Description/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1 recall (ARROW INTERNATIONAL, LLC)
Medical deviceClass IOngoing
Summary: ARROW INTERNATIONAL, LLC is recalling Description/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION KIT/AK-05502D; EPIDURAL CATHETERIZATION KIT/AK-05503; EPIDURAL CATHETERIZATION KIT/AK-05503-L; SPINAL ANES/EPIDURA. Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
What is being recalled
Description/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION KIT/AK-05502D; EPIDURAL CATHETERIZATION KIT/AK-05503; EPIDURAL CATHETERIZATION KIT/AK-05503-L; SPINAL ANES/EPIDURAL CATH KIT/AK-05560; SPINAL ANES/EPIDURAL CATH KIT/ASK-02520-SBMC1; EPIDURAL CATHETERIZATION KIT/ASK-05400-BW1; EPIDURAL CATHETERIZATION KIT/ASK-05400-GH; SPINAL ANES/EPIDURAL CATH KIT/ASK-05401-LMC; EPIDURAL CATHETERIZATION KIT/ASK-05401-NM; SPINAL ANES/EPIDURAL CATH KIT/ASK-05500-BID1; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-JHH1; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-MCV1; EPIDURAL CATHETER KIT: 19GA X 90CM ASK-05500-MVH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05500-UMC; 19 GA X 35.5 IN EPIDURAL CATHETER KIT/ASK-05502-MLH; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-JMH1; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-MKD2; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-OM1; EPIDURAL CATHETERIZATION KIT/ASK-17019-SM; EPIDURAL CATHETERIZATION KIT/BP-05501; EPIDURAL CATHETERIZATION KIT/CK-05401; EPIDURAL CATHETERIZATION KIT/HS-05501; EPIDURAL CATHETERIZATION KIT/MM-05501; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TK; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TKD; EPIDURAL CATHETERIZATION KIT: 19 GA/MP-17019-TKL; EPIDURAL CATHETERIZATION KIT/MSO-05401-JMH; EPIDURAL CATHETERIZATION KIT/NM-05401; EPIDURAL CATHETERIZATION KIT/SJ-05501; EPIDURAL CATHETERIZATION KIT/SM-05401; EPIDURAL CATHETER KIT: 19 GA X 90 CM/TM-05401; EPIDURAL CATHETERIZATION KIT/TM-05502; SPINAL ANES/EPIDURAL CATH KIT/UI-05502; SPINAL ANES/EPIDURAL CATH KIT: 19 GA/ASK-05400-SHM; EPIDURAL CATH KIT: 19 GA/ASK-05400-UOM; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05500-MR; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-05500-TM; EPIDURAL CATHETERIZATION KIT/ASK-05501-BHL1; EPIDURAL CATHETERIZATION KIT/ASK-05501-GH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-SM; EPIDURAL CATHETERIZATION KIT/ASK-05502-BSM; EPIDURAL CATHETERIZATION KIT/ASK-05502-MGH1; EPIDURAL CATHETERIZATION KIT/ASK-05503-BWB1; EPIDURAL CATHETERIZATION KIT/ASK-05503-HMH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05503-SPH; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05503-UPM1; SPINAL ANES/EPIDURAL CATH KIT/ASK-05560-UCI2; EPIDURAL CATHETERIZATION KIT: 19 GA/ASK-17019-UMP; EPIDURAL CATHETERIZATION KIT/MH-05503-1; EPIDURAL CATHETERIZATION KIT/ASK-05401-VB2; Epidural Catheterization Kit: 19GA/ASK-05500-SC1; Epidural Catheterization Kit: 19GA/ASK-05500-SC3; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-AFH1; EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05501-BMC; EPIDURAL CATHETERIZATION KIT/ASK-05502-NY; EPIDURAL CATHETERIZATION KIT/ASK-17019-RMH
Why
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
How serious is it
Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.
Where it was sold
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR. (states named: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, Wyoming)
How to tell if yours is affected
REF(Material)/UDI-DI/Lot(Batch): ASK-05502-VM1/10801902210217, 10801902209938/33F25C0168, 33F25C0880, 33F25E0202, 33F25G0134, 33F25H0126, 33F26C0297, 33F24C0216; AK-05501/)10801902210217, 10801902209501, 10801902209518/33F24E0057, 33F24E0058, 33F24J0496, 33F24K0094, 33F24L0564, 33F25A0460, 33F25A0617, 33F25C0728,33F25E0043, 33F25E0913, 33F25G0081, 33F25H0015, 33F25J0465, 33F25J0822, 33F25K0254, 33F25L0376, 33F26C0300, 33F24B0119; AK-05502/10801902209518, 10801902210217, 10801902209525/33F24C0015, 33F24G0581, 33F24G0582, 33F24G0583, 33F24J0497, 33F24J0678, 33F24J0679, 33F24J0680, 33F24J0681, 33F24K0095, 33F24K0310, 33F24L0580, 33F24L0624, 33F24L0628, 33F24L0630, 33F24L0757, 33F24L0785, 33F25A0461, 33F25A0618, 33F25A0619, 33F25A0620, 33F25C0103, 33F25C0724, 33F25C0725, 33F25C0741, 33F25C0791, 33F25E0044, 33F25E0291, 33F25E0292, 33F25E0293, 33F25E0914, 33F25E1067, 33F25G0082, 33F25G0159, 33F25H0016, 33F25H0197, 33F25H0198, 33F25H0199, 33F25J0466, 33F25J0643, 33F25J0790, 33F25J0805, 33F25J0806, 33F25J0807, 33F25J0808, 33F25K0024, 33F25K0252, 33F25K0358, 33F25K0359, 33F25L0099, 33F25L0266, 33F25L0267, 33F25M0970, 33F25M0978, 33F26A1100, 33F24C0096; AK-05502D/10801902210217, 10801902209532/33F24G0331, 33F24J0498, 33F24K0096, 33F24L0713, 33F25A0462, 33F25C0745, 33F25E0045, 33F25H0017, 33F25J0467, 33F25L0366, 33F24C0017; AK-05503/10801902210217, 10801902209549/33F24J0277, 33F24J0499, 33F24K0097, 33F24L0674, 33F25A0463, 33F25C0949, 33F25E0046, 33F25E0915, 33F25H0018, 33F25J0823, 33F25L0406, 33F25M0912, 33F24C0018; AK-05503-L/10801902209549, 10801902210217, 10801902209549/33F24C0019, 33F24E0066, 33F24J0682, 33F24K0098, 33F24L0613, 33F24L0758, 33F25A0464, 33F25C0104, 33F25C0731, 33F25C0793, 33F25E0047, 33F25E0916, 33F25G0083, 33F25H0019, 33F25H0200, 33F25J0791, 33F25J0809, 33F25K0026, 33F25K0253, 33F25L0057, 33F25L0352, 33F25M0861, 33F26A0989, 33F24J0278; AK-05560/10801902210217, 10801902209556, 30801902121722/33F24J0501, 33F24K0099, 33F24L0552, 33F25A0145, 33F25A0465, 33F25C0743, 33F25E0048, 33F25E0917, 33F25G0084, 33F25H0020, 33F25J0737, 33F25J0824, 33F25M0895, 33F26B0538, 33F26C0592, 33F24A0812; ASK-02520-SBMC1/10801902209570, 10801902210217, 10801902193275/33F24D0103, 33F24F0275, 33F24G0396, 33F25A0137; ASK-05400-BW1/10801902209617, 10801902209631/33F24D0153, 33F24E0474, 33F24G0461, 33F24J0307, 33F24L0556, 33F25B0246, 33F25C0729, 33F25E0178, 33F25G0121, 33F25J0899, 33F25L0042, 33F25L0450, 33F24C0196; ASK-05400-GH/10801902209631, 10801902210217, 10801902202199/33F24D0154, 33F24E0480, 33F24G0292, 33F24J0310, 33F24J0600, 33F24L0597, 33F25A0153, 33F25C0161, 33F25C0754, 33F25E0181, 33F25G0123, 33F25J0532, 33F25L0391, 33F25M0897, 33F24D0155; ASK-05401-LMC/10801902210217, 10801902209662/33F24J0603, 33F25F0240, 33F25G0754, 33F25J0534, 33F24D0031; ASK-05401-NM/10801902210217/33F24E0104, 33F24G0295, 33F24J0311, 33F24J0604, 33F24K0223, 33F24L0559, 33F25A0548, 33F25C0840, 33F25E0184, 33F25G0125, 33F25H0113, 33F25J0535, 33F25L0024, 33F25L0361, 33F24E0027; ASK-05500-BID1/10801902210217/33F24E0490, 33F24G0297, 33F24H0488, 33F24J0607, 33F24L0594, 33F25A0551, 33F25C0823, 33F25E0187, 33F25G0127, 33F25J0536, 33F25L0438, 33F24K0228; ASK-05500-JHH1/10801902210217, 10801902209716, 10801902209723/33F24L0572, 33F25B0379, 33F25D0215, 33F25E0190, 33F25G0129, 33F25H0744, 33F25J0538, 33F25K0407, 33F25M0955, 33S24L0572, 33F24C0203; ASK-05500-MCV1/10801902210217/33F24E0494, 33F24E0495, 33F24G0470, 33F24J0611, 33F24K0229, 33F24L0707, 33F25A0554, 33F25C0784, 33F25E0191, 33F25E1005, 33F25G0761, 33F25H0115, 33F24E0496; ASK-05500-MVH/10801902209730, 10801902210217/33F24D0380, 33F24E0498, 33F24G0472, 33F24L0681; ASK-05500-UMC/10801902209754, 10801902210217/33F24D0382, 33F24J0316, 33F24K0233, 33F24L0085, 33F24L0772, 33F25C0166, 33F25C0737, 33F25E0196, 33F25E0317, 33F25E1008, 33F25H0120, 33F25J0540, 33F25L0397, 33F25M0904, 33F24E0502; ASK-05502-MLH/10801902210217/33F24J0618, 33F24K0239, 33F24L0555, 33F25A0562, 33F25C0996, 33F25E0201, 33F25H0125, 33F25J0544, 33F25L0440, 33F26B0977, 33F24L0583; ASK-05560-JMH1/10801902210217, 10801902209983/33F24G0483, 33F24L0773, 33F25A0566, 33F25C0894, 33F25E1014, 33F25H0130, 33F25J0548, 33F24E0519; ASK-05560-MKD2/10801902193275, 10801902210217/33F24K0246, 33F25G0468, 33F24J0476; ASK-05560-OM1/10801902210217, 10801902210002/33F24K0550, 33F25A0155, 33F25C0171, 33F25F0383, 33F25G0560, 33F25H0131, 33F25H0746, 33F25J0784, 33F25L0006, 33F25M0903, 33F24D0391; ASK-17019-SM/10801902210217, 10801902211221/33F25G0137, 33F25H0136, 33F25J0553, 33F25L0465, 33F24B0348; BP-05501/10801902210217, 10801902193275/33F24E0156, 33F24J0642, 33F24K0270, 33F25B0689, 33F25C0868, 33F25E1034, 33F25G0145, 33F25H0158, 33F25J0568, 33F25L0369, 33F26B0518, 33F25C0444; CK-05401/10801902202144, 10801902210217, 10801902193275/33F24C0256, 33F25A0591, 33F25C0972, 33F25E0256, 33F25E1040, 33F25H0756, 33F25J0935, 33F25L0244, 33F26B0620, 33F25D0674; HS-05501/10801902210217, 10801902202090/33F24E0580, 33F24J0329, 33F24B0389; MM-05501/10801902210217/33F24J0655, 33F24K0287, 33F24L0788, 33F25J0628, 33F25L0445, 33F24C0263; MP-17019-TK/10801902210217/33F24D0409, 33F24J0390, 33F24K0288, 33F24L0546, 33F25A0595, 33F25C0727, 33F25E0262, 33F25E1045, 33F25H0170, 33F25J0629, 33F25L0441, 33F25M0939, 33F24D0410; MP-17019-TKD/10801902210217, 10801902210224/33F24J0657, 33F24L0698, 33F25C0760, 33F25E0263, 33F25H0171, 33F25J0938, 33F24B0152; MP-17019-TKL/10801902210224, 10801902210217/33F24D0038, 33F24K0289, 33F24L0614, 33F25C0190, 33F25C0839, 33F25E0264, 33F25G0149, 33F25J0630, 33F25K0380, 33F24K0290; MSO-05401-JMH/10801902210217/33F25C0193, 33F25H0172, 33F25J0631, 33F25L0375, 33F25M0880, 33F24D0413; NM-05401/10801902210217, 10801902210262, 10801902140644/33F25J0944, 33F25M0968, 33F26B0600, 33F24C0780; SJ-05501/10801902210330, 10801902210217, 10801902202496/33F24C0049, 33F24C0050, 33F24J0382, 33F24J0383, 33F24J0384, 33F24J0385, 33F24J0386, 33F24K0552, 33F24K0553, 33F24K0554, 33F24K0555, 33F24K0556, 33F24K0557, 33F24L0625, 33F24L0721, 33F24L0760, 33F24L0786, 33F25A0605, 33F25A0623, 33F25A0624, 33F25C0202, 33F25C0789, 33F25C0790, 33F25C0945, 33F25C0997, 33F25E0278, 33F25E0295, 33F25E0296, 33F25E1057, 33F25E1068, 33F25G0154, 33F25G0160, 33F25H0183, 33F25H0202, 33F25H0203, 33F25J0639, 33F25J0644, 33F25J0798, 33F25J0811, 33F25J0812, 33F25J0813, 33F26B0601, 33F26C0299, 33F26C0307, 33F24A1336; SM-05401/10801902210217, 10801902210354, 10801902193275/33F24G0561, 33F24K0302, 33F24L0607, 33F25C0204, 33F25C0805, 33F25E0279, 33F25E1058, 33F25H0184, 33F25J0948, 33F26A1030, 33F24G0270; TM-05401/10801902210217, 30801902202155, 10801902210361/33F24D0418, 33F24G0563, 33F24J0669, 33F24L0089, 33F24L0741, 33F25C0207, 33F25C0918, 33F25E0282, 33F25E1059, 33F25J0950, 33F25L0447, 33F26C0308, 33F24B0410; TM-05502/10801902210378/33F24B0411; UI-05502/10801902210217/33F24G0565, 33F24J0670, 33F24L0562, 33F25A0607, 33F25C0881, 33F25E0283, 33F25E1060, 33F25G0752, 33F25H0186, 33F25J0951, 33F24A0359; ASK-05400-SHM/10801902210217/33F24E0482, 33F24G0465, 33F24J0601, 33F24K0221, 33F24L0800, 33F25C0162, 33F25C0888, 33F25E0182, 33F25H0741, 33F25L0457, 33F24E0483; ASK-05400-UOM/10801902210217/33F25G0124, 33F25H0742, 33F25K0263, 33F24G0294; ASK-05500-MR/10801902210217, 10801902209730/33F24E0497, 33F24G0471, 33F24K0230, 33F25A0555, 33F25C0796, 33F25E0192, 33F25H0116, 33F25K0408, 33F25M0854, 33F24C0204; ASK-05500-TM/10801902210217, 10801902209747, 10801902209754/33F25C0165, 33F25E1007, 33F25M0871, 33F26C0256, 33F24C0206; ASK-05501-BHL1/10801902210217/33F24L0566, 33F25A0558, 33F25C0810, 33F25E0197, 33F25E1009, 33F25H0121, 33F25J0541, 33F25L0439, 33F24E0503; ASK-05501-GH/10801902210217, 10801902210392/33F24K0235, 33F25A0559, 33F25E0198, 33F25H0122, 33F25J0542, 33F24D0384; ASK-05501-SM/10801902210217, 10801902210392, 10801902214857/33F24E0505, 33F24K0236, 33F25A0560, 33F25C0808, 33F25E0199, 33F25E1010, 33F25H0123, 33F25J0907, 33F26C0304, 33F24C0210; ASK-05502-BSM/10801902209853, 10801902210217/33F24D0385, 33F24E0508, 33F24G0301, 33F24K0237, 33F25A0154, 33F25B0687, 33F25H0124, 33F25J0543, 33F25K0409, 33F25L0466, 33F26B0592, 33F24E0509; ASK-05502-MGH1/10801902210217, 10801902209860/33F24J0617, 33F24L0569, 33F25B0247, 33F25C0938, 33F25E0304, 33F25G0133, 33F25J0909, 33F25L0098, 33F26B0545, 33F24J0318; ASK-05503-BWB1/10801902210217, 10801902209938, 10801902209952/33F24E0514, 33F24G0481, 33F24J0619, 33F24K0241, 33F24L0690, 33F25C0169, 33F25C0756, 33F25E1011, 33F25J0545, 33F25L0453, 33F26B0584, 33F24C0217; ASK-05503-HMH/10801902210217, 10801902209952, 10801902209969/33F24L0643, 33F25A0563, 33F25C0170, 33F25C0765, 33F25E1012, 33F25G0135, 33F25H0127, 33F25J0546, 33F25L0451, 33F26A1079, 33F24C0218; ASK-05503-SPH/10801902210217, 10801902209969, 10801902211405/33F24E0516, 33F24J0620, 33F24K0243, 33F24L0598, 33F25A0564, 33F25E0204, 33F25L0458, 33F25M0951, 33F24C0219; ASK-05503-UPM1/10801902211405, 10801902210217, 10801902209976/33F24D0389, 33F24E0160, 33F24J0621, 33F24K0244, 33F24L0549, 33F25A0565, 33F25C0992, 33F25E0205, 33F25E1013, 33F25H0129, 33F25J0547, 33F25L0454, 33F24D0390; ASK-05560-UCI2/10801902210217, 10801902193275/33F24G0485, 33F24J0623, 33F24L0593, 33F25A0567, 33F25C0893, 33F25E0206, 33F25E1015, 33F25H0132, 33F25J0911, 33F25L0061; ASK-17019-UMP/10801902210217, 10801902211221, 10801902210057/33F24J0628, 33F24K0250, 33F25B0694, 33F25E0216, 33F25H0137, 33F25J0916, 33F26B1000, 33F24B0350; MH-05503-1/10801902210217/33F25H0169, 33F24G0542 ASK-05401-VB2/10801902214840/33F24C0199, 33F24D0374, 33F24E0486, 33F24G0296, 33F24J0312, 33F24J0605, 33F24K0224, 33F24L0733, 33F25A0549; ASK-05500-SC1 /10801902202250/33F24B0319, 33F24C0205, 33F24D0381, 33F24E0499, 33F24G0299, 33F24J0315, 33F24J0613, 33F24K0232, 33F24L0567, 33F25A0557, 33F25C0884, 33F25E0194; ASK-05500-SC3/10801902223651/33F25G0751, 33F25H0118, 33F26B0644; ASK-05501-AFH1/10801902209761/33F24G0475; ASK-05501-BMC/10801902209785/33F24J0479; ASK-05502-NY/10801902209884/33F24E0512; ASK-17019-RMH/10801902214918/33F24B0346, 33F24E0531
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2371-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class I |
| Status | Ongoing |
| Recall started | April 21, 2026 |
| Reported | June 24, 2026 |
| Company | ARROW INTERNATIONAL, LLC |
| Company location | Morrisville, NC, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 1,525,831 |
More recalls from ARROW INTERNATIONAL, LLC · all medical device recalls · recalls from April 2026