Recall Ledger

Hillrom VOLARA System P.CIRCUIT 5KIT AC recall (Baxter Healthcare Corporation)

Medical deviceClass IOngoing

Summary: Baxter Healthcare Corporation is recalling Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device. Reason: There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

What is being recalled

Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device

Why

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

How serious is it

Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.

Where it was sold

Worldwide distribution - US Nationwide and the country of Canada.

How to tell if yours is affected

UDI: 10887761981499; Lot Numbers: All lot numbers distributed beginning on 4/28/2025

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2551-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass I
StatusOngoing
Recall startedMay 21, 2026
ReportedJuly 8, 2026
CompanyBaxter Healthcare Corporation
Company locationDeerfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity13881 units

More recalls from Baxter Healthcare Corporation · all medical device recalls · recalls from May 2026

Official FDA record