Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number:... recall (TISSIUM SA)
Medical deviceClass IIOngoing
Summary: TISSIUM SA is recalling Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No. Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
What is being recalled
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Why
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA. (states named: California, Florida, Illinois, New Jersey, Ohio, Pennsylvania)
How to tell if yours is affected
Lot Code: Model No NCOCO5010; UDI-DI 03760279380179; Lot Numbers 250700050, 251000043, 260300048, 260300049, and 260300050
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2907-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | June 15, 2026 |
| Reported | August 12, 2026 |
| Company | TISSIUM SA |
| Company location | Paris 12e Arrondissement, France |
| Initiated by | Voluntary: Firm initiated |
More recalls from TISSIUM SA · all medical device recalls · recalls from June 2026