Percutaneous Insertion Kit recall (Maquet Cardiopulmonary Gmbh)

Medical deviceClass IIOngoing

Summary: Maquet Cardiopulmonary Gmbh is recalling Percutaneous Insertion Kit (PIK), Guidewire 150 cm, Model Numbers: 701053069 (USA), 701047385 (OUS);. Reason: The packaging may not consistently maintain sterile barrier integrity.

What is being recalled

Percutaneous Insertion Kit (PIK), Guidewire 150 cm, Model Numbers: 701053069 (USA), 701047385 (OUS);

Why

The packaging may not consistently maintain sterile barrier integrity.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Japan, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom;

How to tell if yours is affected

Model Number (USA): 701053069; UDI-DI: 4037691730677; Lot Numbers: 3000324478, 3000326922, 3000336965, 3000339840, 3000347863, 3000370471, 3000374104, 3000379854, 3000382443, 3000395582, 3000403657, 3000414517, 3000430804, 3000434603, 3000444269, 3000445982, 3000453327, 3000466026, 3000471522, 3000476322, 3000476324, 3000479560, 3000485154, 3000491022, 3000496205, 3000503935, 3000510431, 3000511851, 3000518007, 3000520749, 3000525464, 3000528231, 3000534008; Model Number (OUS): 701047385; UDI-DI: 4037691519876; Lot Numbers: 3000324481, 3000324482, 3000324483, 3000326924, 3000326926, 3000326927, 3000326928, 3000328824, 3000328825, 3000328826, 3000328827, 3000328828, 3000331194, 3000331195, 3000331196, 3000331197, 3000331198, 3000331199, 3000331225, 3000333814, 3000333815, 3000333816, 3000333817, 3000333818, 3000334775, 3000334776, 3000334777, 3000334778, 3000334779, 3000336028, 3000336029, 3000336030, 3000336031, 3000336032, 3000336033, 3000338092, 3000338093, 3000338094, 3000338095, 3000338096, 3000339833, 3000339834, 3000339835, 3000339836, 3000339837, 3000341453, 3000341454, 3000341455, 3000341456, 3000342415, 3000342416, 3000342417, 3000342418, 3000342419, 3000342420, 3000342421, 3000342422, 3000343894, 3000343895, 3000343896, 3000343897, 3000343898, 3000343899, 3000343900, 3000343901, 3000345720, 3000345721, 3000345722, 3000345723, 3000345724, 3000345725, 3000345726, 3000345727, 3000345728, 3000345729, 3000347867, 3000347868, 3000347869, 3000347870, 3000349429, 3000349430, 3000349431, 3000349432, 3000349434, 3000350476, 3000350477, 3000350478, 3000350479, 3000351515, 3000351516, 3000351517, 3000351518, 3000351951, 3000351952, 3000351953, 3000351954, 3000353376, 3000353377, 3000353378, 3000353379, 3000353380, 3000354958, 3000354959, 3000354960, 3000356336, 3000356337, 3000356338, 3000356339, 3000357758, 3000357759, 3000357760, 3000359472, 3000359473, 3000359474, 3000359476, 3000363549, 3000363550, 3000369079, 3000369080, 3000369265, 3000369266, 3000369317, 3000369318, 3000369320, 3000370467, 3000370468, 3000370469, 3000370470, 3000372386, 3000372387, 3000372388, 3000372389, 3000373780, 3000373781, 3000373782, 3000373783, 3000374102, 3000374103, 3000375675, 3000375676, 3000377476, 3000377477, 3000377478, 3000378520, 3000378521, 3000378522, 3000378523, 3000379857, 3000379858, 3000379859, 3000382446, 3000382447, 3000382448, 3000382449, 3000384008, 3000384009, 3000384012, 3000384013, 3000384018, 3000385169, 3000385171, 3000385172, 3000385173, 3000387896, 3000387897, 3000387898, 3000387899, 3000389001, 3000389002, 3000389003, 3000389004, 3000389005, 3000390764, 3000390765, 3000390766, 3000390767, 3000392517, 3000392518, 3000392519, 3000392521, 3000392523, 3000393469, 3000393471, 3000393472, 3000395577, 3000395578, 3000395579, 3000395580, 3000397025, 3000397027, 3000397028, 3000397029, 3000397033, 3000397034, 3000398275, 3000398786, 3000398787, 3000398788, 3000400244, 3000400245, 3000400246, 3000400247, 3000400248, 3000400249, 3000404886, 3000404887, 3000404888, 3000404889, 3000404890, 3000406842, 3000406846, 3000406847, 3000406849, 3000408180, 3000408181, 3000408182, 3000408183, 3000408184, 3000408185, 3000410261, 3000410262, 3000410263, 3000410264, 3000411200, 3000412146, 3000412642, 3000412643, 3000413597, 3000413598, 3000413599, 3000413600, 3000414520, 3000414521, 3000414522, 3000414523, 3000414524, 3000417082, 3000417083, 3000417084, 3000417085, 3000417086, 3000417087, 3000418479, 3000418480, 3000418481, 3000418482, 3000420376, 3000420377, 3000420378, 3000423668, 3000423669, 3000423671, 3000423672, 3000423673, 3000424270, 3000424271, 3000424272, 3000424273, 3000424274, 3000425770, 3000425771, 3000425772, 3000425773, 3000431822, 3000431823, 3000431824, 3000431825, 3000431826, 3000431827, 3000431828, 3000434599, 3000434600, 3000434601, 3000434602, 3000438012, 3000438013, 3000441388, 3000441389, 3000441824, 3000444275, 3000444276, 3000444277, 3000444718, 3000444719, 3000444720, 3000444721, 3000445645, 3000445646, 3000445647, 3000445648, 3000445978, 3000445979, 3000445980, 3000445981, 3000448173, 3000448174, 3000448175, 3000448176, 3000448177, 3000449727, 3000449728, 3000449729, 3000449730, 3000451984, 3000451985, 3000451986, 3000451987, 3000451988, 3000453331, 3000453332, 3000455124, 3000455125, 3000455126, 3000455127, 3000458528, 3000458529, 3000458530, 3000458531, 3000458532, 3000461212, 3000461213, 3000461214, 3000461215, 3000461217, 3000464623, 3000464624, 3000464625, 3000464626, 3000466030, 3000466031, 3000466032, 3000466033, 3000468423, 3000468424, 3000468425, 3000471528, 3000471529, 3000471530, 3000471531, 3000471536, 3000471541, 3000471542, 3000471544, 3000475595, 3000475596, 3000475597, 3000477179, 3000477180, 3000477181, 3000478746, 3000478747, 3000478748, 3000478749, 3000478836, 3000479564, 3000479565, 3000479567, 3000481004, 3000481005, 3000481006, 3000481945, 3000481946, 3000481947, 3000481948, 3000484178, 3000484179, 3000484180, 3000484181, 3000485143, 3000486784, 3000486785, 3000486786, 3000488308, 3000488309, 3000488310, 3000489895, 3000489896, 3000490871, 3000490874, 3000491430, 3000491431, 3000491432, 3000491433, 3000493065, 3000493066, 3000493067, 3000493068, 3000496199, 3000496200, 3000496201, 3000496202, 3000496203, 3000498673, 3000498674, 3000498834, 3000501303, 3000501304, 3000501305, 3000501307, 3000502840, 3000502841, 3000502843, 3000502844, 3000503939, 3000503942, 3000503943, 3000510425, 3000510426, 3000510427, 3000511848, 3000511849, 3000511850, 3000513501, 3000513502, 3000513503, 3000513504, 3000513505, 3000514609, 3000514610, 3000514611, 3000514612, 3000515696, 3000515697, 3000515698, 3000515699, 3000517998, 3000518004, 3000518005, 3000519528, 3000519529, 3000519530, 3000519531, 3000519532, 3000520745, 3000520746, 3000520747, 3000520748, 3000522411, 3000522412, 3000522413, 3000522414, 3000522415, 3000523510, 3000523511, 3000524539, 3000524540, 3000524541, 3000525039, 3000525040, 3000525041, 3000525042, 3000525043, 3000525460, 3000525461, 3000525462, 3000525463, 3000526919, 3000526920, 3000526921, 3000526922, 3000528714, 3000528715, 3000528716, 3000530042, 3000530043, 3000530044, 3000530045, 3000534004, 3000534005, 3000534006, 3000535444, 3000535445, 3000535446, 3000535447, 3000537045, 3000537046, 3000537047, 3000537048, 3000539525, 3000539526, 3000539527, 3000539528, 3000541228, 3000541229, 3000541230, 3000541231, 3000541232, 3000543979, 3000543986, 3000543989, 3000547001, 3000547002, 3000547003, 3000547004, 3000549189, 3000549190, 3000549191, 3000550936, 3000550937, 3000550938, 3000550939, 3000553331, 3000553332, 3000553333, 3000553334, 3000553335, 3000554727, 3000554729, 3000554730, 3000556412, 3000556413, 3000556414, 3000556415;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3145-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJuly 31, 2026
ReportedSeptember 23, 2026
CompanyMaquet Cardiopulmonary Gmbh
Company locationRastatt, N/A, Germany
Initiated byVoluntary: Firm initiated
Quantity185617 units (5926 US, 179691 OUS)

More recalls from Maquet Cardiopulmonary Gmbh · all medical device recalls · recalls from July 2026

Official FDA record