Recall Ledger

SmartPath to dStream for 1.5T recall (Philips North America Llc)

Medical deviceClass IIOngoing

Summary: Philips North America Llc is recalling SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, SmartPath to Ingenia Elition X; Model Numbers (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782145 (3.0T), 5. 781270 (XR and 3.0T), 6. 782113 (XR and 3.0T), 7. 782129 (XR and 3.0T),. Reason: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

What is being recalled

SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, SmartPath to Ingenia Elition X; Model Numbers (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782145 (3.0T), 5. 781270 (XR and 3.0T), 6. 782113 (XR and 3.0T), 7. 782129 (XR and 3.0T), 8. 782118 (Ingenia Elition X), 9. 782144 (Ingenia Elition X);

Why

A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia; (states named: Georgia)

How to tell if yours is affected

1. Model Number (System): 781260 (1.5T) UDI-DI: 884838095076; Serial Numbers: 32953, 8750, 8554, 9013, 18574, 18801, 18988, 20211, 32923, 32778, 8156, 8752, 10139, 10164, 21379, 21175, 32445, 11100, 18583, 18746, 18761, 18853, 18965, 18986, 20003, 21138, 21349, 21384, 21748, 22144, 22266, 30037, 30139, 30146, 32223, 32458, 33044, 33045, 33220, 33246, 33716, 72250, 5572, 8240, 8734, 8808, 8985, 10183, 12030, 20282, 21466, 32643, 8713, 8917, 10087, 10130, 18673, 18701, 18803, 32283, 22220, 8087, 8215, 8923, 13026, 22127, 32467, 8427, 10528, 8862, 21333, 21710, 21768, 22255, 22256, 22257, 21554, 22070, 32049, 5425, 5447, 5484, 5536, 5539, 5550, 8001, 8084, 8096, 8110, 8175, 8263, 8315, 8318, 8321, 8387, 8401, 8407, 8476, 8486, 8559, 8601, 8611, 8642, 8704, 8804, 8885, 9020, 9024, 10008, 10033, 10061, 10075, 10117, 10170, 10174, 10191, 10717, 11000, 11031, 11062, 11102, 11108, 12121, 18013, 18565, 18631, 18690, 18697, 18894, 18906, 18938, 18957, 20057, 20070, 20160, 21015, 21018, 21020, 21082, 21453, 21814, 21932, 21979, 22194, 30097, 32099, 32134, 32163, 32164, 32176, 32333, 32334, 32340, 32341, 32454, 32636, 32710, 32752, 39018, 39031, 39103, 5499, 5505, 8657, 21296, 32680, 10029, 18987, 8054, 8350, 11064, 11072, 12053, 13011, 18575, 18718, 18720, 18940, 20032, 20055, 20206, 20295, 21004, 21087, 21091, 21141, 21184, 21200, 21210, 21223, 21247, 21261, 21312, 21347, 21451, 21555, 21576, 21582, 21611, 21680, 21691, 21694, 21737, 21739, 21755, 21863, 21878, 21912, 21942, 22017, 22018, 22086, 22093, 22103, 22171, 22240, 30132, 30239, 32140, 32158, 32285, 32441, 32449, 32822, 33023, 33051, 33106, 33126, 33179, 39005, 72081, 8213, 5466, 5537, 5546, 8045, 8311, 8395, 8424, 8431, 8641, 8667, 8712, 8786, 8840, 10102, 10185, 10220, 10457, 10552, 10554, 11017, 11046, 12036, 13001, 18069, 18101, 18509, 18511, 18548, 18655, 18788, 18970, 20111, 20132, 20165, 20196, 20413, 21032, 21362, 21551, 21565, 21623, 21746, 21809, 21890, 21892, 22044, 22055, 22182, 22296, 22299, 30033, 32264, 32316, 32478, 32492, 32506, 32670, 32805, 32849, 32990, 33100, 39037, 72239, 72243, 5511, 8012, 8057, 8060, 8094, 8309, 8481, 8517, 8531, 8561, 8711, 8724, 8843, 8857, 8887, 8896, 8949, 10007, 10151, 10498, 10513, 10719, 10906, 10913, 11005, 11042, 11091, 12055, 12114, 12120, 13029, 13046, 13047, 13062, 13073, 18017, 18107, 18524, 18670, 18704, 18929, 18954, 20002, 20011, 20035, 20187, 20195, 20209, 20237, 20263, 20278, 20288, 20302, 21089, 21129, 21182, 21209, 21213, 21222, 21249, 21278, 21326, 21355, 21403, 21448, 21482, 21519, 21560, 21579, 21583, 21640, 21697, 21759, 21780, 21792, 21969, 21987, 22013, 22031, 22098, 22123, 22157, 22175, 22184, 32132, 32217, 32236, 32258, 39009, 21804, 13069, 33352, 22237, 33062, 8091, 10230, 11022, 12169, 18610, 18800, 18948, 21067, 21481, 32709, 33215, 8639, 8686, 18663, 18699, 21019, 21023, 21359, 32119, 32742, 8086, 8292, 10543, 21591, 22189, 32573, 32145, 21994, 21995, 10086, 10556, 8983, 18985, 8166, 8723, 10161, 10530, 22301, 22342, 32391, 8168, 8451, 21498, 18512, 21929, 22152, 21143, 21612, 22233, 22273, 22285, 8616, 20112, 32311, 32513, 33198, 33366, 10570, 11018, 8354, 8694, 9006, 32200, 5479, 8485, 8550, 8746, 8779, 8799, 9032, 10461, 10716, 12047, 21133, 21294, 21577, 32839, 32845, 11085, 12134, 21438, 22100, 22137, 32094, 32147, 32221, 32296, 32374, 32405, 32059, 32180, 8392, 12115, 20182, 22107, 32359, 32245, 8492, 8895, 10079, 20151, 20152, 20198, 21084, 21454, 21880, 32807, 21880, 5475, 8100, 8352, 8560, 8578, 8630, 8760, 8772, 8835, 8906, 8971, 8975, 9001, 10036, 10196, 10482, 10703, 12033, 12034, 18558, 18726, 20226, 20303, 21108, 21290, 21664, 21665, 21788, 32131, 32174, 33385; 2. Model Number (System): 782112 (1.5T) UDI-DI: 884838098886; Serial Numbers: 18857, 21252, 21684, 30313, 32250, 8266, 8552, 18776, 21771, 21803, 22048, 30058, 30089, 32106, 32400, 32487, 32827, 32897, 33000, 33103, 33189, 33275, 33371, 33404, 33494, 33545, 33547, 33565, 33568, 33749, 33751, 18942, 20224, 21495, 21608, 21726, 22247, 32295, 32673, 32762, 33019, 32724, 30054, 32229, 32265, 32418, 32419, 32558, 32596, 32803, 33033, 33248, 33249, 33470, 33647, 72204, 72208, 22141, 8798, 10169, 10190, 10474, 10521, 12029, 13021, 13024, 18810, 21077, 21090, 21158, 21685, 21740, 22027, 39106, 8298, 18714, 18510, 32469, 21940, 32539, 8573, 10128, 18995, 20099, 20159, 21433, 21893, 22290, 32457, 32577, 32580, 32661, 32719, 32725, 32798, 33082, 75012, 8109, 8876, 10222, 18520, 18651, 18892, 18950, 20171, 21205, 21212, 21298, 21513, 21906, 21939, 22328, 32244, 32348, 32355, 32375, 22051, 22298, 30145, 32412, 32690, 33218, 33635, 33061, 8433, 18707, 30273, 32287, 32624, 33639, 32133, 21164, 32327, 33217, 10028, 21708, 13023, 21127, 21297, 33048, 72275, 32188, 32363, 32364, 32426, 22267, 22173, 32610, 32613, 32799, 32859, 32654, 39070, 21463, 32519, 33338, 10221, 20204, 39021, 22045; 3. Model Number (System): 782146 (1.5T) UDI-DI: 884838108691; Serial Numbers: 32679, 33002, 33480, 33570, 30308, 33611, 32251, 22251, 13013, 21990, 32407, 32683, 8935, 20414, 32116; 4. Model Number (System): 782145 (3.0T) UDI-DI: 884838108684; Serial Numbers: 34256, 38295, 74041, 34144, 17377, 38127, 38313, 73004, 23029, 24062, 34208, 17435, 38276; 5. Model Number (System): 781270 (XR and 3.0T) UDI-DI: 884838095083; Serial Numbers: 17536, 17212, 17508, 24013, 38184, 23011, 17290, 34060, 38366, 17434, 38013, 38109, 17302, 17429, 17006, 17378, 17386, 35073, 38062, 17533, 17590, 34111, 38235, 17204, 17292, 17339, 17415, 17416, 17436, 17447, 23030, 24055, 34032, 34142, 34145, 34152, 34211, 35044, 38037, 38040, 38192, 38202, 38317, 17247, 17265, 17268, 17286, 17353, 17367, 17389, 17425, 17509, 17517, 17572, 17580, 24022, 24036, 34001, 34058, 35020, 35078, 38001, 38020, 38135, 38224, 17524, 23026, 24095, 34083, 34107, 34236, 34249, 35060, 17222, 17242, 17245, 17250, 17299, 17308, 17342, 17357, 17388, 17399, 17493, 17499, 17556, 17559, 17564, 17594, 23034, 24047, 24109, 24110, 24111, 24117, 24134, 34036, 34090, 34096, 34221, 35027, 35046, 35048, 35053, 38005, 38006, 38015, 38032, 38041, 38161, 38168, 17344, 17495, 17570, 17615, 24020, 24133, 34067, 34076, 34292, 38263, 17428, 23036, 38026, 38131, 38134, 17550, 24145, 38217, 38225, 17237, 34250, 38138, 17352, 17274, 34037, 34080, 35047, 38116, 17326, 17172, 17490, 24007, 38126, 17356, 38111, 24107, 38065, 38259, 38300, 38023, 38102, 74011, 17206, 17240, 17284, 17313, 17375, 17459, 17516, 24058, 24076, 34019, 34223, 35019, 35031, 35040; 6. Model Number (System): 782113 (XR and 3.0T) UDI-DI: 884838098909; Serial Numbers: 17449, 17455, 24154, 17470, 34059, 34262, 34056, 24052, 38106, 38066, 38082, 24097, 34172, 38049, 38070, 38142, 17238, 17372, 24135, 34143, 38354, 74036, 35070, 34072, 35041, 38105, 17271, 17346, 24021, 24093, 24146, 34117, 34127, 34128, 34129, 34153, 34202, 35045, 35059, 38094, 38290, 38311, 38310, 17560, 38114, 38166, 38261, 38042, 38281, 34110, 38117, 17521, 38010, 34130, 34304, 38219, 38222, 17307, 38075, 17384, 17461; 7. Model Number (System): 782129 (XR and 3.0T) UDI-DI: 884838105805; Serial Numbers: 34055, 38195; 8. Model Number (System): 782118 (Ingenia Elition X) UDI-DI: 884838099746; Serial Numbers: 71019, 71025, 42072, 42054, 42194, 85676, 42113, 42085, 42049, 42098, 42158, 42159, 85655, 85665; 9. Model Number (System): 782144 (Ingenia Elition X); UDI-DI: 884838108677; Serial Numbers: 71278, 42250, 71508, 71366;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3243-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 21, 2026
ReportedSeptember 30, 2026
CompanyPhilips North America Llc
Company locationCambridge, MA, United States
Initiated byVoluntary: Firm initiated
Quantity965

More recalls from Philips North America Llc · all medical device recalls · recalls from August 2026

Official FDA record