Recall Ledger

GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL recall (GE Medical Systems Information Technologies Inc)

Medical deviceClass IIOngoing

Summary: GE Medical Systems Information Technologies Inc is recalling GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product Code SS9; used in conjunction with ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). Reason: GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

What is being recalled

GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product Code SS9; used in conjunction with ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Why

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution.

How to tell if yours is affected

UDI (01)00840682144773(10)v5.0.2R1C1, Serial Number or Sales Order Number: SS924160026SA, SS921080002SA, SS923080001SA, SS923080006SA, SS922460008SA, SS921320004SA, SS921320010SA, SS921320012SA, SS922120001SA, SS921210005SA, SS921060024SA, SS921060026SA, SS924290077SA, SS923290011SA, SS923290012SA, SS923290013SA, SS923290014SA, SS921520004SA, SS922080001SA, SS922080002SA, SS922120002SA, SS922120003SA, SS924200022SA, SS924450085SA, SS924220065SA, SS921210002SA, SS921210003SA, SS924220066SA, SS924050001SA, SS922170005SA, SS924160027SA, SS924160028SA, SS924160029SA, SS924160031SA, SS924160032SA, SS924160033SA, SS921300003SA, SS921420015SA, SS921420016SA, SS921520002SA, SS920520004SA, SS922250001SA, SS923510002SA, SS923510003SA, SS924160023SA, SS924170017SA, SS924200031SA, SS924200032SA, SS924090011SA, SS924090016SA, SS924090018SA, SS924090020SA, SS924090021SA, SS924090024SA, SS924140003SA, SS924140006SA, SS924140011SA, SS924140012SA, SS924140014SA, SS924170006SA, SS924170007SA, SS924170008SA, SS924170009SA, SS924170010SA, SS924170011SA, SS924170012SA, SS924170013SA, SS924170014SA, SS924170015SA, SS924170016SA, SS921110009SA, SS921420013SA, SS922170004SA, SS922340005SA, SS922290002SA, SS923220003SA, SS923290008SA, SS924020004SA, SS921040002SA, SS921050002SA, SS921060023SA, SS921060025SA, SS921060020SA, SS921060022SA, SS921060016SA, SS921060017SA, SS921060018SA, SS921060021SA, SS921110004SA, SS921110001SA, SS921240007SA, SS920490011SA, SS924430083SA, SS924050023SA, SS924060001SA, SS922080004SA, SS922080005SA, SS922080006SA, SS923240002SA, SS923520001SA, SS922120004SA, SS922390002SA, SS923400001SA, SS922340008SA, SS922340009SA, SS922340010SA, SS922020001SA, SS923230004SA, SS923230005SA, SS923230006SA, SS923470001SA, SS923470002SA, SS923470003SA, SS923470006SA, SS923470007SA, SS923480001SA, SS924060011SA, SS922240002SA, SS924060012SA, SS924070006SA, SS924070010SA, SS921110005SA, SS921180006SA, SS921420001SA, SS921470002SA, SS921470004SA, SS922250003SA, SS922340004SA, SS922460004SA, SS922460007SA, SS924060008SA, SS924060009SA, SS924060010SA, SS922130001SA, SS921300001SA, SS922410002SA, SS920500009SA, SS920500008SA, SS921240009SA, SS923430002SA, SS922010003SA, SS923370001SA, SS924050013SA, SS924050025SA, SS924050009SA, SS924050010SA, SS924050017SA, SS924290078SA, SS920430005SA, SS920430006SA, SS920460004SA, SS921140001SA, SS924040005SA, SS920450002SA, SS920450003SA, SS923230017SA, SS924070007SA, SS924070005SA, SS924090013SA, SS924090014SA, SS923250002SA, SS923250003SA, SS920520001SA, SS921180010SA, SS922340011SA, SS920490001SA, SS923030004SA, SS923240001SA, SS924140001SA, SS923110002SA, SS924040006SA, SS922310001SA, SS921080001SA, SS924050007SA, SS921180008SA, SS923520002SA, SS922340006SA, SS924050016SA, SS924050012SA, SS924150001SA, SS924150002SA, SS921320001SA, SS921320002SA, SS921320003SA, SS921320006SA, SS921320007SA, SS921320008SA, SS921320009SA, SS921320011SA, SS921300005SA, SS921240001SA, SS920450001SA, SS921300006SA, SS923450001SA, SS921520003SA, SS924130001SA, SS924130003SA, SS922410003SA, SS922460003SA, SS921520008SA, SS924130004SA, SS920430001SA, SS920510003SA, SS920510007SA, SS920510010SA, SS923070001SA, SS921300002SA, SS924070009SA, SS922040008SA, SS920510009SA, SS924050003SA, SS922090001SA, SS923290007SA, SS923300001SA, SS923300002SA, SS924050006SA, SS924050011SA, SS924050008SA, SS922080003SA, SS923030003SA, SS921130001SA, SS921150001SA, SS921150002SA, SS921460009SA, SS921320005SA, SS922180001SA, SS920490002SA, SS921460007SA, SS924050005SA, SS925270092SA, SS925270093SA, SS925270095SA, SS925270094SA, SS924160022SA, SS924050015SA, SS925330096SA, SS920430002SA, SS920430004SA, SS924070015SA, SS925260090SA, SS925260091SA, SS921460005SA, SS922250002SA, SS923030002SA, SS923030006SA, SS924050014SA, SS924430084SA, SS921470005SA, SS921210001SA, SS921210007SA, SS921210008SA, SS921210009SA, SS921210010SA, SS924050004SA, SS923230026SA, SS921420017SA, SS921150003SA, SS921050001SA, SS924070003SA, SS924070004SA, SS922120005SA, SS922250004SA, SS924200024SA, SS924200025SA, SS924200026SA, SS924200027SA, SS924200028SA, SS924200029SA, SS924200030SA, SS921110003SA, SS923110003SA, SS922390001SA, SS922290001SA, SS924140004SA, SS923200001SA, SS921150005SA, SS924090017SA, SS923290006SA, SS921240004SA, SS921240005SA, SS921240006SA, SS921240010SA, SS921460003SA, SS921420018SA, SS921420019SA, SS924170004SA, SS924170005SA, SS924170003SA, SS924090015SA, SS924140016SA, SS924090027SA, SS924090026SA, SS924090023SA, SS924090006SA, SS922040001SA, SS924090012SA, SS921110006SA, SS924170001SA, SS924170002SA, SS924090025SA, SS924090019SA, SS923070002SA, SS923070003SA, SS922040006SA, SS923400003SA, SS921300007SA, SS922380001SA, SS921460001SA, SS921460004SA, SS920430003SA, SS921250001SA, SS923220001SA, SS922050001SA, SS922050002SA, SS922050003SA, SS922050004SA, SS924200023SA, SS924010001SA, SS924070011SA, SS920500007SA, SS920500010SA, SS922010001SA, SS923320001SA, SS923230011SA, SS923230012SA, SS923230013SA, SS923230015SA, SS923230016SA, SS923230014SA, SS923230007SA, SS923230008SA, SS923230009SA, SS923230010SA, SS922290003SA, SS921460006SA, SS922300001SA, SS923230021SA, SS923230023SA, SS923230022SA, SS923230020SA, SS923230019SA, SS923230018SA, SS924220059SA, SS924220060SA, SS920490010SA, SS923080003SA, SS923080002SA, SS922460006SA, SS922460005SA, SS924040001SA, SS921040001SA, SS922210001SA, SS924150006SA, SS924470086SA, SS924070012SA, SS924070014SA, SS924150004SA, SS924070013SA, SS921520001SA, SS924160015SA, SS924160016SA, SS924160001SA, SS924160002SA, SS924160003SA, SS924160004SA, SS924160005SA, SS924160006SA, SS924160007SA, SS924160008SA, SS924160009SA, SS924160010SA, SS924160011SA, SS924160012SA, SS924160013SA, SS924160014SA, SS921460010SA, SS924160024SA, SS923290002SA, SS923290003SA, SS922220001SA, SS922220002SA, SS922220003SA, SS921130003SA, SS921130004SA, SS925140088SA, SS924130002SA, SS923290004SA, SS923290005SA, SS922110001SA, SS922110002SA, SS921520007SA, SS921520009SA, SS921520005SA, SS924420082SA, SS924420081SA, SS922010002SA, SS924090022SA, SS925120087SA, SS922040002SA, SS922040004SA, SS922040005SA, SS921180009SA, SS922040003SA, SS922040007SA, SS920490006SA, SS920490009SA, SS924040011SA, SS922460002SA, SS922050005SA, SS921490002SA, SS923080005SA, SS924070001SA, SS924040002SA, SS921470003SA, SS920490004SA, SS920510005SA, SS921210006SA, SS920520003SA, SS920510008SA, SS920490005SA, SS921240008SA, SS924040009SA, SS921150004SA, SS922300002SA, SS921420011SA, SS920490007SA, SS920490008SA, SS921420002SA, SS920490003SA, SS920510012SA, SS920510011SA, SS923250001SA, SS924040007SA, SS922390003SA, SS922420001SA, SS921130002SA, SS921350001SA, SS923110004SA, SS924050018SA, SS924050019SA, SS924050020SA, SS924050021SA, SS924050022SA, SS924050024SA, SS923210002SA, SS923210003SA, SS924050002SA, SS921300004SA, SS922340007SA, SS923400002SA, SS923420001SA, SS924020002SA, SS920510004SA, SS924060002SA, SS921520006SA, SS923080004SA, SS922240001SA, SS923030005SA, SS921460011SA, SS920510002SA, SS924020001SA, SS924020003SA, SS923030001SA, SS921210004SA, SS924150003SA, SS921060027SA, SS921030011SA, SS923170001SA, SS922170001SA, SS922170002SA, SS922170003SA, SS923230024SA, SS920510001SA, SS922310002SA, SS923290001SA, SS922300003SA.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2242-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedApril 24, 2026
ReportedJune 3, 2026
CompanyGE Medical Systems Information Technologies Inc
Company locationWaukesha, WI, United States
Initiated byVoluntary: Firm initiated
Quantity476 units

More recalls from GE Medical Systems Information Technologies Inc · all medical device recalls · recalls from April 2026

Official FDA record