1431, (01)00884838054202(21)2044, (01)00884838054202(21)2405, (01)00884838054202(21)1185, (01)00884838054202(21)1647, (01)00884838054202(21)735, (01)00884838054202(21)737, (01)00884838054202(21)2631, (01)00884838054202(21)1470, (01)00884838054202(21)2271, (01)00884838054202(21)2381, (01)00884838054202(21)1537, (01)00884838054202(21)2193, (01)00884838054202(21)1090, (01)00884838054202(21)1839, (01)00884838054202(21)1497, (01)00884838054202(21)2227, (01)00884838054202(21)2557, (01)00884838054202(21)1208, (01)00884838054202(21)2130, (01)00884838054202(21)2477, (01)00884838054202(21)2476, (01)00884838054202(21)1987, (01)00884838054202(21)2018, (01)00884838054202(21)2127, (01)00884838054202(21)2223, (01)00884838054202(21)1194, (01)00884838054202(21)2064, (01)00884838054202(21)1760, (01)00884838054202(21)1298, (01)00884838054202(21)1564, (01)00884838054202(21)1589, (01)00884838054202(21)1471, (01)00884838054202(21)2279, (01)00884838054202(21)2188, (01)00884838054202(21)2042, (01)00884838054202(21)1806, (01)00884838054202(21)2366, (01)00884838054202(21)2367, (01)00884838054202(21)1055, (01)00884838054202(21)1474, (01)00884838054202(21)2220, (01)00884838054202(21)1096, (01)00884838054202(21)1581, (01)00884838054202(21)2387, (01)00884838054202(21)2536, (01)00884838054202(21)1336, (01)00884838054202(21)2359, (01)00884838054202(21)2363, (01)00884838054202(21)1561, (01)00884838054202(21)2032, (01)00884838054202(21)1057, (01)00884838054202(21)2125, (01)00884838054202(21)2464, (01)00884838054202(21)1770, (01)00884838054202(21)1048, (01)00884838054202(21)1315, (01)00884838054202(21)848, (01)00884838054202(21)2072, (01)00884838054202(21)1885, (01)00884838054202(21)2201, (01)00884838054202(21)1933, (01)00884838054202(21)1512, (01)00884838054202(21)1777, (01)00884838054202(21)2137, (01)00884838054202(21)2369, (01)00884838054202(21)2432, (01)00884838054202(21)2358, (01)00884838054202(21)2542, (01)00884838054202(21)2147, (01)00884838054202(21)2031, (01)00884838054202(21)1531, (01)00884838054202(21)2210, (01)00884838054202(21)2396, (01)00884838054202(21)1106, (01)00884838054202(21)1767, (01)00884838054202(21)1121, (01)00884838054202(21)1941, (01)00884838054202(21)2159, (01)00884838054202(21)2287, (01)00884838054202(21)2323, (01)00884838054202(21)1377, (01)00884838054202(21)1681, (01)00884838054202(21)1780, (01)00884838054202(21)1289, (01)00884838054202(21)1906, (01)00884838054202(21)1936, (01)00884838054202(21)2504, (01)00884838054202(21)2592, (01)00884838054202(21)2166, (01)00884838054202(21)2428, (01)00884838054202(21)2040, (01)00884838054202(21)2600, (01)00884838054202(21)2389, (01)00884838054202(21)2552, (01)00884838054202(21)2300, (01)00884838054202(21)2131, (01)00884838054202(21)1706, (01)00884838054202(21)2411, (01)00884838054202(21)1804, (01)00884838054202(21)1674, (01)00884838054202(21)1245, (01)00884838054202(21)1867, (01)00884838054202(21)1711, (01)00884838054202(21)2439, (01)00884838054202(21)1696, (01)00884838054202(21)2512, (01)00884838054202(21)2138, (01)00884838054202(21)2393, (01)00884838054202(21)1536, (01)00884838054202(21)2257, (01)00884838054202(21)1541, (01)00884838054202(21)1240, (01)00884838054202(21)1402, (01)00884838054202(21)815, (01)00884838054202(21)1990, (01)00884838054202(21)1138, (01)00884838054202(21)1801, (01)00884838054202(21)2392, (01)00884838054202(21)1229, (01)00884838054202(21)1903, (01)00884838054202(21)2265, (01)00884838054202(21)2234, (01)00884838054202(21)1882, (01)00884838054202(21)2285, (01)00884838054202(21)1803, (01)00884838054202(21)2397, (01)00884838054202(21)2114, (01)00884838054202(21)2068, (01)00884838054202(21)1998, (01)00884838054202(21)1764, (01)00884838054202(21)1217, (01)00884838054202(21)2436, (01)00884838054202(21)936, (01)00884838054202(21)2613, (01)00884838054202(21)2346, (01)00884838054202(21)2409, (01)00884838054202(21)1778, (01)00884838054202(21)2567, (01)00884838054202(21)2465, (01)00884838054202(21)1455, (01)00884838054202(21)1921, (01)00884838054202(21)1871, (01)00884838054202(21)1468, (01)00884838054202(21)1091, (01)00884838054202(21)2161, (01)00884838054202(21)2468, (01)00884838054202(21)2075, (01)00884838054202(21)2182, (01)00884838054202(21)2165, (01)00884838054202(21)2333, (01)00884838054202(21)2023, (01)00884838054202(21)1360, (01)00884838054202(21)2228, (01)00884838054202(21)1908, (01)00884838054202(21)1196, (01)00884838054202(21)1752, (01)00884838054202(21)2509, (01)00884838054202(21)1427, (01)00884838054202(21)2214, (01)00884838054202(21)2303, (01)00884838054202(21)1962, (01)00884838054202(21)2523, (01)00884838054202(21)2184, (01)00884838054202(21)2628, (01)00884838054202(21)2027, (01)00884838054202(21)2175, (01)00884838054202(21)1250, (01)00884838054202(21)1942, (01)00884838054202(21)1899, (01)00884838054202(21)2283, (01)00884838054202(21)2454, (01)00884838054202(21)1862, (01)00884838054202(21)1574, (01)00884838054202(21)2301, (01)00884838054202(21)2146, (01)00884838054202(21)1929, (01)00884838054202(21)2092, (01)00884838054202(21)1876, (01)00884838054202(21)1664, (01)00884838054202(21)2378, (01)00884838054202(21)1599, (01)00884838054202(21)2492, (01)00884838054202(21)1765, (01)00884838054202(21)1739, (01)00884838054202(21)1725, (01)00884838054202(21)2022, (01)00884838054202(21)2203, (01)00884838054202(21)1290, (01)00884838054202(21)2142, (01)00884838054202(21)2457, (01)00884838054202(21)2202, (01)00884838054202(21)2440, (01)00884838054202(21)2574, (01)00884838054202(21)1935, (01)00884838054202(21)2185, (01)00884838054202(21)1415, (01)00884838054202(21)1912, (01)00884838054202(21)2288, (01)00884838054202(21)2141, (01)00884838054202(21)1542, (01)00884838054202(21)2593, (01)00884838054202(21)1991, (01)00884838054202(21)2264, (01)00884838054202(21)1528, (01)00884838054202(21)1744, (01)00884838054202(21)2164, (01)00884838054202(21)2326, (01)00884838054202(21)2325, (01)00884838054202(21)2469, (01)00884838054202(21)1888, (01)00884838054202(21)2453, (01)00884838054202(21)2309, (01)00884838054202(21)1976, (01)00884838054202(21)2497, (01)00884838054202(21)2564, (01)00884838054202(21)1905, (01)00884838054202(21)2282, (01)00884838054202(21)1742, (01)00884838054202(21)2329, (01)00884838054202(21)1796, (01)00884838054202(21)2394, (01)00884838054202(21)943, (01)00884838054202(21)1793, (01)00884838054202(21)2014, (01)00884838054202(21)2343, (01)00884838054202(21)2013, (01)00884838054202(21)2498, (01)00884838054202(21)796, (01)00884838054202(21)1161, (01)00884838054202(21)1504, (01)00884838054202(21)2109, (01)00884838054202(21)2128, (01)00884838054202(21)2307, (01)00884838054202(21)2499, (01)00884838054202(21)2096, (01)00884838054202(21)1910, (01)00884838054202(21)2242, (01)00884838054202(21)1841, (01)00884838054202(21)1643, (01)00884838054202(21)2052, (01)00884838054202(21)1944, (01)00884838054202(21)2406, (01)00884838054202(21)1727, (01)00884838054202(21)2337, (01)00884838054202(21)1465, (01)00884838054202(21)2382, (01)00884838054202(21)2313, (01)00884838054202(21)1923, (01)00884838054202(21)1216, (01)00884838054202(21)2596, (01)00884838054202(21)1642, (01)00884838054202(21)1573, (01)00884838054202(21)2462, (01)00884838054202(21)1575, (01)00884838054202(21)1761, (01)00884838054202(21)1680, (01)00884838054202(21)1665, (01)00884838054202(21)2448, (01)00884838054202(21)1530, (01)00884838054202(21)2494, (01)00884838054202(21)2530, (01)00884838054202(21)1965, (01)00884838054202(21)2554, (01)00884838054202(21)1904, (01)00884838054202(21)2599, (01)00884838054202(21)1552, (01)00884838054202(21)1469, (01)00884838054202(21)1683, (01)00884838054202(21)2524, (01)00884838054202(21)1019, (01)00884838054202(21)2611, (01)00884838054202(21)2118, (01)00884838054202(21)2365, (01)00884838054202(21)1473, (01)00884838054202(21)1544, (01)00884838054202(21)1253, (01)00884838054202(21)1808, (01)00884838054202(21)1963, (01)00884838054202(21)1153, (01)00884838054202(21)1795, (01)00884838054202(21)2412, (01)00884838054202(21)2336, (01)00884838054202(21)2067, (01)00884838054202(21)1762, (01)00884838054202(21)1738, (01)00884838054202(21)1958, (01)00884838054202(21)1593, (01)00884838054202(21)1355, (01)00884838054202(21)2341, (01)00884838054202(21)1223, (01)00884838054202(21)2260, (01)00884838054202(21)1158, (01)00884838054202(21)1661, (01)00884838054202(21)1849, (01)00884838054202(21)1663, (01)00884838054202(21)1521, (01)00884838054202(21)1847, (01)00884838054202(21)2449, (01)00884838054202(21)2395, (01)00884838054202(21)2270, (01)00884838054202(21)1992, (01)00884838054202(21)2219, (01)00884838054202(21)999, (01)00884838054202(21)1518, (01)00884838054202(21)1997, (01)00884838054202(21)2225, (01)00884838054202(21)2136, (01)00884838054202(21)2399, (01)00884838054202(21)1981, (01)00884838054202(21)1773, (01)00884838054202(21)1659, (01)00884838054202(21)1409, (01)00884838054202(21)1276, (01)00884838054202(21)2272, (01)00884838054202(21)2437, (01)00884838054202(21)2474, (01)00884838054202(21)2278, (01)00884838054202(21)2268, (01)00884838054202(21)1510, (01)00884838054202(21)2156, (01)00884838054202(21)2263, (01)00884838054202(21)1961, (01)00884838054202(21)2520, (01)00884838054202(21)2003, (01)00884838054202(21)1943, (01)00884838054202(21)2452, (01)00884838054202(21)2401, (01)00884838054202(21)1502, (01)00884838054202(21)2144, (01)00884838054202(21)2331, (01)00884838054202(21)2380, (01)00884838054202(21)2078, (01)00884838054202(21)2011, (01)00884838054202(21)1954, (01)00884838054202(21)1763, (01)00884838054202(21)1697, (01)00884838054202(21)1162, (01)00884838054202(21)1398, (01)00884838054202(21)2582, (01)00884838054202(21)1746, (01)00884838054202(21)2289, (01)00884838054202(21)2508, (01)00884838054202(21)2442, (01)00884838054202(21)2269, (01)00884838054202(21)1671, (01)00884838054202(21)1670, (01)00884838054202(21)2308, (01)00884838054202(21)1227, (01)00884838054202(21)2431, (01)00884838054202(21)2299, (01)00884838054202(21)2563, (01)00884838054202(21)1668, (01)00884838054202(21)2471, (01)00884838054202(21)1097, (01)00884838054202(21)1258, (01)00884838054202(21)802, (01)00884838054202(21)1555, (01)00884838054202(21)1292, (01)00884838054202(21)817, (01)00884838054202(21)764, (01)00884838054202(21)1302, (01)00884838054202(21)941, (01)00884838054202(21)1044, (01)00884838054202(21)1451, (01)00884838054202(21)2102, (01)00884838054202(21)1159, (01)00884838054202(21)1025, (01)00884838054202(21)1638, (01)00884838054202(21)1088, (01)00884838054202(21)1312, (01)00884838054202(21)2207, (01)00884838054202(21)1403, (01)00884838054202(21)1134, (01)00884838054202(21)1135, (01)00884838054202(21)850, (01)00884838054202(21)1802, (01)00884838054202(21)2485, (01)00884838054202(21)1342, (01)00884838054202(21)1535, (01)00884838054202(21)1703, (01)00884838054202(21)1603, (01)00884838054202(21)2077, (01)00884838054202(21)2495, (01)00884838054202(21)1539, (01)00884838054202(21)2390, (01)00884838054202(21)1206, (01)00884838054202(21)1986, (01)00884838054202(21)1438, (01)00884838054202(21)1086, (01)00884838054202(21)1087, (01)00884838054202(21)1937, (01)00884838054202(21)2076, (01)00884838054202(21)951, (01)00884838054202(21)2256, (01)00884838054202(21)1816, (01)00884838054202(21)1817, (01)00884838054202(21)467, (01)00884838054202(21)2169, (01)00884838054202(21)1818, (01)00884838054202(21)2104, (01)00884838054202(21)926, (01)00884838054202(21)709, (01)00884838054202(21)1648, (01)00884838054202(21)2398, (01)00884838054202(21)1532, (01)00884838054202(21)822, (01)00884838054202(21)2445, (01)00884838054202(21)2163, (01)00884838054202(21)118 4, (01)00884838054202(21)1442, (01)00884838054202(21)1596, (01)00884838054202(21)2410, (01)00884838054202(21)1972, (01)00884838054202(21)2450, (01)00884838054202(21)1123, (01)00884838054202(21)1124, (01)00884838054202(21)2353, (01)00884838054202(21)1879, (01)00884838054202(21)1977, (01)00884838054202(21)1660, (01)00884838054202(21)2062, (01)00884838054202(21)1968, (01)00884838054202(21)1654, (01)00884838054202(21)832, (01)00884838054202(21)1851, (01)00884838054202(21)1554, (01)00884838054202(21)1105, (01)00884838054202(21)1346, (01)00884838054202(21)1309, (01)00884838054202(21)1361, (01)00884838054202(21)1689, (01)00884838054202(21)1792, (01)00884838054202(21)1209, (01)00884838054202(21)1221, (01)00884838054202(21)1884, (01)00884838054202(21)1825, (01)00884838054202(21)2546, (01)00884838054202(21)2503, (01)00884838054202(21)1913, (01)00884838054202(21)1320, (01)00884838054202(21)1064, (01)00884838054202(21)2531, (01)00884838054202(21)1960, (01)00884838054202(21)2015, (01)00884838054202(21)1595, (01)00884838054202(21)1901, (01)00884838054202(21)2357, (01)00884838054202(21)2356, (01)00884838054202(21)1073, (01)00884838054202(21)2274, (01)00884838054202(21)1602, (01)00884838054202(21)1372, (01)00884838054202(21)1045, (01)00884838054202(21)2101, (01)00884838054202(21)1037, (01)00884838054202(21)732, (01)00884838054202(21)1548, (01)00884838054202(21)1426, (01)00884838054202(21)919, (01)00884838054202(21)2467, (01)00884838054202(21)1265, (01)00884838054202(21)1268, (01)00884838054202(21)2496, (01)00884838054202(21)2082, (01)00884838054202(21)1547, (01)00884838054202(21)1546, (01)00884838054202(21)2294, (01)00884838054202(21)1858, (01)00884838054202(21)1429, (01)00884838054202(21)1101, (01)00884838054202(21)1926, (01)00884838054202(21)1145, (01)00884838054202(21)1791, (01)00884838054202(21)975, (01)00884838054202(21)2006, (01)00884838054202(21)2558, (01)00884838054202(21)1757, (01)00884838054202(21)1367, (01)00884838054202(21)1358, (01)00884838054202(21)1356, (01)00884838054202(21)2514, (01)00884838054202(21)2049, (01)00884838054202(21)2198, (01)00884838054202(21)2004, (01)00884838054202(21)2267, (01)00884838054202(21)1886, (01)00884838054202(21)2386, (01)00884838054202(21)1627, (01)00884838054202(21)2413, (01)00884838054202(21)2122, (01)00884838054202(21)1164, (01)00884838054202(21)2423, (01)00884838054202(21)1503, (01)00884838054202(21)2338, (01)00884838054202(21)2657, (01)00884838054202(21)1838, (01)00884838054202(21)2354, (01)00884838054202(21)791, (01)00884838054202(21)1784, (01)00884838054202(21)1874, (01)00884838054202(21)2149, (01)00884838054202(21)1585, (01)00884838054202(21)1615, (01)00884838054202(21)1934, (01)00884838054202(21)2046, (01)00884838054202(21)1868, (01)00884838054202(21)1896, (01)00884838054202(21)2434, (01)00884838054202(21)2330, (01)00884838054202(21)1712, (01)00884838054202(21)1982, (01)00884838054202(21)847, (01)00884838054202(21)1118, (01)00884838054202(21)1183, (01)00884838054202(21)2189, (01)00884838054202(21)1918, (01)00884838054202(21)1625, (01)00884838054202(21)1715, (01)00884838054202(21)2150, (01)00884838054202(21)2029, (01)00884838054202(21)1062, (01)00884838054202(21)1684, (01)00884838054202(21)2074, (01)00884838054202(21)1788, (01)00884838054202(21)2073, (01)00884838054202(21)429, (01)00884838054202(21)1685, (01)00884838054202(21)1731, (01)00884838054202(21)1799, (01)00884838054202(21)2170, (01)00884838054202(21)1102, (01)00884838054202(21)1557, (01)00884838054202(21)974, (01)00884838054202(21)2171, (01)00884838054202(21)1170, (01)00884838054202(21)2026, (01)00884838054202(21)1241, (01)00884838054202(21)2190, (01)00884838054202(21)1297, (01)00884838054202(21)1996, (01)00884838054202(21)2065, (01)00884838054202(21)2199, (01)00884838054202(21)1550, (01)00884838054202(21)1453, (01)00884838054202(21)1975, (01)00884838054202(21)1714, (01)00884838054202(21)898, (01)00884838054202(21)752, (01)00884838054202(21)1257, (01)00884838054202(21)2778, (01)00884838054202(21)2697, (01)00884838054202(21)2783, (01)00884838054202(21)2458, (01)00884838054202(21)2735, (01)00884838054202(21)2762, (01)00884838054202(21)2769, (01)00884838054202(21)2852, (01)00884838054202(21)2694; 12. Allura Xper FD20 Biplane; Model: 722008; UDI-DI: N/A; Serial Numbers: 445, 178, 420, 267, 245, 516, 67, 71, 379, 258, 124, 54, 142, 81, 33, 57, 447, 215, 122, 61, 58, 17, 31; 13. Allura Xper FD20 Biplane; Model: 722013; UDI: (01)00884838059061(21)441, (01)00884838059061(21)249, (01)00884838059061(21)410, (01)00884838059061(21)268, (01)00884838059061(21)124, (01)00884838059061(21)531, (01)00884838059061(21)316, (01)00884838059061(21)78, (01)00884838059061(21)564, (01)00884838059061(21)164, (01)00884838059061(21)422, (01)00884838059061(21)428, (01)00884838059061(21)634, (01)00884838059061(21)327, (01)00884838059061(21)313, (01)00884838059061(21)309, (01)00884838059061(21)35, (01)00884838059061(21)266, (01)00884838059061(21)633, (01)00884838059061(21)417, (01)00884838059061(21)390, (01)00884838059061(21)525, (01)00884838059061(21)542, (01)00884838059061(21)344, (01)00884838059061(21)465, (01)00884838059061(21)394; 14. Allura Xper FD20 OR Table; Model: 722023; UDI: (01)00884838059085(21)127, (01)00884838059085(21)177, (01)00884838059085(21)9, (01)00884838059085(21)38, (01)00884838059085(21)56, (01)00884838059085(21)126, (01)00884838059085(21)144; 15. Allura Xper FD20 OR Table; Model: 722035; UDI: (01)00884838054240(21)119, (01)00884838054240(21)126, (01)00884838054240(21)169, (01)00884838054240(21)124, (01)00884838054240(21)161, (01)00884838054240(21)183, (01)00884838054240(21)246, (01)00884838054240(21)245, (01)00884838054240(21)242, (01)00884838054240(21)196, (01)00884838054240(21)236, (01)00884838054240(21)172, (01)00884838054240(21)212, (01)00884838054240(21)222, (01)00884838054240(21)121, (01)00884838054240(21)131, (01)00884838054240(21)145, (01)00884838054240(21)127, (01)00884838054240(21)187, (01)00884838054240(21)243, (01)00884838054240(21)143, (01)00884838054240(21)244, (01)00884838054240(21)125, (01)00884838054240(21)117, (01)00884838054240(21)247, (01)00884838054240(21)165, (01)00884838054240(21)252, (01)00884838054240(21)192, (01)00884838054240(21)128, (01)00884838054240(21)111, (01)00884838054240(21)227, (01)00884838054240(21)144, (01)00884838054240(21)174, (01)00884838054240(21)130, (01)00884838054240(21)263, (01)00884838054240(21)132, (01)00884838054240(21)134, (01)00884838054240(21)226, (01)00884838054240(21)190, (01)00884838054240(21)249, (01)00884838054240(21)109, (01)00884838054240(21)251, (01)00884838054240(21)232, (01)00884838054240(21)207, (01)00884838054240(21)158, (01)00884838054240(21)160, (01)00884838054240(21)237, (01)00884838054240(21)163, (01)00884838054240(21)248, (01)00884838054240(21)253, (01)00884838054240(21)154, (01)00884838054240(21)170, (01)00884838054240(21)195, (01)00884838054240(21)166, (01)00884838054240(21)235, (01)00884838054240(21)230, (01)00884838054240(21)164, (01)00884838054240(21)221, (01)00884838054240(21)255, (01)00884838054240(21)229, (01)00884838054240(21)256, (01)00884838054240(21)254, (01)00884838054240(21)261, (01)00884838054240(21)218, (01)00884838054240(21)238, (01)00884838054240(21)113, (01)00884838054240(21)258, (01)00884838054240(21)241, (01)00884838054240(21)188, (01)00884838054240(21)259, (01)00884838054240(21)167, (01)00884838054240(21)204, (01)00884838054240(21)233, (01)00884838054240(21)153, (01)00884838054240(21)191, (01)00884838054240(21)225, (01)00884838054240(21)193, (01)00884838054240(21)185, (01)00884838054240(21)186, (01)00884838054240(21)129, (01)00884838054240(21)151, (01)00884838054240(21)92, (01)00884838054240(21)201, (01)00884838054240(21)149, (01)00884838054240(21)211, (01)00884838054240(21)198, (01)00884838054240(21)136, (01)00884838054240(21)194, (01)00884838054240(21)200, (01)00884838054240(21)116, (01)00884838054240(21)206, (01)00884838054240(21)168, (01)00884838054240(21)203, (01)00884838054240(21)146, (01)00884838054240(21)156, (01)00884838054240(21)173, (01)00884838054240(21)171, (01)00884838054240(21)184, (01)00884838054240(21)215, (01)00884838054240(21)231, (01)00884838054240(21)202; 16. Allura Xper FD20 Biplane OR Table; Model: 722025; UDI: (01)00884838059108(21)2; 17. Allura Xper FD20/10; Model: 722029; UDI: (01)00884838054219(21)88, (01)00884838054219(21)95, (01)00884838054219(21)126, (01)00884838054219(21)94, (01)00884838054219(21)97, (01)00884838054219(21)91, (01)00884838054219(21)110, (01)00884838054219(21)100, (01)00884838054219(21)114, (01)00884838054219(21)128, (01)00884838054219(21)108, (01)00884838054219(21)103, (01)00884838054219(21)107, (01)00884838054219(21)81, (01)00884838054219(21)119, (01)00884838054219(21)118, (01)00884838054219(21)80, (01)00884838054219(21)106, (01)00884838054219(21)123, (01)00884838054219(21)127, (01)00884838054219(21)122, (01)00884838054219(21)101, (01)00884838054219(21)116, (01)00884838054219(21)102, (01)00884838054219(21)109, (01)00884838054219(21)120, (01)00884838054219(21)92, (01)00884838054219(21)69, (01)00884838054219(21)93, (01)00884838054219(21)51, (01)00884838054219(21)124, (01)00884838054219(21)87, (01)00884838054219(21)104, (01)00884838054219(21)59, (01)00884838054219(21)79, (01)00884838054219(21)96, (01)00884838054219(21)82, (01)00884838054219(21)75, (01)00884838054219(21)98, (01)00884838054219(21)89; 18. Allura Xper FD20/15; Model: 722058; UDI: (01)00884838059115(21)239, (01)00884838059115(21)305, (01)00884838059115(21)203, (01)00884838059115(21)373, (01)00884838059115(21)290, (01)00884838059115(21)124, (01)00884838059115(21)274, (01)00884838059115(21)366, (01)00884838059115(21)340, (01)00884838059115(21)162, (01)00884838059115(21)186, (01)00884838059115(21)405, (01)00884838059115(21)196, (01)00884838059115(21)156, (01)00884838059115(21)146, (01)00884838059115(21)285, (01)00884838059115(21)277, (01)00884838059115(21)168, (01)00884838059115(21)178, (01)00884838059115(21)138, (01)00884838059115(21)212, (01)00884838059115(21)257, (01)00884838059115(21)337, (01)00884838059115(21)265, (01)00884838059115(21)134, (01)00884838059115(21)238, (01)00884838059115(21)190, (01)00884838059115(21)219, (01)00884838059115(21)197, (01)00884838059115(21)325, (01)00884838059115(21)123, (01)00884838059115(21)319, (01)00884838059115(21)378, (01)00884838059115(21)205, (01)00884838059115(21)149, (01)00884838059115(21)264, (01)00884838059115(21)167, (01)00884838059115(21)365, (01)00884838059115(21)199, (01)00884838059115(21)278, (01)00884838059115(21)215, (01)00884838059115(21)288, (01)00884838059115(21)254, (01)00884838059115(21)304, (01)00884838059115(21)279, (01)00884838059115(21)171, (01)00884838059115(21)202, (01)00884838059115(21)182, (01)00884838059115(21)200, (01)00884838059115(21)129, (01)00884838059115(21)59, (01)00884838059115(21)332, (01)00884838059115(21)142, (01)00884838059115(21)335, (01)00884838059115(21)151, (01)00884838059115(21)232, (01)00884838059115(21)292, (01)00884838059115(21)183, (01)00884838059115(21)349, (01)00884838059115(21)261, (01)00884838059115(21)375, (01)00884838059115(21)161, (01)00884838059115(21)368, (01)00884838059115(21)372, (01)00884838059115(21)252, (01)00884838059115(21)320, (01)00884838059115(21)213, (01)00884838059115(21)293, (01)00884838059115(21)389, (01)00884838059115(21)169, (01)00884838059115(21)336, (01)00884838059115(21)143, (01)00884838059115(21)133, (01)00884838059115(21)192, (01)00884838059115(21)201, (01)00884838059115(21)163, (01)00884838059115(21)170, (01)00884838059115(21)77, (01)00884838059115(21)273, (01)00884838059115(21)222, (01)00884838059115(21)229, (01)00884838059115(21)253, (01)00884838059115(21)248, (01)00884838059115(21)334, (01)00884838059115(21)384, (01)00884838059115(21)268, (01)00884838059115(21)240, (01)00884838059115(21)289, (01)00884838059115(21)258, (01)00884838059115(21)269, (01)00884838059115(21)226, (01)00884838059115(21)244, (01)00884838059115(21)141, (01)00884838059115(21)391, (01)00884838059115(21)241, (01)00884838059115(21)164, (01)00884838059115(21)381, (01)00884838059115(21)281, (01)00884838059115(21)154, (01)00884838059115(21)246, (01)00884838059115(21)315, (01)00884838059115(21)256, (01)00884838059115(21)233, (01)00884838059115(21)348, (01)00884838059115(21)204, (01)00884838059115(21)396, (01)00884838059115(21)185, (01)00884838059115(21)95, (01)00884838059115(21)131, (01)00884838059115(21)377, (01)00884838059115(21)206, (01)00884838059115(21)150, (01)00884838059115(21)333, (01)00884838059115(21)155, (01)00884838059115(21)313, (01)00884838059115(21)344, (01)00884838059115(21)376, (01)00884838059115(21)382, (01)00884838059115(21)360, (01)00884838059115(21)370, (01)00884838059115(21)387, (01)00884838059115(21)356, (01)00884838059115(21)392, (01)00884838059115(21)358, (01)00884838059115(21)385, (01)00884838059115(21)367, (01)00884838059115(21)331, (01)00884838059115(21)379, (01)00884838059115(21)369, (01)00884838059115(21)397, (01)00884838059115(21)357, (01)00884838059115(21)388, (01)00884838059115(21)211, (01)00884838059115(21)125, (01)00884838059115(21)301, (01)00884838059115(21)354, (01)00884838059115(21)355, (01)00884838059115(21)159, (01)00884838059115(21)139, (01)00884838059115(21)96, (01)00884838059115(21)343, (01)00884838059115(21)177, (01)00884838059115(21)311, (01)00884838059115(21)386, (01)00884838059115(21)132, (01)00884838059115(21)291, (01)00884838059115(21)327, (01)00884838059115(21)220, (01)00884838059115(21)214, (01)00884838059115(21)145, (01)00884838059115(21)223, (01)00884838059115(21)137, (01)00884838059115(21)136, (01)00884838059115(21)119, (01)00884838059115(21)103, (01)00884838059115(21)188, (01)00884838059115(21)350, (01)00884838059115(21)347, (01)00884838059115(21)55, (01)00884838059115(21)194, (01)00884838059115(21)189, (01)00884838059115(21)237, (01)00884838059115(21)230, (01)00884838059115(21)221, (01)00884838059115(21)247, (01)00884838059115(21)242, (01)00884838059115(21)276, (01)00884838059115(21)345, (01)00884838059115(21)166, (01)00884838059115(21)179, (01)00884838059115(21)303, (01)00884838059115(21)90, (01)00884838059115(21)112, (01)00884838059115(21)245, (01)00884838059115(21)148, (01)00884838059115(21)234, (01)00884838059115(21)208, (01)00884838059115(21)193, (01)00884838059115(21)236, (01)00884838059115(21)266, (01)00884838059115(21)130, (01)00884838059115(21)235, (01)00884838059115(21)251, (01)00884838059115(21)195, (01)00884838059115(21)144, (01)00884838059115(21)330, (01)00884838059115(21)295, (01)00884838059115(21)187, (01)00884838059115(21)361, (01)00884838059115(21)153, (01)00884838059115(21)284, (01)00884838059115(21)181, (01)00884838059115(21)351, (01)00884838059115(21)210, (01)00884838059115(21)243, (01)00884838059115(21)218; 19. Allura Xper FD20/15 OR Table; Model: 722059; UDI: (01)00884838059122(21)33, (01)00884838059122(21)7, (01)00884838059122(21)35, (01)00884838059122(21)38, (01)00884838059122(21)37, (01)00884838059122(21)2; 20. Allura Xper FD20/20; Model: 722038; UDI: (01)00884838054226(21)108, (01)00884838054226(21)173, (01)00884838054226(21)119, (01)00884838054226(21)126, (01)00884838054226(21)130, (01)00884838054226(21)157, (01)00884838054226(21)141, (01)00884838054226(21)137, (01)00884838054226(21)167, (01)00884838054226(21)120, (01)00884838054226(21)169, (01)00884838054226(21)134, (01)00884838054226(21)143, (01)00884838054226(21)133, (01)00884838054226(21)166, (01)00884838054226(21)147, (01)00884838054226(21)127, (01)00884838054226(21)135, (01)00884838054226(21)123, (01)00884838054226(21)129, (01)00884838054226(21)148, (01)00884838054226(21)153, (01)00884838054226(21)164, (01)00884838054226(21)111, (01)00884838054226(21)160, (01)00884838054226(21)172, (01)00884838054226(21)171, (01)00884838054226(21)159, (01)00884838054226(21)155, (01)00884838054226(21)144, (01)00884838054226(21)163, (01)00884838054226(21)168, (01)00884838054226(21)161, (01)00884838054226(21)158, (01)00884838054226(21)156, (01)00884838054226(21)138, (01)00884838054226(21)162, (01)00884838054226(21)150, (01)00884838054226(21)146, (01)00884838054226(21)154, (01)00884838054226(21)124, (01)00884838054226(21)136, (01)00884838054226(21)140, (01)00884838054226(21)170; 21. Cardio Vascular-Allura (722400); UDI-DI: N/A: Serial Numbers: 486, 480, 322, 280, 506, 507, 497, 502, 429, 413, 222, 510, 439, 382, 364, 209, 347, 552, 482, 414, 407, 338, 424, 273, 245, 307, 476, 533, 314, 432, 458, 491, 361, 272, 250, 252, 308, 321, 525, 519, 352, 434, 232, 471, 470, 440, 545, 279, 346, 317, 369, 495, 547, 348, 380, 438, 437, 341, 513, 459, 379, 355, 485, 290, 400, 498, 278, 283, 512, 487, 455, 448, 251, 319, 381, 504, 535, 479, 423, 473, 296, 539, 285, 419, 305, 430, 449, 329, 472, 356, 292, 398, 548, 523, 299, 494, 286, 300, 488, 541, 543, 427, 447, 344, 542, 262, 536, 469, 554, 555, 546, 375, 521, 433, 463, 443, 483, 357, 374, 277, 530, 337, 549, 441, 320, 445, 287, 556, 446, 527, 526, 532, 418, 544, 444, 511, 267, 393, 303, 550, 524, 540, 284, 557, 353, 553, 496, 368, 531, 297, 538, 415, 227, 528, 259, 330, 264, 386, 462, 551, 396, 460, 522, 412, 395, 478, 499, 515, 306, 271, 435, 501, 518, 226, 360, 370, 477;
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2228-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | April 28, 2026 |
| Reported | May 27, 2026 |
| Company | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. |
| Company location | Best, Netherlands |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 3314 units (816 US, 2498 OUS) |
More recalls from PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · all medical device recalls · recalls from April 2026