PowerLED 300 Concealed Lighting System recall (Maquet SAS)
Medical deviceClass IIOngoing
Summary: Maquet SAS is recalling PowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm) Model Numbers: 1. ARD568474114A, 2. ARD567501286;. Reason: The potential for the device to detach at the junction between the spring arm and suspension arm and fall causing potential injury.
What is being recalled
PowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm) Model Numbers: 1. ARD568474114A, 2. ARD567501286;
Why
The potential for the device to detach at the junction between the spring arm and suspension arm and fall causing potential injury.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution.
How to tell if yours is affected
1. Model Number: ARD568474114A; UDI-DI: N/A; Serial Numbers: 500190, 500194, 500201, 500203, 500209, 500216, 500218, 500197, 500204, 500210, 500211, 500212, 500213, 500214, 500217, 500206, 500390, 500393, 500397, 500414, 500416, 500417, 500418, 500451, 500266, 500284, 500291, 500293, 500294, 500299, 500300, 500301, 500302, 500306, 500307, 500308, 500117, 500116, 500273, 500274, 500399, 500400, 500402, 500168, 500188, 500192, 500196, 500208, 500219, 500221, 500228, 500233, 500237, 500240, 500247, 500159, 500207, 500215, 500396, 500415, 500419, 500429, 500430, 500433, 500434, 500437, 500439, 500440, 500443, 500478, 500394, 500398, 500403, 500407, 500391, 500404, 500060, 500061, 500062, 500063, 500080, 500081, 500086, 500089, 500094, 500095, 500096, 500099, 500105, 500106, 500111, 500112, 500118, 500119, 500392, 500395, 500405, 500406, 500409, 500427, 500428, 500435, 500436, 500238, 500249, 500251, 500254, 500256, 500258, 500259, 500260, 500261, 500262, 500277, 500278, 500285, 500296, 500304, 500309, 500311, 500312, 500316, 500323, 500328, 500331, 500332, 500338, 500198, 500200, 500067, 500073, 500074, 500076, 500079, 500082, 500069, 500161, 500078, 500051, 500053, 500093, 500100, 500107, 500108, 500113, 500114, 500065, 500070, 500071, 500072, 500075, 500077, 500084, 500085, 500087, 500143, 500145, 500152, 500164, 500167, 500178, 500179, 500253, 500255, 500267, 500268, 500286, 500287, 500292, 500224, 500225, 500229, 500235, 500236, 500128, 500134, 500137, 500148, 500149, 500129, 500173, 500177, 500182, 500184, 500185, 500189, 500195, 500199, 500205, 500169, 500170, 500180, 500142, 500144, 500166, 500172, 500174, 500175, 500181, 500242, 500160, 500162, 500165, 500176, 500157, 500158, 500163, 500171, 500068, 500088, 500131, 500132, 500133, 500151, 500220, 500223, 500226, 500241, 500245, 500248, 500064, 500313, 500315, 500317, 500318, 500320, 500324, 500325, 500326, 500327, 500329, 500330, 500333, 500334, 500335, 500337, 500092, 500097, 500098, 500101, 500102, 500103, 500104, 500109, 500110, 500115, 500120, 500121, 500408, 500359, 500361, 500362, 500363, 500364, 500365, 500368, 500369, 500370, 500372, 500380, 500232, 500239, 500263, 500264, 500269, 500271, 500272, 500373, 500374, 500375, 500386, 500387, 500453, 500454, 500455, 500456, 500461, 500462, 500492, 500493, 500279, 500303, 500305, 500413, 500422, 500425, 500426, 500431, 500438, 500444, 500445, 500446, 500130, 500135, 500136, 500138, 500141, 500146, 500150, 500153, 500154, 500280, 500281, 500282, 500297, 500028, 500029, 500031, 500032, 500033, 500036, 500040, 500041, 500042, 500049, 500056, 500183, 500193, 500257, 500270, 500275, 500276, 500401, 500410, 500411, 500412, 500423, 500424, 500441, 500442, 500447, 500448, 500420, 500421, 500432, 500243, 500037, 500083, 500125, 500126, 500127, 500139, 500140, 500147, 500250, 500252, 500265, 500283, 500288, 500289, 500290, 500295, 500298, 500030, 500034, 500035, 500038, 500039, 500043, 500044, 500045, 500046, 500047, 500048, 500050, 500052, 500054, 500055, 500057, 500227, 500230, 500234, 500244, 500246, 500066, 500222, 500231, 500314, 500319, 500321, 500322, 500336, 500360, 500366, 500367, 500371, 500376, 500377, 500378, 500379, 500381, 500382, 500383, 500384, 500385, 500388, 500186; 2. Model Number: ARD567501286; UDI-DI: 03700712403478; Serial Numbers: 1314H000217400, 4014H000279430, 4014H000279432, 4014H000279433, 4014H000279434, 4014H000279437, 4014H000279438, 4314H000288228, 4314H000288229, 4314H000288232, 4314H000288233, 4314H000288234, 4314H000288237, 4314H000288239, 4314H000288257, 4314H000288258, 4314H000288259, 4714H000296081, 4717H000296095, 1315H000382641, 1715H000392775, 2115H000409030, 2115H000409038, 2115H000409040, 2115H000409062, 2113H000123425, 3613H000150570, 2115H000409063, 3015H000436557, 3015H000436569, 3015H000436571, 2115H000409065, 2115H000409067, 2615H000423256, 3015H000436561, 3015H000436566, 3015H000436570, 1314H000215902, 1314H000215904, 0116H000537022, 0716H000208366, 2816H000647716, 0615H000334882, 0615H000334887, 0615H000334891, 0815H000351763, 0815H000351768, 0815H000351780, 1314H000215893, 1314H000215894, 1314H000215896, 1314H000215900, 1314H000215903, 1314H000215907, 1314H000215885, 2114H000236788, 2613H000132823, 2613H000132825, 2613H000132826, 2613H000132827, 2613H000132828, 2614H000244892, 3613H000150571, 3613H000150572, 4014H000279429, 4014H000279431, 4314H000288230, 4314H000288235, 1314H000215886, 1314H000215895, 1314H000215898, 1314H000215899, 1314H000215908, 2114H000236778, 2114H000236779, 2114H000236787, 2114H000236789, 2114H000236790, 2614H000244886, 2115H000409051, 2115H000409058, 3015H000436556, 3015H000436558, 3015H000436568, 4314H000288231, 1314H000215888, 1314H000215889, 1314H000215906, 1314H000215879, 1314H000215880, 1314H000215881, 1314H000215882, 1314H000215890, 1314H000215892, 1314H000215897, 1314H000215901, 2614H000244885, 2113H000123420, 2113H000123421, 2113H000123423, 2113H000123424, 2113H000123426, 2113H000123427, 2113H000123429, 2613H000132824, 2816H000647710, 2614H000244884, 2614H000244889, 2114H000236777, 2114H000236782, 2114H000236785, 2614H000244877, 2614H000244878, 2614H000244883, 2614H000244890, 1314H000215905, 2115H000409039, 0116H000537017, 0116H000537020, 0116H000537021, 0116H000537023, 0116H000537024, 0116H000537025, 0116H000537026, 0615H000334878, 0615H000334881, 0715H000391839, 0716H000261478, 0716H000561475, 0716H000561479, 0716H000561493, 1314H000217399, 2215H000537018, 2615H000423253, 2816H000647709, 2816H000647712, 2816H000647714, 2916H000647713, 3015H000436560, 3015H000436564, 4314H000288260, 4714H000296097, 1314H000215883, 1314H000215884, 1314H000215887, 0815H000351782, 2115H000409029, 2114H000236780, 2114H000236781, 2114H000236783, 2114H000236784, 2114H000236786, 2614H000244879, 2614H000244880, 2614H000244887, 2614H000244888, 2614H000244891, 2115H000409027, 2115H000409055, 2115H000409032, 2115H000409034, 2115H000409053, 2115H000409064, 2115H000409070, 2615H000423252, 2615H000423254, 3015H000436563, 0116H000537019, 0716H000561481, 3613H000150569, 0615H000334888, 0615H000334889, 0716H000561477, 0716H000561480, 0615H000334883, 0615H000334884, 2115H000409068, 2115H000409069, 2615H000423258, 3015H000436559, 3015H000436562, 3015H000436565, 3015H000436567, 4314H000288236, 4314H000288240, 0615H000334885, 0615H000334886, 0615H000334890, 0815H000351765, 0815H000351767, 0815H000351771, 0815H000351776, 0815H000351777, 0815H000351779, 1314H000215891, 2614H000244881, 2614H000244882, 4014H000279436, 4314H000288238, 4314H000288261, 1813H000118088, 1813H000118093, 2613H000132815, 2613H000132818, 3613H000150533, 3613H000150534, 3613H000150537, 3613H000150538, 3613H000150541, 2613H000132811, 3613H000150544, 3613H000150545, 3613H000150546, 3613H000150547, 3613H000150548, 3613H000150565, 3613H000150566, 3613H000150567, 3613H000150568, 0815H000351774, 4014H000279428, 2115H000409054, 0815H000351773, 1315H000382643, 1315H000382644, 2115H000409028, 2115H000409033, 2115H000409035, 2115H000409036, 2115H000409037, 2115H000409052, 2115H000409056, 2115H000409057, 2115H000409059, 2115H000409060, 2115H000409061, 2816H000647715, 0815H000351775, 0815H000351778, 4014H000279435, 3613H000150535, 3613H000150536, 3613H000150539, 3613H000150540, 3613H000150542, 3613H000150543, 2115H000409031, 0815H000351764, 0815H000351766, 0815H000351769, 0815H000351770, 0815H000351772, 0815H000351781, 1813H000118090, 2113H0001231, 2113H000123428, 2613H000132800, 2613H000132801, 2613H000132802, 2613H000132803, 2613H000132804, 2613H000132806, 2613H000132807, 2613H000132808, 2613H000132809, 2613H000132810, 2613H000132812, 2613H000132813, 2613H000132814, 2813H000132805;
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-3224-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | August 7, 2026 |
| Reported | September 30, 2026 |
| Company | Maquet SAS |
| Company location | Orleans Cedex 2, France |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 667 |
More recalls from Maquet SAS · all medical device recalls · recalls from August 2026