Recall Ledger

GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000... recall (Instrumentation Laboratory)

Medical deviceClass IIOngoing

Summary: Instrumentation Laboratory is recalling GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859. Reason: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

What is being recalled

GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859

Why

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

GEM Premier 5000 Part number: 00024019255, UDI: 08426950870227 GEM Premier 7000 Part number: 00000015279, UDI: 0195226000924 Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2821-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMay 8, 2026
ReportedAugust 5, 2026
CompanyInstrumentation Laboratory
Company locationBedford, MA, United States
Initiated byVoluntary: Firm initiated
Quantity122

More recalls from Instrumentation Laboratory · all medical device recalls · recalls from May 2026

Official FDA record