Impella Optical Controller recall (Abiomed, Inc.)
Medical deviceClass IOngoing
Summary: Abiomed, Inc. is recalling Impella Optical Controller, Automated Impella Controller, Impella Connect Package with the below product descriptions and corresponding Product Codes. 1. Optical, AIC, Impella Connect, Packaged, AU; Product Code: 0042-0040-AU. 2. Optical, AIC, Impella Connect, Packaged, CA; Product Code: 0042-0040-C. Reason: The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
What is being recalled
Impella Optical Controller, Automated Impella Controller, Impella Connect Package with the below product descriptions and corresponding Product Codes. 1. Optical, AIC, Impella Connect, Packaged, AU; Product Code: 0042-0040-AU. 2. Optical, AIC, Impella Connect, Packaged, CA; Product Code: 0042-0040-CA. 3. Optical, AIC, Impella Connect, Packaged, EU; Product Code: 0042-0040-EU. 4. Optical, AIC, Impella Connect, Packaged, JP; Product Code: 0042-0040-JP. 5. Optical, AIC, Impella Connect, Packaged, UK; Product Code: 0042-0040-UK. 6. Optical, AIC, Impella Connect, Packaged, US; Product Code: 0042-0040-US. 7. Optical, AIC, Impella Connect, Packaged, KR; Product Code: 0042-0040-KR;
Why
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
How serious is it
Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.
Where it was sold
Domestic: Nationwide Distribution. International: United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Ireland, Israel, India, Italy, Japan, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Pakistan, Poland, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan;
How to tell if yours is affected
1. Optical, AIC, Impella Connect, Packaged, AU; Product Code: 0042-0040-AU; UDI-DI: 00813502012972; Serial Numbers: IC12674, IC12675, IC13066, IC10162, IC10163, IC10387, IC10388, IC10747, IC10748, IC11472, IC11487, IC11489, IC11490, IC11544, IC11546, IC11566, IC12008, IC12054, IC12151, IC12583, IC12584, IC12804, IC13070, IC13071, IC13085, IC13093, IC13094, IC13095, IC13096, IC13233, IC13234, IC13561, IC13566, IC13749, IC13753; 2. Optical, AIC, Impella Connect, Packaged, CA; Product Code: 0042-0040-CA; UDI-DI: 00813502012958; Serial Numbers: IC10306, IC10312, IC11171, IC11172, IC12604, IC12605, IC12891, IC12892, IC12894, IC12895, IC13266, IC13267, IC13503, IC13504, IC13617, IC13618, IC13713, IC13714, IC13715, IC13716, IC13726, IC13801, IC13802; 3. Optical, AIC, Impella Connect, Packaged, EU; Product Code: 0042-0040-EU; UDI-DI: 00813502012217 Serial Numbers: IC10382, IC10419, IC10681, IC10683, IC11123, IC11125, IC11257, IC11258, IC11262, IC11264, IC11265, IC11266, IC11282, IC12012, IC12015, IC12447, IC12483, IC12814, IC13029, IC13030, IC13031, IC9010, IC9011, IC9949, IC9950, IC10146, IC10180, IC10371, IC10372, IC10396, IC10397, IC10665, IC10670, IC10671, IC10673, IC10721, IC10772, IC10828, IC10835, IC11072, IC11116, IC11119, IC11132, IC11138, IC11140, IC11146, IC11679, IC11706, IC11712, IC11713, IC11721, IC11878, IC11951, IC11958, IC11967, IC11969, IC11978, IC11986, IC11992, IC11997, IC12260, IC12261, IC12446, IC12454, IC12458, IC12459, IC12463, IC12465, IC12466, IC12544, IC12546, IC12555, IC12565, IC12566, IC12586, IC12593, IC12594, IC12595, IC12596, IC12666, IC12670, IC12744, IC12747, IC12749, IC12750, IC12761, IC12764, IC12771, IC12810, IC12822, IC12828, IC12845, IC12929, IC13034, IC13035, IC13041, IC13047, IC13053, IC13118, IC13122, IC13135, IC13199, IC13226, IC13245, IC13246, IC13360, IC13364, IC13627, IC13635, IC13640, IC13642, IC12768, IC13023, IC13322, IC9071, IC9073, IC10126, IC10139, IC10169, IC10174, IC10392, IC10556, IC10558, IC10628, IC10632, IC10674, IC10677, IC10678, IC10680, IC11215, IC11217, IC11219, IC11220, IC11281, IC11283, IC11289, IC11358, IC11879, IC11904, IC11911, IC11930, IC11945, IC11952, IC11959, IC11960, IC11961, IC11966, IC11970, IC11973, IC11974, IC11976, IC11989, IC11991, IC12195, IC12206, IC12210, IC12213, IC12216, IC12217, IC12257, IC12258, IC12262, IC12265, IC12267, IC12269, IC12270, IC12272, IC12273, IC12274, IC12276, IC12277, IC12307, IC12309, IC12310, IC12319, IC12455, IC12456, IC12457, IC12467, IC12481, IC12484, IC12485, IC12488, IC12491, IC12494, IC12548, IC12551, IC12553, IC12765, IC12767, IC12812, IC12816, IC12824, IC12835, IC12942, IC12951, IC13177, IC13192, IC13291, IC13298, IC13328, IC13329, IC13330, IC13331, IC13352, IC13354, IC13363, IC13365, IC13366, IC13628, IC13636, IC9726, IC9727, IC9729, IC9730, IC9761, IC9830, IC9959, IC9965, IC10154, IC10159, IC10171, IC10175, IC10394, IC10400, IC10938, IC10952, IC11305, IC11715, IC11722, IC12001, IC12220, IC12221, IC12524, IC12542, IC12671, IC12672, IC8840, IC9252, IC9757, IC9759, IC10779, IC10782, IC10181, IC10184, IC10185, IC10186, IC10326, IC10360, IC10369, IC10540, IC10541, IC10544, IC10551, IC10555, IC10559, IC10627, IC10639, IC10679, IC10682, IC10707, IC10710, IC10711, IC10713, IC10809, IC11007, IC11010, IC11011, IC11090, IC11221, IC11223, IC11234, IC11363, IC11368, IC11717, IC11718, IC11719, IC11910, IC11913, IC11931, IC11948, IC11949, IC11955, IC11980, IC11983, IC12259, IC12263, IC12264, IC12266, IC12281, IC12282, IC12290, IC12311, IC12314, IC12486, IC12490, IC12496, IC12497, IC12576, IC12577, IC12695, IC12696, IC12697, IC12698, IC12699, IC12700, IC12702, IC12703, IC12704, IC12753, IC12754, IC12755, IC12756, IC12757, IC12802, IC12809, IC13069, IC13072, IC13120, IC13121, IC13133, IC13134, IC13136, IC13138, IC13165, IC13213, IC13248, IC13249, IC13281, IC13282, IC13293, IC13297, IC13371, IC13372, IC13565, IC13638, IC13639, IC13643, IC13645, IC13650, IC13653, IC8821, IC8822, IC8823, IC8825, IC8826, IC9002A, IC9015, IC9023, IC9024, IC9051, IC9052, IC9072, IC9076, IC9077, IC9078, IC9106, IC9107, IC9108, IC9109, IC9110, IC9111, IC9112, IC9113, IC9114, IC9115, IC9169, IC9171, IC9176, IC9216, IC9217, IC9251, IC9253, IC9257, IC9260, IC9342, IC9345, IC9352, IC9424, IC9427, IC9762, IC9763, IC9764, IC9765, IC9766, IC9824, IC9945, IC9948, IC13709, IC13710, IC13711, IC13712, IC10123, IC10124, IC10127, IC10138, IC10140, IC10141, IC10142, IC10143, IC10144, IC10148, IC10152, IC10155, IC10157, IC10158, IC10161, IC10166, IC10173, IC10176, IC10256, IC10257, IC10258, IC10259, IC10260, IC10261, IC10262, IC10263, IC10264, IC10265, IC10329, IC10347, IC10370, IC10389, IC10393, IC10395, IC10398, IC10401, IC10404, IC10408, IC10415, IC10416, IC10423, IC10424, IC10425, IC10433, IC10434, IC10437, IC10438, IC10439, IC10440, IC10442, IC10468, IC10506, IC10510, IC10542, IC10543, IC10552, IC10591, IC10621, IC10633, IC10634, IC10635, IC10636, IC10638, IC10720, IC10738, IC10767, IC10768, IC10778, IC10781, IC10819, IC10858, IC10870, IC10883, IC10884, IC10898, IC10909, IC10917, IC10919, IC10939, IC10943, IC10944, IC10945, IC10947, IC10948, IC10949, IC10953, IC10954, IC10955, IC11003, IC11004, IC11005, IC11009, IC11095, IC11120, IC11122, IC11124, IC11127, IC11133, IC11136, IC11137, IC11141, IC11142, IC11149, IC11193, IC11209, IC11214, IC11224, IC11225, IC11226, IC11228, IC11233, IC11235, IC11277, IC11278, IC11279, IC11280, IC11286, IC11287, IC11288, IC11322, IC11328, IC11332, IC11334, IC11360, IC11361, IC11364, IC11367, IC11699, IC11705, IC11707, IC11708, IC11709, IC11710, IC11711, IC11900, IC11909, IC11928, IC11929, IC11934, IC11937, IC11942, IC11943, IC11946, IC11947, IC11950, IC11953, IC11954, IC11956, IC11957, IC11962, IC11963, IC11971, IC11975, IC11977, IC11981, IC11984, IC11985, IC11987, IC11988, IC11993, IC11995, IC12002, IC12014, IC12208, IC12212, IC12219, IC12222, IC12223, IC12271, IC12275, IC12280, IC12287, IC12289, IC12291, IC12308, IC12312, IC12313, IC12320, IC12442, IC12451, IC12453, IC12464, IC12470, IC12487, IC12493, IC12495, IC12523, IC12536, IC12538, IC12541, IC12550, IC12554, IC12556, IC12558, IC12567, IC12568, IC12574, IC12575, IC12585, IC12590, IC12591, IC12597, IC12640, IC12648, IC12667, IC12668, IC12673, IC12751, IC12759, IC12760, IC12763, IC12805, IC12807, IC12811, IC12821, IC12827, IC12829, IC12833, IC12834, IC12854, IC12865, IC12939, IC12940, IC12941, IC12944, IC12946, IC12947, IC12948, IC13003, IC13026, IC13036, IC13081, IC13097, IC13098, IC13100, IC13101, IC13103, IC13104, IC13105, IC13106, IC13107, IC13119, IC13137, IC13178, IC13183, IC13186, IC13292, IC13295, IC13296, IC13318, IC13319, IC13321, IC13332, IC13353, IC13355, IC13359, IC13367, IC13369, IC13370, IC13374, IC13731, IC13741, IC13751, IC13752, IC13762, IC13775, IC13777, IC13778, IC13779, IC13780, IC13806, IC8818, IC8819, IC8820, IC8824, IC8827, IC8836, IC8837, IC8838, IC8839, IC8841, IC8842, IC8865, IC8866, IC8868, IC8869, IC8870, IC8871, IC8872, IC8873, IC8874, IC8876, IC8898, IC8899, IC8900, IC8901, IC8902, IC8910, IC8911, IC8912, IC8913, IC8914, IC9000A, IC9001A, IC9003A, IC9004A, IC9009, IC9012, IC9013, IC9014, IC9016, IC9017, IC9019, IC9020, IC9021, IC9022, IC9039, IC9050, IC9069, IC9070, IC9075, IC9167, IC9168, IC9170, IC9172, IC9173, IC9174, IC9175, IC9179, IC9180, IC9189, IC9190, IC9191, IC9193, IC9194, IC9215, IC9254, IC9255, IC9256, IC9258, IC9259, IC9281, IC9330, IC9343, IC9389, IC9396, IC9398, IC9401, IC9425, IC9426, IC9429, IC9552, IC9553, IC9554, IC9555, IC9556, IC9643, IC9687, IC9689, IC9692, IC9693, IC9758, IC9760, IC9828, IC9868, IC9942, IC9944, IC9951, IC9954, IC9955, IC9956, IC9957, IC9958, IC9960, IC9966, IC9967, IC9968, IC12826, IC12831, IC13629, IC13647, IC9192, IC10560, IC10590, IC10698, IC10706, IC10773, IC10775, IC10799, IC10820, IC10885, IC10886, IC10890, IC10891, IC10892, IC10916, IC10920, IC10921, IC10940, IC10942, IC10946, IC10951, IC11074, IC11121, IC11229, IC11240, IC11302, IC11366, IC11714, IC11716, IC11914, IC11915, IC11964, IC11965, IC11968, IC11998, IC12000, IC12215, IC12218, IC12284, IC12288, IC12443, IC12444, IC12445, IC12452, IC12545, IC12552, IC12587, IC12762, IC12803, IC12815, IC12931, IC13017, IC13024, IC13091, IC13092, IC13099, IC13108, IC13109, IC13187, IC13294, IC9420, IC9428, IC9430, IC9431, IC9690, IC9827, IC9962, IC9964, IC10125, IC10156, IC10167, IC10168, IC10172, IC10183, IC10950, IC11999, IC12278, IC12480, IC12543, IC12766, IC12830, IC12933, IC13191, IC13247, IC13320, IC13356, IC13113, IC13127, IC13180, IC13193, IC10402, IC10403, IC10441, IC10505, IC10507, IC10508, IC10675, IC10676, IC12801, IC12808, IC9399, IC9402, IC9623, IC9626, IC9644, IC9645, IC9647, IC10153, IC10160, IC10580, IC10629, IC10887, IC10889, IC11006, IC11321, IC11323, IC11324, IC11326, IC11330, IC11331, IC11726, IC12013, IC12224, IC12549, IC12649, IC12786, IC12988, IC13020, IC13102, IC13132, IC13181, IC13184, IC9403, IC9404, IC9624, IC9646, IC9896, IC9946, IC9728, IC13142, IC13143, IC13144, IC13146, IC13148, IC13150, IC13151, IC13156, IC13157, IC13158; 4. Optical, AIC, Impella Connect, Packaged, JP; Product Code: 0042-0040-JP; UDI-DI: 00813502011814; Serial Numbers: IC10016, IC10018, IC10019, IC10020, IC10021, IC10022, IC10068, IC10107, IC10119, IC10120, IC10272, IC10273, IC10274, IC10275, IC10276, IC10280, IC10290, IC10354, IC10355, IC10356, IC10357, IC10359, IC10363, IC10368, IC10405, IC10409, IC10413, IC10426, IC10427, IC10503, IC10504, IC10536, IC10537, IC10538, IC10539, IC10563, IC10583, IC10699, IC10700, IC10705, IC10806, IC10807, IC10810, IC10869, IC10906, IC10907, IC10928, IC10930, IC10931, IC10932, IC11088, IC11192, IC11195, IC11218, IC11253, IC11339, IC11340, IC11370, IC11376, IC11378, IC11405, IC11417, IC11453, IC11454, IC11459, IC11499, IC11570, IC11623, IC11636, IC11666, IC11672, IC11678, IC11687, IC11689, IC11690, IC11694, IC11720, IC11725, IC11727, IC11738, IC11743, IC11750, IC11751, IC11777, IC11778, IC12026, IC12028, IC12029, IC12030, IC12031, IC12404, IC12405, IC12434, IC12437, IC12438, IC12515, IC12516, IC12517, IC12518, IC12609, IC12610, IC12612, IC12613, IC12614, IC12617, IC12618, IC12619, IC12620, IC12752, IC12769, IC12775, IC12776, IC12841, IC12843, IC12846, IC12849, IC12852, IC12853, IC13258, IC13259, IC13324, IC13326, IC13392, IC13393, IC8196, IC8197, IC8237, IC8238, IC8239, IC8240, IC8241, IC8245, IC8246, IC8298, IC8299, IC8301, IC8302, IC8303, IC8304, IC8306, IC8309, IC8310, IC8364, IC8369, IC8370, IC8371, IC8372, IC8374, IC8375, IC8491, IC8492, IC8523, IC8524, IC8526, IC8531, IC8571, IC8573, IC8583, IC8595, IC8597, IC8599, IC8601, IC8619, IC8626, IC8666, IC8670, IC8679, IC8680, IC8681, IC8682, IC8684, IC8685, IC8687, IC8731, IC8732, IC8733, IC8734, IC8735, IC8779, IC8780, IC8782, IC8784, IC8785, IC8886, IC8887, IC8888, IC8889, IC8890, IC8892, IC8893, IC8894, IC8896, IC8897, IC8915, IC8916, IC8917, IC8918, IC8919, IC8921, IC8922, IC8923, IC8924, IC8927, IC8937, IC9125, IC9126, IC9127, IC9128, IC9129, IC9130, IC9131, IC9132, IC9134, IC9135, IC9136, IC9137, IC9138, IC9139, IC9140, IC9288, IC9289, IC9290, IC9291, IC9292, IC9294, IC9295, IC9298, IC9321, IC9322, IC9323, IC9325, IC9326, IC9329, IC9331, IC9332, IC9411, IC9462, IC9463, IC9485, IC9489, IC9490, IC9491, IC9492, IC9494, IC9495, IC9496, IC9497, IC9498, IC9499, IC9540, IC9559, IC9560, IC9561, IC9562, IC9563, IC9580, IC9694, IC9702, IC9703, IC9704, IC9705, IC9707, IC9708, IC9709, IC9786, IC9787, IC9790, IC9791, IC9811, IC9838, IC9894, IC9923, IC9925, IC9926, IC9927, IC9928, IC9929, IC9930, IC9931; 5. Optical, AIC, Impella Connect, Packaged, UK; Product Code: 0042-0040-UK; UDI-DI: 00813502012484; Serial Numbers: IC10008, IC10011, IC10025, IC10240, IC10241, IC10592, IC10593, IC10598, IC10601, IC11196, IC11197, IC11198, IC11640, IC11642, IC11645, IC11646, IC12714, IC12717, IC12725, IC12727, IC12728, IC9990, IC12255, IC10164, IC10165, IC12315, IC12318, IC12561, IC12563, IC12726, IC13338, IC10224, IC10225, IC10788, IC10789, IC10790, IC10792, IC10793, IC11723, IC11724, IC12256, IC12716, IC12718, IC13235, IC11682, IC11683; 6. Optical, AIC, Impella Connect, Packaged, US; Product Code: 0042-0040-US; UDI-DI: 00813502011401; Serial Numbers: IC10000, IC10001, IC10002, IC10003, IC10004, IC10005, IC10006, IC10007, IC10009, IC10010, IC10026, IC10027, IC10028, IC10029, IC10030, IC10031, IC10036, IC10038, IC10041, IC10042, IC10045, IC10046, IC10047, IC10048, IC10050, IC10051, IC10052, IC10053, IC10054, IC10055, IC10056, IC10057, IC10058, IC10059, IC10060, IC10061, IC10062, IC10065, IC10067, IC10069, IC10070, IC10071, IC10072, IC10073, IC10074, IC10075, IC10076, IC10077, IC10078, IC10079, IC10080, IC10081, IC10082, IC10083, IC10084, IC10085, IC10086, IC10087, IC10088, IC10089, IC10090, IC10091, IC10092, IC10093, IC10094, IC10095, IC10096, IC10097, IC10098, IC10099, IC10100, IC10101, IC10104, IC10110, IC10111, IC10112, IC10113, IC10114, IC10115, IC10116, IC10117, IC10118, IC10128, IC10129, IC10130, IC10131, IC10132, IC10133, IC10134, IC10135, IC10136, IC10137, IC10147, IC10151, IC10187, IC10188, IC10189, IC10190, IC10191, IC10192, IC10193, IC10194, IC10195, IC10196, IC10197, IC10198, IC10199, IC10200, IC10201, IC10202, IC10203, IC10204, IC10205, IC10206, IC10207, IC10208, IC10209, IC10210, IC10211, IC10212, IC10213, IC10214, IC10215, IC10216, IC10217, IC10218, IC10219, IC10220, IC10221, IC10222, IC10223, IC10226, IC10227, IC10228, IC10229, IC10230, IC10231, IC10232, IC10233, IC10235, IC10237, IC10242, IC10246, IC10247, IC10248, IC10249, IC10250, IC10251, IC10252, IC10253, IC10254, IC10255, IC10266, IC10269, IC10271, IC10277, IC10279, IC10282, IC10283, IC10284, IC10285, IC10286, IC10287, IC10288, IC10289, IC10291, IC10292, IC10293, IC10294, IC10295, IC10296, IC10297, IC10298, IC10299, IC10301, IC10303, IC10304, IC10307, IC10308, IC10309, IC10310, IC10311, IC10313, IC10314, IC10315, IC10316, IC10317, IC10321, IC10322, IC10323, IC10324, IC10325, IC10327, IC10328, IC10330, IC10331, IC10332, IC10333, IC10334, IC10335, IC10336, IC10337, IC10339, IC10340, IC10342, IC10343, IC10344, IC10345, IC10346, IC10348, IC10349, IC10350, IC10351, IC10352, IC10358, IC10361, IC10373, IC10374, IC10375, IC10376, IC10377, IC10378, IC10379, IC10380, IC10381, IC10383, IC10384, IC10385, IC10386, IC10390, IC10391, IC10407, IC10410, IC10412, IC10418, IC10420, IC10421, IC10422, IC10428, IC10429, IC10430, IC10431, IC10432, IC10435, IC10436, IC10443, IC10444, IC10445, IC10446, IC10447, IC10448, IC10449, IC10450, IC10451, IC10452, IC10453, IC10454, IC10455, IC10459, IC10461, IC10462, IC10466, IC10470, IC10471, IC10472, IC10474, IC10475, IC10476, IC10477, IC10478, IC10479, IC10480, IC10481, IC10482, IC10484, IC10485, IC10486, IC10487, IC10509, IC10511, IC10512, IC10514, IC10515, IC10516, IC10519, IC10520, IC10521, IC10522, IC10523, IC10524, IC10525, IC10526, IC10527, IC10528, IC10529, IC10530, IC10531, IC10532, IC10533, IC10534, IC10535, IC10545, IC10546, IC10548, IC10549, IC10550, IC10553, IC10554, IC10557, IC10561, IC10562, IC10564, IC10565, IC10566, IC10567, IC10568, IC10569, IC10570, IC10571, IC10572, IC10578, IC10581, IC10582, IC10596, IC10597, IC10599, IC10600, IC10602, IC10603, IC10604, IC10605, IC10606, IC10607, IC10608, IC10609, IC10610, IC10611, IC10612, IC10613, IC10614, IC10615, IC10616, IC10617, IC10618, IC10619, IC10620, IC10622, IC10623, IC10624, IC10625, IC10626, IC10637, IC10642, IC10643, IC10644, IC10647, IC10653, IC10654, IC10655, IC10656, IC10657, IC10658, IC10659, IC10660, IC10661, IC10662, IC10669, IC10684, IC10685, IC10687, IC10688, IC10690, IC10691, IC10692, IC10693, IC10694, IC10695, IC10696, IC10697, IC10704, IC10708, IC10709, IC10712, IC10714, IC10715, IC10716, IC10717, IC10718, IC10719, IC10722, IC10723, IC10724, IC10725, IC10726, IC10728, IC10729, IC10730, IC10731, IC10732, IC10733, IC10734, IC10736, IC10737, IC10739, IC10740, IC10741, IC10742, IC10743, IC10744, IC10745, IC10746, IC10749, IC10750, IC10751, IC10752, IC10753, IC10754, IC10756, IC10757, IC10758, IC10759, IC10760, IC10763, IC10764, IC10769, IC10770, IC10774, IC10776, IC10777, IC10783, IC10787, IC10791, IC10795, IC10796, IC10800, IC10811, IC10812, IC10813, IC10814, IC10815, IC10816, IC10817, IC10818, IC10821, IC10822, IC10823, IC10824, IC10825, IC10826, IC10830, IC10831, IC10832, IC10833, IC10838, IC10839, IC10842, IC10845, IC10847, IC10848, IC10850, IC10852, IC10855, IC10856, IC10857, IC10859, IC10860, IC10861, IC10873, IC10875, IC10876, IC10877, IC10878, IC10879, IC10880, IC10881, IC10888, IC10895, IC10897, IC10899, IC10900, IC10901, IC10908, IC10910, IC10911, IC10912, IC10913, IC10915, IC10918, IC10922, IC10923, IC10924, IC10925, IC10926, IC10927, IC10929, IC10934, IC10935, IC10936, IC10937, IC10941, IC10957, IC10958, IC10959, IC10962, IC10971, IC10972, IC10974, IC10975, IC10978, IC10980, IC10981, IC10982, IC10998, IC10999, IC11001, IC11002, IC11008, IC11012, IC11013, IC11014, IC11016, IC11017, IC11018, IC11021, IC11022, IC11024, IC11030, IC11031, IC11032, IC11033, IC11034, IC11035, IC11036, IC11037, IC11038, IC11039, IC11040, IC11041, IC11042, IC11043, IC11044, IC11045, IC11046, IC11047, IC11051, IC11052, IC11053, IC11054, IC11056, IC11058, IC11059, IC11060, IC11062, IC11063, IC11064, IC11065, IC11066, IC11068, IC11069, IC11070, IC11073, IC11075, IC11076, IC11077, IC11078, IC11079, IC11080, IC11081, IC11082, IC11085, IC11087, IC11092, IC11093, IC11094, IC11096, IC11103, IC11106, IC11107, IC11108, IC11109, IC11117, IC11129, IC11139, IC11147, IC11150, IC11154, IC11155, IC11156, IC11157, IC11158, IC11159, IC11160, IC11161, IC11163, IC11164, IC11165, IC11166, IC11167, IC11168, IC11169, IC11170, IC11173, IC11175, IC11176, IC11178, IC11179, IC11180, IC11181, IC11184, IC11185, IC11187, IC11188, IC11189, IC11191, IC11194, IC11200, IC11203, IC11205, IC11206, IC11207, IC11208, IC11241, IC11242, IC11243, IC11244, IC11245, IC11246, IC11247, IC11249, IC11251, IC11268, IC11269, IC11270, IC11271, IC11272, IC11273, IC11274, IC11275, IC11276, IC11284, IC11285, IC11290, IC11291, IC11292, IC11293, IC11294, IC11295, IC11296, IC11297, IC11298, IC11310, IC11312, IC11313, IC11314, IC11315, IC11316, IC11318, IC11319, IC11320, IC11327, IC11333, IC11335, IC11347, IC11348, IC11349, IC11350, IC11351, IC11352, IC11353, IC11354, IC11355, IC11356, IC11357, IC11359, IC11362, IC11371, IC11373, IC11374, IC11375, IC11380, IC11381, IC11382, IC11383, IC11384, IC11385, IC11386, IC11387, IC11388, IC11389, IC11390, IC11391, IC11392, IC11394, IC11395, IC11396, IC11397, IC11398, IC11402, IC11403, IC11406, IC11408, IC11409, IC11411, IC11412, IC11415, IC11418, IC11419, IC11420, IC11421, IC11422, IC11423, IC11424, IC11425, IC11426, IC11427, IC11428, IC11429, IC11430, IC11431, IC11432, IC11433, IC11434, IC11435, IC11437, IC11438, IC11439, IC11443, IC11446, IC11447, IC11452, IC11457, IC11458, IC11460, IC11461, IC11464, IC11465, IC11466, IC11467, IC11469, IC11471, IC11473, IC11474, IC11477, IC11478, IC11479, IC11480, IC11481, IC11482, IC11484, IC11486, IC11488, IC11492, IC11493, IC11495, IC11503, IC11510, IC11511, IC11512, IC11513, IC11514, IC11518, IC11519, IC11521, IC11522, IC11523, IC11524, IC11525, IC11526, IC11528, IC11529, IC11530, IC11531, IC11532, IC11533, IC11534, IC11535, IC11536, IC11537, IC11538, IC11539, IC11540, IC11549, IC11550, IC11551, IC11552, IC11555, IC11556, IC11557, IC11558, IC11559, IC11560, IC11561, IC11563, IC11564, IC11565, IC11567, IC11568, IC11569, IC11572, IC11573, IC11574, IC11581, IC11582, IC11583, IC11584, IC11585, IC11586, IC11597, IC11606, IC11609, IC11610, IC11613, IC11614, IC11615, IC11617, IC11618, IC11621, IC11622, IC11624, IC11625, IC11626, IC11627, IC11628, IC11629, IC11631, IC11632, IC11633, IC11634, IC11635, IC11637, IC11638, IC11639, IC11641, IC11644, IC11654, IC11662, IC11680, IC11684, IC11686, IC11692, IC11695, IC11696, IC11697, IC11700, IC11702, IC11704, IC11753, IC11754, IC11755, IC11756, IC11757, IC11758, IC11759, IC11760, IC11761, IC11762, IC11763, IC11764, IC11765, IC11766, IC11767, IC11768, IC11769, IC11770, IC11773, IC11774, IC11776, IC11779, IC11781, IC11783, IC11784, IC11785, IC11786, IC11787, IC11788, IC11790, IC11791, IC11792, IC11793, IC11795, IC11796, IC11797, IC11798, IC11799, IC11800, IC11801, IC11802, IC11803, IC11804, IC11805, IC11806, IC11807, IC11808, IC11809, IC11810, IC11811, IC11812, IC11813, IC11814, IC11815, IC11816, IC11817, IC11818, IC11821, IC11822, IC11823, IC11824, IC11825, IC11826, IC11827, IC11828, IC11829, IC11830, IC11831, IC11832, IC11833, IC11834, IC11835, IC11836, IC11837, IC11838, IC11839, IC11840, IC11841, IC11842, IC11843, IC11844, IC11845, IC11846, IC11847, IC11848, IC11849, IC11850, IC11851, IC11852, IC11853, IC11854, IC11856, IC11857, IC11860, IC11861, IC11862, IC11863, IC11864, IC11866, IC11868, IC11869, IC11870, IC11871, IC11872, IC11873, IC11874, IC11875, IC11876, IC11877, IC11881, IC11882, IC11883, IC11884, IC11885, IC11886, IC11887, IC11888, IC11889, IC11890, IC11891, IC11892, IC11894, IC11896, IC11897, IC11898, IC11899, IC11901, IC11902, IC11903, IC11905, IC11906, IC11907, IC11908, IC11912, IC11916, IC11917, IC11918, IC11919, IC11920, IC11921, IC11923, IC11925, IC11926, IC11927, IC11932, IC11933, IC11935, IC11936, IC11938, IC11939, IC11940, IC11944, IC11979, IC11982, IC11990, IC11994, IC11996, IC12003, IC12004, IC12005, IC12006, IC12007, IC12009, IC12010, IC12011, IC12016, IC12017, IC12018, IC12019, IC12020, IC12021, IC12022, IC12023, IC12024, IC12025, IC12033, IC12034, IC12035, IC12036, IC12037, IC12038, IC12039, IC12040, IC12041, IC12042, IC12044, IC12045, IC12046, IC12047, IC12048, IC12049, IC12050, IC12051, IC12052, IC12053, IC12055, IC12056, IC12057, IC12058, IC12059, IC12060, IC12061, IC12062, IC12064, IC12065, IC12066, IC12070, IC12072, IC12073, IC12074, IC12075, IC12076, IC12077, IC12078, IC12079, IC12080, IC12081, IC12082, IC12083, IC12084, IC12085, IC12086, IC12087, IC12088, IC12089, IC12090, IC12091, IC12092, IC12093, IC12094, IC12095, IC12096, IC12097, IC12098, IC12099, IC12100, IC12102, IC12103, IC12104, IC12105, IC12106, IC12107, IC12109, IC12110, IC12111, IC12112, IC12113, IC12114, IC12115, IC12116, IC12117, IC12118, IC12119, IC12120, IC12121, IC12122, IC12124, IC12125, IC12126, IC12128, IC12129, IC12130, IC12131, IC12132, IC12133, IC12134, IC12135, IC12136, IC12137, IC12138, IC12139, IC12140, IC12141, IC12142, IC12143, IC12144, IC12145, IC12146, IC12147, IC12148, IC12149, IC12150, IC12152, IC12153, IC12154, IC12155, IC12156, IC12157, IC12158, IC12159, IC12160, IC12161, IC12162, IC12164, IC12165, IC12166, IC12167, IC12168, IC12169, IC12170, IC12171, IC12172, IC12173, IC12174, IC12175, IC12176, IC12177, IC12178, IC12179, IC12180, IC12181, IC12182, IC12183, IC12184, IC12185, IC12186, IC12188, IC12189, IC12190, IC12191, IC12192, IC12193, IC12194, IC12196, IC12197, IC12198, IC12199, IC12200, IC12201, IC12202, IC12203, IC12204, IC12205, IC12207, IC12209, IC12211, IC12225, IC12226, IC12227, IC12228, IC12229, IC12230, IC12231, IC12232, IC12233, IC12234, IC12235, IC12236, IC12237, IC12238, IC12239, IC12240, IC12241, IC12242, IC12243, IC12244, IC12246, IC12247, IC12249, IC12250, IC12251, IC12252, IC12253, IC12254, IC12268, IC12283, IC12285, IC12286, IC12292, IC12293, IC12294, IC12295, IC12296, IC12297, IC12298, IC12299, IC12300, IC12301, IC12302, IC12303, IC12304, IC12305, IC12306, IC12347, IC12348, IC12349, IC12350, IC12351, IC12352, IC12353, IC12354, IC12355, IC12356, IC12357, IC12358, IC12360, IC12361, IC12362, IC12363, IC12364, IC12365, IC12366, IC12367, IC12368, IC12369, IC12370, IC12371, IC12372, IC12373, IC12374, IC12375, IC12376, IC12377, IC12381, IC12382, IC12383, IC12384, IC12385, IC12386, IC12387, IC12388, IC12389, IC12390, IC12391, IC12396, IC12399, IC12400, IC12406, IC12426, IC12428, IC12430, IC12433, IC12435, IC12436, IC12440, IC12441, IC12448, IC12449, IC12450, IC12460, IC12461, IC12462, IC12468, IC12469, IC12472, IC12473, IC12474, IC12475, IC12476, IC12477, IC12478, IC12557, IC12592, IC12601, IC12602, IC12611, IC12621, IC12622, IC12623, IC12624, IC12625, IC12626, IC12628, IC12629, IC12630, IC12631, IC12632, IC12633, IC12634, IC12635, IC12636, IC12637, IC12638, IC12639, IC12641, IC12642, IC12643, IC12644, IC12645, IC12646, IC12647, IC12650, IC12651, IC12652, IC12653, IC12654, IC12655, IC12656, IC12658, IC12659, IC12660, IC12669, IC12676, IC12677, IC12678, IC12679, IC12680, IC12681, IC12682, IC12683, IC12684, IC12685, IC12686, IC12687, IC12688, IC12689, IC12690, IC12691, IC12692, IC12693, IC12694, IC12701, IC12705, IC12706, IC12707, IC12708, IC12709, IC12710, IC12711, IC12712, IC12729, IC12730, IC12731, IC12732, IC12733, IC12734, IC12735, IC12736, IC12737, IC12738, IC12739, IC12740, IC12741, IC12742, IC12743, IC12745, IC12746, IC12748, IC12777, IC12778, IC12779, IC12780, IC12781, IC12782, IC12783, IC12784, IC12785, IC12787, IC12788, IC12790, IC12791, IC12792, IC12793, IC12794, IC12795, IC12796, IC12797, IC12798, IC12799, IC12800, IC12806, IC12836, IC12842, IC12847, IC12848, IC12850, IC12851, IC12860, IC12861, IC12862, IC12863, IC12868, IC12871, IC12872, IC12873, IC12874, IC12875, IC12876, IC12877, IC12878, IC12879, IC12880, IC12886, IC12887, IC12893, IC12896, IC12897, IC12898, IC12899, IC12901, IC12902, IC12903, IC12904, IC12906, IC12907, IC12908, IC12909, IC12910, IC12911, IC12912, IC12913, IC12916, IC12918, IC12919, IC12920, IC12921, IC12922, IC12927, IC12928, IC12932, IC12934, IC12945, IC12967, IC12972, IC12979, IC12985, IC12986, IC12987, IC12989, IC12992, IC12995, IC12997, IC12998, IC13004, IC13005, IC13006, IC13007, IC13008, IC13009, IC13010, IC13012, IC13014, IC13015, IC13016, IC13018, IC13019, IC13025, IC13028, IC13032, IC13033, IC13043, IC13044, IC13050, IC13052, IC13054, IC13055, IC13056, IC13058, IC13067, IC13068, IC13073, IC13074, IC13075, IC13076, IC13124, IC13125, IC13126, IC13128, IC13129, IC13131, IC13141, IC13161, IC13162, IC13163, IC13164, IC13166, IC13173, IC13175, IC13179, IC13185, IC13189, IC13194, IC13195, IC13196, IC13217, IC13222, IC13224, IC13314, IC13316, IC13317, IC13361, IC13373, IC13375, IC13376, IC13377, IC13378, IC13379, IC13380, IC13381, IC13382, IC13383, IC13384, IC13385, IC13386, IC13387, IC13388, IC13389, IC13410, IC13411, IC13412, IC13413, IC13415, IC13416, IC13417, IC13418, IC13419, IC13420, IC13421, IC13427, IC13428, IC13429, IC13430, IC13431, IC13432, IC13433, IC13434, IC13435, IC13437, IC13438, IC13439, IC13440, IC13441, IC13442, IC13454, IC13456, IC13470, IC13472, IC13473, IC13474, IC13475, IC13476, IC13477, IC13478, IC13481, IC13483, IC13485, IC13488, IC13490, IC13491, IC13492, IC13495, IC13496, IC13497, IC13498, IC13501, IC13502, IC13506, IC13507, IC13508, IC13509, IC13510, IC13511, IC13512, IC13513, IC13514, IC13515, IC13516, IC13517, IC13519, IC13521, IC13533, IC13534, IC13535, IC13536, IC13537, IC13538, IC13543, IC13548, IC13549, IC13551, IC13552, IC13553, IC13554, IC13555, IC13556, IC13557, IC13558, IC13559, IC13562, IC13563, IC13564, IC13571, IC13572, IC13573, IC13575, IC13578, IC13589, IC13591, IC13592, IC13593, IC13594, IC13595, IC13596, IC13597, IC13598, IC13599, IC13600, IC13601, IC13602, IC13603, IC13604, IC13605, IC13606, IC13607, IC13608, IC13609, IC13610, IC13619, IC13620, IC13621, IC13622, IC13623, IC13624, IC13625, IC13626, IC13631, IC13632, IC13634, IC13637, IC13648, IC13649, IC13651, IC13652, IC13654, IC13655, IC13656, IC13657, IC13658, IC13659, IC13660, IC13661, IC13662, IC13664, IC13665, IC13666, IC13668, IC13669, IC13670, IC13671, IC13672, IC13673, IC13674, IC13675, IC13676, IC13680, IC13685, IC13692, IC13693, IC13694, IC13695, IC13697, IC13699, IC13700, IC13701, IC13704, IC13734, IC13736, IC13737, IC13738, IC13739, IC13740, IC13742, IC13743, IC13744, IC13745, IC13746, IC13747, IC13754, IC13755, IC13756, IC13757, IC13758, IC13759, IC13760, IC13765, IC13770, IC13771, IC13772, IC13773, IC13830, IC13831, IC13832, IC7182, IC7183, IC7184, IC7185, IC7186, IC7187, IC7188, IC7189, IC7190, IC7191, IC7339, IC7340, IC7341, IC7342, IC7343, IC7519, IC7520, IC7521, IC7560, IC7562, IC7563, IC7565, IC7566, IC7567, IC7568, IC7569, IC7571, IC7572, IC7631, IC7632, IC7633, IC7634, IC7635, IC7641, IC7643, IC7656, IC7684, IC7685, IC7696, IC7697, IC7698, IC7766, IC7767, IC7768, IC7769, IC7770, IC7831, IC7832, IC7833, IC7834, IC7835, IC7840, IC7841, IC7842, IC7843, IC7844, IC7845, IC7856, IC7857, IC7858, IC7859, IC7870, IC7871, IC7872, IC7873, IC7874, IC7875, IC7876, IC7881, IC7891, IC7892, IC7895, IC7907, IC7909, IC7912, IC7913, IC7914, IC7915, IC7916, IC7917, IC7918, IC7921, IC7923, IC7924, IC7925, IC7936, IC7938, IC7948, IC7949, IC7950, IC7951, IC7952, IC7960, IC7961, IC7962, IC7963, IC7986, IC7987, IC7988, IC7989, IC7990, IC7991, IC7992, IC7993, IC7994, IC7995, IC7997, IC7998, IC7999, IC8000, IC8001, IC8005, IC8007, IC8009, IC8010, IC8014, IC8015, IC8017, IC8018, IC8019, IC8020, IC8021, IC8022, IC8023, IC8024, IC8025, IC8027, IC8028, IC8030, IC8033, IC8036, IC8038, IC8039, IC8042, IC8043, IC8044, IC8045, IC8086, IC8087, IC8088, IC8089, IC8090, IC8091, IC8092, IC8093, IC8094, IC8095, IC8096, IC8097, IC8098, IC8099, IC8100, IC8108, IC8109, IC8110, IC8111, IC8112, IC8135, IC8136, IC8137, IC8138, IC8139, IC8140, IC8141, IC8142, IC8143, IC8144, IC8148, IC8149, IC8150, IC8151, IC8152, IC8154, IC8155, IC8156, IC8163, IC8176, IC8177, IC8178, IC8179, IC8180, IC8181, IC8182, IC8183, IC8184, IC8185, IC8191, IC8192, IC8193, IC8194, IC8195, IC8198, IC8199, IC8200, IC8211, IC8212, IC8213, IC8214, IC8215, IC8216, IC8231, IC8232, IC8233, IC8234, IC8235, IC8236, IC8247, IC8265, IC8270, IC8271, IC8272, IC8273, IC8274, IC8275, IC8286, IC8287, IC8288, IC8289, IC8290, IC8291, IC8292, IC8293, IC8305, IC8307, IC8312, IC8313, IC8314, IC8317, IC8318, IC8319, IC8320, IC8321, IC8322, IC8323, IC8324, IC8325, IC8326, IC8327, IC8328, IC8329, IC8330, IC8346, IC8347, IC8348, IC8349, IC8350, IC8351, IC8352, IC8353, IC8354, IC8355, IC8356, IC8357, IC8358, IC8359, IC8360, IC8361, IC8362, IC8363, IC8365, IC8366, IC8367, IC8380, IC8382, IC8383, IC8384, IC8385, IC8386, IC8390, IC8391, IC8392, IC8393, IC8394, IC8398, IC8399, IC8406, IC8407, IC8408, IC8411, IC8412, IC8413, IC8414, IC8415, IC8416, IC8417, IC8418, IC8429, IC8430, IC8431, IC8432, IC8433, IC8434, IC8435, IC8436, IC8437, IC8438, IC8440, IC8452, IC8453, IC8454, IC8455, IC8456, IC8458, IC8459, IC8460, IC8464, IC8466, IC8468, IC8469, IC8470, IC8471, IC8472, IC8473, IC8474, IC8475, IC8477, IC8478, IC8479, IC8480, IC8481, IC8483, IC8484, IC8485, IC8497, IC8500, IC8507, IC8508, IC8510, IC8517, IC8518, IC8519, IC8520, IC8521, IC8522, IC8528, IC8529, IC8530, IC8532, IC8533, IC8534, IC8535, IC8536, IC8537, IC8538, IC8539, IC8546, IC8552, IC8553, IC8562, IC8564, IC8566, IC8598, IC8602, IC8603, IC8604, IC8611, IC8613, IC8615, IC8617, IC8618, IC8620, IC8624, IC8627, IC8631, IC8632, IC8633, IC8634, IC8635, IC8636, IC8637, IC8638, IC8656, IC8667, IC8668, IC8669, IC8671, IC8672, IC8673, IC8674, IC8675, IC8676, IC8678, IC8688, IC8689, IC8695, IC8702, IC8709, IC8710, IC8711, IC8712, IC8713, IC8715, IC8716, IC8717, IC8718, IC8719, IC8720, IC8721, IC8722, IC8724, IC8730, IC8738, IC8742, IC8743, IC8744, IC8745, IC8746, IC8747, IC8756, IC8757, IC8758, IC8759, IC8760, IC8766, IC8769, IC8770, IC8771, IC8772, IC8773, IC8776, IC8777, IC8778, IC8783, IC8786, IC8787, IC8789, IC8790, IC8791, IC8793, IC8794, IC8795, IC8796, IC8797, IC8798, IC8799, IC8800, IC8801, IC8802, IC8803, IC8804, IC8805, IC8806, IC8807, IC8808, IC8809, IC8810, IC8828, IC8829, IC8830, IC8831, IC8832, IC8833, IC8834, IC8835, IC8847, IC8848, IC8849, IC8850, IC8851, IC8852, IC8853, IC8862, IC8867, IC8878, IC8879, IC8881, IC8882, IC8883, IC8884, IC8885, IC8903, IC8904, IC8905, IC8906, IC8907, IC8908, IC8909, IC8925, IC8926, IC8928, IC8934, IC8935, IC8936, IC8938, IC8939, IC8940, IC8941, IC8942, IC8943, IC8944, IC8945, IC8946, IC8947, IC8948, IC8949, IC8950, IC8951, IC8952, IC8953, IC8954, IC8955, IC8956, IC8957, IC8958, IC8959, IC8960, IC8961, IC8962, IC8963, IC8964, IC8965, IC8966, IC8967, IC8968, IC8969, IC8970, IC8971, IC8972, IC8973, IC8974, IC8975, IC8976, IC8977, IC8978, IC8979, IC8981, IC8992, IC8993, IC8994, IC8995, IC8996, IC8997, IC8998, IC8999, IC9026, IC9029, IC9030, IC9032, IC9033, IC9034, IC9035, IC9037, IC9038, IC9040, IC9041, IC9042, IC9043, IC9044, IC9045, IC9046, IC9047, IC9048, IC9049, IC9068, IC9074, IC9079, IC9080, IC9081, IC9082, IC9083, IC9084, IC9086, IC9087, IC9088, IC9089, IC9097, IC9098, IC9099, IC9100, IC9101, IC9102, IC9103, IC9104, IC9116, IC9117, IC9118, IC9119, IC9120, IC9121, IC9122, IC9123, IC9124, IC9141, IC9142, IC9143, IC9144, IC9145, IC9146, IC9147, IC9148, IC9149, IC9150, IC9151, IC9152, IC9153, IC9154, IC9155, IC9156, IC9157, IC9158, IC9159, IC9160, IC9161, IC9162, IC9164, IC9165, IC9166, IC9184, IC9185, IC9186, IC9187, IC9188, IC9195, IC9196, IC9197, IC9198, IC9199, IC9204 , IC9205, IC9206, IC9207, IC9208, IC9209, IC9210, IC9211, IC9212, IC9213, IC9214, IC9218, IC9219, IC9220, IC9221, IC9222, IC9223, IC9224, IC9225, IC9226, IC9227, IC9231, IC9232, IC9243, IC9244, IC9245, IC9246, IC9247, IC9248, IC9249, IC9250, IC9266, IC9267, IC9268, IC9269, IC9270, IC9271, IC9272, IC9273, IC9274, IC9275, IC9276, IC9277, IC9278, IC9279, IC9280, IC9282, IC9283, IC9284, IC9285, IC9286, IC9287, IC9300, IC9301, IC9302, IC9303, IC9304, IC9305, IC9306, IC9307, IC9308, IC9309, IC9310, IC9311, IC9312, IC9313, IC9314, IC9315, IC9316, IC9317, IC9318, IC9319, IC9327, IC9333, IC9334, IC9335, IC9337, IC9338, IC9339, IC9350, IC9351, IC9353, IC9355, IC9357, IC9359, IC9360, IC9361, IC9362, IC9373, IC9374, IC9375, IC9376, IC9377, IC9378, IC9379, IC9380, IC9381, IC9382, IC9405, IC9406, IC9408, IC9409, IC9410, IC9412, IC9418, IC9421, IC9422, IC9423, IC9451, IC9452, IC9453, IC9454, IC9455, IC9456, IC9457, IC9458, IC9460, IC9461, IC9465, IC9466, IC9467, IC9468, IC9469, IC9470, IC9471, IC9472, IC9473, IC9474, IC9475, IC9476, IC9477, IC9478, IC9479, IC9480, IC9481, IC9482, IC9483, IC9484, IC9486, IC9488, IC9493, IC9510, IC9511, IC9512, IC9513, IC9514, IC9515, IC9516, IC9517, IC9518, IC9519, IC9520, IC9521, IC9522, IC9523, IC9524, IC9525, IC9526, IC9527, IC9528, IC9530, IC9532, IC9533, IC9534, IC9535, IC9536, IC9537, IC9538, IC9539, IC9541, IC9542, IC9543, IC9544, IC9545, IC9546, IC9547, IC9548, IC9549, IC9550, IC9551, IC9567, IC9571, IC9572, IC9573, IC9574, IC9575, IC9576, IC9577, IC9578, IC9579, IC9582, IC9583, IC9585, IC9586, IC9587, IC9588, IC9589, IC9594, IC9596, IC9597, IC9598, IC9606, IC9607, IC9608, IC9609, IC9611, IC9612, IC9613, IC9614, IC9617, IC9618, IC9619, IC9620, IC9621, IC9625, IC9649, IC9650, IC9651, IC9652, IC9653, IC9654, IC9655, IC9656, IC9657, IC9658, IC9659, IC9660, IC9661, IC9662, IC9663, IC9665, IC9666, IC9667, IC9668, IC9669, IC9670, IC9671, IC9672, IC9673, IC9675, IC9676, IC9677, IC9678, IC9679, IC9680, IC9681, IC9682, IC9683, IC9684, IC9685, IC9686, IC9688, IC9691, IC9697, IC9698, IC9699, IC9700, IC9701, IC9710, IC9711, IC9712, IC9713, IC9714, IC9715, IC9716, IC9717, IC9718, IC9719, IC9720, IC9721, IC9722, IC9723, IC9724, IC9725, IC9731, IC9733, IC9734, IC9735, IC9736, IC9737, IC9738, IC9739, IC9740, IC9741, IC9742, IC9743, IC9747, IC9748, IC9749, IC9750, IC9751, IC9752, IC9753, IC9754, IC9755, IC9756, IC9768, IC9769, IC9770, IC9771, IC9772, IC9774, IC9775, IC9776, IC9778, IC9779, IC9792, IC9794, IC9795, IC9806, IC9807, IC9808, IC9809, IC9810, IC9812, IC9823, IC9825, IC9826, IC9829, IC9831, IC9836, IC9837, IC9839, IC9840, IC9841, IC9842, IC9843, IC9844, IC9845, IC9846, IC9847, IC9848, IC9849, IC9850, IC9851, IC9852, IC9853, IC9854, IC9855, IC9856, IC9857, IC9858, IC9859, IC9860, IC9861, IC9862, IC9863, IC9864, IC9869, IC9870, IC9871, IC9872, IC9873, IC9874, IC9875, IC9876, IC9877, IC9878, IC9879, IC9880, IC9881, IC9882, IC9883, IC9884, IC9885, IC9886, IC9887, IC9888, IC9899, IC9900, IC9903, IC9904, IC9906, IC9907, IC9909, IC9910, IC9911, IC9912, IC9913, IC9914, IC9916, IC9917, IC9918, IC9920, IC9924, IC9932, IC9933, IC9934, IC9935, IC9936, IC9937, IC9938, IC9939, IC9940, IC9941, IC9963, IC9975, IC9976, IC9977, IC9978, IC9979, IC9980, IC9982, IC9983, IC9984, IC9985, IC9986, IC9987, IC9988, IC9991, IC9992, IC9993, IC9994, IC9995, IC9997, IC9999, IC11516, IC11819, IC12615, IC12627, IC13313, IC13315, IC13696, IC10547, IC10844, IC11055, IC11089, IC11091, IC11097, IC11151, IC11152, IC11494, IC12245, IC12248, IC13505, IC13560, IC13630, IC13633, IC8880, IC9036, IC9464; 7. Product Code: 0042-0040-KR; UDI-DI: 00813502013160; Serial Numbers: IC13611, IC13612, IC13613, IC13614, IC13615, IC13616, IC13722, IC13723, IC13724, IC13725;
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-3150-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class I |
| Status | Ongoing |
| Recall started | August 12, 2026 |
| Reported | September 23, 2026 |
| Company | Abiomed, Inc. |
| Company location | Danvers, MA, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 4100 units (2785 US, 1315 OUS) 11 10 |