Medline Convenience Kits: 1) CDS840396X recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) CDS840396X, Catalog Number: CDS840396X; 2) CDS840396Y, Catalog Number: CDS840396Y; 3) CDS983348AA, Catalog Number: CDS983348AA; 4) CDS983348AB, Catalog Number: CDS983348AB; 5) CDS983672O, Catalog Number: CDS983672O; 6) CDS983672P, Catalog Number: CDS983672P; . Reason: The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.

What is being recalled

Medline Convenience Kits: 1) CDS840396X, Catalog Number: CDS840396X; 2) CDS840396Y, Catalog Number: CDS840396Y; 3) CDS983348AA, Catalog Number: CDS983348AA; 4) CDS983348AB, Catalog Number: CDS983348AB; 5) CDS983672O, Catalog Number: CDS983672O; 6) CDS983672P, Catalog Number: CDS983672P; 7) DYKM1982, Catalog Number: DYKM1982; 8) DYNDA1080, Catalog Number: DYNDA1080; 9) DYNJ67490C, Catalog Number: DYNJ67490C; 10) DYNJ907400D, Catalog Number: DYNJ907400D; 11) DYNJ908990A, Catalog Number: DYNJ908990A

Why

The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

1) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 24ABC933; 2) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 24ABK089; 3) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 24ABL893; 4) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 25BBG163; 5) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24KBK977; 6) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24BBO585; 7) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24CBR417; 8) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24FBS930; 9) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24IBK539; 10) CDS983348AA, UDI-DI: 10195327385088(each), 40195327385089(case), Lot Number: 24ABM288; 11) CDS983348AA, UDI-DI: 10195327385088(each), 40195327385089(case), Lot Number: 24ABR888; 12) CDS983348AA, UDI-DI: 10195327385088(each), 40195327385089(case), Lot Number: 24EBB218; 13) CDS983348AA, UDI-DI: 10195327385088(each), 40195327385089(case), Lot Number: 24EBM057; 14) CDS983348AA, UDI-DI: 10195327385088(each), 40195327385089(case), Lot Number: 24FBH873; 15) CDS983348AA, UDI-DI: 10195327385088(each), 40195327385089(case), Lot Number: 24HBG929; 16) CDS983348AB, UDI-DI: 10198459136986(each), 40198459136987(case), Lot Number: 25ABK077; 17) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 24BBC413; 18) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 24CBQ574; 19) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 24IBC602; 20) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 24JBC919; 21) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 24JBK787; 22) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 24JBL738; 23) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 24KBB600; 24) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 24KBE775; 25) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 23JBR840; 26) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 23KBM370; 27) CDS983672O, UDI-DI: 10195327207625(each), 40195327207626(case), Lot Number: 23LBD735; 28) CDS983672P, UDI-DI: 10198459149139(each), 40198459149130(case), Lot Number: 25ABI289; 29) CDS983672P, UDI-DI: 10198459149139(each), 40198459149130(case), Lot Number: 24LBO787; 30) DYKM1982, UDI-DI: 10193489417357(each), 40193489417358(case), Lot Number: 25BBB356; 31) DYKM1982, UDI-DI: 10193489417357(each), 40193489417358(case), Lot Number: 23LBU657; 32) DYKM1982, UDI-DI: 10193489417357(each), 40193489417358(case), Lot Number: 24CBF023; 33) DYKM1982, UDI-DI: 10193489417357(each), 40193489417358(case), Lot Number: 24CBL355; 34) DYKM1982, UDI-DI: 10193489417357(each), 40193489417358(case), Lot Number: 24KBA176; 35) DYNDA1080, UDI-DI: 10884389388679(each), 40884389388670(case), Lot Number: 25BBB185; 36) DYNDA1080, UDI-DI: 10884389388679(each), 40884389388670(case), Lot Number: 24EBE506; 37) DYNDA1080, UDI-DI: 10884389388679(each), 40884389388670(case), Lot Number: 24GBK362; 38) DYNJ67490C, UDI-DI: 10195327631710(each), 40195327631711(case), Lot Number: 24BBN881; 39) DYNJ907400D, UDI-DI: 10195327538552(each), 40195327538553(case), Lot Number: 24BBL947; 40) DYNJ908990A, UDI-DI: 10195327505219(each), 40195327505210(case), Lot Number: 25ALA775; 41) DYNJ908990A, UDI-DI: 10195327505219(each), 40195327505210(case), Lot Number: 25CLA341; 42) DYNJ908990A, UDI-DI: 10195327505219(each), 40195327505210(case), Lot Number: 24ILA196; 43) DYNJ908990A, UDI-DI: 10195327505219(each), 40195327505210(case), Lot Number: 24JLA337; 44) DYNJ908990A, UDI-DI: 10195327505219(each), 40195327505210(case), Lot Number: 24LLA178

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3049-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 12, 2026
ReportedSeptember 9, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity4636 kits

More recalls from Medline Industries, LP · all medical device recalls · recalls from June 2026