Recall Ledger

Oncology sets with drip chambers recall (ICU Medical, Inc.)

Medical deviceClass IIOngoing

Summary: ICU Medical, Inc. is recalling Oncology sets with drip chambers. Reason: Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.

What is being recalled

Oncology sets with drip chambers

Why

Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR. (states named: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, Wyoming)

How to tell if yours is affected

All products manufactured on or before 7/4/2026. UDI-DI: 10840619054486, 10887709091914

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3158-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMay 12, 2026
ReportedSeptember 23, 2026
CompanyICU Medical, Inc.
Company locationSan Clemente, CA, United States
Initiated byVoluntary: Firm initiated
Quantity107,800

More recalls from ICU Medical, Inc. · all medical device recalls · recalls from May 2026