IntelliCuff recall (Hamilton Medical AG)
Medical deviceClass IIOngoing
Summary: Hamilton Medical AG is recalling IntelliCuff, Portable automatic cuff pressure controller REF: 951001. Reason: Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
What is being recalled
IntelliCuff, Portable automatic cuff pressure controller REF: 951001
Why
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MA, MI, ND, NE, NE, NY, OH, PA, PR, SD, TX, VA, WA, WI, WV, WY and the country of Canada. (states named: Arizona, California, Colorado, Florida, Georgia, Hawaii, Illinois, Kentucky, Massachusetts, Michigan, North Dakota, Nebraska, New York, Ohio, Pennsylvania, Puerto Rico, South Dakota, Texas, Virginia, Washington, Wisconsin, West Virginia, Wyoming)
How to tell if yours is affected
GTIN: 07630002800839/ Lot # range: 00001 -19732
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2625-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | May 22, 2026 |
| Reported | July 8, 2026 |
| Company | Hamilton Medical AG |
| Company location | Domat/Ems, N/A, Switzerland |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 962 units |