Description/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA... recall (ARROW INTERNATIONAL, LLC)

Medical deviceClass IOngoing

Summary: ARROW INTERNATIONAL, LLC is recalling Description/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT: 2L 5.5 FR X 55 CM CG+ VPS/ASK-45552-RH AGBA PICC NaviCurve: 2L 5.5FR x 55CM BP/DLX-45552-CURVB AGBA PICC NaviCur. Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

What is being recalled

Description/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT: 2L 5.5 FR X 55 CM CG+ VPS/ASK-45552-RH AGBA PICC NaviCurve: 2L 5.5FR x 55CM BP/DLX-45552-CURVB AGBA PICC NaviCurve: 2L 5.5FR x 55CM TCG/DLX-45552-CURVC AGBA PICC: 2L 5.5FR x 55CM w Biopatch/DLX-45552-HPKB AGBA PICC NaviCurve: 3L 6FR x 55CM BP/DLX-45563-CURVB PI PICC KIT: 4 FR X 50 CM/ASK-35041-NH PI PICC KIT: 2-L 5 FR X 50 CM/ASK-35052-NH PI PICC G4 Stylet: 1L 4FR x 55CM w CHG/DLX-35541-VPSC PI PICC NaviCurve: 2L 5FR x 55CM w BP/DLX-35552-CURVB PI PICC NaviCurve: 3L 6FR x 55CM w BP/DLX-35563-CURVB PI PICC G4 Stylet: 3L 6FR x 55CM w BP/DLX-35563-VPSB AGBA PI PICC KIT: 2L 5.5 FR X 55CM/CDC-45552-HPK1A PI AGBA PICC KIT: 3-L 6 FR X 55 CM/CDC-45563-HPK1A PI AGBA PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UOI1 PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-VFH1 PI AGBA PICC KIT: 2L 5.5 FR X 55 CM VPS/ASK-45552-UOI1 PICC KIT: 2L 5.5 FR X 55 CM VPS/ASK-45552-VFH1 PI AGBA PICC KIT: 3L 6 FR X 55 CM W/VPS/ASK-45563-UWH1 AGBA PICC NaviCurve: 1L 4.5FR x 55CM BP/DLX-45541-CURVB AGBA PICC NaviCurve: 1L 4.5FR x 55CM TCG/DLX-45541-CURVC AGBA PICC G4 Stylet: 1L 4.5FR x 55CM TCG/DLX-45541-VPSC AGBA PICC G4 Stylet: 2L 5.5FR x 55CM BP/DLX-45552-VPSB AGBA PICC G4 Stylet: 2L 5.5FR x 55CM TCG/DLX-45552-VPSC PI PICC:1L 4FR X 50CM PRELOAD VPS STYLET/CDC-35041-VPS PICC KIT: 3 FR X 55 CM/CDC-05531-PK1A

Why

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

How serious is it

Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.

Where it was sold

US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR. (states named: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, Wyoming)

How to tell if yours is affected

REF(Material)/UDI-DI/Lot(Batch): ASK-44063-NS/10801902193275/33F25F0100; ASK-45052-NS/10801902193275/33F24L0864 ASK-45541-MHBV/10801902193275, 10801902167696/33F25D0410, 33F26C0012, 33S24A0690, 33F25B0574; ASK-45552-RH/10801902193275/33F25B0575, 33F25D0361, 33F25F0197, 33F25K0373, 33F25A0082; DLX-45552-CURVB/10801902193275, 10801902208702/33F25E0506, 33F25E0833, 33F26C0038, 33S24B0518, 33F24L0982; DLX-45552-CURVC/10801902193275, 10801902208733/33F25B0386, 33F25D0381, 33F25D0624, 33F25E0765, 33F25F0225, 33F25H0602, 33F26C0182, 33F25B0390; DLX-45552-HPKB/10801902193275, 10801902208825, 10801902208856/33F25D0454, 33F25D0677, 33F25E0834, 33F25J0446, 33F25M0172, 33F26B0867, 33S24A0586, 33S24A1370, 33F24F0097; DLX-45563-CURVB/10801902193275/33F25C0339, 33F25E0836, 33F25A0321; ASK-35041-NH/10801902221077, 10801902193275/33F26B0140 33F25G0460; ASK-35052-NH/10801902193275/33F25H0412, 33F24L1080; DLX-35541-VPSC/10801902193275/33F25C0449, 33F25D0245, 33F25H0501, 33F24L1148; DLX-35552-CURVB/10801902193275, 10801902208764/33F25C0214, 33F25E0511 ,33F25E0832 ,33F26C0120 33F26C0732, 33F25C0373; DLX-35563-CURVB/10801902193275, 10801902208771/33F25C0337, 33F26B0155, 33F25D0453; DLX-35563-VPSB/10801902193275/33F25C0395; CDC-45552-HPK1A/10801902193275, 10801902188462/33F25D0067, 33F25D0243, 33F25E0452, 33F25E0658, 33F25E0829, 33F26B0891, 33F26C0202, 33S23K0168, 33S24A0419, 33S24A0523, 33S24A1387, 33F25K0317; CDC-45563-HPK1A/10801902193275, 10801902188479/33F25G0283, 33F25H0626, 33S24A0543, 33F24D0497; ASK-45541-UOI1/10801902193275, 10801902221664/33F25H0411, 33F25J0399, 33S24A1368, 33F26A0226; ASK-45541-VFH1/10801902193275, 10801902218886/33F25B0655, 33F25E0573, 33F25G0210, 33F26A1533, 33F26B0841, 33F25G0584; ASK-45552-UOI1/10801902193275/33F26A0227; ASK-45552-VFH1/10801902193275/33F25D0362, 33F25E0866, 33F25J0226, 33F25M0180, 33F26A1455, 33F25H0408; ASK-45563-UWH1/10801902193275/33F25D0660, 33F25J0227, 33F24K0066; DLX-45541-CURVB/10801902193275/33F25B0385, 33F25D0250, 33F25E0760, 33F25J0336, 33F25K0194, 33F24L1115; DLX-45541-CURVC/10801902193275/33F25B0507, 33F25C0216, 33F25E0761, 33F25F0223, 33F25G0362, 33F25J0337, 33F25E0762; DLX-45541-VPSC/10801902193275, 10801902208610/33F24L0390, 33F24L1130, 33F25C0247, 33F25D0251, 33F25D0380, 33F25D0623, 33F25E0764, 33F25E0900, 33F25F0224, 33F25H0604, 33F25J0445, 33F26A1541, 33F26B0460, 33F26C0220, 33F26C0717, 33F25B0387; DLX-45552-VPSB/10801902193275/33F25C0249, 33F25D0253, 33F25D0314, 33F25D0382, 33F24L0392; DLX-45552-VPSC/10801902193275, 10801902208634/33F24L0939, 33F25B0508, 33F25D0254, 33F25D0315, 33F25D0383,33F25D0626, 33F25E0581, 33F25E0709, 33F25E0835, 33F25J0129, 33F25J0257, 33F25M0167, 33F26B0712, 33F26B0937, 33F25B0388; CDC-35041-VPS/10801902186741, 10801902193275/33X25K0001, 33F25E0425; CDC-05531-PK1A/10801902193275, 60801902123420/33F25C0641, 33F25D0435, 33F25E0646, 33F25J0416, 33F26B0711, 33F25D0765

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2370-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass I
StatusOngoing
Recall startedApril 21, 2026
ReportedJune 24, 2026
CompanyARROW INTERNATIONAL, LLC
Company locationMorrisville, NC, United States
Initiated byVoluntary: Firm initiated
Quantity37,738