Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2.... recall (Insulet Corporation)

Medical deviceClass IOngoing

Summary: Insulet Corporation is recalling Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6720. 6. BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod).. Reason: External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

What is being recalled

Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6720. 6. BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod).

Why

External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.

How serious is it

Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.

Where it was sold

Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.

How to tell if yours is affected

1. Model/REF Number: POD-OMNI-I1-6720. Pack UDI-DI: 20385083000579. Single Pod UDI-DI: 10385083000527. Lot Numbers: PH1U01032521, PH1U01102521, PH1U12102421, PH1U12172421, PH1U11212421, PH1U11252421, PH1U11222421, PH1U12092421, PH1U09112421, PH1U11262421, PH1U04032521, PH1U12022421, PH1U12182421, PH1U04302524, PH1U05022521, PH1U01222521, PR1U11202533, PH1U09122421, PH1U10102421, PH1U12122421, PH1U01302522, PH1U02222521, PH1U05052521, PH1U10042421, PH1U12102412, PH1U01312521, PH1U03312521, PH1U05012521, PH1U09122422, PH1U09162421, PH1U09192421, PH1U09212421, PH1U09262411, PH1U09262421, PH1U10112422, PH1U10122421, PH1U10172421, PH1U12132422, PH1U01132521, PH1U01172521, PH1U02132511, PH1U03202521, PH1U03262521, PH1U04042521, PR1U11192531, PH1U11192511, PR1U01162621, PH1U09172421, PH1U09182421, PH1U09202421, PH1U10142421, PH1U10152421, PH1U10162421, PH1U01232522, PH1U01312511, PH1U02032521, PH1U02042521, PH1U02212521, PH1U03282521, PH1U04092521, PH1U04102521, PH1U04112521, PH1U10172522, PH1U10182521, PH1U10202521, PH1U10212521, PH1U10222521, PH1U10232521, PH1U10242521, PH1U10252521, PH1U11152521, PH1U11172511, PH1U11172521, PH1U11182521, PR1U11182531, PR1U11182541, PR1U11192541, PH1U11192521, PR1U11192532, PR1U11192542, PR1U11202521, PR1U11202532, PR1U11202541, PR1U11202542, PR1U11212521, PR1U11222531, PR1U11222541, PR1U11222521, PR1U11232541, PR1U11242521, PR1U11242531, PR1U11252541, PR1U11252512, PR1U11252521, PR1U11262521, PR1U12012521, PR1U12022521, PR1U12032521, PR1U12102522, PR1U12112521, PR1U12122521, PR1U12132511, PR1U12132521, PR1U12152521, PR1U12162521, PR1U12172521, PR1U01132621, PR1U01152621, PR1U01172621, PR1U01202621, PR1U01212621, PR1U01222621, PR1U01232621, PR1U01272621, PR1U01302641, PR1U01312631, PR1U02022641, PR1U02032631, PR1U02032641, PR1U02042641, PR1U02052631, PR1U02052641, PR1U02062631, PR1U02112641, PR1U02132611, PR1U02132641, PR1U11252531, PR1U11262541, PH1U06282422. 2. Model/REF Number: POD-OMNI-I1-6220. Pack UDI-DI: 20385083000425. Single Pod UDI-DI: 10385083000411. Lot Numbers: PH1U10182422, PH1U02202521, PH1U04222521, PH1U04232511, PH1U04232521, PH1U11142521, PH1U12042521, PH1U12052521, PH1U01092621. 3. Model/REF Number: OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: PH1U01222522. 4. Model/REF Number: OMNI-I1-6220. Pack UDI-DI: 2038508300043. Single Pod UDI-DI: 10385083000411. Lot Numbers: PH1U04182521, PH1U04212521, PH1U12192411. 5. Model/REF Number: OMNI-I1-6720. Pack UDI-DI: 20385083000548. Single Pod UDI-DI: 10385083000527. Lot Numbers: PH1U01022521, PH1U12202421, PH1U12192421, PH1U10212421, PH1U10232421. 6. Model/REF Number: BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod). Pack UDI-DI: 20385083000135. Single Pod UDI-DI: 10385083000114. Lot Numbers: PH1U01072521, PH1U01242521, PH1U03082521, PH1U04172521, PH1U01272513.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2603-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass I
StatusOngoing
Recall startedMay 20, 2026
ReportedJuly 8, 2026
CompanyInsulet Corporation
Company locationActon, MA, United States
Initiated byVoluntary: Firm initiated
Quantity4,936,056 units