Recall Ledger

Philips IntraSight Plus REF: 797423 and 797417 100-120v-220-240v recall (Volcano Corp)

Medical deviceClass IIOngoing

Summary: Volcano Corp is recalling Philips IntraSight Plus REF: 797423 and 797417 100-120v-220-240v,50/60hZ, 1000 VA. Reason: Software system may associate image data from one patient with another patient's record, potentially resulting in incorrect diagnosis or treatment.

What is being recalled

Philips IntraSight Plus REF: 797423 and 797417 100-120v-220-240v,50/60hZ, 1000 VA

Why

Software system may associate image data from one patient with another patient's record, potentially resulting in incorrect diagnosis or treatment.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution in the states of MA, MN, TX, AL, GA, CA and the countries of Belgium, Denmark, Germany, Netherlands, Poland, Spain, Sweden, United Kingdom. (states named: Alabama, California, Georgia, Massachusetts, Minnesota, Texas)

How to tell if yours is affected

UDI: REF: 797423:(01)00845225012946/REF:797417: (01)00845225012939 all Philips IntraSight Plus Systems with Software Version 6.0.4

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3276-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 6, 2026
ReportedOctober 7, 2026
CompanyVolcano Corp
Company locationSan Diego, CA, United States
Initiated byVoluntary: Firm initiated
Quantity19 systems

More recalls from Volcano Corp · all medical device recalls · recalls from August 2026