Medline convenience kits: CMH CENTRAL LINE DRESSING CHANGE KIT EBSI1705A ... recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline convenience kits: CMH CENTRAL LINE DRESSING CHANGE KIT EBSI1705A CVC/PICC/MIDLINE DRESSING CHANGE KIT EBSI1057A DIALYSIS DRESSING CHANGE EBSI1498 EBSI CVC DRESSING CHANGE KIT EBSI1585 PICC DRESSING CHANGE TRAY EBSI1621 PORT ACCESS KIT EBSI1837. Reason: Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
What is being recalled
Medline convenience kits: CMH CENTRAL LINE DRESSING CHANGE KIT EBSI1705A CVC/PICC/MIDLINE DRESSING CHANGE KIT EBSI1057A DIALYSIS DRESSING CHANGE EBSI1498 EBSI CVC DRESSING CHANGE KIT EBSI1585 PICC DRESSING CHANGE TRAY EBSI1621 PORT ACCESS KIT EBSI1837
Why
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution. OUS distribution pending.
How to tell if yours is affected
EBSI1705A UDI-DI 10653160382850 Lot 2025103090 EBSI1057A UDI-DI 10653160356875 Lot 2025120890 EBSI1498 UDI-DI 10193489111538 Lot 2025110690 EBSI1585 UDI-DI 10653160354796 Lot 2025102190 EBSI1621 UDI-DI 10653160358336 Lot 2025110390 EBSI1837 UDI-DI 10653160388555 Lot 2025102990
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2438-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | April 27, 2026 |
| Reported | June 24, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 49,654 kits total |