Recall Ledger

Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number:... recall (The Binding Site Group, Ltd.)

Medical deviceClass IIOngoing

Summary: The Binding Site Group, Ltd. is recalling Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction . Reason: It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.

What is being recalled

Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.

Why

It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2362-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedApril 16, 2026
ReportedJune 17, 2026
CompanyThe Binding Site Group, Ltd.
Company locationBirmingham, N/A, United Kingdom
Initiated byVoluntary: Firm initiated
Quantity1

More recalls from The Binding Site Group, Ltd. · all medical device recalls · recalls from April 2026

Official FDA record