Photera 600 3-D recall (Daavlin Distributing Company)
Medical deviceClass IIOngoing
Summary: Daavlin Distributing Company is recalling Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders. Reason: Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
What is being recalled
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Why
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution.
How to tell if yours is affected
Serial Numbers: 6-09278NB8CL, 6-09271NB8CL, 6-09055NB8CL, 6-09113NB8CL, 6-09204NB8CL, 6-09207NB8CL, 6-09116NB8CL, 6-09263NB8CL, 6-09272NB8CL, 6-09118NB8CL, 6-09275NB8CL, 6-09264NB8CL, 6-09060NB8CL, 6-09122NB8CL, 6-09216NB8CL, 6-09125NB8CL, 6-09353NB8CL, 6-09346NB8CL, 6-09128NB8CL, 6-09129NB8CL, 6-09133NB8CL, 6-09134NB8CL, 6-09266NB8CL, 6-09215NB8CL, 6-09355NB8CL, 6-09137NB8CL, 6-09359NB8CL, 6-09343NB8CL, 6-09140NB8CL, 6-09142NB8CL, 6-09143NB8CL, 6-09273NB8CL, 6-09065NB8CL, 6-09144NB8CL, 6-09237NB8CL, 6-09225NB8CL, 6-09145NB8CL, 6-09147NB8CL, 6-09148NB8CL, 6-09378NB8CL, 6-09252NB8CL, 6-09051NB8CL, 6-09260NB8CL, 6-09250NB8CL, 6-09149NB8CL, 6-09238NB8CL, 6-09212NB8CL, 6-09277NB8CL, 6-09048NB8CL, 6-09274NB8CL, 6-09259NB8CL, 6-09224NB8CL, 6-09057NB8CL, 6-09436NB12CL, 6-09438NB8CL, 6-09434NB12CL, 6-09270NB8CL, 6-09115NB8CL, 6-09111NB8CL, 6-09435NB12CL, 6-09124NB8CL, 6-09332NB12CL, 6-09223NB8CL, 6-09219NB8CL, 6-09234NB8CL, 6-09063NB8CL, 6-09232NB8CL, 6-09479NB12CL, 6-09373NB8CL, 6-09127NB8CL, 6-09247NB8CL, 6-09135NB8CL, 6-09344NB8CL, 6-09221NB8CL, 6-09050NB8CL, 6-09336NB12CL, 6-09049NB8CL, 6-09068NB8CL, 6-09331NB12CL, 6-09240NB8CL, 6-09235NB8CL, 6-09365NB8CL, 6-09205NB8CL, 6-09345NB8CL, 6-09202NB8CL, 6-09136NB8CL, 6-09201NB8CL, 6-09249NB8CL, 6-09437NB8CL, 6-09208NB8CL, 6-09434NB12CL, 6-09503NB8CL, 6-09132NB8CL, 6-09228NB8CL, 6-09257NB8CL, 6-09058NB8CL, 6-09210NB8CL, 6-09054NB8CL, 6-09062NB8CL, 6-09254NB8CL, 6-09256NB8CL, 6-09255NB8CL, 6-09229NB8CL, 6-09062NB8CL, 6-09203NB8CL, 6-09364NB8CL, 6-09236NB8CL, 6-09141NB8CL, 6-09357NB8CL, 6-09241NB8CL, 6-09213NB8CL, 6-09330NB12CL, 6-09226NB8CL, 6-09245NB8CL, 6-09379NB8CL, 6-09276NB8CL, 6-09381NB8CL, 6-09358NB8CL, 6-09110NB8CL, 6-09067NB8CL, 6-09380NB8CL, 6-09112NB8CL, 6-09261NB8CL, 6-09371NB8CL, 6-09114NB8CL, 6-09504NB8CL, 6-09514NB8CL, 6-09524NB8CL, 6-09117NB8CL, 6-09119NB8CL, 6-09377NB8CL, 6-09268NB8CL, 6-09269NB8CL, 6-09262NB8CL, 6-09120NB8CL, 6-09121NB8CL, 6-09123NB8CL, 6-09126NB8CL, 6-09356NB8CL, 6-09138NB8CL, 6-09246NB8CL, 6-09214NB8CL, 6-09265NB8CL, 6-09239NB8CL, 6-09374NB8CL, 6-09046NB8CL, 6-09347NB8CL, 6-09231NB8CL, 6-09146NB8CL, 6-09505NB8CL, 6-09244NB8CL, 6-09248NB8CL, 6-09369NB8CL, 6-09502NB8CL, 6-09372NB8CL, 6-09508NB8CL, 6-09047NB8CL, 6-09342NB8CL, 6-09501NB8CL, 6-09349NB8CL, 6-09360NB8CL, 6-09053NB8CL, 6-09375NB8CL, 6-09253NB8CL, 6-09066NB8CL, 6-09370NB8CL, 6-09333NB12CL, 6-09329NB12CL, 6-09334NB12CL, 6-09242NB8CL, 6-09350NB8CL, 6-09354NB8CL, 6-09064NB8CL, 6-09227NB8CL, 6-09243NB8CL, 6-09341NB8CL, 6-09130NB8CL, 6-09061NB8CL, 6-09131NB8CL, 6-09222NB8CL, 6-09206NB8CL, 6-09200NB8CL, 6-09139NB8CL, 6-09363NB8CL, 6-09217NB8CL, 6-09059NB8CL, 6-09506NB8CL, 6-09509NB8CL, 6-09376NB8CL, 6-09368NB8CL, 6-09362NB8CL, 6-09052NB8CL, 6-09220NB8CL, 6-09258NB8CL, 6-09209NB8CL, 6-09352NB8CL, 6-09218NB8CL, 6-09361NB8CL, 6-09211NB8CL, 6-09233NB8CL, 6-09230NB8CL, 6-09562NB12CL, 6-09335NB12CL, 6-09348NB8CL, 6-09513NB8CL, 6-09267NB8CL, 6-09351NB8CL, 6-09344NB8CL, 6-09365NB8CL, 6-09234NB8CL, 6-09331NB12CL, 6-09221NB8CL.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2744-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | March 5, 2026 |
| Reported | July 22, 2026 |
| Company | Daavlin Distributing Company |
| Company location | Bryan, OH, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 212 units |
More recalls from Daavlin Distributing Company · all medical device recalls · recalls from March 2026