Recall Ledger

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number:... recall (TISSIUM SA)

Medical deviceClass IIOngoing

Summary: TISSIUM SA is recalling Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No. Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

What is being recalled

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Why

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA. (states named: California, Florida, Illinois, New Jersey, Ohio, Pennsylvania)

How to tell if yours is affected

Lot Code: Model No NCOCO2075; UDI-DI ¿03760279380100; Lot Number 250700016, 251000048, 251000049, and 260300043

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2904-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 15, 2026
ReportedAugust 12, 2026
CompanyTISSIUM SA
Company locationParis 12e Arrondissement, France
Initiated byVoluntary: Firm initiated

More recalls from TISSIUM SA · all medical device recalls · recalls from June 2026