Recall Ledger

FARADRIVE Steerable Sheath Clear IDE recall (Boston Scientific Corporation)

Medical deviceClass IIOngoing

Summary: Boston Scientific Corporation is recalling FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01. Reason: Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

What is being recalled

FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01

Why

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Global Distribution. US Nationwide.

How to tell if yours is affected

UDI-DI: 00810087181062; Lots: CL12910, CL12978, CL15153

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2754-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 11, 2026
ReportedJuly 29, 2026
CompanyBoston Scientific Corporation
Company locationSaint Paul, MN, United States
Initiated byVoluntary: Firm initiated
Quantity400 units

More recalls from Boston Scientific Corporation · all medical device recalls · recalls from June 2026

Official FDA record