FARADRIVE STEERABLE SHEATH recall (Boston Scientific Corporation)
Medical deviceClass IIOngoing
Summary: Boston Scientific Corporation is recalling FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01. Reason: Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
What is being recalled
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
Why
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Global Distribution. US Nationwide.
How to tell if yours is affected
UDI-DI: 00191506043131; Lots: CL12695, CL12696, CL12699, CL12700, CL12703, CL12808, CL12816, CL12817, CL12823, CL12834, CL12835, CL12836, CL12853, CL12860, CL12882, CL12883, CL12884, CL12891, CL12907, CL12908, CL12909, CL12919, CL12949, CL12950, CL12969, CL12976, CL12979, CL12980, CL13005, CL13010, CL13011, CL13012, CL13018, CL13032, CL13033, CL13034, CL13037, CL13056, CL13057, CL13058, CL13074, CL13075, CL13076, CL13077, CL13097, CL13098, CL13099, CL13100, CL13104, CL13115, CL13116, CL13117, CL13118, CL13119, CL13135, CL13136, CL13137, CL13138, CL13139, CL13156, CL13157, CL13191, CL13192, CL13193, CL13194, CL13197, CL13208, CL13209, CL13210, CL13215, CL13216, CL13217, CL13224, CL13229, CL13230, CL13257, CL13258, CL13288, CL13289, CL13337, CL13338, CL13365, CL13366, CL13384, CL13398, CL13399, CL13419, CL13420, CL13423, CL13467, CL13469, CL13470, CL13471, CL13472, CL13517, CL13525, CL13526, CL13527, CL13528, CL13535, CL13536, CL13541, CL13550, CL13551, CL13552, CL13553, CL13555, CL13564, CL13565, CL13566, CL13571, CL13574, CL13575, CL13576, CL13598, CL13599, CL13616, CL13637, CL13642, CL13643, CL13644, CL13665, CL13666, CL13667, CL13668, CL13704, CL13705, CL13706, CL13715, CL13716, CL13721, CL13734, CL13735, CL13736, CL13761, CL13762, CL13763, CL13798, CL13799, CL13800, CL13822, CL13825, CL13826, CL13835, CL13836, CL13837, CL13858, CL13859, CL13860, CL13912, CL13913, CL13914, CL13927, CL13928, CL13929, CL13942, CL13944, CL13945, CL13974, CL13975, CL13981, CL13990, CL13991, CL13992, CL14027, CL14028, CL14029, CL14040, CL14041, CL14042, CL14053, CL14054, CL14055, CL14079, CL14080, CL14081, CL14122, CL14123, CL14124, CL14139, CL14140, CL14141, CL14169, CL14172, CL14173, CL14174, CL14188, CL14189, CL14190, CL14227, CL14228, CL14229, CL14250, CL14251, CL14252, CL14288, CL14289, CL14290, CL14347, CL14348, CL14377, CL14378, CL14398, CL14399, CL14400, CL14440, CL14441, CL14442, CL14489, CL14490, CL14492, CL14493, CL14496, CL14502, CL14505, CL14506, CL14520, CL14583, CL14584, CL14585, CL14634, CL14636, CL14637, CL14671, CL14672, CL14673, CL14674, CL14698, CL14699, CL14700, CL14712, CL14713, CL14714, CL14715, CL14751, CL14752, CL14753, CL14771, CL14772, CL14773, CL14774, CL14815, CL14816, CL14817, CL14818, CL14856, CL14857, CL14858, CL14859, CL14872, CL14873, CL14874, CL14930, CL14931, CL14932, CL14936, CL14937, CL14938, CL14970, CL14971, CL14972, CL15006, CL15007, CL15010, CL15011, CL15068, CL15069, CL15070, CL15080, CL15089, CL15110, CL15130, CL15131, CL15132, CL15194, CL15195, CL15196, CL15241, CL15242, CL15243, CL15273, CL15274, CL15275, CL15276, CL15319, CL15320, CL15321, CL15322, CL15359, CL15360, CL15361, CL15362, CL15398, CL15399, CL15400, CL15401, CL15428, CL15429, CL15430, CL15431, CL15451, CL15475, CL15476, CL15477, CL15491, CL15492, CL15493, CL15494, CL15525, CL15540, CL15541, CL15548, CL15549, CL15570, CL15583, CL15584, CL15599, CL15603, CL15617, CL15641, CL15642, CL15718, CL15756, CL15769, CL15770, CL15777
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2756-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | June 11, 2026 |
| Reported | July 29, 2026 |
| Company | Boston Scientific Corporation |
| Company location | Saint Paul, MN, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 508,151 units |
More recalls from Boston Scientific Corporation · all medical device recalls · recalls from June 2026