Octopus 4 Tissue Stabilizer recall (Medtronic Perfusion Systems)

Medical deviceClass IIOngoing

Summary: Medtronic Perfusion Systems is recalling Octopus 4 Tissue Stabilizer, Model 29400. Reason: During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

What is being recalled

Octopus 4 Tissue Stabilizer, Model 29400

Why

During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution.

How to tell if yours is affected

UDI-DI: 00763000543679; Serial Numbers: 0231266463 0231468923 0231478249 0231478253 0231478260 0231651627 0231651634 0231651636 0231651637 0231651638 0231651639 0231651641 0231651642

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2577-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMay 12, 2026
ReportedJuly 1, 2026
CompanyMedtronic Perfusion Systems
Company locationBrooklyn Park, MN, United States
Initiated byVoluntary: Firm initiated
Quantity590 units