Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR ... recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903465K OPEN HEART DYNJ909090G SILVER CROSS OPEN HEART DYNJ906102M VPH OPEN HEART DYNJ902664M DYNJ902664N. Reason: Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
What is being recalled
Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903465K OPEN HEART DYNJ909090G SILVER CROSS OPEN HEART DYNJ906102M VPH OPEN HEART DYNJ902664M DYNJ902664N
Why
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI. (states named: Arizona, California, Florida, Illinois, Missouri, New Jersey, Ohio, Oklahoma, Tennessee, Texas, Wisconsin)
How to tell if yours is affected
DYNJ901328G UDI-DI 10198459601538 Lots 26ABD786 26CBD768 26DBJ434 DYNJ903465K UDI-DI 10198459292743 Lots 25GDB257 26ADA464 26BDA587 26DDA021 DYNJ909090G UDI-DI 10198459406027 Lots 25LBH690 25LBP697 DYNJ906102M UDI-DI 10198459406409 Lots 25JBE315 25KBI700 25KBM827 25LBB028 26ABH106 26CBQ206 DYNJ902664M UDI-DI 10198459349126 Lots 25ILA177 25KLA500 DYNJ902664N UDI-DI 10198459675287 Lot 26BLA783
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2607-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | May 28, 2026 |
| Reported | July 8, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 9,272 kits |