Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F recall (ARROW INTERNATIONAL, LLC)
Medical deviceClass IOngoing
Summary: ARROW INTERNATIONAL, LLC is recalling Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 CM/ASK-15123-UOC1, AHDC KIT: 3-L 12 FR X 20 CM/ASK-15123-UPM, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-P1A, AHDC KIT: 3-L 12. Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
What is being recalled
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 CM/ASK-15123-UOC1, AHDC KIT: 3-L 12 FR X 20 CM/ASK-15123-UPM, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-P1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XP1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPCN1A, AHDC KIT: 3-L 12 FR X 16 CM/CDC-12123-XPN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-P1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPCN1A, AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XPN1A, AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XPCN1A, HEMODIALYSIS KIT 2L: 12 FR X 20CM AGB/AK-25122-CDC, HEMODIALYSIS KIT: 2-LUMEN 14 FR X 20 CM/AK-25142-F, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM1, HEMODIALYSIS KIT: 2L 12 FR X 6 IN/ASK-22122-MM2, AHDC KIT: 2L 14 FR X 15 CM/ASK-22142-UOC1, AHDC AGB KIT: 2-L 14 FR X 15 CM/ASK-22142-UPM, HEMODIALYSIS KIT: 2L 14 FR X 20 CM/ASK-25142-UNC1, AHDC KIT: 2L 14 FR X 20 CM/ASK-25142-UOC1, AHDC AGB KIT: 2L 14 FR X 20 CM/ASK-25142-UPM, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/ASK-26142-UNC1, HEMODIALYSIS KIT: 2LUMEN 12 FR X 16 CM/AU-22122-F, HEMODIALYSIS KIT: 2L 12 FR X 16 CM AGB/CDC-22122-1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-XN1A-U, HEMODIALYSIS KIT 2L: 14 FR X 15 CM AGB/CDC-22142-1A, HEMODIALYSIS KIT: 2L 14 FR X 15 CM AGB/CDC-22142-XC1A, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-XN1A-C, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XCN1A-U, AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-XN1A-U, HEMODIALYSIS KIT: 2L 12 FR X 13 CM AGB/CDC-23122-XU1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-X1A, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A, AHDC AGB U-BEND KIT: 2-L 12 FR X 20 CM/CDC-25122-XCN1A-U, AHDC AGB KIT: 2-L 12 FR X 20 CM/CDC-25122-XN1A, HEMODIALYSIS KIT: 2L 12 FR X 20 CM AGB/CDC-25122-XU1A, HEMODIALYSIS KIT 2L: 14 FR X 20 CM AGB/CDC-25142-1A, HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB/CDC-25142-XC1A, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XCN1A-C, AHDC AGB KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A, AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-XN1A-C, HEMODIALYSIS KIT: 2L 12 FR X 25 CM AGB/CDC-26122-1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XCN1A, AHDC AGB KIT: 2-L 12 FR X 25 CM/CDC-26122-XN1A, HEMODIALYSIS KIT 2L: 14 FR X 25 CM AGB/CDC-26142-1A, HEMODIALYSIS KIT: 2L 14 FR X 25 CM AGB/CDC-26142-X1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XCN1A, AHDC AGB KIT: 2-L 14 FR X 25 CM/CDC-26142-XN1A, HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-25122-F, AHDC AGB KIT: 2-L 12 FR X 16 CM/CDC-22122-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 16 CM/CDC-22123-XCN1A, LBCVC AGB KIT: 3-L 12 FR X 20 CM/CDC-25123-XCN1A, LBCVC AGB KIT:3-L 12 FR X 20 CM/CDC-25123-XN1A
Why
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
How serious is it
Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.
Where it was sold
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR. (states named: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, Wyoming)
How to tell if yours is affected
REF(Material)/UDI-DI/Lot(Batch): AK-15122-F/10801902197136, 10801902193275, 10801902168525/33F24E0348, 33F24J0280, 33F24K0119, 33F24L0601, 33F25A0477, 33F25C0908, 33F24A0850; ASK-12123-UOC1/10801902193275/33F25B0430, 33F25D0119; ASK-12123-UPM/10801902193275, 10801902222463/33F25E0535, 33F25G0188, 33F25G0189, 33F26C0333, 33F25D0262; ASK-15123-UOC1/10801902193275/33F25C0378, 33F25D0121; ASK-15123-UPM/10801902193275, 10801902159950/33F25E0718, 33F25H0447, 33F26A1389, 33F24E0649; CDC-12123-P1A/40801902111768, 10801902193275/33F24D0632, 33F24F0758, 33F24L0465, 33F25A0281, 33F25E0794; CDC-12123-XP1A/10801902193275, 30801902111785, 10801902213416/33F25J0723, 33F26A1382, 33F26B0433, 36F25A0007, 36F25B0009, 36F25E0018, 33F24D0299; CDC-12123-XPCN1A/10801902213416, 10801902193275/33F24D0950, 33F24K0417, 33F24L0466, 33F25A0216, 33F25B0667, 33F25D0436, 33F25D0666, 33F25E0693, 33F25G0239, 33F25J0247, 33F26A1556, 33F26B0801, 33F26C0767, 33F25B0668; CDC-12123-XPN1A/10801902193275, 10801902213409/33F25D0554, 33F25G0240, 33F25J0312, 33F26B0214,33F26C0199, 33F25A0181; CDC-15123-P1A/10801902193275/33F25D0437, 33F25H0547,33F25B0613; CDC-15123-XPCN1A/10801902213430, 10801902193275/33F24F0107, 33F24F0998, 33F24K0403, 33F24L0834, 33F25A0283, 33F25B0527, 33F25E0650, 33F25E0745, 33F25F0085, 33F25H0596, 33F25J0314, 33F26A1464, 33F26B0883, 33F25B0669; CDC-15123-XPN1A/0801902193275, 10801902213423, 10801902116793/33F25D0439, 33F25G0238, 33F25H0597, 33F25J0315, 33F26A1468, 33F26B0826, 33F24C0461; CDC-16123-XPCN1A/10801902193275/33F25B0671, 33F25D0440, 33F25E0651, 33F25G0236, 33F25H0529, 33F25J0316, 33F25G0237, AK-25122-CDC/10801902193275/33F24G0370, 33F24K0135, 33F25C0120, 33F25C0969, 33F25E0942, 33F24G0371; AK-25142-F/10801902193275, 70801902123205, 10801902167795/33F25C0863, 33F25E0087, 33F25H0038, 33F25J0846, 33F26B0531, 33F24D0089; ASK-22122-MM1/10801902193275/33F25E0546; ASK-22122-MM2/10801902193275, 10801902221923, 10801902144956/33F25G0172, 33F25L0031, 33F25M0310, 33F26A1652, 33F26B0151, 33F26B0450, 33F24F0854; ASK-22142-UOC1/10801902193275/33F25C0545, 33F25D0122; ASK-22142-UPM/10801902193275, 10801902222487, 10801902202281/33F25D0594, 33F25F0116, 33F25G0187, 33F26A1390, 33F26B0205, 33F24D0396; ASK-25142-UNC1/10801902193275/33F25C0655, 33F24K0561; ASK-25142-UOC1/10801902193275/33F25B0362, 33F25D0123; ASK-25142-UPM/10801902193275/33F25E0595, 33F25H0446, 33F25J0381, 33F26A1446, 33F25J0092; ASK-26142-UNC1/10801902193275/33F25B0427, 33F25K0058; AU-22122-F/10801902193275/33F24K0269, 33F25A0584, 33F25C0862, 33F25E0245, 33F25J0567; CDC-22122-1A/20801902123484, 10801902193275/, CDC-22122-XCN1A-U/33F24F0996, 33F24L0374, 33F25C0461; CDC-22122-XN1A/10801902193275/33F25C0239, 33F25E0875, 33F25J0419; CDC-22122-XN1A-U/10801902193275, 10801902213621/33F25A0309, 33F25E0876, 33F26B0217; CDC-22142-1A/50801902123515, 10801902193275/33F24D0928, 33F24F0103, 33F24F0896, 33F24L0111, 33F24L0955, 33F25B0500, 33F25D0447, 33F25E0654, 33F25G0278, 33F25H0577, 33F25J0319; CDC-22142-XC1A/10801902193275/33F25A0284, 33F25J0180; CDC-22142-XCN1A/10801902193275, 10801902213539/33F25A0285, 33F25B0529, 33F25D0449, 33F25E0655, 33F25E0818, 33F25F0214, 33F25G0233, 33F26A1477, 33F26B0898, 33R25J0455; CDC-22142-XCN1A-C/10801902193275/33F24L0473, 33F25B0615, 33F25F0213; CDC-22142-XN1A/10801902193275, 10801902213522/33F25C0240, 33F25E0578, 33F25F0215, 33F25H0528, 33F25J0321, 33F26A1383, 33F26B0215, 33F26C0016; CDC-22142-XN1A-C/10801902193275/33F25B0485, 33F25E0879; CDC-23122-XCN1A-U/10801902193275/33F24L0279, 33F25D0492; CDC-23122-XN1A-U/10801902193275/33F25E0579; CDC-23122-XU1A/10801902193275/33F25D0493, 33F25H0432, 33F25J0248; CDC-25122-1A/10801902193275/33F25B0619, 33F25E0881; CDC-25122-X1A/10801902193275/33F25E0882, 33F25J0422; CDC-25122-XCN1A/10801902213492, 10801902193275/33F24F0929, 33F25C0242, 33F25D0498, 33F25E0883, 33F25G0234, 33F26B0439, 33F26C0214; CDC-25122-XCN1A-U/10801902193275/33F24L1041, 33F25B0620, 33F25F0121, 33F25G0226; CDC-25122-XN1A/10801902193275, 10801902213485/33F25C0243, 33F25E0884, 33F25K0183, 33R25E0884; CDC-25122-XU1A/10801902193275/33F25B0621, 33F25E0886, 33F25H0568; CDC-25142-1A/50801902123584, 10801902193275/33F24F0837, 33F24K0457, 33F24L0113, 33F24L1031, 33F25B0622, 33F25D0503, 33F25E0889, 33F25F0106, 33F25J0127; CDC-25142-XC1A/10801902193275/33F25B0467; CDC-25142-XCN1A/10801902193275, 10801902213553/33F25A0312, 33F25E0657, 33F25E0821, 33F25G0231, 33F25H0527, 33F25J0428, 33F26B0452, 33F26C0133; CDC-25142-XCN1A-C/10801902193275/33F24L0377, 33F25D0238; CDC-25142-XN1A/10801902193275, 10801902213546/33F25B0624, 33F25D0504, 33F25E0823, 33F25G0232, 33F26B0437; CDC-25142-XN1A-C/10801902193275/33F24L0485, 33F25B0676; CDC-26122-1A/60801902123611, 10801902193275/33F24F0084, 33F24K0475, 33F25G0389; CDC-26122-XCN1A/10801902213515, 10801902193275/33F24F1028, 33F24L0117, 33F25D0308, 33F25D0557, 33F25F0219; CDC-26122-XN1A/10801902193275/33F25E0451; CDC-26142-1A/10801902193275/33F24L0488, 33F24L1025, 33F25C0368, 33F25E0891, 33F25H0406, 33F25J0116; CDC-26142-X1A/10801902193275/33F25D0558, 33F25E0892; CDC-26142-XCN1A/10801902193275, 10801902213577/33F24L0941, 33F25B0530, 33F25D0507, 33F25E0699, 33F25H0525, 33F26B0396, 33F26B0895, 33F26C0780; CDC-26142-XN1A/10801902193275, 10801902213560/33F24L0927, 33F25E0580, 33F25H0526, 33F26B0393; AK-25122-F/ CDC-22122-XCN1A/10801902193275/33F24G0371, 33F24L0742, 33F25C0822, 33F25E0084, 33F25J0483, 33F25L0400; CDC-22123-XCN1A/10801902213591, 10801902193275/33F24D0842,33F24F1018, 33F24L0110, 33F25C0205, 33F25D0613, 33F25E0878, 33F26B0455; CDC-25123-XCN1A/10801902213614, 10801902193275/33F24D0818, 33F24L0483, 33F25C0244, 33F25D0502, 33F25E0888, 33F25G0228, 33F25J0425, 33F26B0395, 33F26B0894; CDC-25123-XN1A/10801902193275/33F25C0332, 33F25G0229
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2388-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class I |
| Status | Ongoing |
| Recall started | April 21, 2026 |
| Reported | June 24, 2026 |
| Company | ARROW INTERNATIONAL, LLC |
| Company location | Morrisville, NC, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 62,719 |
More recalls from ARROW INTERNATIONAL, LLC · all medical device recalls · recalls from April 2026