Medline Convenience Kits: 1) BLOOD DRAW KIT recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD PHLEBOTOMY KIT, Catalog Number: DYNDB3005D; 3) QP KIT, Catalog Number: DYNDB3005F; 4) DOUBLE RED CELL PHLEBOTOMY KIT, Catalog Number: DYNDB3006C; 5) PLATELET PHLEBOTOMY KIT, Catalog Number: DYNDB3007C; . Reason: The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.

What is being recalled

Medline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD PHLEBOTOMY KIT, Catalog Number: DYNDB3005D; 3) QP KIT, Catalog Number: DYNDB3005F; 4) DOUBLE RED CELL PHLEBOTOMY KIT, Catalog Number: DYNDB3006C; 5) PLATELET PHLEBOTOMY KIT, Catalog Number: DYNDB3007C; 6) WHOLE BLOOD KIT, Catalog Number: DYNDB3041D; 7) PHLEBOTOMY, Catalog Number: DYNDB3044D; 8) PHLEBOTOMY, Catalog Number: DYNDB3044F; 9) DONATION - WB, Catalog Number: DYNDB3086D; 10) DONATION - ALYX/TRIMA W/O TOUR, Catalog Number: DYNDB3088B; 11) THERAPEUTIC NFT, Catalog Number: DYNDB3100; 12) BLOOD CULTURE TRAY ADULT, Catalog Number: DYNDH1194A; 13) ADULT BLOOD CULTURE KIT, Catalog Number: DYNDH1444D; 14) PEDS BLOOD CULTURE KIT, Catalog Number: DYNDH1445C; 15) BLOOD CULTURE KIT, Catalog Number: DYNDH1462B; 16) BLOOD CULTURE ADULT, Catalog Number: DYNDH1619A; 17) BLOOD CULTURE PEDI, Catalog Number: DYNDH1620A; 18) BLOOD CULTURE KIT, Catalog Number: DYNDH1622C; 19) BLOOD CULTURE KIT, Catalog Number: DYNDH1622D; 20) BLOOD CULTURE COLLECTION KIT, Catalog Number: DYNDH1626B; 21) ADULT BLOOD CULTURE KIT, Catalog Number: DYNDH1694A; 22) ADULT BLOOD CULTURE KIT, Catalog Number: DYNDH1695; 23) BLOOD CULTURE KIT, Catalog Number: DYNDH1708A; 24) BLOOD CULTURE COLLECTION KIT, Catalog Number: DYNDH1714A; 25) BLOOD CULTURE KIT, Catalog Number: DYNDH1715B; 26) BLOOD CULTURE KIT, Catalog Number: DYNDH1723; 27) NS BLOOD CULTURE 23G KIT, Catalog Number: DYNDH1740A; 28) NS BLOOD CULTURE 23G KIT, Catalog Number: DYNDH1740B; 29) NS BLOOD CULTURE 21G KIT, Catalog Number: DYNDH1741A; 30) NS BLOOD CULTURE 21G KIT, Catalog Number: DYNDH1741B; 31) PEDS BLOOD CULTURE KIT, Catalog Number: DYNDH1771A; 32) BLOOD CULTURE KIT, Catalog Number: DYNDH1805B; 33) BLOOD CULTURE KIT, Catalog Number: DYNDH1817A; 34) BLOOD CULTR KT DIVRSN DEVICE, Catalog Number: DYNDH1866; 35) BLOOD CULTURE KIT, Catalog Number: DYNDH1867; 36) BLOOD CULTURE KIT, Catalog Number: DYNDH1884; 37) PHLEBOTOMY BLOOD CULTURE KIT, Catalog Number: DYNDH1908; 38) BLOOD CULTURE KIT, Catalog Number: DYNDH1912; 39) BLOOD CULTURE KIT, Catalog Number: DYNDH1940; 40) BLOOD CULTURE KIT W/O BOTTLES, Catalog Number: DYNDH1985; 41) BLOOD CULTURE KIT, Catalog Number: DYNDH1988; 42) BLOOD CULTURE, Catalog Number: DYNDH2002; 43) BLOOD CULTURE W/NO IV, Catalog Number: DYNDH2016; 44) BLOOD CULTURE W/IV, Catalog Number: DYNDH2017; 45) BLOOD CULTURE KIT, Catalog Number: DYNDH2023; 46) BLOOD CULTURE KIT, Catalog Number: DYNDH2028; 47) BLOOD CULTURE KIT, Catalog Number: DYNDH2030; 48) BLOOD CULTURE KIT, Catalog Number: DYNDH2048; 49) BLOOD CULTURE KIT, Catalog Number: DYNDH2059; 50) BLOOD CULTURE KIT 450, Catalog Number: DYNDH2074; 51) BLOOD CULTURE-NB, Catalog Number: DYNDH2089

Why

The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

1) DYLAB1081, UDI-DI: 10198459199608(each), 40198459199609(case), Lot Number: 25CMC514; 2) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 24CBG614; 3) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 24CBT912; 4) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 23HBH298; 5) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 23JBN521; 6) DYNDB3005D, UDI-DI: 10195327038700(each), 40195327038701(case), Lot Number: 23KBT270; 7) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 25BBD241; 8) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 25BBP477; 9) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 25CBI161; 10) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24LBM458; 11) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24FBL961; 12) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24FBS200; 13) DYNDB3005F, UDI-DI: 10198459025068(each), 40198459025069(case), Lot Number: 24IBO183; 14) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 24ABB637; 15) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 24ABY532; 16) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 24CBH932; 17) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 23JBN434; 18) DYNDB3006C, UDI-DI: 10195327038717(each), 40195327038718(case), Lot Number: 23LBH909; 19) DYNDB3007C, UDI-DI: 10195327038724(each), 40195327038725(case), Lot Number: 24CBL044; 20) DYNDB3007C, UDI-DI: 10195327038724(each), 40195327038725(case), Lot Number: 23JBC957; 21) DYNDB3007C, UDI-DI: 10195327038724(each), 40195327038725(case), Lot Number: 23KBT835; 22) DYNDB3007C, UDI-DI: 10195327038724(each), 40195327038725(case), Lot Number: 23LBI822; 23) DYNDB3041D, UDI-DI: 10195327170738(each), 40195327170739(case), Lot Number: 24KBS893; 24) DYNDB3044D, UDI-DI: 10195327028565(each), 40195327028566(case), Lot Number: 24ABH596; 25) DYNDB3044D, UDI-DI: 10195327028565(each), 40195327028566(case), Lot Number: 23KBB360; 26) DYNDB3044D, UDI-DI: 10195327028565(each), 40195327028566(case), Lot Number: 23LBI693; 27) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 25BBA987; 28) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24LBE163; 29) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24DBO764; 30) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24DBO786; 31) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24EBD140; 32) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24GBR601; 33) DYNDB3044F, UDI-DI: 10195327661236(each), 40195327661237(case), Lot Number: 24IBR933; 34) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24LBR503; 35) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 23LBM024; 36) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24BBM148; 37) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24CBN122; 38) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24FBC084; 39) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24JBV595; 40) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 23JBL111; 41) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 25ABB075; 42) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24BBO180; 43) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24CBN068; 44) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24DBW006; 45) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24IBC763; 46) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24JBA762; 47) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 24JBX189; 48) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 23JBY503; 49) DYNDB3088B, UDI-DI: 10195327086275(each), 40195327086276(case), Lot Number: 23LBM098; 50) DYNDB3100, UDI-DI: 10198459028915(each), 40198459028916(case), Lot Number: 24GBF854; 51) DYNDH1194A, UDI-DI: 10195327432300(each), 40195327432301(case), Lot Number: 23KBQ229; 52) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 25ABW691; 53) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 25DBO475; 54) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 25EBD598; 55) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 25FBQ279; 56) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 23LBU409; 57) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 24CBQ394; 58) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 24EBI903; 59) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 24GBH145; 60) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 23JBI911; 61) DYNDH1444D, UDI-DI: 10195327502492(each), 40195327502493(case), Lot Number: 23JBQ284; 62) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24LBU139; 63) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 25BBE915; 64) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 25EBD601; 65) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24ABT708; 66) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24CBH112; 67) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24DBS252; 68) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 24FBH711; 69) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 23KBH644; 70) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 23KBV432; 71) DYNDH1445C, UDI-DI: 10195327502614(each), 40195327502615(case), Lot Number: 23LBJ808; 72) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 25ABK267; 73) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 25BBB147; 74) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24CBU138; 75) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24DBJ003; 76) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24DBP521; 77) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24EBA062; 78) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24EBU164; 79) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24FBD281; 80) DYNDH1462B, UDI-DI: 24FBK195(each), 10195327662462(case), Lot Number: 40195327662463; 81) DYNDH1462B, UDI-DI: 24FBS941(each), 10195327662462(case), Lot Number: 40195327662463; 82) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24GBH144; 83) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24GBX678; 84) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24HBG621; 85) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24HBJ952; 86) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24HBT054; 87) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 24JBA557; 88) DYNDH1462B, UDI-DI: 10195327662462(each), 40195327662463(case), Lot Number: 25ABK267Z; 89) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 25ABC488; 90) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 25ABO279; 91) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24LBD413; 92) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24ABL303; 93) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24ABY205; 94) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24CBJ142; 95) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24CBT682; 96) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24EBI894; 97) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24FBS943; 98) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24GBT335; 99) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24IBL862; 100) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24IBV950; 101) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24JBB865; 102) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24JBI309; 103) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24KBE071; 104) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 23IBL873; 105) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 23JBH261; 106) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 23KBQ211; 107) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24LBU144; 108) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24LBI541; 109) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24DBO997; 110) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24GBC034; 111) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24KBH535; 112) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 23KBV431; 113) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 25EBD604; 114) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24ABK078; 115) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24BBB587; 116) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24BBV398; 117) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24CBJ143; 118) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24CBU661; 119) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24EBU004; 120) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24FBK924; 121) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24GBH140; 122) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24HBT637; 123) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24JBB866; 124) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 24JBF575; 125) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 23JBQ525; 126) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 23KBW274; 127) DYNDH1622C, UDI-DI: 10195327509873(each), 40195327509874(case), Lot Number: 23LBL909; 128) DYNDH1622D, UDI-DI: 10198459234965(each), 40198459234966(case), Lot Number: 25GBA905; 129) DYNDH1622D, UDI-DI: 10198459234965(each), 40198459234966(case), Lot Number: 25HBN203; 130) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24LBI550; 131) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24ABI955; 132) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24BBB590; 133) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24CBJ146; 134) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24DBJ006; 135) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24EBA605; 136) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24GBH142; 137) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24IBL883; 138) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24JBI311; 139) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 24JBW009; 140) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 23IBI210; 141) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 23KBH642; 142) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 23KBL369; 143) DYNDH1626B, UDI-DI: 10195327323332(each), 40195327323333(case), Lot Number: 23LBD097; 144) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24LBU146; 145) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 25ABK273; 146) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 25BBB149; 147) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24LBI542; 148) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24ABF793; 149) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24ABT705; 150) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24BBI818; 151) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24BBU666; 152) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24DBE739; 153) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24GBH139; 154) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24IBB026; 155) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24IBE383; 156) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24IBL900; 157) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24IBV952; 158) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24JBF571; 159) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24JBW011; 160) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24KBH537; 161) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23JBC264; 162) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23JBH262; 163) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23KBI725; 164) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23KBQ208; 165) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23LBD095; 166) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 23LBF186; 167) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 25AMI450; 168) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 25BMD501; 169) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 25CME092; 170) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24GMH831; 171) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24IMD773; 172) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24IMI239; 173) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24JMC905; 174) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24JMG344; 175) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24KMH750; 176) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24LMC717; 177) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24BBB589; 178) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24BBR118; 179) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24CBR792; 180) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24EBA607; 181) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 23JBP690; 182) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 23LBG924; 183) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 25ABT228; 184) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 25BBM180; 185) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24KBN935; 186) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24LBI551; 187) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24ABM740; 188) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24DBG375; 189) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24DBH805; 190) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24EBA011; 191) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24GBH136; 192) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 23IBG944; 193) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 23KBL592; 194) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 23LBE163; 195) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 25ABT229; 196) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 25BBI372; 197) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24KBN938; 198) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24LBI548; 199) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24ABG625; 200) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24BBB584; 201) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24DBA250; 202) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24FBB687; 203) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24GBL430; 204) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24IBL892; 205) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24JBG988; 206) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24JBI307; 207) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24JBW010; 208) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 23KBC138; 209) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 23KBT360; 210) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 23LBF190; 211) DYNDH1715B, UDI-DI: 10195327407094(each), 40195327407095(case), Lot Number: 24IMF148; 212) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 25ABK268; 213) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 25BBG725; 214) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24LBN286; 215) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24ABX116; 216) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24CBP727; 217) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24CBR793; 218) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24IBL881; 219) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24JBN261; 220) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24JBW018; 221) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 23JBO541; 222) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 23KBI724; 223) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 23LBF189; 224) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 24ABM651; 225) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 24ABW198; 226) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 24DBQ460; 227) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23JBL839; 228) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23JBS096; 229) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23JBW106; 230) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23KBH643; 231) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23KBM453; 232) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23KBV433; 233) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23LBK910; 234) DYNDH1740A, UDI-DI: 10195327192105(each), 40195327192106(case), Lot Number: 23LBM124; 235) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 25ABC762; 236) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 25ABF383; 237) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 25BBC863; 238) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 25JBA059; 239) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24LBC961; 240) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24EBA610; 241) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24GBT336; 242) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24IBK137; 243) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24JBW017; 244) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 24ABM652; 245) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 24BBH982; 246) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 24CBQ396; 247) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 24CBR795; 248) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23IBS931; 249) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23JBD149; 250) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23KBJ502; 251) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23LBF192; 252) DYNDH1741A, UDI-DI: 10195327192099(each), 40195327192090(case), Lot Number: 23LBM125; 253) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 25ABC763; 254) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 25ABK266; 255) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 25ABT230; 256) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24KBN936; 257) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24LBI543; 258) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24IBK132; 259) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24JBS020; 260) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24KBH536; 261) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 24LBU145; 262) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 25ABK265; 263) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 25ABO639; 264) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 25BBG724; 265) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 25CBG204; 266) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 24JBG779; 267) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 24JBQ113; 268) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 24JBS994; 269) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25AMD341; 270) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25AMG536; 271) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25BMB721; 272) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25BMH602; 273) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 25CME260; 274) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24DMK121; 275) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24EMK057; 276) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24FMH384; 277) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24GMB446; 278) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24GMG550; 279) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24IMD770; 280) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24IMI240; 281) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24JMA923; 282) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24JMC906; 283) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24JMG338; 284) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24LMA311; 285) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24LMG806; 286) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24ABX115; 287) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 24CBI261; 288) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 23LBD096; 289) DYNDH1805B, UDI-DI: 10195327518455(each), 40195327518456(case), Lot Number: 23LBF187; 290) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24LBD415; 291) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 23LBF191; 292) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24CBQ393; 293) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24EBA611; 294) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24GBT338; 295) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24IBK136; 296) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 24JBW014; 297) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 23JBQ200; 298) DYNDH1817A, UDI-DI: 10195327450755(each), 40195327450756(case), Lot Number: 23LBE911; 299) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 25ABC487; 300) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 25ABO282; 301) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 25BBC864; 302) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 23LBU410; 303) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24ABL891; 304) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24BBI806; 305) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24CBB996; 306) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24CBR796; 307) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24DBP522; 308) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24EBA612; 309) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24EBI902; 310) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24EBT285; 311) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24FBH707; 312) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24FBN853; 313) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24GBB459; 314) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24GBT339; 315) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24HBB568; 316) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24JBB782; 317) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 24JBI308; 318) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 23JBP541; 319) DYNDH1866, UDI-DI: 10195327400958(each), 40195327400959(case), Lot Number: 23LBB120; 320) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 23LBU408; 321) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 24BBR119; 322) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 24CBJ147; 323) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 24DBJ005; 324) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 24EBA588; 325) DYNDH1867, UDI-DI: 10195327400965(each), 40195327400966(case), Lot Number: 23JBM365; 326) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25AMD344; 327) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25AMG374; 328) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25BMB195; 329) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25BMD502; 330) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 25BMH148; 331) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24FMC615; 332) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24FMH383; 333) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24GMG552; 334) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24GMI903; 335) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24HMI304; 336) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24IMD766; 337) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24IMG695; 338) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24IMI237; 339) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24JMA921; 340) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24JMG345; 341) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24KMD105; 342) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24LMC718; 343) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24LMG808; 344) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24ABF789; 345) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24ABT706; 346) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24BBU664; 347) DYNDH1884, UDI-DI: 10195327421021(each), 40195327421022(case), Lot Number: 24DBO706; 348) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24KBU495; 349) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24ABF388; 350) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24ABI775; 351) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24ABO844; 352) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24CBJ647; 353) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24GBL009; 354) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 23JBL322; 355) DYNDH1912, UDI-DI: 10195327476939(each), 40195327476930(case), Lot Number: 23JBK266; 356) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 25BMD500; 357) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 25BMH149; 358) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 25CMH786; 359) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24DMF413; 360) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24EMC278; 361) DYNDH1940, UDI-DI: 10195327526535( each), 40195327526536(case), Lot Number: 24EMK058; 362) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24FME926; 363) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24FMF294; 364) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24GMB440; 365) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24GMF833; 366) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24GMG551; 367) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24GMI904; 368) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24HMC740; 369) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24HMI305; 370) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24IMD767; 371) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24IMG698; 372) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24IMI238; 373) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24JMA922; 374) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24JMG343; 375) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24JMI885; 376) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24LMA312; 377) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24LMC719; 378) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24LMG811; 379) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24ABF792; 380) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24BBB873; 381) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24BBH981; 382) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24CBI260; 383) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 23KBC202; 384) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 23KBI726; 385) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 23KBQ207; 386) DYNDH1985, UDI-DI: 10195327623043(each), 40195327623044(case), Lot Number: 24DBP083; 387) DYNDH1988, UDI-DI: 10195327628772(each), 40195327628773(case), Lot Number: 24LBI546; 388) DYNDH1988, UDI-DI: 10195327628772(each), 40195327628773(case), Lot Number: 24CBQ392; 389) DYNDH1988, UDI-DI: 10195327628772(each), 40195327628773(case), Lot Number: 24IBK134; 390) DYNDH1988, UDI-DI: 10195327628772(each), 40195327628773(case), Lot Number: 24JBW016; 391) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 25ABO281; 392) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 25ABT231; 393) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24LBD416; 394) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24EBU007; 395) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24IBK135; 396) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24JBS021; 397) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24KBE074; 398) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 26AMC986; 399) DYNDH2016, UDI-DI: 10198459022166(each), 40198459022167(case), Lot Number: 24FBS944; 400) DYNDH2016, UDI-DI: 10198459022166(each), 40198459022167(case), Lot Number: 24GBH203; 401) DYNDH2016, UDI-DI: 10198459022166(each), 40198459022167(case), Lot Number: 24IBR097; 402) DYNDH2017, UDI-DI: 10198459022159(each), 40198459022150(case), Lot Number: 24FBS945; 403) DYNDH2017, UDI-DI: 10198459022159(each), 40198459022150(case), Lot Number: 24GBH204; 404) DYNDH2023, UDI-DI: 10198459050954(each), 40198459050955(case), Lot Number: 24GBQ903; 405) DYNDH2023, UDI-DI: 10198459050954(each), 40198459050955(case), Lot Number: 24IBB027; 406) DYNDH2028, UDI-DI: 10198459035715(each), 40198459035716(case), Lot Number: 24HBI441; 407) DYNDH2030, UDI-DI: 10198459043703(each), 40198459043704(case), Lot Number: 24GBG703; 408) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25ABP888; 409) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25DBP574; 410) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25DBV279; 411) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25EBB578; 412) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 25EBI293; 413) DYNDH2048, UDI-DI: 10198459106125(each), 40198459106126(case), Lot Number: 24IBV953; 414) DYNDH2059, UDI-DI: 10198459238406(each), 40198459238407(case), Lot Number: 25BBG684; 415) DYNDH2074, UDI-DI: 10198459205828(each), 40198459205829(case), Lot Number: 25BBB182; 416) DYNDH2089, UDI-DI: 10198459244070(each), 40198459244071(case), Lot Number: 25DBM846

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3063-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJune 12, 2026
ReportedSeptember 9, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity996692 kits