Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel ... recall (PHILIPS MEDICAL SYSTEMS)

Medical deviceClass IIOngoing

Summary: PHILIPS MEDICAL SYSTEMS is recalling Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 Upgrades. Reason: The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.

What is being recalled

Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 Upgrades

Why

The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide Distribution including Puerto Rico (states named: Puerto Rico)

How to tell if yours is affected

FCO72800812; FCO72800814 728305 728306 DI (01)00884838059474 728307 DI (01)00884838104600 728311 728231 (01)00884838083325 728232

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2650-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 11, 2026
ReportedAugust 5, 2026
CompanyPHILIPS MEDICAL SYSTEMS
Company locationCambridge, MA, United States
Initiated byFDA Mandated
Quantity128