Recall Ledger

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04 recall (GE Medical Systems Information Technologies Inc)

Medical deviceClass IIOngoing

Summary: GE Medical Systems Information Technologies Inc is recalling Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV. Reason: Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

What is being recalled

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV

Why

Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution.

How to tell if yours is affected

5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3143-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJuly 30, 2026
ReportedSeptember 23, 2026
CompanyGE Medical Systems Information Technologies Inc
Company locationWaukesha, WI, United States
Initiated byVoluntary: Firm initiated
Quantity29,570 (23,632 US; 5,938 OUS)

More recalls from GE Medical Systems Information Technologies Inc · all medical device recalls · recalls from July 2026