NOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I. recall (NOXBOX LTD)

Medical deviceClass IIOngoing

Summary: NOXBOX LTD is recalling NOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I.. Reason: Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.

What is being recalled

NOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I.

Why

Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the state of Georgia. (states named: Georgia)

How to tell if yours is affected

Model Number: NOXBOX-I. UDI-DI: 05060541640009. Serial Numbers: NI100825, NI100833, NI100839, NI100840, NI100841, NI100847, NI100848, NI100849, NI100850, NI100851, NI100852, NI100855, NI100858, NI100859, NI100861, NI100862, NI100865, NI100866, NI100868, NI100869, NI100872, NI100873, NI100875, NI100876, NI100878, NI100879, NI100880, NI100881, NI100882, NI100883, NI100884, NI100885, NI100886, NI100887, NI100888, NI100895, NI100932, NI100935, NI100942, NI100948, NI100949, NI100953, NI100961, NI100975, NI100976, NI100978, Serial Number, NI100980, NI100983, NI100985, NI100986, NI100987, NI100989, NI100990, NI100991, NI100993, NI100994, NI100996, NI100997, NI100999, NI101001, NI101002, NI101003, NI101004, NI101005, NI101006, NI101008, NI101009, NI101010, NI101011, NI101012, NI101013, NI101014, NI101015, NI101017, NI101018, NI101019, NI101024, NI101026, NI101028, NI101029, NI101030, NI101031, NI101033, NI101034, NI101035, NI101038, NI101039, NI101041, NI101042, NI101044, NI101048, NI101049, Serial Number, NI101052, NI101057, NI101059, NI101060, NI101061, NI101063, NI101065, NI101067, NI101069, NI101071, NI101072, NI101073, NI101077, NI101079, NI101080, NI101081, NI101083, NI101084, NI101086, NI101088, NI101089, NI101090, NI101091, NI101092, NI101093, NI101094, NI101095, NI101099, NI101100, NI101101, NI101105, NI101106, NI101107, NI101108, NI101109, NI101110, NI101111, NI101112, NI101113, NI101114, NI101115, NI101116, NI101117, NI101125, NI101130, NI101131, Serial Number, NI101132, NI101133, NI101134, NI101135, NI101136, NI101137, NI101138, NI101141, NI101142, NI101144, NI101145, NI101146, NI101147, NI101148, NI101149, NI101150, NI101151, NI101153, NI101154, NI101155, NI101156, NI101158, NI101159, NI101160, NI101161, NI101162, NI101163, NI101164, NI101165, NI101167, NI101168, NI101170, NI101173, NI101174, NI101175, NI101176, NI101177, NI101178, NI101179, NI101180, NI101181, NI101182, NI101183, NI101184, NI101190, NI101191, Serial Number, NI101192, NI101199, NI101202, NI101204, NI101205, NI101207, NI101208, NI101209, NI101211, NI101212, NI101213, NI101214, NI101215, NI101216, NI101217, NI101218, NI101219, NI101220, NI101221, NI101222, NI101223, NI101224, NI101225, NI101227, NI101228, NI101229, NI101230, NI101231, NI101232, NI101233, NI101234, NI101236, NI101237, NI101241, NI101243, NI101244, NI101245, NI101246, NI101249, NI101250, NI101251, NI101252, NI101253, NI101254, NI101255, NI101256, Serial Number, NI101257, NI101258, NI101259, NI101263, NI101264, NI101265, NI101266, NI101268, NI101270, NI101271, NI101273, NI101278, NI101282, NI101284, NI101285, NI101286, NI101287, NI101288, NI101289, NI101292, NI101294, NI101295, NI101297, NI101298, NI101299, NI101300, NI101301, NI101303, NI101305, NI101307, NI101308, NI101309, NI101313, NI101314, NI101317, NI101321, NI101322, NI101323, NI101324, NI101326, NI101327, NI101328, NI101329, NI101330, NI101331, NI101333, Serial Number, NI101335, NI101336, NI101337, NI101338, NI101340, NI101342, NI101344, NI101346, NI101347, NI101349, NI101350, NI101351, NI101353, NI101354, NI101355, NI101356, NI101357, NI101359, NI101360, NI101361, NI101368, NI101372, NI101373, NI101376, NI101377, NI101379, NI101380, NI101381, NI101384, NI101386, NI101387, NI101388, NI101390, NI101392, NI101393, NI101394, NI101395, NI101396, NI101402, NI101404, NI101405, NI101406, NI101407, NI101408, NI101409, NI101410, Serial Number, NI101411, NI101412, NI101413, NI101414, NI101415, NI101416, NI101423, NI101425, NI101426, NI101428, NI101429, NI101430, NI101431, NI101432, NI101433, NI101434, NI101435, NI101436, NI101437, NI101439, NI101441, NI101442, NI101443, NI101444, NI101446, NI101448, NI101449, NI101451, NI101452, NI101454, NI101456, NI101457, NI101459, NI101460, NI101461, NI101462, NI101463, NI101464, NI101467, NI101468, NI101469, NI101472, NI101473, NI101474, NI101475, NI101478, Serial Number, NI101479, NI101482, NI101483, NI101484, NI101485, NI101486, NI101487, NI101489, NI101490, NI101491, NI101492, NI101496, NI101505, NI101506, NI101508, NI101509, NI101512, NI101514, NI101515, NI101516, NI101520, NI101521, NI101522, NI101524, NI101525, NI101526, NI101527, NI101529, NI101530, NI101532, NI101533, NI101540, NI101550, NI101551, NI101552, NI101553, NI101554, NI101557, NI101560, NI101563, NI101565, NI101566, NI101569, NI101570, NI101572, NI101573, Serial Number, NI101574, NI101575, NI101576, NI101577, NI101578, NI101579, NI101581, NI101584, NI101585, NI101588, NI101590, NI101591, NI101592, NI101593, NI101594, NI101595, NI101596, NI101597, NI101598, NI101599, NI101600, NI101601, NI101602, NI101604, NI101605, NI101606, NI101607, NI101608, NI101609, NI101610, NI101611, NI101612, NI101613, NI101614, NI101615, NI101616, NI101617, NI101620, NI101622, NI101623, NI101624, NI101625, NI101626, NI101627, NI101628, NI101629, Serial Number, NI101635, NI101636, NI101638, NI101639, NI101640, NI101641, NI101643, NI101644, NI101645, NI101651, NI101654, NI101655, NI101656, NI101657, NI101658, NI101659, NI101660, NI101662, NI101664, NI101665, NI101666, NI101667, NI101668, NI101669, NI101670, NI101671, NI101672, NI101674, NI101675, NI101676, NI101677, NI101678, NI101679, NI101680, NI101681, NI101682, NI101683, NI101684, NI101685, NI101686, NI101687, NI101688, NI101689, NI101690, NI101691, NI101692, Serial Number, NI101693, NI101694, NI101695, NI101696, NI101697, NI101699, NI101700, NI101703, NI101704, NI101705, NI101706, NI101707, NI101708, NI101709, NI101710, NI101711, NI101712, NI101713, NI101714, NI101715, NI101716, NI101717, NI101718, NI101719, NI101720, NI101721, NI101722, NI101723, NI101724, NI101726, NI101727, NI101729, NI101731, NI101732, NI101733, NI101735, NI101736, NI101737, NI101739, NI101740, NI101742, NI101743, NI101744, NI101746, NI101747, NI101748, Serial Number, NI101749, NI101750, NI101751, NI101753, NI101756, NI101759, NI101760, NI101761, NI101763, NI101765, NI101766, NI101767, NI101768, NI101769, NI101770, NI101771, NI101775, NI101776, NI101777, NI101778, NI101781, NI101782, NI101783, NI101784, NI101787, NI101791, NI101793, NI101794, NI101795, NI101796, NI101797, NI101798, NI101799, NI101804, NI101805, NI101806, NI101807, NI101808, NI101809, NI101810, NI101811, NI101812, NI101813, NI101814, NI101815, NI101816, Serial Number, NI101817, NI101818, NI101819, NI101820, NI101821, NI101822, NI101823, NI101824, NI101825, NI101826, NI101828, NI101829, NI101831, NI101832, NI101834, NI101835, NI101836, NI101837, NI101838, NI101840, NI101841, NI101842, NI101843, NI101844, NI101845, NI101846, NI101848, NI101849, NI101851, NI101854, NI101857, NI101859, NI101861, NI101862, NI101864, NI101865, NI101867, NI101868, NI101869, NI101872, NI101873, NI101874, NI101875, NI101876, NI101878, NI101879, Serial Number, NI101880, NI101881, NI101882, NI101883, NI101884, NI101885, NI101886, NI101887, NI101889, NI101890, NI101891, NI101892, NI101895, NI101896, NI101898, NI101899, NI101901, NI101902, NI101903, NI101904, NI101905, NI101906, NI101907, NI101908, NI101909, NI101910, NI101912, NI101913, NI101914, NI101916, NI101919, NI101920, NI101922, NI101924, NI101925, NI101926, NI101928, NI101936, NI101937, NI101939, NI101940, NI101941, NI101942, NI101943, NI101944, NI101945, Serial Number, NI101946, NI101947, NI101948, NI101952, NI101953, NI101954, NI101955, NI101956, NI101957, NI101958, NI101959, NI101960, NI101961, NI101962, NI101963, NI101964, NI101965, NI101966, NI101967, NI101968, NI101970, NI101971, NI101972, NI101973, NI101974, NI101975, NI101976, NI101977, NI101978, NI101979, NI101981, NI101982, NI101983, NI101984, NI101985, NI101986, NI101987, NI101988, NI101989, NI101990, NI101991, NI101992, NI101993, NI101994, NI101995, NI101996, Serial Number, NI101997, NI101998, NI101999, NI102000, NI102001, NI102002, NI102003, NI102004, NI102006, NI102007, NI102008, NI102009, NI102010, NI102011, NI102012, NI102013, NI102014, NI102015, NI102017, NI102018, NI102019, NI102020, NI102021, NI102023, NI102024, NI102025, NI102026, NI102027, NI102028, NI102031, NI102032, NI102033, NI102035, NI102036, NI102037, NI102038, NI102040, NI102043, NI102044, NI102045, NI102046, NI102047, NI102048, NI102049, NI102051, NI102052, Serial Number, NI102053, NI102055, NI102056, NI102057, NI102059, NI102060, NI102061, NI102062, NI102063, NI102065, NI102067, NI102068, NI102069, NI102070, NI102071, NI102072, NI102074, NI102075, NI102076, NI102077, NI102078, NI102079, NI102081, NI102082, NI102083, NI102086, NI102089, NI102090, NI102091, NI102092, NI102093, NI102094, NI102095, NI102096, NI102097, NI102098, NI102099, NI102102, NI102103, NI102104, NI102105, NI102106, NI102107, NI102108, NI102109, NI102110, Serial Number, NI102111, NI102113, NI102114, NI102115, NI102116, NI102117, NI102118, NI102119, NI102120, NI102121, NI102123, NI102124, NI102125, NI102126, NI102127, NI102128, NI102129, NI102131, NI102132, NI102133, NI102134, NI102135, NI102136, NI102137, NI102138, NI102139, NI102140, NI102141, NI102142, NI102144, NI102145, NI102146, NI102147, NI102148, NI102149, NI102150, NI102152, NI102155, NI102160, NI102161, NI102163, NI102164, NI102165, NI102166, NI102167, NI102168, Serial Number, NI102169, NI102170, NI102171, NI102172, NI102173, NI102174, NI102175, NI102176, NI102177, NI102180, NI102182, NI102183, NI102184, NI102186, NI102188, NI102189, NI102190, NI102191, NI102192, NI102193, NI102194, NI102195, NI102196, NI102197, NI102198, NI102199, NI102200, NI102205, NI102206, NI102208, NI102209, NI102212, NI102213, NI102214, NI102215, NI102216, NI102219, NI102221, NI102222, NI102226, NI102227, NI102228, NI102230, NI102231, NI102232, NI102233, Serial Number, NI102234, NI102235, NI102237, NI102238, NI102239, NI102240, NI102241, NI102244, NI102245, NI102246, NI102247, NI102249, NI102250, NI102253, NI102254, NI102255, NI102256, NI102259, NI102263, NI102265, NI102266, NI102268, NI102269, NI102270, NI102271, NI102272, NI102273, NI102274, NI102275, NI102276, NI102277, NI102278, NI102279, NI102280, NI102281, NI102282, NI102283, NI102284, NI102285, NI102286, NI102287, NI102288, NI102290, NI102291, NI102292, NI102293, Serial Number, NI102295, NI102296, NI102297, NI102298, NI102299, NI102300, NI102304, NI102305, NI102306, NI102307, NI102308, NI102309, NI102310, NI102311, NI102327, NI102329, NI102330, NI102331, NI102332, NI102338, NI102339, NI102340, NI102341, NI102342, NI102343, NI102345, NI102346, NI102347, NI102349, NI102350, NI102351, NI102355, NI102356, NI102357, NI102358, NI102359, NI102360, NI102361, NI102362, NI102363, NI102364, NI102366, NI102377, NI102380, NI102381, NI102383, Serial Number, NI102384, NI102385, NI102387, NI102389, NI102390, NI102391, NI102392, NI102393, NI102394, NI102395, NI102396, NI102397, NI102398, NI102399, NI102400, NI102402, NI102403, NI102404, NI102405, NI102406, NI102407, NI102408, NI102410, NI102411, NI102413, NI102414, NI102415, NI102417, NI102418, NI102422, NI102423, NI102424, NI102425, NI102426, NI102427, NI102428, NI102429, NI102431, NI102432, NI102433, NI102434, NI102435, NI102436, NI102437, NI102438, NI102444, Serial Number, NI102445, NI102447, NI102463, NI102466, NI102467, NI102472, NI102473, NI102474, NI102475, NI102476, NI102477, NI102478, NI102479, NI102480, NI102481, NI102482, NI102483, NI102484, NI102485, NI102486, NI102487, NI102488, NI102489, NI102490, NI102492, NI102494, NI102495, NI102496, NI102497, NI102498, NI102499, NI102500, NI102501, NI102502, NI102503, NI102504, NI102506, NI102507, NI102509, NI102510, NI102511, NI102512, NI102513, NI102517, NI102518, NI102519, Serial Number, NI102520, NI102522, NI102525, NI102528, NI102530, NI102531, NI102535, NI102536, NI102537, NI102538, NI102539, NI102540, NI102541, NI102542, NI102543, NI102544, NI102545, NI102546, NI102547, NI102548, NI102550, NI102551, NI102552, NI102553, NI102554, NI102555, NI102556, NI102557, NI102558, NI102560, NI102561, NI102562, NI102563, NI102564, NI102565, NI102566, NI102567, NI102568, NI102569, NI102570, NI102571, NI102574, NI102575, NI102576, NI102577, NI102578, Serial Number, NI102579, NI102580, NI102581, NI102582, NI102583, NI102584, NI102585, NI102586, NI102587, NI102588, NI102590, NI102591, NI102592, NI102593, NI102594, NI102595, NI102596, NI102597, NI102598, NI102599, NI102600, NI102601, NI102602, NI102604, NI102605, NI102606, NI102607, NI102608, NI102610, NI102611, NI102613, NI102614, NI102615, NI102616, NI102617, NI102618, NI102619, NI102620, NI102621, NI102622, NI102623, NI102624, NI102625, NI102627, NI102628, NI102629, Serial Number, NI102630, NI102631, NI102632, NI102634, NI102635, NI102637, NI102638, NI102639, NI102640, NI102641, NI102642, NI102643, NI102645, NI102647, NI102648, NI102649, NI102651, NI102652, NI102653, NI102655, NI102656, NI102657, NI102658, NI102659, NI102660, NI102661, NI102662, NI102663, NI102664, NI102665, NI102666, NI102667, NI102668, NI102669, NI102670, NI102671, NI102672, NI102673, NI102674, NI102675, NI102676, NI102680, NI102681, NI102683, NI102684, NI102685, Serial Number, NI102686, NI102687, NI102688, NI102689, NI102691, NI102692, NI102693, NI102695, NI102696, NI102697, NI102698, NI102699, NI102700, NI102701, NI102702, NI102703, NI102704, NI102705, NI102706, NI102707, NI102708, NI102709, NI102710, NI102711, NI102713, NI102714, NI102715, NI102717, NI102720, NI102722, NI102724, NI102725, NI102726, NI102727, NI102728, NI102729, NI102730, NI102731, NI102732, NI102733, NI102734, NI102735, NI102736, NI102737, NI102738, NI102739, Serial Number, NI102740, NI102741, NI102742, NI102743, NI102744, NI102745, NI102746, NI102747, NI102748, NI102751, NI102752, NI102753, NI102755, NI102756, NI102757, NI102758, NI102759, NI102760, NI102762, NI102763, NI102764, NI102765, NI102766, NI102767, NI102768, NI102769, NI102770, NI102771, NI102772, NI102773, NI102774, NI102776, NI102777, NI102778, NI102781, NI102783, NI102784, NI102785, NI102786, NI102789, NI102790, NI102791, NI102792, NI102793, NI102794, NI102795, Serial Number, NI102798, NI102799, NI102800, NI102801, NI102802, NI102803, NI102804, NI102805, NI102806, NI102807, NI102808, NI102809, NI102810, NI102811, NI102812, NI102813, NI102814, NI102815, NI102816, NI102817, NI102818, NI102819, NI102821, NI102823, NI102824, NI102825, NI102828, NI102830, NI102832, NI102833, NI102834, NI102836, NI102837, NI102838, NI102839, NI102840, NI102841, NI102842, NI102843, NI102844, NI102845, NI102846, NI102847, NI102848, NI102849, NI102850, Serial Number, NI102851, NI102852, NI102853, NI102854, NI102855, NI102856, NI102858, NI102859, NI102860, NI102861, NI102862, NI102863, NI102865, NI102866, NI102868, NI102870, NI102871, NI102872, NI102873, NI102874, NI102876, NI102877, NI102878, NI102879, NI102880, NI102881, NI102882, NI102883, NI102884, NI102885, NI102886, NI102887, NI102888, NI102889, NI102890, NI102891, NI102892, NI102893, NI102894, NI102897, NI102898, NI102899, NI102900, NI102901, NI102902, NI102904, Serial Number, NI102905, NI102906, NI102907, NI102908, NI102910, NI102911, NI102913, NI102914, NI102915, NI102916, NI102917, NI102918, NI102919, NI102920, NI102921, NI102922, NI102923, NI102926, NI102927, NI102928, NI102929, NI102930, NI102931, NI102932, NI102933, NI102934, NI102935, NI102936, NI102937, NI102939, NI102942, NI102944, NI102947, NI102950, NI102955, NI102957, NI102958, NI102961, NI102962, NI102963, NI102964, NI102966, NI102967, NI102968, NI102970, NI102973, Serial Number, NI102977, NI102979, NI102986, NI102987, NI102989, NI102990, NI102991, NI102992, NI102996, NI102997, NI102998, NI102999, NI103001, NI103002, NI103003, NI103007, NI103011, NI103012, NI103014, NI103016, NI103017, NI103018, NI103019, NI103020, NI103021, NI103022, NI103023, NI103024, NI103074, NI103079, NI103082, NI103083, NI103084, NI103085, NI103087, NI103088, NI103089, NI103091, NI103093, NI103094, NI103095, NI103104, NI103116, NI103117, NI103120, NI103122, Serial Number, NI103123, NI103126, NI103127, NI103131, NI103136, NI103139, NI103144, NI103145, NI103146, NI103147, NI103148, NI103151, NI103152, NI103153, NI103155, NI103157, NI103159, NI103168, NI103170, NI103181, NI103182, NI103183, NI103189, NI103196, NI103203, NI103204, NI103208, NI103215, NI103220. Includes NOxBOXi with Oxygen sensor lots manufactured before April 13, 2026

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3306-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedSeptember 14, 2026
ReportedOctober 7, 2026
CompanyNOXBOX LTD
Company locationSittingbourne, United Kingdom
Initiated byVoluntary: Firm initiated
Quantity1,667 units