Medline Convenience Kits: 1) ANTERIOR SPINE PACK recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) ANTERIOR SPINE PACK, Catalog Number: DYNJT10197; 2) CRANIOTOMY #IMF 56646-LF, Catalog Number: DYNJVB91001; 3) CRANIOTOMY PACK-LF, Catalog Number: DYNJ0369348O; 4) SLUH CRANIOTOMY PACK-LF, Catalog Number: DYNJ0369348P; 5) LAMINECTOMY PACK-LF, Catalog Number: DYNJ. Reason: The kits contain neurological sponges with the potential for unexpected variability of endotoxin levels. Potentially elevated endotoxin levels may present health risks that could require medical intervention, such as neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
What is being recalled
Medline Convenience Kits: 1) ANTERIOR SPINE PACK, Catalog Number: DYNJT10197; 2) CRANIOTOMY #IMF 56646-LF, Catalog Number: DYNJVB91001; 3) CRANIOTOMY PACK-LF, Catalog Number: DYNJ0369348O; 4) SLUH CRANIOTOMY PACK-LF, Catalog Number: DYNJ0369348P; 5) LAMINECTOMY PACK-LF, Catalog Number: DYNJ04342M; 6) LAMINECTOMY PACK-LF, Catalog Number: DYNJ43218G; 7) CRANIOTOMY PACK, Catalog Number: DYNJ46024F; 8) LAMINECTOMY PACK-LF, Catalog Number: DYNJ46070C; 9) CRANI PACK, Catalog Number: DYNJ46672B; 10) SPINE TRAY II, Catalog Number: DYNJ52961I; 11) SPINE PACK, Catalog Number: DYNJ53518C; 12) SPINE PACK, Catalog Number: DYNJ54479C; 13) CRANI ADJUNCT PACK, Catalog Number: DYNJ57815; 14) WC SPINE PACK-LF, Catalog Number: DYNJ57971C; 15) PEDI CRANI MAJOR, Catalog Number: DYNJ59338C; 16) LAMINECTOMY PACK-LF, Catalog Number: DYNJ66742C; 17) SPINAL PACK-LF, Catalog Number: DYNJ67543B; 18) LAMINECTOMY PACK, Catalog Number: DYNJ69214A; 19) EPCH LAMINECTOMY I PACK, Catalog Number: DYNJ83098; 20) SPINE, Catalog Number: DYNJ900269I; 21) PACK,CRANIOTOMY, Catalog Number: DYNJ906893C; 22) PACK,CRANIOTOMY, Catalog Number: DYNJ906893D; 23) PACK,CRANIOTOMY, Catalog Number: DYNJ906893F; 24) UNIVERSITY CRANIOTOMY, Catalog Number: DYNJ907192B; 25) UNIVERSITY CRANIOTOMY, Catalog Number: DYNJ907192C; 26) DR MOSHEL CRANI, Catalog Number: DYNJ907920C; 27) SPINE-LF, Catalog Number: DYNJ910956; 28) TOURO OR LAMINECTOMY PACK, Catalog Number: DYNJ912687
Why
The kits contain neurological sponges with the potential for unexpected variability of endotoxin levels. Potentially elevated endotoxin levels may present health risks that could require medical intervention, such as neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide.
How to tell if yours is affected
1) DYNJT10197, UDI-DI: 10198459778865(each), 40198459778866(case), Lot Number: 26GBA998; 2) DYNJVB91001, UDI-DI: 10193489801170(each), 40193489801171(case), Lot Number: 23HLA649; 3) DYNJVB91001, UDI-DI: 10193489801170(each), 40193489801171(case), Lot Number: 23GLA935; 4) DYNJVB91001, UDI-DI: 10193489801170(each), 40193489801171(case), Lot Number: 23GLA081; 5) DYNJVB91001, UDI-DI: 10193489801170(each), 40193489801171(case), Lot Number: 23CLA457; 6) DYNJVB91001, UDI-DI: 10193489801170(each), 40193489801171(case), Lot Number: 23BLA516; 7) DYNJ0369348O, UDI-DI: 10195327025892(each), 40195327025893(case), Lot Number: 22IBO169; 8) DYNJ0369348O, UDI-DI: 10195327025892(each), 40195327025893(case), Lot Number: 22GBD964; 9) DYNJ0369348O, UDI-DI: 10195327025892(each), 40195327025893(case), Lot Number: 22DBU156; 10) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 24KBH955; 11) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 24JBB595; 12) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 24GMF624; 13) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 24CMJ263; 14) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 24BMK244; 15) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 24AMF230; 16) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 23JBJ151; 17) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 23FBG216; 18) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 23EBC638; 19) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 23CBI768; 20) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 23ABM980; 21) DYNJ0369348P, UDI-DI: 10195327295417(each), 40195327295418(case), Lot Number: 23ABI593; 22) DYNJ04342M, UDI-DI: 10193489914672(each), 40193489914673(case), Lot Number: 23HMB096; 23) DYNJ43218G, UDI-DI: 10195327421977(each), 40195327421978(case), Lot Number: 23HMC253; 24) DYNJ46024F, UDI-DI: 10195327237011(each), 40195327237012(case), Lot Number: 23HBJ773; 25) DYNJ46070C, UDI-DI: 10193489853230(each), 40193489853231(case), Lot Number: 23HMC669; 26) DYNJ46672B, UDI-DI: 10889942278909(each), 40889942278900(case), Lot Number: 23HMA338; 27) DYNJ52961I, UDI-DI: 10195327346119(each), 40195327346110(case), Lot Number: 23HBD216; 28) DYNJ53518C, UDI-DI: 10195327054502(each), 40195327054503(case), Lot Number: 23CBG270; 29) DYNJ53518C, UDI-DI: 10195327054502(each), 40195327054503(case), Lot Number: 23BBB578; 30) DYNJ53518C, UDI-DI: 10195327054502(each), 40195327054503(case), Lot Number: 22LBP140; 31) DYNJ54479C, UDI-DI: 10193489460254(each), 40193489460255(case), Lot Number: 23HBA720; 32) DYNJ57815, UDI-DI: 10889942779437(each), 40889942779438(case), Lot Number: 23HBP930; 33) DYNJ57971C, UDI-DI: 10195327225209(each), 40195327225200(case), Lot Number: 23GBT243; 34) DYNJ59338C, UDI-DI: 10195327021238(each), 40195327021239(case), Lot Number: 23HLA933; 35) DYNJ66742C, UDI-DI: 10195327343156(each), 40195327343157(case), Lot Number: 23GBS430; 36) DYNJ67543B, UDI-DI: 10195327471606(each), 40195327471607(case), Lot Number: 23HBG744; 37) DYNJ69214A, UDI-DI: 10195327184056(each), 40195327184057(case), Lot Number: 23GBR467; 38) DYNJ83098, UDI-DI: 10195327231316(each), 40195327231317(case), Lot Number: 23HMA427; 39) DYNJ900269I, UDI-DI: 10198459569524(each), 40198459569525(case), Lot Number: 26CMI592; 40) DYNJ906893C, UDI-DI: 10193489974713(each), 40193489974714(case), Lot Number: 22IBC626; 41) DYNJ906893C, UDI-DI: 10193489974713(each), 40193489974714(case), Lot Number: 22HBS240; 42) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 24JMH504; 43) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 24JMD199; 44) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 24IMG055; 45) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 24GMJ361; 46) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 24EMD234; 47) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 24DMG428; 48) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 24BMI349; 49) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 24AMG705; 50) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 23LMF389; 51) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 23JMK268; 52) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 23JMB370; 53) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 23IMC078; 54) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 23HMH387; 55) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 23GMD961; 56) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 23DME252; 57) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 23DMB260; 58) DYNJ906893D, UDI-DI: 10195327303372(each), 40195327303373(case), Lot Number: 23CMG273; 59) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 26GMG394; 60) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 26FMD545; 61) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 26DMD616; 62) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 26DMA441; 63) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 26DMA421; 64) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 25KMJ807; 65) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 25KMB625; 66) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 25HMI954; 67) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 25FMC018; 68) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 25EMF909; 69) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 25CMF056; 70) DYNJ906893F, UDI-DI: 10198459168734(each), 40198459168735(case), Lot Number: 25AMF815; 71) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 25GBW221; 72) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 25GBD210; 73) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 25EBM992; 74) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 25DBN524; 75) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 25BBQ529; 76) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 24LBP184; 77) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 24KBH441; 78) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 24JBD802; 79) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 24IBK611; 80) DYNJ907192B, UDI-DI: 24GBX565(each), 10193489946581(case), Lot Number: 40193489946582; 81) DYNJ907192B, UDI-DI: 24FBB829(each), 10193489946581(case), Lot Number: 40193489946582; 82) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 24DBV145; 83) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 24CBM496; 84) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 24ABP283; 85) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 23KBN917; 86) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 23JBR304; 87) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 23IBU169; 88) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 23IBH841; 89) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 23FBH149; 90) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 23EBL017; 91) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 23CBP523; 92) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 23BBF825; 93) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 23BBF826; 94) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 22JBL437; 95) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 22IBA578; 96) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 22IBA287; 97) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 22GBI511; 98) DYNJ907192B, UDI-DI: 10193489946581(each), 40193489946582(case), Lot Number: 22GBH553; 99) DYNJ907192C, UDI-DI: 10198459479588(each), 40198459479589(case), Lot Number: 26FBR970; 100) DYNJ907192C, UDI-DI: 10198459479588(each), 40198459479589(case), Lot Number: 26EBL570; 101) DYNJ907192C, UDI-DI: 10198459479588(each), 40198459479589(case), Lot Number: 26DBL401; 102) DYNJ907192C, UDI-DI: 10198459479588(each), 40198459479589(case), Lot Number: 26CBD787; 103) DYNJ907192C, UDI-DI: 10198459479588(each), 40198459479589(case), Lot Number: 25LBB786; 104) DYNJ907192C, UDI-DI: 10198459479588(each), 40198459479589(case), Lot Number: 25JBQ411; 105) DYNJ907192C, UDI-DI: 10198459479588(each), 40198459479589(case), Lot Number: 25JBH229; 106) DYNJ907920C, UDI-DI: 10198459124839(each), 40198459124830(case), Lot Number: 26DBQ234; 107) DYNJ910956, UDI-DI: 10198459210884(each), 40198459210885(case), Lot Number: 25GMA962; 108) DYNJ910956, UDI-DI: 10198459210884(each), 40198459210885(case), Lot Number: 25CMJ170; 109) DYNJ912687, UDI-DI: 10198459758485(each), 40198459758486(case), Lot Number: 26EBH822
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-3269-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | August 10, 2026 |
| Reported | September 30, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 4013 kits |