Medline convenience kits: LOOP PACK DYNJ61460C OPEN HEART CDS840150Y ... recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline convenience kits: LOOP PACK DYNJ61460C OPEN HEART CDS840150Y PACEMAKER DYNJ56297C PACEMAKER PK DYNJT3454 PACEMAKER/ICD PACK DYNJ33822L PAD OPEN HEART DYNJ900962O PEDIATRIC OPEN HEART DYNJ54238L DYNJT6766 STAZ BASIC VASCULAR DYNJ909794C VASCULAR ACCESS KIT . Reason: Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
What is being recalled
Medline convenience kits: LOOP PACK DYNJ61460C OPEN HEART CDS840150Y PACEMAKER DYNJ56297C PACEMAKER PK DYNJT3454 PACEMAKER/ICD PACK DYNJ33822L PAD OPEN HEART DYNJ900962O PEDIATRIC OPEN HEART DYNJ54238L DYNJT6766 STAZ BASIC VASCULAR DYNJ909794C VASCULAR ACCESS KIT CVI5323A VASCULAR CDS CDS983040M
Why
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution. OUS distribution pending.
How to tell if yours is affected
DYNJ61460C UDI-DI 10198459422133 lot 25KBA098 CDS840150Y UDI-DI 10198459303029 lots 25JBW220 26BBD028 DYNJ56297C UDI-DI 10198459513954 lot 25LMJ874 DYNJT3454 UDI-DI 10198459325663 lot 25KBG761 DYNJ33822L UDI-DI 10198459385582 lot 26AMA072 DYNJ900962O UDI-DI 10198459452581 lot 26CBC302 DYNJ54238L UDI-DI 10198459443114 lot 25LMH549 DYNJT6766 UDI-DI 10198459580888 lot 25KBM870 DYNJ909794C UDI-DI 10198459528002 lot 25LBF702 CVI5323A UDI-DI 10198459582035 lot 25KBM758 CDS983040M UDI-DI 10198459579813 lots 25KBO682 26BBB926
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2418-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | April 27, 2026 |
| Reported | June 24, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 49,654 kits total |
More recalls from Medline Industries, LP · all medical device recalls · recalls from April 2026