Recall Ledger

Impulsator recall (Percussionaire Corporation)

Medical deviceClass IIOngoing

Summary: Percussionaire Corporation is recalling Impulsator, Product Number: S00010, Phasitron 5, Product Number: P5 used with Instruction Manual. Reason: Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation.

What is being recalled

Impulsator, Product Number: S00010, Phasitron 5, Product Number: P5 used with Instruction Manual

Why

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa. (states named: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia)

How to tell if yours is affected

Product Number/UDI-DI: S00010/00849436000389, P5/00849436000501, Instruction Manual: P20047 Rev E

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-3323-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 18, 2026
ReportedOctober 7, 2026
CompanyPercussionaire Corporation
Company locationSandpoint, ID, United States
Initiated byVoluntary: Firm initiated
Quantity228,277

More recalls from Percussionaire Corporation · all medical device recalls · recalls from August 2026

Official FDA record