Summary: Howmedica Osteonics Corp. is recalling TRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400.. Reason: Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
What is being recalled
TRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400.
Why
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution in the states of AZ, CA, GA, MI, NC, NJ, NY, WV. (states named: Arizona, California, Georgia, Michigan, North Carolina, New Jersey, New York, West Virginia)
How to tell if yours is affected
Part Number: 5521-B-400. UDI: (01)07613327026061(17)260921(10) TYX7OB. Lot Number: TYX7OB.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2462-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | May 19, 2026 |
| Reported | June 24, 2026 |
| Company | Howmedica Osteonics Corp. |
| Company location | Mahwah, NJ, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 7 units |