Recall Ledger

MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C. Reason: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

What is being recalled

MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C

Why

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI. (states named: California)

How to tell if yours is affected

UDI/DI 10197344149001 (each), 20197344149008 (case); ALL LOTS

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2484-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMay 4, 2026
ReportedJune 24, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity10980 units

More recalls from Medline Industries, LP · all medical device recalls · recalls from May 2026

Official FDA record