5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6 recall (Fresenius Medical Care Renal Therapies Group, LLC)
Medical deviceClass IIOngoing
Summary: Fresenius Medical Care Renal Therapies Group, LLC is recalling 5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6. Reason: Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
What is being recalled
5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6
Why
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania. (states named: Alabama, California, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Massachusetts, Maryland, Minnesota, Missouri, North Carolina, New Hampshire, New York, Ohio, Pennsylvania, Tennessee, Texas, Virginia)
How to tell if yours is affected
Lots 25KR01227 25NR01257 UDI-DI 00840861102556
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-3084-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | August 10, 2026 |
| Reported | September 9, 2026 |
| Company | Fresenius Medical Care Renal Therapies Group, LLC |
| Company location | Waltham, MA, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 120 units |