Recall Ledger

GETINGE Y SENSOR Module recall (Maquet Critical Care AB)

Medical deviceClass IIOngoing

Summary: Maquet Critical Care AB is recalling GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge.. Reason: The module did not undergo electrical safety tests and functional tests during production.

What is being recalled

GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge.

Why

The module did not undergo electrical safety tests and functional tests during production.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of CA, GA, MN, NY. (states named: California, Georgia, Minnesota, New York)

How to tell if yours is affected

DUI/DI 07325710000304, Serial/Lot Numbers: 21049, 21088, 21089, 21093, 21100, 21103 21108, 21114, 21115.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2963-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 30, 2026
ReportedAugust 26, 2026
CompanyMaquet Critical Care AB
Company locationSolna, Sweden
Initiated byVoluntary: Firm initiated
Quantity9 units

More recalls from Maquet Critical Care AB · all medical device recalls · recalls from June 2026