Recall Ledger

MEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling MEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C; 2) PK CUST CV A&B CABRINI CO, Medline Kit Number/SKU DYNJ60583C.. Reason: During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

What is being recalled

MEDLINE Medical Procedure Kits labeled as: 1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C; 2) PK CUST CV A&B CABRINI CO, Medline Kit Number/SKU DYNJ60583C.

Why

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI. (states named: California)

How to tell if yours is affected

Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21KME435; Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21JMC247; Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21JMB678; Medline Product Number/SKU DYNJ53973C: UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21IMH919; Medline Product Number/SKU DYNJ60583C: UDI/DI 10195327605582 (each), 40195327605583 (case), Lot Number 24CBO389; Medline Product Number/SKU DYNJ60583C: UDI/DI 10195327605582 (each), 40195327605583 (case), Lot Number 24ABR872.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2504-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMay 4, 2026
ReportedJune 24, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity168 units

More recalls from Medline Industries, LP · all medical device recalls · recalls from May 2026