MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT . Reason: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

What is being recalled

MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.

Why

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI. (states named: California)

How to tell if yours is affected

Medline Kit Number/SKU DYNJ912271: UDI/DI 10198459688324 (each), 40198459688325 (case), Lot Number 26CBQ109; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 25FBO720; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBO387; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBO015; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBD819; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBA736; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24FBG144; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24DBO838; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24BBC453; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24ABM118; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 23KBS950; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 23IBC670; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22LBN883; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22KBF121; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22IBN293; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22FBO695; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22OBH349; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 21JBY498; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 21JBY507; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 21JBH105; Medline Kit Number/SKU DYNJ905368D: UDI/DI 10198459698149 (each), 40198459698140 (case), Lot Number 26DBE022; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25GBD045; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25EBN367; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25CBR397; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25BBK435; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 25ABR151; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24JBU152; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24IBD106; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24HBI890; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24GBS879; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24EBB182; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24CBT638; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 24CBN624; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23KBU577; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23IBS119; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23GBV028; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23HBO351; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23BBL739; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23BBL740; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23ABS355; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 23ABG430; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22GBG375; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22GBG376; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22GBG377; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22EBM264; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 22EBL230; Medline Kit Number/SKU DYNJ905368B: UDI/DI 10195327047917 (each), 40195327047918 (case), Lot Number 21LBV309; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21IMB152; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21HMD185; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21DMA957; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21BME627; Medline Kit Number/SKU DYNJ46530I: UDI/DI 10193489338843 (each), 40193489338844 (case), Lot Number 21BMC286.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2488-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMay 4, 2026
ReportedJune 24, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity1248 units