Recall Ledger

Boston Scientific TheraSphere 360 Y-90 Management Platform recall (Boston Scientific Corporation)

Medical deviceClass IIOngoing

Summary: Boston Scientific Corporation is recalling Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.. Reason: Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.

What is being recalled

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

Why

Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY. (states named: Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Illinois, Indiana, Louisiana, Massachusetts, Minnesota, Nebraska, New York, Ohio, Oklahoma, Oregon, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington, Wisconsin, West Virginia, Wyoming)

How to tell if yours is affected

UDI/DI 00191506068264, All released software versions (V1.1.2, V1.2.1, V1.2.2. V2.0.1, V2.0.2)

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2771-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 19, 2026
ReportedJuly 29, 2026
CompanyBoston Scientific Corporation
Company locationMaple Grove, MN, United States
Initiated byVoluntary: Firm initiated
Quantity68 units

More recalls from Boston Scientific Corporation · all medical device recalls · recalls from June 2026