Recall Ledger

VITEK 2 AST-N447 TEST KIT Product reference 424620 recall (Biomerieux Inc)

Medical deviceClass IIOngoing

Summary: Biomerieux Inc is recalling VITEK 2 AST-N447 TEST KIT Product reference 424620. Reason: Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.

What is being recalled

VITEK 2 AST-N447 TEST KIT Product reference 424620

Why

Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

International distribution to the countries of Belgium and the Netherlands.

How to tell if yours is affected

Lot 0853293604 UDI-DI 03573026634551 Product is not registered in the US. OUS only.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2715-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 4, 2026
ReportedJuly 15, 2026
CompanyBiomerieux Inc
Company locationHazelwood, MO, United States
Initiated byVoluntary: Firm initiated
Quantity45,900 cards

More recalls from Biomerieux Inc · all medical device recalls · recalls from June 2026

Official FDA record