Phothera 200 recall (Daavlin Distributing Company)

Medical deviceClass IIOngoing

Summary: Daavlin Distributing Company is recalling Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis, Vitiligo, and Atopic Dermatitis (Eczema) under the direction of a physician, Model Numbers 210NDT4 and 210NCL4. Reason: Some of the Calibration/Output Certificates had the outputs reversed.

What is being recalled

Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis, Vitiligo, and Atopic Dermatitis (Eczema) under the direction of a physician, Model Numbers 210NDT4 and 210NCL4

Why

Some of the Calibration/Output Certificates had the outputs reversed.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

Serial Numbers: 2-03409NB4DT 2-03560NB4DT 2-03798NB4DT 2-03472NB4DT 2-03315NB4DT 2-03726NB4DT 2-03396NB4DT 2-03320NB4DT 2-03447NB4DT 2-03564NB4DT 2-03516NB4DT 2-03467NB4DT 2-03292NB4DT 2-03371NB4DT 2-03433NB4DT 2-03446NB4DT 2-03343NB4DT 2-03400NB4DT 2-03342NB4DT 2-03547NB4DT 2-03449NB4DT 2-03685NB4DT 2-03574NB4DT 2-03392NB4DT 2-03345NB4DT 2-03431NB4DT 2-03608NB4DT 2-03530NB4DT 2-03501NB4DT 2-03362NB4DT 2-03603NB4DT 2-03713NB4DT 2-03508NB4DT 2-03341NB4DT 2-03418NB4DT 2-03721NB4DT 2-03662NB4DT 2-03679NB4DT 2-03591NB4DT 2-03319NB4DT 2-03411NB4DT 2-03505NB4DT 2-03335NB4DT 2-03585NB4DT 2-03556NB4DT 2-03355NB4DT 2-03664NB4DT 2-03515NB4DT 2-03424NB4DT 2-03470NB4DT 2-03386NB4DT 2-03455NB4DT 2-03401NB4DT 2-03391NB4DT 2-03578NB4DT 2-03388NB4DT 2-03632NB4DT 2-03615NB4DT 2-03434NB4DT 2-03454NB4DT 2-03387NB4DT 2-03390NB4DT 2-03569NB4DT 2-03541NB4DT 2-03509NB4DT 2-03789NB4DT 2-03804NB4DT 2-03439NB4DT 2-03367NB4DT 2-03442NB4DT 2-03750NB4DT 2-03349NB4DT 2-03510NB4DT 2-03538NB4DT 2-03559NB4DT 2-03539NB4DT 2-03491NB4DT 2-03502NB4DT 2-03629NB4DT 2-03383NB4DT 2-03496NB4DT 2-03619NB4DT 2-03511NB4DT 2-03670NB4DT 2-03565NB4DT 2-03581NB4DT 2-03479NB4DT 2-03348NB4DT 2-03534NB4DT 2-03484NB4DT 2-03802NB4DT 2-03504NB4DT 2-03457NB4DT 2-03417NB4DT 2-03482NB4DT 2-03318NB4DT 2-03503NB4DT 2-03518NB4DT 2-03350NB4DT 2-03686NB4DT 2-03317NB4DT 2-03687NB4DT 2-03507NB4DT 2-03663NB4DT 2-03658NB4DT 2-03330NB4DT 2-03489NB4DT 2-03451NB4DT 2-03486NB4DT 2-03465NB4DT 2-03346NB4DT 2-03598NB4DT 2-03377NB4DT 2-03414NB4DT 2-03422NB4DT 2-03531NB4DT 2-03329NB4DT 2-03579NB4DT 2-03817NB4DT 2-03613NB4DT 2-03322NB4DT 2-03570NB4DT 2-03397NB4DT 2-03326NB4DT 2-03347NB4DT 2-03469NB4DT 2-03483NB4DT 2-03359NB4DT 2-03344NB4DT 2-03436NB4DT 2-03711NB4DT 2-03437NB4DT 2-03459NB4DT 2-03384NB4DT 2-03389NB4DT 2-03592NB4DT 2-03432NB4DT 2-03823NB4DT 2-03616NB4DT 2-03498NB4DT 2-03512NB4DT 2-03809NB4DT 2-03647NB4DT 2-03606NB4DT 2-03722NB4DT 2-03818NB4DT 2-03589NB4DT 2-03356NB4DT 2-03583NB4DT 2-03799NB4DT 2-03487NB4DT 2-03398NB4DT 2-03775NB4DT 2-03427NB4DT 2-03533NB4DT 2-03676NB4DT 2-03468NB4DT 2-03684NB4DT 2-03413NB4DT 2-03716NB4DT 2-03374NB4DT 2-03430NB4DT 2-03473NB4DT 2-03575NB4DT 2-03339NB4DT 2-03666NB4DT 2-03323NB4DT 2-03471NB4DT 2-03334NB4DT 2-03310NB4DT 2-03631NB4DT 2-03405NB4DT 2-03720NB4DT 2-03372NB4DT 2-03463NB4DT 2-03353NB4DT 2-03563NB4DT 2-03419NB4DT 2-03725NB4DT 2-03621NB4DT 2-03336NB4DT 2-03312NB4DT 2-03806NB4DT 2-03572NB4DT 2-03597NB4DT 2-03370NB4DT 2-03435NB4DT 2-03354NB4DT 2-03555NB4DT 2-03440NB4DT 2-03376NB4DT 2-03610NB4DT 2-03653NB4DT 2-03749NB4DT 2-03380NB4DT 2-03558NB4DT 2-03462NB4DT 2-03313NB4DT 2-03517NB4DT 2-03790NB4DT 2-03513NB4DT 2-03602NB4DT 2-03683NB4DT 2-03561NB4DT 2-03514NB4DT 2-03604NB4DT 2-03445NB4DT 2-03423NB4DT 2-03338NB4DT 2-03375NB4DT 2-03361NB4DT 2-03543NB4DT 2-03492NB4DT 2-03801NB4DT 2-03497NB4DT 2-03601NB4DT 2-03824NB4DT 2-03477NB4DT 2-03545NB4DT 2-03813NB4DT 2-03557NB4DT 2-03426NB4DT 2-03428NB4DT 2-03394NB4DT 2-03645NB4DT 2-03842NB4DT 2-03528NB4DT 2-03321NB4DT 2-03478NB4DT 2-03438NB4DT 2-03369NB4DT 2-03661NB4DT 2-03682NB4DT 2-03412NB4DT 2-03351NB4DT 2-03673NB4DT 2-03453NB4DT 2-03605NB4DT 2-03461NB4DT 2-03495NB4DT 2-03582NB4DT 2-03540NB4DT 2-03594NB4DT 2-03612NB4DT 2-03810NB4DT 2-03364NB4DT 2-03609NB4DT 2-03672NB4DT 2-03529NB4DT 2-03379NB4DT 2-03554NB4DT 2-03490NB4DT 2-03407NB4DT 2-03586NB4DT 2-03600NB4DT 2-03618NB4DT 2-03421NB4DT 2-03659NB4DT 2-03671NB4DT 2-03448NB4DT 2-03607NB4DT 2-03402NB4DT 2-03485NB4DT 2-03480NB4DT 2-03420NB4DT 2-03328NB4DT 2-03408NB4DT 2-03481NB4DT 2-03625NB4DT 2-03331NB4DT 2-03614NB4DT 2-03646NB4DT 2-03668NB4DT 2-03800NB4DT 2-03499NB4DT 2-03753NB4DT 2-03520NB4DT 2-03464NB4DT 2-03316NB4DT 2-03580NB4DT 2-03675NB4DT 2-03665NB4DT 2-03546NB4DT 2-03590NB4DT 2-03544NB4DT 2-03415NB4DT 2-03719NB4DT 2-03456NB4DT 2-03532NB4DT 2-03385NB4DT 2-03452NB4DT 2-03593NB4DT 2-03366NB4DT 2-03759NB4DT 2-03506NB4DT 2-03425NB4DT 2-03660NB4DT 2-03360NB4DT 2-03519NB4DT 2-03688NB4DT 2-03537NB4DT 2-03429NB4DT 2-03406NB4DT 2-03340NB4DT 2-03311NB4DT 2-03562NB4DT 2-03474NB4DT 2-03655NB4DT 2-03466NB4DT 2-03595NB4DT 2-03476NB4DT 2-03403NB4DT 2-03752NB4DT 2-03332NB4DT 2-03314NB4DT 2-03309NB4DT 2-03678NB4DT 2-03410NB4DT 2-03460NB4DT 2-03717NB4DT 2-03521NB4DT 2-03674NB4DT 2-03363NB4DT 2-03667NB4DT 2-03475NB4DT 2-03724NB4DT 2-03493NB4DT 2-03617NB4DT 2-03680NB4DT 2-03494NB4DT 2-03450NB4DT 2-03754NB4DT 2-03378NB4DT 2-03712NB4DT 2-03840NB4DT 2-03441NB4DT 2-03416NB4DT 2-03393NB4DT 2-03577NB4DT 2-03751NB4DT 2-03681NB4DT 2-03368NB4DT 2-03333NB4DT 2-03640NB4DT 2-03373NB4DT 2-03358NB4DT 2-03500NB4DT 2-03596NB4DT 2-03811NB4DT 2-03308NB4DT 2-03669NB4DT 2-03337NB4DT 2-03381NB4DT 2-03352NB4DT 2-03488NB4DT 2-03382NB4DT 2-03443NB4DT 2-03357NB4DT 2-03444NB4DT 2-03542NB4DT 2-03599NB4DT 2-03404NB4DT 2-03395NB4DT 2-03365NB4DT 2-03399NB4DT 2-03576NB4DT 2-03458NB4DT 2-03611NB4DT

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2295-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 10, 2025
ReportedJune 10, 2026
CompanyDaavlin Distributing Company
Company locationBryan, OH, United States
Initiated byVoluntary: Firm initiated
Quantity368 units