Fresenius Medical Care recall (Fresenius Medical Care Renal Therapies Group, LLC)

Medical deviceClass IIOngoing

Summary: Fresenius Medical Care Renal Therapies Group, LLC is recalling Fresenius Medical Care, 2008T BlueStar Hemodialysis Machine, Model Numbers: 191126, 191130.. Reason: Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) has received complaints regarding flow-, no water-, or temperature-related alarms with the 2008T¿ BlueStar Hemodialysis machines.

What is being recalled

Fresenius Medical Care, 2008T BlueStar Hemodialysis Machine, Model Numbers: 191126, 191130.

Why

Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) has received complaints regarding flow-, no water-, or temperature-related alarms with the 2008T¿ BlueStar Hemodialysis machines.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

191126, UDI/DI 00840861102099, Serial Numbers: 5T0S291116, 5T0S291119, 5T0S291125, 5T0S291135, 5T0S291137, 5T0S291138, 5T0S291139, 5T0S291140, 5T0S291141, 5T0S291143, 5T0S291144, 5T0S291145, 5T0S291146, 5T0S291148, 5T0S291149, 5T0S291150, 5T0S291151, 5T0S291152, 5T0S291154, 5T0S291155, 5T0S291156, 5T0S291157, 5T0S291158, 5T0S291159, 5T0S291160, 5T0S291161, 5T0S291162, 5T0S291163, 5T0S291164, 5T0S291165, 5T0S291166, 5T0S291168, 5T0S291169, 5T0S291170, 5T0S291173, 5T0S291177, 5T0S291178, 5T0S291179, 5T0S291180, 5T0S291181, 5T0S291182, 5T0S291183, 5T0S291184, 5T0S291185, 5T0S291186, 5T0S291187, 5T0S291188, 5T0S291189, 5T0S291190, 5T0S291191, 5T0S291192, 5T0S291193, 5T0S291194, 5T0S291195, 5T0S291196, 5T0S291197, 5T0S291198, 5T0S291199, 5T0S291200, 5T0S291201, 5T0S291202, 5T0S291203, 5T0S291204, 5T0S291205, 5T0S291206, 5T0S291207, 5T0S291208, 5T0S291209, 5T0S291210, 5T0S291211, 5T0S291212, 5T0S291213, 5T0S291214, 5T0S291215, 5T0S291216, 5T0S291217, 5T0S291218, 5T0S291219, 5T0S291220, 5T0S291221, 5T0S291222, 5T0S291223, 5T0S291224, 5T0S291225, 5T0S291226, 5T0S291228, 5T0S291229, 5T0S291230, 5T0S291231, 5T0S291232, 5T0S291234, 5T0S291235, 5T0S291242, 5T0S291243, 5T0S291244, 5T0S291245, 5T0S291247, 5T0S291248, 5T0S291249, 5T0S291250, 5T0S291251, 5T0S291252, 5T0S291253, 5T0S291314, 5T0S291315, 5T0S291316, 5T0S291317, 5T0S291318, 5T0S291319, 5T0S291320, 5T0S291321, 5T0S291322, 5T0S291323, 5T0S291324, 5T0S291325, 5T0S291326, 5T0S291327, 5T0S291328, 5T0S291329, 5T0S291330, 5T0S291332, 5T0S291333, 5T0S291334, 5T0S291335, 5T0S291337, 5T0S291338, 5T0S291339, 5T0S291340, 5T0S291341, 5T0S291342, 5T0S291343, 5T0S291346, 5T0S291403, 5T0S291407, 5T0S291408, 5T0S291409, 5T0S291410, 5T0S291411, 5T0S291412, 5T0S291413, 5T0S291414, 5T0S291415, 5T0S291416, 5T0S291417, 5T0S291418, 5T0S291421, 5T0S291423, 5T0S291424, 5T0S291425, 5T0S291426, 5T0S291427, 5T0S291428, 5T0S291433, 5T0S291435, 5T0S291436, 5T0S291438, 5T0S291440, 5T0S291441, 5T0S291442, 5T0S291444, 5T0S291446, 5T0S291447, 5T0S291450, 5T0S291452, 5T0S291458, 5T0S291460, 5T0S291461, 5T0S291463, 5T0S291464, 5T0S291466, 5T0S291467, 5T0S291472, 5T0S291474, 5T0S291476, 5T0S291479, 5T0S291480, 5T0S291526, 5T0S291556, 5T0S291584, 6T0S292446, 6T0S292448, 6T0S292460, 6T0S292461, 6T0S292463, 6T0S292464, 6T0S292465, 6T0S292466, 6T0S292467, 6T0S292468, 6T0S292469, 6T0S292470, 6T0S292471, 6T0S292472, 6T0S292473, 6T0S292474, 6T0S292475, 6T0S292476, 6T0S292477, 6T0S292478, 6T0S292479, 6T0S292480, 6T0S292481, 6T0S292482, 6T0S292483, 6T0S292484, 6T0S292485, 6T0S292486, 6T0S292487, 6T0S292488, 6T0S292489, 6T0S292490, 6T0S292491, 6T0S292492, 6T0S292493, 6T0S292494, 6T0S292495, 6T0S292496, 6T0S292498, 6T0S292499, 6T0S292500, 6T0S292501, 6T0S292502, 6T0S292503, 6T0S292504, 6T0S292505, 6T0S292506, 6T0S292507, 6T0S292508, 6T0S292509, 6T0S292510, 6T0S292511, 6T0S292512, 6T0S292513, 6T0S292514, 6T0S292515, 6T0S292516, 6T0S292517, 6T0S292518, 6T0S292519, 6T0S292520, 6T0S292521, 6T0S292522, 6T0S292523, 6T0S292524, 6T0S292525, 6T0S292526. 191130, UDI/DI 00840861102112, Serial Numbers: 6T0S291719, 6T0S291816, 6T0S291842, 6T0S291844, 6T0S291850, 6T0S291856, 6T0S291857, 6T0S291859, 6T0S291861, 6T0S291874, 6T0S291884, 6T0S291893, 6T0S291899, 6T0S291901, 6T0S291902, 6T0S291903, 6T0S291904, 6T0S291905, 6T0S291906, 6T0S291907, 6T0S291908, 6T0S291909, 6T0S291910, 6T0S291911, 6T0S291912, 6T0S291913, 6T0S291914, 6T0S291915, 6T0S291916, 6T0S291917, 6T0S291918, 6T0S291919, 6T0S291920, 6T0S291921, 6T0S291922, 6T0S291923, 6T0S291924, 6T0S291925, 6T0S291926, 6T0S291927, 6T0S291928, 6T0S291929, 6T0S291930, 6T0S291931, 6T0S291932, 6T0S291933, 6T0S291934, 6T0S291935, 6T0S291936, 6T0S291937, 6T0S291938, 6T0S291939, 6T0S291940, 6T0S291941, 6T0S291942, 6T0S291943, 6T0S291944, 6T0S291945, 6T0S291946, 6T0S291947, 6T0S291948, 6T0S291949, 6T0S291950, 6T0S291951, 6T0S291952, 6T0S291953, 6T0S291954, 6T0S291955, 6T0S291956, 6T0S291957, 6T0S291958, 6T0S291959, 6T0S291960, 6T0S291961, 6T0S291962, 6T0S291963, 6T0S291964, 6T0S291965, 6T0S291966, 6T0S291967, 6T0S291968, 6T0S291969, 6T0S291970, 6T0S291971, 6T0S291972, 6T0S291973, 6T0S291974, 6T0S291975, 6T0S291976, 6T0S291977, 6T0S291978, 6T0S291979, 6T0S291980, 6T0S291981, 6T0S291982, 6T0S291983, 6T0S291984, 6T0S291985, 6T0S291986, 6T0S291987, 6T0S291988, 6T0S291989, 6T0S291990, 6T0S291991, 6T0S291992, 6T0S291993, 6T0S291994, 6T0S291995, 6T0S291996, 6T0S291997, 6T0S291998, 6T0S291999, 6T0S292000, 6T0S292001, 6T0S292002, 6T0S292003, 6T0S292004, 6T0S292005, 6T0S292006, 6T0S292007, 6T0S292008, 6T0S292009, 6T0S292010, 6T0S292042, 6T0S292043, 6T0S292044, 6T0S292045, 6T0S292046, 6T0S292047, 6T0S292048, 6T0S292049, 6T0S292050, 6T0S292051, 6T0S292052, 6T0S292053, 6T0S292054, 6T0S292055, 6T0S292056, 6T0S292057, 6T0S292058, 6T0S292059, 6T0S292060, 6T0S292061, 6T0S292062, 6T0S292063, 6T0S292064, 6T0S292065, 6T0S292066, 6T0S292067, 6T0S292069, 6T0S292074, 6T0S292075, 6T0S292076, 6T0S292077, 6T0S292079, 6T0S292080, 6T0S292081, 6T0S292082, 6T0S292083, 6T0S292084, 6T0S292085, 6T0S292087, 6T0S292088, 6T0S292089, 6T0S292090, 6T0S292091, 6T0S292092, 6T0S292093, 6T0S292094, 6T0S292095, 6T0S292096, 6T0S292097, 6T0S292098, 6T0S292099, 6T0S292100, 6T0S292101, 6T0S292102, 6T0S292103, 6T0S292104, 6T0S292105, 6T0S292106, 6T0S292107, 6T0S292108, 6T0S292109, 6T0S292110, 6T0S292111, 6T0S292112, 6T0S292113, 6T0S292114, 6T0S292115, 6T0S292116, 6T0S292117, 6T0S292118, 6T0S292119, 6T0S292120, 6T0S292121, 6T0S292122, 6T0S292123, 6T0S292124, 6T0S292125, 6T0S292126, 6T0S292127, 6T0S292128, 6T0S292129, 6T0S292130, 6T0S292131, 6T0S292132, 6T0S292133, 6T0S292134, 6T0S292135, 6T0S292136, 6T0S292137, 6T0S292138, 6T0S292139, 6T0S292140, 6T0S292141, 6T0S292142, 6T0S292143, 6T0S292144, 6T0S292145, 6T0S292146, 6T0S292147, 6T0S292148, 6T0S292149, 6T0S292150, 6T0S292151, 6T0S292152, 6T0S292153, 6T0S292154, 6T0S292155, 6T0S292156, 6T0S292157, 6T0S292158, 6T0S292159, 6T0S292160, 6T0S292161, 6T0S292162, 6T0S292163, 6T0S292164, 6T0S292165, 6T0S292166, 6T0S292167, 6T0S292168, 6T0S292169, 6T0S292170, 6T0S292171, 6T0S292172, 6T0S292173, 6T0S292174, 6T0S292175, 6T0S292176, 6T0S292177, 6T0S292178, 6T0S292179, 6T0S292180, 6T0S292181, 6T0S292182, 6T0S292183, 6T0S292184, 6T0S292185, 6T0S292186, 6T0S292187, 6T0S292188, 6T0S292189, 6T0S292190, 6T0S292191, 6T0S292192, 6T0S292193, 6T0S292194, 6T0S292195, 6T0S292196, 6T0S292197, 6T0S292198, 6T0S292199, 6T0S292200, 6T0S292201, 6T0S292202, 6T0S292203, 6T0S292204, 6T0S292205, 6T0S292206, 6T0S292207, 6T0S292208, 6T0S292209, 6T0S292210, 6T0S292211, 6T0S292212, 6T0S292213, 6T0S292214, 6T0S292215, 6T0S292216, 6T0S292217, 6T0S292218, 6T0S292219, 6T0S292220, 6T0S292221, 6T0S292222, 6T0S292223, 6T0S292224, 6T0S292234, 6T0S292235, 6T0S292236, 6T0S292237, 6T0S292238, 6T0S292239, 6T0S292240, 6T0S292241, 6T0S292242, 6T0S292243, 6T0S292244, 6T0S292245, 6T0S292246, 6T0S292247, 6T0S292248, 6T0S292249, 6T0S292250, 6T0S292251, 6T0S292252, 6T0S292253, 6T0S292254, 6T0S292255, 6T0S292256, 6T0S292257, 6T0S292258, 6T0S292259, 6T0S292260, 6T0S292261, 6T0S292262, 6T0S292263, 6T0S292264, 6T0S292265, 6T0S292266, 6T0S292267, 6T0S292268, 6T0S292269, 6T0S292270, 6T0S292271, 6T0S292272, 6T0S292273, 6T0S292274, 6T0S292275, 6T0S292276, 6T0S292277, 6T0S292278, 6T0S292279, 6T0S292280, 6T0S292281, 6T0S292282, 6T0S292283, 6T0S292284, 6T0S292285, 6T0S292286, 6T0S292287, 6T0S292288, 6T0S292289, 6T0S292290, 6T0S292291, 6T0S292292, 6T0S292293, 6T0S292294, 6T0S292295, 6T0S292296, 6T0S292297, 6T0S292298, 6T0S292299, 6T0S292300, 6T0S292301, 6T0S292302, 6T0S292303, 6T0S292304, 6T0S292305, 6T0S292306, 6T0S292307, 6T0S292308, 6T0S292309, 6T0S292310, 6T0S292311, 6T0S292312, 6T0S292313, 6T0S292314, 6T0S292315, 6T0S292316, 6T0S292317, 6T0S292318, 6T0S292339, 6T0S292340, 6T0S292341, 6T0S292342, 6T0S292343, 6T0S292344, 6T0S292345, 6T0S292346, 6T0S292347, 6T0S292348, 6T0S292349, 6T0S292350, 6T0S292351, 6T0S292352, 6T0S292353, 6T0S292354, 6T0S292355, 6T0S292356, 6T0S292357, 6T0S292358, 6T0S292359, 6T0S292360, 6T0S292361, 6T0S292362, 6T0S292363, 6T0S292364, 6T0S292365, 6T0S292366, 6T0S292367, 6T0S292368, 6T0S292369, 6T0S292370, 6T0S292371, 6T0S292372, 6T0S292373, 6T0S292374, 6T0S292375, 6T0S292376, 6T0S292377, 6T0S292378, 6T0S292379, 6T0S292380, 6T0S292381, 6T0S292382, 6T0S292383, 6T0S292384, 6T0S292385, 6T0S292386, 6T0S292387, 6T0S292388, 6T0S292389, 6T0S292390, 6T0S292391, 6T0S292392, 6T0S292393, 6T0S292394, 6T0S292395, 6T0S292396, 6T0S292397, 6T0S292398, 6T0S292399, 6T0S292400, 6T0S292401, 6T0S292402, 6T0S292404, 6T0S292405, 6T0S292406, 6T0S292407, 6T0S292408, 6T0S292409, 6T0S292410, 6T0S292411, 6T0S292412, 6T0S292413, 6T0S292414, 6T0S292415, 6T0S292416, 6T0S292417, 6T0S292418, 6T0S292419, 6T0S292420, 6T0S292421, 6T0S292422, 6T0S292423, 6T0S292424, 6T0S292425, 6T0S292426, 6T0S292427, 6T0S292428, 6T0S292429, 6T0S292430, 6T0S292431, 6T0S292432, 6T0S292433, 6T0S292434, 6T0S292435, 6T0S292436, 6T0S292437, 6T0S292438, 6T0S292439, 6T0S292440, 6T0S292441, 6T0S292442, 6T0S292443, 6T0S292444, 6T0S292445, 6T0S292447, 6T0S292449, 6T0S292450, 6T0S292452, 6T0S292453, 6T0S292454, 6T0S292455, 6T0S292456, 6T0S292457, 6T0S292459, 6T0S292462, 6T0S292497, 6T0S292527, 6T0S292528, 6T0S292529, 6T0S292530, 6T0S292531, 6T0S292532, 6T0S292533, 6T0S292534, 6T0S292535, 6T0S292536, 6T0S292537, 6T0S292538, 6T0S292539, 6T0S292540, 6T0S292541, 6T0S292542, 6T0S292543, 6T0S292544, 6T0S292545, 6T0S292546, 6T0S292547, 6T0S292548, 6T0S292549, 6T0S292551, 6T0S292552, 6T0S292553, 6T0S292554, 6T0S292556, 6T0S292557, 6T0S292558, 6T0S292560, 6T0S292561, 6T0S292562, 6T0S292564, 6T0S292566, 6T0S292567, 6T0S292569, 6T0S292570, 6T0S292571, 6T0S292573, 6T0S292575, 6T0S292576, 6T0S292577, 6T0S292578, 6T0S292579, 6T0S292580, 6T0S292581, 6T0S292582, 6T0S292583, 6T0S292584, 6T0S292585, 6T0S292588, 6T0S292589, 6T0S292591, 6T0S292592, 6T0S292594, 6T0S292596, 6T0S292597, 6T0S292598, 6T0S292599, 6T0S292600, 6T0S292601, 6T0S292605, 6T0S292607, 6T0S292608, 6T0S292609, 6T0S292611, 6T0S292612, 6T0S292613, 6T0S292614, 6T0S292615, 6T0S292616, 6T0S292618, 6T0S292619, 6T0S292621, 6T0S292622, 6T0S292623, 6T0S292626, 6T0S292627, 6T0S292628, 6T0S292629, 6T0S292630, 6T0S292633, 6T0S292634, 6T0S292635, 6T0S292636, 6T0S292637, 6T0S292638, 6T0S292639, 6T0S292640, 6T0S292641, 6T0S292642, 6T0S292643, 6T0S292644, 6T0S292645, 6T0S292646, 6T0S292647, 6T0S292648, 6T0S292649, 6T0S292650, 6T0S292651, 6T0S292652, 6T0S292653, 6T0S292654, 6T0S292655, 6T0S292656, 6T0S292657, 6T0S292658, 6T0S292659, 6T0S292660, 6T0S292661, 6T0S292662, 6T0S292663, 6T0S292664, 6T0S292665, 6T0S292666, 6T0S292667, 6T0S292668, 6T0S292669, 6T0S292670, 6T0S292671, 6T0S292672, 6T0S292673, 6T0S292674, 6T0S292675, 6T0S292676, 6T0S292677, 6T0S292678, 6T0S292679, 6T0S292680, 6T0S292681, 6T0S292682, 6T0S292683, 6T0S292684, 6T0S292685, 6T0S292686, 6T0S292687, 6T0S292688, 6T0S292689, 6T0S292690, 6T0S292691, 6T0S292692, 6T0S292693, 6T0S292694, 6T0S292695, 6T0S292696, 6T0S292697, 6T0S292698, 6T0S292699, 6T0S292700, 6T0S292701, 6T0S292702, 6T0S292703, 6T0S292704, 6T0S292705, 6T0S292706, 6T0S292707, 6T0S292708, 6T0S292709, 6T0S292710, 6T0S292711, 6T0S292712, 6T0S292713, 6T0S292714, 6T0S292715, 6T0S292716, 6T0S292717, 6T0S292721, 6T0S292722, 6T0S292723, 6T0S292724, 6T0S292725, 6T0S292727, 6T0S292728, 6T0S292729, 6T0S292730, 6T0S292731, 6T0S292732, 6T0S292733, 6T0S292734, 6T0S292735, 6T0S292737, 6T0S292739, 6T0S292742, 6T0S292745, 6T0S292747, 6T0S292748, 6T0S292749, 6T0S292751, 6T0S292752, 6T0S292753, 6T0S292754, 6T0S292755, 6T0S292756, 6T0S292757, 6T0S292758, 6T0S292759, 6T0S292760, 6T0S292763, 6T0S292764, 6T0S292765, 6T0S292766, 6T0S292767, 6T0S292768, 6T0S292769, 6T0S292770, 6T0S292771, 6T0S292772, 6T0S292773, 6T0S292774, 6T0S292775, 6T0S292776, 6T0S292777, 6T0S292778, 6T0S292779, 6T0S292780, 6T0S292781, 6T0S292782, 6T0S292783, 6T0S292785, 6T0S292786, 6T0S292787, 6T0S292788, 6T0S292789, 6T0S292790, 6T0S292791, 6T0S292792, 6T0S292793, 6T0S292794, 6T0S292795, 6T0S292796, 6T0S292797, 6T0S292798, 6T0S292799, 6T0S292800, 6T0S292801, 6T0S292802, 6T0S292803, 6T0S292804, 6T0S292805, 6T0S292807, 6T0S292808, 6T0S292809, 6T0S292810, 6T0S292811, 6T0S292812, 6T0S292813, 6T0S292814, 6T0S292815, 6T0S292816, 6T0S292817, 6T0S292818, 6T0S292819, 6T0S292820, 6T0S292821, 6T0S292822, 6T0S292823, 6T0S292824, 6T0S292825, 6T0S292826, 6T0S292827, 6T0S292828, 6T0S292829, 6T0S292830, 6T0S292831, 6T0S292832, 6T0S292833, 6T0S292834, 6T0S292835, 6T0S292836, 6T0S292837, 6T0S292838, 6T0S292839, 6T0S292840, 6T0S292841, 6T0S292842, 6T0S292843, 6T0S292845, 6T0S292871, 6T0S292872, 6T0S292873, 6T0S292874, 6T0S292875, 6T0S292876, 6T0S292877, 6T0S292878, 6T0S292879, 6T0S292880, 6T0S292882, 6T0S292883, 6T0S292886, 6T0S292888, 6T0S292890, 6T0S292891, 6T0S292892, 6T0S292894, 6T0S292895, 6T0S292896, 6T0S292898, 6T0S292900, 6T0S292901, 6T0S292902, 6T0S292903, 6T0S292904, 6T0S292905, 6T0S292906, 6T0S292909, 6T0S292910, 6T0S292912, 6T0S292913, 6T0S292917, 6T0S292919, 6T0S292922, 6T0S292925, 6T0S292927, 6T0S292928, 6T0S292929, 6T0S292931, 6T0S292932, 6T0S292935, 6T0S292937, 6T0S292938, 6T0S292940, 6T0S292943, 6T0S292945, 6T0S292946, 6T0S292947, 6T0S292950, 6T0S292951, 6T0S292955, 6T0S292964, 6T0S292967, 6T0S292985, 6T0S293042, 6T0S293059, 6T0S293063, 6T0S293067, 6T0S293082, 6T0S293092, 6T0S293093, 6T0S293094, 6T0S293097, 6T0S293098, 6T0S293099, 6T0S293100, 6T0S293101, 6T0S293102, 6T0S293103, 6T0S293107, 6T0S293108, 6T0S293109, 6T0S293111, 6T0S293112, 6T0S293114, 6T0S293115, 6T0S293116, 6T0S293117, 6T0S293118, 6T0S293119, 6T0S293120, 6T0S293121, 6T0S293122, 6T0S293123, 6T0S293124, 6T0S293125, 6T0S293126, 6T0S293127, 6T0S293128, 6T0S293129, 6T0S293131, 6T0S293132, 6T0S293133, 6T0S293134, 6T0S293135, 6T0S293136, 6T0S293137, 6T0S293138, 6T0S293139, 6T0S293140, 6T0S293141, 6T0S293142, 6T0S293144, 6T0S293145, 6T0S293146, 6T0S293147, 6T0S293148, 6T0S293190, 6T0S293191, 6T0S293198, 6T0S293200, 6T0S293311, 6T0S293317.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3017-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 30, 2026
ReportedSeptember 2, 2026
CompanyFresenius Medical Care Renal Therapies Group, LLC
Company locationWaltham, MA, United States
Initiated byVoluntary: Firm initiated
Quantity1140 systems